Page 28

GAEBRIDGE ENCODER SIMULATOR This piece of test equipment can also be used for testing RGO32Bit counter modules. Although using the BEI incremental encoder in my experience is a better method, this unit can be set up to send a given number of pulses to the encoder module. This is useful if missing bits are suspected as it is very difficult to move the BEI encoder shaft to generate a known number of counts. The simulator is very easy to use. The cable plugs directly into the 37way D connector on the counter module.

Figure 22 GAEBRIDGE encoder simulator In the photo, the simulator is setup to send 8192 pulses to the counter module. The thumb wheel switch can be set to any value up to 9999, but its common sense to select a value which equals a standard binary value, e.g. 1024, 2048, 4096 and the highest value with this unit - 8192. The COARSE and FINE knobs control the velocity of the bit rate sent. These I set usually to their maximum values. The two toggle switches are: 1. Count direction - UP (U) or DOWN (D) 2. Mode of operation - CONTINUOUS (C) or BURST (B) The push buttons ON and OFF will start or stop the counts being sent to the encoder module. In the photo, the settings correspond to an UP count of 8192 with BURST mode selected. Thus when the ON button is pushed, a single stream of 8192 pulse will be sent at full velocity rate to the module under test.

John Mills

CAMAC TEST PROCEDURES

26

Page 29

USING THE SIMULATOR To use the simulator, after the cable has been connected to the module, take the crate off line and perform a C/Z. This resets all the counter registers to zero. Make sure the RGO32Bit counter is receiving the 20Hz clock signal from the frequency generator. Put the crate back on line and push the ON button. With the value of 8192 selected, the LED’s on the module will count up in binary fashion, but finally stopping with just one displayed. This being bit 15. See photo below.

Figure 23 Bit 15 displayed Another point regarding the operation of the simulator and demonstrated in the image is the CLOCK and ZEROSET green Leds are illuminated but not the 1PPR indicator. There is a push button on the simulator labelled Z.M. (Zero Marker) for emulating the ZEROSET pulse, but there is no switch or push button for a 1PPR test. For this reason, it is best in the first instance to use the BEI encoder for testing as this sends out a REAL 1ppr signal. If the simulator is put in CONTINUOUS mode with the 8192 count selected, this is a good method of testing for changes in the MSB bits as it can be left running for a long period which would be laborious if rotating the BEI encoder shaft by hand.

This completes the use of the encoder simulator and RGO32Bit counter tests.

TESTING MISCELLANEOUS MODULES MULTIDAC (Digital to Analogue Converter) The 9085 MULTIDAC modules as used in the INT/JKT are easy to test. All that is required is a LEMO cable with an open end which can be connected to a MULTIMETER. This is set on a low voltage range. The DAC uses 12 bit sampling and generates a +5 and –5V output voltage. These modules are setup as such that a binary word of 2048 gives a 0V (zero volts) output. A data word of 4096 gives a +5V output and 0000 a minus -5V output. Use the WGR2403 word generator (in slot N4) to produce the binary word for the DAC to output these voltages. The DAC under test would be setup (say in slot N8) with an F16 or A0, A1, A2 or A3 set on the SC-TST-1 to write to the appropriate output channel. There are 4 channels used on the INT There is a trim-pot on the module for output voltage calibration. The user manual will explain how to adjust this. It should be noted the 9085 is no longer made and a HYTEC 640VSL substitutes for this. Refer to the User Manual for more information.

John Mills

CAMAC TEST PROCEDURES

27

Page 30

HYTEC ADC1232 (Analogue to Digital) Module There are several versions of ADC modules in use, but they all have similar characteristics and use the same CAMAC commands for addressing them. The HYTEC ADC1232 being the most common one in use. There are 32 input channels. Channels 1 to 16 are on the upper 37way D connector and channels 17 to 32 on the lower 37way D connector. The ADC1232 samples to 12 bits resolution. These ADC modules are primarily used for measuring the temperature sensors fitted on the trusses of the telescopes and on the mirror. To test these modules, there is a special Break Out Box (BOB). This allows input voltages from the Precision Voltage Generator (which has a thumbwheel switch for selecting voltages in small increments) which can be applied to the input channels of the ADC module. The BOB cables are fitted with 37way standard D type connectors (one male the other female) so the box can also be used in series with the ADC to measure for example the voltages coming in from the temperature sensors. The terminal pins on the BOB are arranged as such that pins 1 to 16 are the ADC input channels (+Ve) and pins 20 to 36 are the OV returns.

Figure 24 Break out box for ADC modules Pins 1 and 20 are Channel 1 Pins 2 and 21 are Channel 2 Pins 3 and 22 are Channel 3 Pins 4 and 23 are Channel 4 Etc...

F=0 A=0 F=0 A=1 F=0 A=2 F=0 A=3

To test channels 17 to 32 (on the lower connector) the F bit is changed to F=1 Pins 1 and 20 are Channel 17 Pins 2 and 21 are Channel 18 Pins 3 and 22 are Channel 19 Pins 4 and 23 are Channel 20 Etc...

F=1 A=0 F=1 A=1 F=1 A=2 F=1 A=3

To test the ADC, as with previous tests - Insert the module into slot N8 e.g. To test ADC Channel 3. Toggle the switch A2 on the SC-TST-1 to ‘1’. All other A and F switches are set to ‘0’ Operation 1 - N8 A2 F0 (read ADC channel 3) Operation 2 - N1 A0 F16 (write to the DTM4) 1. Connect the +Ve and – Ve leads from the Precision Voltage Source across pins 3 and 22 of the BOB. 2. The User Manual will give the information regarding binary counts for a particular voltage input. I think 10mV (LSB - bit 0) is the lowest voltage change that the ADC will detect. 3. Step through the voltage increments using the thumbwheel switches on the Precision Voltage Generator and check what bits are displayed on the DTM4. The binary counts should match the voltage selected +/- 1 bit. The Users manual gives instructions on how to calibrate the ADC if required. To test Channels 17 to 32, use the lower connector and toggle the F1 switch to ‘1’ and set the A switches to the number of the channel to be tested - This completes the ADC test.

John Mills

CAMAC TEST PROCEDURES

28

Page 31

In this document, I have covered the testing of most of the CAMAC modules in use at ING. Those outstanding are the ED012 pulse generator, the LP34 millisecond generator and the 3340 RS232 communications modules (as used in the INT/JKT). The ED012 and LP34 modules are easy to test and the User Manuals for these modules are readily available. The 3340 RS232 serial communications modules are somewhat difficult to test and require a VT220 terminal or similar to be connected to the 25way D socket on this module. If time is available at a later date, I will expand the information in this document to cover the testing of these other modules. John Mills - June 2010

---------------------------------------------------------------------------------------------------------------------------------------

John Mills

CAMAC TEST PROCEDURES

29

Page 32 read from a scan by OCR; it may hold reading errors

CAMAC TESTING

TESTING THE SYSTEM CRATE

IN y ARBITRATION BUS (4PIN LEM) | OUT

Figure 25 Testing the System crate

John Mills

DTM WORD VOR BR CPR 4-1 TEST CONTROLLER eee 4 WGR-241 SC -TST-1 MX-CTR-3 LAM @L ; BRANCH BRANCH R s COUPLER OPERATION 1 2 e-7 =) 2 1 2 cS> z [>> TRANSFER 4 @ @ SSeS MODE Gy ioe © @ RS] eH cD e ar Switches on modules are shown in correct positions to produce an ee 4 e a) [> BR2 [1 > || CAMAC cs) 4 alternate 1010... bit pattern @ @ @ PRESENT @06 > & > @? $ $ BRANCH O@O Lao eR) RTE ~5 i] OPERATION 1 - READ FROM WORD GENERATOR IN SLOT 4 ee | DEAN ote So S| © me | Address - BO CO N4 AO FO 1@ @ @ ENABLED @0@ SS SD WATCHDOG Si sr | dt te TSN pao} © rac § OPERATION 2 - WRITE TO DATAWAY MONITOR IN SLOT 1 OFF | 3214 [35> N8 [5D os Address - BO CO N1 AO F16 9 . a ena S55 5 crate © [35> N2 [355 e! OFF 40HZ [SD NI Ce] =) ; A © MAN [> As [o> G ON BRANCH CS> a4 [> MASTER =) 4 LINE |NITIALISE cycle [5 A2 [55 INIT [3 At [35> (ep 5 [S> Fé Co] > 8 CAMAC RESET [> F8 [3 G Ss 7 DATAWAY 6 [CS> F4 PS> 6 [S> F2 [3> Zz Gis LINK tre => Ft Ca> cD 4 ane [> ¢ > ar TO ©) 0 4 cS 2 BRANCH DEMAND = <¢# - ss > CRATE ci 0 4 =P, 123 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 DRAWING No : System crate test REV: 1.0 |AUTHOR: E. J. Mills FILENAME : test system crate.vsd SYSTEM : CAMAC DATE: 30/5/2010

CAMAC TEST PROCEDURES

30

Page 33 read from a scan by OCR; it may hold reading errors

CAMAC TESTING BRANCH CRATE DT™M "CEN, Reeeaan BT 6502 || Cc 2089 4 WGR-241 FREQUENCY A2 CRATE @. Lau @N GENERATOR BRANCH CONTROLLER TERMINATOR 6 POs a EXAMINE re z ~ eeee CAMAC e C5) P| CRATE SETUP FOR TESTING ee e@ @ DATAWAY ae ©, | THE © stotete @ | urcae] @ se CD 5 rl RGO32BIT COUNTER {MHz esse Nome | ° ee & rg MODULE cee © eo0o qi i (0) e008 ee Z j P| eee ee ove 2 i Me e383 33/1 a eee ryt @0@ C0 | cme ON eee ee ADDRESS UNS eco ©, Cee e@ eee =) 20Hz Y) e eee ? ~)) EXAMINE 2 eee ey, x . . . A ° ©) © CAMAC CAMAC eee DATAWAY DATAWAY | @ ©00@ va 8 \ 10Hz W 7 eee CS) 6 © coe @s e 00 [D> 4 1mS/Sec 333 e eee CD 2 eee (Gal R o 4 | | 1 2 3 4 8 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 DRAWING No: Branch crate REV: 1.0 |AUTHOR: E. J. Mills FILENAME : test branch crate.vsd SYSTEM: CAMAC DATE: 7/5/2010 Figure 26 Branch crate John Mills CAMAC TEST PROCEDURES

31