Browse
(440) 299-7778
Menu
(440) 299-7778
One-year parts and labor warranty30-day money-back guaranteeContact

The text of a manual · part 9 of 47

Bulletin 2803 VIM Vision Input Module User Manual: 5 - Connection and Power-up, pages 79 to 87

Pages 79 to 87 of 404, from Bulletin 2803 VIM Vision Input Module User Manual

The original document is the PDF (26 MB). This is its text as extracted, unedited; tables and drawings read best in the PDF, and each page heading opens it at that page.

For the Allen-Bradley 2803-VIM2

Page 79

Connection and Power- up

Chapter 5

Swinganm

Connections

5-79

Trigger Input - Terminals 3 and 4

(continued) Figure 5.34 ipical Wiring for Current Sourcing Sensor Trigger when sensor goes ON

,t 03 Trigger - 04

0 8 3

I SUPPlY r +

Source output

Power

SuPPlY

Sensor

-

Gnd

Swingarm

Trigger when sensor goes OFF

SUPPlY m

.

+

‘V* 4. 03 Triggter -- 04

Source output

\; J ,I mm-m-2

SUPPlY

Sensor Rl -

Power

Gnd

Swingarm

RI = Resistor, 470 Ohms, typical

A

Page 80

Connection and Power-up

Chapter 5,

5-20

Trigger Input - Terminals 3 and 4

Swingarm Connections (continued)

Figure 5.15 ypical Wiring for Current Sinking Sensor Trigger when sensor goes OFF

1 SUPPlY +

V+ Sensor ------

+ 03 Trigger - 04 output

.I

1

SUPPlY r’! I -Gnd 1

L

I

Power 1 I

Swingarm Rl = Resistor, 470 Ohms, typical

Trigger when sensor goes ON lol -

1 SUPPlY

0 v+

+ 03 Trigger - 04

Sensor

I Sink output

. I 1 I A i -Gnd

L

Swinga,rm

------

Power SUPPlY

Page 81

Chapter

Connection and Power-up 5

5 wingarm Connections

5-27

Trigger

Input - Terminals

3 and 4

(continued)

Pushbutton Triggering: You may want to trigger the VIM2 module using a push button connected to the swingarm. This is useful for lab testing and system integration. Two examples of triggering circuits are provided in Figures 5.16 and 5.17. One is a “single shot” trigger which initiates a single inspection cycle and the other is a continuous trigger which repeats cycles as long as the button is pressed (Note: The module must be set to “SWG TRIG” mode - see Chapter 7 - to use the pushbutton trigger as shown). “Single Shot” Push Button -This circuit diagram describes the connections for a push-button switch which you may attach to the swingarm to manually control triggering of the VIM2 module. The module must be set to “SWG TRIG” mode to use the pushbutton trigger as shown. Figure 5.16 “Single Shot” Push-button

Circuit

Swingarm + 01 Camera Pwr. - 02 + 03 Trigger - 04 +05 Strobe -06 +07 Decision - 08

+09 Busy -10

CLH 12

1

51 = Momentary push button (normally open). Such as A-8 Cat. No. 800-AMZAR or 800-AMZAG.

Page 82

Chapter 5-22

Swingam

Connections

5

Connection and Power-up

Trigger Input-Terminals

3 and 4

(con timed)

“Continuous” Push Button -The continuous trigger will repeatedly cycle the VIM2 module as long as the push button is pressed. The BUSYoutput is used to loop back a Trigger signal. Figure 5.17 “Cnntinlinllc” --......----

Pllrh.huttnn . --.. --_.-..

Circuit -.. --.-

Swingarm

+ 01 Camera Pwr. - 02 + 03 Trigger - 04 +05 Strobe -06 +07 Decision -08

CLL 11 CLH 12

Rl = Resistor, 560 Ohms, 20% 51 = Momentary Push Button. Normally open. A-B Cat No. 800-AMZAR or 800-AMZAG Open = No Trigger Closed = Trigger Oscillations

Page 83

Chapter

Swingar,m Connections (continued)

5

Connection and Power-up 5-23

Strobe Light Output - Terminals

5 and 6

This output provides a 5 volt ‘M’L strobe light (positive edge) trigger. The strobe trigger cable must be of shielded type and terminated (to 7.5k ohms, typical). Shielded cables reduce noise and interference and are highly recommended.

Note: Refer to Figure 5.18 for typical wiring to a strobe light trigger input.

Figure 5.18 Wirina for Strobe Outtxit to Strobe Triaaer Input

TyDid

0 Rrc\ u l-l-

Strobe Light

Strobe Trigger Input

+ 05 Strobe Trigger - 06

I Rl = Resistor, 7.5 K Ohms, typical

Page 84

Chapter 5-24

Swingarm

Connections (con tin ued)

5

Connection and Power-up

Decision Output-Terminals 7 and 8 The VIM2 module communicates summary accept/reject results through the Decision discrete output bit, and through the Decision terminal on the swingarm. The Decision transmits a reject signal if any of the acceptance range tests for the line gages, windows, probe, or math tools fail. The swingarm Decision output toggle ON/OFF according to the inspection result. This is an open-collector transistor type output, rated for 3 to 32 VDC, 1 ampere. Accept = Decision LED OFF, high impedance (approximately 1 kohm). REJECT = Decision LED ON, low impedance to common. Note: You can use the 12.5 VDC camera power as a current source for the Decision output. Refer to Figure 5.19 for typical wiring to a 1771 input board. Figure 5.19 Typical Wiring for Decision Output (to 1771~IB input Board) r Swingarm + 01 Camera Power -02

17714B Input .

I

- DC common

Input +07 Decision -08

Page 85

Connection and Power-up

Chapter 5

Swingarm Connections (continued)

5-25

Busy Output-Terminals

9 and 10

The Busy signal indicates that the module is busy and a decision is pending. It comes on as soon as the trigger has been received and goes off after trigger has been reset and the Decision output is set/reset. This is an open-collector transistor type output, rated for 3 to 32 VDC, 1 ampere. Busy = low impedance to common and Busy LED ON. Not Busy or Ready = high impedance (approximately 1 kohm) and B my LED OFF.

A t l

WARNING: Do not use the Busy output (terminals 9 and 10) to directly energize external equipment. The Busy output is energized during power-up and reset to indicate that the module is not yet ready to perform an inspection.

Note: You can use the 12.5 VDC camera power as a current source for the Busy output. Refer to Figure 5.20 for typical wiring to a 1771 input board. Figure 5.20 TvDical Wirina for Busv Outr>ut (to 17714B InDut Board) Swingarm + 01 Camera Power -02

17714B Input

0

I

- DC common

0 0 0 0 +09 Busy -10

7 input

Page 86

Chapter S-26

Swingarm

Connections (continued)

5

Connection and Power-up

RS-485 Port - Terminals

11 and 12

Connect your external host computer, if you intend to use one, to the RS-485 port at swingarm terminals 11 and 12. You will need to supply the communications cable to link your host device to the VIM2 module. Use shielded, twistedpair cabling (such as Belden type 9841) for noise immunity. The VIM2 RS-485 port has been designed to comply with Electronic Industries Association (EIA) standards for RS-485 balanced digital point-to-point systems. Consult the EIA RS-485 standards, if necessary, for additional information. Also see Chapter 15 for more information regarding the use of the RS-485 port. Communication terminals (11 and 12) are marked as follows (see Figure 5.21 f or typical wiring to the RS-485 port): Terrninal 11 - CLL (communication line low)

Terminal 12 - CLH (communication line high)

Note: Reversing the connections of the RS-485 port to the host will not damage the unit, but will prevent any communication from occurring. For connecting host devices with RS-232 ports, use of an RS232/RS-485 adapter is required. Figure 5.21 pical Wirina for RS-485 Port

Shielded Cable

To host

-

Page 87

Chapter

Indicdt~r LEDs

5

Connection and Power-up 5-27

The VIM2 module is equipped with seven indicator LEDs. The LEDs are located on the front panel of the VIM2 module (see Figure 5.6) and depict the current status of the module. PWR - This Power Indicator LED goes ON when the chassis power is on, OFF when the chassis power is offi CPU FAULT-This LED goes ON when a possible hardware failure has been detected. During normal operation it will be OFF. When this light is ON, the module does not respond to triggers, and the Decision output goes OFF (Pow). Note: At power-up, the CPU FAULT LED goes ON until the reset sequence is complete. CONFIG FAULT -This Configuration Fault LED indicates that the module is not properly configured. This can be caused by a loss of memory, an invalid configuration download, a power loss during configuration transfer or by the loss of a “train-throughthe-lens” mask due to a power outage. When this light is ON, the module will not respond to triggers, and the Decision output goes OFF. Clear the configuration fault by either of two means: l

Configure the VIM2 module with the light pen, or l

Download a configuration from the master host device (PLC or serial host), if there is one.

The CONFIG FAULT LED goes ON at power-up until the reset sequence is complete. ACQ ERROR -This Acquisition Error LED indicates that the camera is not acquiring a satisfactory image. The LED goes ON when any of these conditions occur: l

The brightness probe is out of range. This can be caused by a disconnected camera or loss of illumination, such as a burnt-out light bulb.

l

Either the X or Y float gage values are out of range or in error, such as if, for example, the X or Y-gage fails to find a blob edge.

l

Any window or line gage is floated too far so that it collides with the edge of the screen. “Through-thelens” windows will not cause an ACQ ERROR.

The “ACQ ERROR” conditions cause a reject decision, and cause the Decision LED to go ON.