What this file really is

The old site listed this file as a manual for the Camac Output register Driver CAMAC. The file is named as a CAMAC output register driver manual, but it is an IAEA book on troubleshooting nuclear instruments (IAEA-TECDOC-426) that mentions CAMAC only in passing.

Download the PDF12 MB

Filed with · 1 model

Also filed with these models on the old site

Summary

What is in it

A technical document issued by the International Atomic Energy Agency in 1987 on servicing and repairing nuclear instruments. It is written for electronics engineers, senior technicians and scientists who must keep their own instruments running without good service manuals. It covers setting up a service laboratory, tools and test instruments, troubleshooting whole systems, and then power supplies, preamplifiers and amplifiers, discriminators, scalers and ratemeters, multichannel analyzers, interfaces, dedicated instruments and radiation detectors.

Its parts

  1. Introduction
  2. Organization of the Laboratory
  3. Tools, Instruments, Accessories, Components, Skills
  4. Troubleshooting in Systems
  5. Power Supplies
  6. Preamplifiers, Amplifiers
  7. Discriminators, Single Channel Analyzers, Timing Circuits
  8. Scalers, Timers, Ratemeters
  9. Multichannel Analyzers
  10. Interfaces
  11. Dedicated Instruments
  12. Radiation Detectors

From the manual

Key facts from this manual

Some pages are scans or cannot be read; this says only what the readable text supports. Each line names its page in the PDF.

IAEA-TECDOC-426, Troubleshooting Nuclear Instruments, a 1987 technical book from the International Atomic Energy Agency on servicing and repairing nuclear electronics. It is a general repair guide, not a manual for one CAMAC module, and it uses ORTEC, CANBERRA and NARDEAUX instruments as worked examples. Its chapters cover the repair laboratory, tools and test instruments, system troubleshooting, power supplies, amplifiers, scalers, multichannel analyzers, interfaces and radiation detectors.

Read for this summary: pages 1 to 10, 62 to 70, 97 and 196 to 199 of 273.

Document
IAEA-TECDOC-426, issued by the International Atomic Energy Agency, Vienna, June 1987 p. 1, 2
Intended readers
Young electronics engineers, senior technicians, and scientists who must repair their own instruments p. 5
Chapters
Laboratory organization; tools, instruments, accessories, components, skills; troubleshooting in systems; power supplies; preamplifiers, amplifiers; discriminators, single channel analyzers, timing circuits; scalers, timers, ratemeters; multichannel analyzers; interfaces; dedicated instruments; radiation detectors p. 7
Worked examples
Circuit diagrams and service manuals of ORTEC, CANBERRA and NARDEAUX instruments p. 6
Module types
Desktop or rack instruments with built-in supplies, or modules for NIM, CAMAC or EUROCARD crates without them p. 97
System check tools
Multimeter; 0-15/20 MHz dual trace oscilloscope; pulse generator; radioactive test source of known activity p. 64
Interfaces covered
RS232 C/V24 serial, Centronics parallel, IEC 625 (IEEE-488, GPIB), I/O ports, motor control p. 197
RS-232 pins
Pin 1 protective ground, 2 transmitted data, 3 received data, 4 request to send, 5 clear to send, 6 data set ready, 7 signal ground, 8 data carrier detect, 20 data terminal ready p. 199
Null modem cable
Pins 1 and 7 straight through; pin 2 on one end crossed to pin 3 on the other p. 199
Dead instrument checks
Power switch, AC outlet, AC plug and cable, AC and DC fuses, DC supplies, signal lamps p. 64, 65
No-count checks
Test function, HV connected to the detector, amplifier input polarity, SCA threshold and window, DC fuses and supplies, signal cables p. 65
Resolution loss causes
HV ripple, detector colouration or broken crystal, bad optical joint, unstable supplies or amplifiers, light leaks, worn photo-cathode, bad earth p. 70
Lab stock
Documentation library, components and spare parts, spare instruments if possible, maintenance kits p. 10

What this manual walks you through

  • Checking an instrument that does not work at all p. 64, 65
  • Finding why a counting system does not count p. 65
  • Tracing count rates lower or higher than usual p. 65, 66
  • Finding why pulse height has changed p. 67, 68
  • Tracking down spurious counts p. 68, 69
  • Diagnosing loss of spectrum resolution p. 70
  • Choosing an interface for a system p. 197, 198
  • Wiring an RS-232 null modem cable p. 199
  • Setting up a maintenance and repair laboratory p. 10

Good to know

  • Check the instrument settings first, to tell operator errors from instrument failure. p. 63
  • Many troubleshooting questions can only be answered if settings and earlier test results are kept in a logbook. p. 63
  • The book states it cannot replace a good service manual. p. 5
  • Amplifier overshoots may be counted too; watch for signal reflections in long cables. p. 67
  • When connecting RS-232 devices, an output signal must go to an input and the reverse. p. 198

Questions this manual answers

Is this a manual for the CAMAC output register driver?

No. It is the IAEA general book on troubleshooting nuclear instruments; it names CAMAC as one of the crate types that plug-in modules fit into. p. 1, 97

What should I check when a nuclear counting system does not count?

Check the test function, that HV is on and connected, the amplifier input polarity, the SCA threshold and window, the DC fuses and supplies, and the signal cables. p. 65

How is an RS-232 null modem cable wired?

Protective ground (pin 1) and signal ground (pin 7) run straight through, and pin 2 at one end is crossed to pin 3 at the other. p. 199

What test equipment is needed to check a detector system?

A multimeter, a 0-15/20 MHz dual trace oscilloscope, hand tools, a pulse generator and a radioactive test source of known activity. p. 64

The whole text · 28 parts

Read the manual here as text

Its text as extracted from the PDF, unedited, in parts; each page opens the PDF at that page.

  1. Pages 1 to 101-10
  2. Pages 11 to 2011-20
  3. Pages 21 to 3021-30
  4. Pages 31 to 4031-40
  5. Pages 41 to 5041-50
  6. Pages 51 to 6051-60
  7. Pages 61 to 7061-70
  8. Pages 71 to 8071-80
  9. Pages 81 to 9081-90
  10. Pages 91 to 10091-100
  11. Pages 101 to 110101-110
  12. Pages 111 to 120111-120
  13. Pages 121 to 130121-130
  14. Pages 131 to 140131-140
  15. Pages 141 to 150141-150
  16. Pages 151 to 160151-160
  17. Pages 161 to 170161-170
  18. Pages 171 to 180171-180
  19. Pages 181 to 190181-190
  20. Pages 191 to 200191-200
  21. Pages 201 to 210201-210
  22. Pages 211 to 220211-220
  23. Pages 221 to 230221-230
  24. Pages 231 to 240231-240
  25. Pages 241 to 250241-250
  26. Pages 251 to 260251-260
  27. Pages 261 to 270261-270
  28. Pages 271 to 273271-273

The file

Revision and size

Part no.IAEA-TECDOC-426
RevisionJune 1987
PublisherInternational Atomic Energy Agency
Pages273
PDF12 MB
TextText from the PDF itself; 20 pages are scans, read by OCR, which may hold reading errors.
IAEA-TECDOC-426

The first pages · 3 of 273

Look before you download

Each page opens the PDF at that page.

Page 1
Page 1of 273
Page 2
Page 2of 273
Page 3
Page 3of 273

More documents · 1

For the same models