Manual · Camac
CAMAC Testing Procedures: A Guide on How to Manually Test CAMAC Modules
Manual33 pagesPDF 1.1 MB
Guide to testing CAMAC crates and modules by hand, including input registers and output drivers such as the OD48 and OD2407.
Filed with · 1 model
Also filed with these models on the old site
Summary
What is in it
An in-house guide from the Isaac Newton Group of Telescopes on testing CAMAC modules manually, without a computer, using a system crate, a branch crate and the SC-TST-1 test controller. It explains CAMAC and the BCNAF command convention, the test rack layout, testing branch couplers and crate controllers, and step-by-step tests with switch settings for input register and output driver modules (IR48, OD48, OD2411, Hytec 450-4, PR2402/PR2403, OD2407), the RGO 32 bit counter, a MultiDAC and the Hytec ADC1232.
Its parts
- Overview
- Layout of Rack for CAMAC Testing and Repairs
- Basic Settings for the SC-TST-1 Test Controller
- WGR2403 Word Generator and DTM4 Dataway Monitor Testing
- Testing Branch Couplers (BC) and Crate Controllers (CC)
- Read and Write Tests, System to Branch, Branch to System
- Testing Input Register and Output Driver Modules
- Testing the RGO 32bit Counter Module
- Testing Miscellaneous Modules
From the manual
Key facts from this manual
Each line names its page in the PDF.
A June 2010 guide from the Isaac Newton Group of Telescopes on testing CAMAC crates and modules by hand, without a computer. It covers output driver and output register modules (OD48, OR48, OD2411, OD2407) as well as input registers (IR48, HY450-4, PR2402, PR2403), counters, DACs and ADCs. It explains the test rack, test controller switch settings, test boxes and the bit patterns used to find faults.
- Document
- CAMAC module test procedure, Rev 1.0, June 2010 p. 1
- OD48 / OR48
- Dual 24-bit output driver; the OR48 output register has less current drive p. 12
- OD48 connectors
- OD48: two 24-bit output registers on high density 31-way D connectors p. 16
- OD48 addressing
- OD48: REG 0 (upper connector) A0 F16; REG 1 (lower connector) A1 F16 p. 16
- OD2411
- 48-channel 40mA/30V output register with handshake; standard 37-way D connectors p. 16
- OD2411 addressing
- OD2411: Output A = A0 F16; Output B = A1 F16 p. 16
- OD2407
- Dual 24-bit output driver with two 38-way Hughes connectors p. 21, 22
- OD2407 addressing
- OD2407: OUTPUT A (upper) A0 F16; OUTPUT B (lower) A1 F16 p. 22
- OD2407 use
- Drives relays and commands for moving the dome, focus and instrument turntables p. 21
- Test equipment
- SC-TST-1 test controller, WGR2403 word generator in slot N4, DTM4 p. 15, 16
- Test box power
- Special power cable from the crate's 9-pin front panel D connector, nominally +6V, reduced to +5 in the box p. 15
- Test patterns
- All ones, all zeros, alternate bits, reverse bits p. 16
- Test cables
- OD/OR48, IR48, OD2411, PR2402/3, OD2407 p. 13
- CAMAC bus
- 24 read/write lines; module slots 1 to 23, slots 24/25 reserved for the crate controller p. 4
What this manual walks you through
- Setting up the system and branch test crates p. 5, 6
- Setting the SC-TST-1 branch and crate switches p. 6
- Choosing the right test box for each module p. 12
- Testing an OD48 or OR48 output driver p. 15, 16
- Testing an OD2411 output register p. 16
- Testing an OD2407 output driver p. 21, 22
- Testing an IR48 input register p. 14
- Testing a HYTEC 450-4 input/output register p. 17, 18
- Testing PR2402 and PR2403 input registers p. 19, 20
Good to know
- The LEMO link cable between the executive controller and the test controller must be present, or the test system does not work. p. 5
- The module's own LEDs can look right while the data leaving its connector is faulty, so check the test box LEDs too. p. 16
- Unlit LEDs on the test box may be a bad cable connection; check the connectors are seated before calling the module faulty. p. 16
- Hughes connector pins are easily damaged; align the test cable connector squarely before clamping the finger screw. p. 21
- The opto-isolated input facility of the metal test box no longer works; use the plastic box for PR2403 modules. p. 12
Questions this manual answers
How do I test a CAMAC OD48 output driver without a computer?
Put the OD48 in slot N8, write bit patterns to it with the WGR2403 word generator in slot N4 using F16 at A0 or A1, and check the module and test box LEDs agree. p. 15, 16
What is the difference between the OD48 and the OR48?
For testing they are the same; the OD48 can sink more current. p. 15
What connectors does the OD2407 output driver use?
Two 38-way Hughes connectors, OUTPUT A and OUTPUT B, with the polarising sleeve reversed between them. p. 22
What sub-addresses do the OD2411 outputs use?
Output A is A0 F16 and Output B is A1 F16. p. 16
What does CAMAC stand for?
Computer Automated Measurement And Control, a control and data acquisition system built from crates and modules. p. 4
The whole text · 4 parts
Read the manual here as text
Its text as extracted from the PDF, unedited, in parts; each page opens the PDF at that page.
The file