Page 61
-45-
Chapter 3
In the case of missing drawings and documentation, functional diagram preparation is recommended as an efficient in repair work vs. the t r a d i t i o n a l d r a f t s m a n a p p r o a c h .
the aid
(65)
Page 62
Chapter 4 TROUBLESHOOTING IN SYSTEMS
Page 63
-1-
Chapter 4
TROUBLESHOOTING IN SYSTEMS
4. 1
INTRODUCTION
A Nal detector is used in many fields of application of nuclear techniques, e.g. in nuclear medicine, radioimmunoassay, agronomy, radiation protection and uranium prospection. Therefore, a Nal detector system was chosen as an example of troubleshooting in a d e t e c t i o n system. Such a system consists of a detector, usually surrounded by a radiation absorbing shield and a collimator, a p r e a m p l i f i e r , an amplifier, a single channel analyzer (SCA) and a sealer-timer or a ratemeter, or, instead of the last three, a multichannel analyzer (MCA). Furthermore, the system contains a high voltage supply (HV) and several low voltage DC supplies (LV). The units composing the system and their functions are given in Fig. 4.1. For other detectors, a similar set of electronic units is used with characteristics adapted to the particular d e t e c t o r . In the following t e x t , examples are taken from medical applications. The reader should be able to find equivalent ones in his own field of application.
Fig.
4.1: A typical Nal system
Most of the instrument users, applying nuclear techniques, work in fixed geometry, and with fixed HV, amplifier and single-channel-analyzer settings. Most of their measurements are relative measurements: a comparison is made with radioactive standards. Thus when the user or operator says, "My instrument does not behave as it should", one expects that they have not changed their instrumental settings. However, checking those settings is the first part of troubleshooting, since a clear distinction must be made between user/operator errors and instrument failure. Many of the questions which must be asked during troubleshooting can only be answered correctly when instrument settings and previous test results are available in a logbook.
(69)
Page 64
Chapter 4 4.2
-2-
INSTRUMENTS AND SKILLS
The following instruments and tools are required for a system check (for d e t a i l e d information, see Chapter 3): - a m u l t i m e t e r (MM);
- an oscilloscope (0) 0-15/20 MHz, with dual beam or dual trace, sensitive trigger properties (it must be possible to trigger both channels at the same time or alternatively with the signal on one channel) and with signal delay lines to enable visualization of the front edge of pulses; - a set of screwdrivers, Alien keys, plyers, flat keys;
- (a) pulse generator(s) (PG) with which nuclear pulses can be simulated, both for analog pulses and for digital pulses; and
- a radioactive test source, with known a c t i v i t y , the same as used during acceptance testing.
preferably
In order to do troubleshooting one should know: (1) the usual composition of a Nal detection system and the functions as given in Fig. 4.1; (2) the forms of the signals at the output of the PMT, the p r e a m p l i f i e r , amplifier and single channel analyzer; (3) why one does not observe a signal at the input of a charge sensitive amplifier; (4) that the rise time of the pulse at the output of the PMT depends on the half life of the scintillation process in the d e t e c t o r ; and (5) that the pulse heights and shapes at the o u t p u t of the pulse amplifier are so chosen as to fit best to the SCA (or MCA) and to guarantee optimum s p e c t r u m resolution. Each of the following paragraphs gives a possible indication of failures as may be given by instrument users/operators and a list of possible origins or reasons for these failures a n d , when not obvious, actions to be taken. There may be among the operators very unexperienced ones. This may lead to some reasons which seem ridiculous; nevertheless, they are based on experience in the field. 4.3
THE INSTRUMENT DOES NOT WORK
To be checked: 1.
Is the instrument switched on?
2.
Is there power on the AC outlet?
3.
Are the connections in the cable all right?
4.
Are the fuses on the instrument all right (AC and DC)?
5.
Are the DC power supplies all right?
(70)
AC plug all right?
Is the power
Page 65
-36. A.A
Chapter A
Are the signal lamps faulty? THE INSTRUMENT DOES NOT COUNT
To be checked:
1.
Is the test sealer in test function?
2.
Is the KV switched on and connected to the detector?
3.
Is the input polarity of the amplifier correctly set?
A.
Has the threshold of the SCA been set to maximum or the window to 0?
5.
Are the DC fuses all right?
6.
Are the DC supplies correct? Check voltages of HV and LV.
7.
Are all signal cables alright? Check the presence of signals at the output or preamplifier, the input and output of the amplifier, the input and output of the SCA and at the input of the sealer. Check whether the plugs are correctly mounted, e.g. the pin in a male BNC may be moving backwards when connected to the socket.
8.
Are the cables connected to correct inputs and outputs?
A. 5
THE COUNT RATE IS LOWER THAN USUAL
Possible reasons:
1.
There is no radioactive source in the measuring position, or a wrong source, a wrong patient or sample, a patient wrongly injected.
2.
The instrument is set incorrectly for the radioisotope used.
3.
There is more absorbing medium between source and detector than usual: different source (test-tube) shapes, or the test tube has thicker walls or is made of material other than what is usual.
A.
The source-detector geometry has changed, not fully reach the measuring position.
5.
A wrong colliraator is used (particularly for a scanner).
6.
The AC frequency is too high (in case the timer uses frequency as time base). Check with wristwatch.
7.
The HV setting is incorrect.
or the sample did the
AC
(71)
Page 66
Chapter 4
-4-
8.
The settings of the amplifier have been changed.
9.
There is a change in the SCA setting.
10.
There is a change in the sealer scale factor or input polarity.
11.
The
12.
The preamplifier or amplifier is faulty, check pulse shapes.
13.
There is a change in pulse height or shape.
See section 4.7
14.
The SCA is faulty: incorrect.
or
15.
The sealer has an intermittent failure or is faulty.
16.
The time clock is defective.
17.
A broken resistor in the dynode chain, a bad phot omul t i p 1 i e r , a bad optical coupling between scintillator and PM, a shorted capacitor in the dynode chain.
4.6
THE COUNT-RATE IS HIGHER THAN USUAL
LV threshold s e t t i n g , in the power supplies are faulty; the HV supply output pulse height is faulty.
pulse duration
Check with wristwatch.
Possible reasons:
1.
There is an extra background source: detector directed to source preparation or store rooms, injected patients near to detector, radioactive contamination, a test source is not placed in its lead container, people in neighbouring laboratories are using radioactive sources, e t c .
2.
The patient is injected with the wrong patient wrong source.
3.
The patient was previously injected with another radioisotope.
4.
The instrument is set incorrectly for the radioisotope used.
5.
There is less absorbing medium between the source and the detector than usual; a different source (test tube) shape, the test tube has a thinner wall, or is made of a different material than usual.
6.
The source-detector geometry has changed.
7.
A wrong collimator is used.
8.
The lead shield or collimator has been removed or changed position.
9.
There is a change in amplifier, SCA or HV setting.
(72)
radioisotope , wrong in
Page 67
-5-
Chapter 4
10.
The AC frequency is too low (in case the timer uses the AC frequency as time base). Check with wristwatch.
11.
The sealer has a wrong setting of the scale factor and/or input p o l a r i t y , or the setting of input discriminator has been changed.
12.
For m e d i c a l laboratories: the activity meter (dose calibrator) is f a u l t y .
13.
There are parasitic pulses, spurious counts or section 4.7
14.
The o u t p u t pulses of the amplifier have overshoots. Overshoots may be counted too! Pay attention to signal reflections in long cables. Overshoots may cause ghost peaks on MCA systems w i t h a fast ADC. Check signal shapes. See section 4.7
15.
The earthing has changed or ground connections have loosened.
16.
The DC power supplies are not correct (LV, HV).
17.
The a m p l i f i e r or preamplifier is faulty.
18.
There is a change in pulse height or shape.
19.
The SCA is faulty.
20.
The sealer is defective.
21.
The timer is defective.
22.
There is & light leak in the PMT and/or the canning of the d e t e c t o r resulting in extra pulses in the low energy part of the spectra and peak widening.
4.7 noise.
See
Check pulse shapes. See section 4.7
THE PULSE HEIGHT IS LOWER OR HIGHER THAN USUAL
Possible reasons are: A.
W i t h o u t change in pulse shape;
1.
Incorrect radioactive source.
2.
Incorrect HV setting.
3.
Incorrect amplifier settings:
amplification factor, pulse polarity. 4.
The detector assembly has been interchanged.
5.
The amplification factor of the preamplifier has been changed.
6.
A channel shift or wrong setting of the SCA comes up only when no oscilloscope is used).
(this question
(73)
Page 68
Chapter 4
-6-
7.
The d e t e c t o r is broken or yellow-brown (this cannot be checked for an integral line assembly).
8.
A bad o p t i c a l joint between the d e t e c t o r and the PMT (this cannot be checked for an integral line assembly).
9.
A HV s u p p l y failure.
B.
W i t h change in pulse shape (rise time, fall time, overundershoots) :
10.
Incorrect s e t t i n g of the amp 1 if ier .
11.
The pulse height and/or shape p r e a m p l i f i e r has changed. A faulty pulse-shaping amplifier.
12.
differentiation at and and integration time of the output of the
13.
The ground connection to the d e t e c t o r canning is loosened or i n t e r r u p t e d , or in g e n e r a l , a change in earthing.
14.
The PMT is faulty or has changed its c h a r a c t e r i s t i c s . The d y n o d e s may have changed position or shape. E s p e c i a l l y in portable instruments, rectilinear scanners and other i n s t r u m e n t s of which the d e t e c t o r heads are moving or are exposed to m e c h a n i c a l shock.
15.
Bad c o n t a c t s in or f a u l t y HV v o l t a g e d i v i d e r , or bad c o n t a c t s between PMT and base. A t t e n t i o n should be paid to c a p a c i t o r s at the dynodes near the anode.
16.
The u - m e t a l s h i e l d has changed its p r o p e r t i e s or is dislocated (in f a c t , such a deficiency mainly influences the h e i g h t of the pulse).
17.
A very low AC voltage.
18.
A failure in one of the LV supplies.
19.
F a u l t y p l u g - c a b l e connections.
4.8
SPURIOUS COUNTS
1.
Take all radioactive sources background is higher than usual.
2.
Check whether background of (parasitic) pulses.
3.
If this is the case, observe pulses on oscilloscope or sealer and try to detect relations between pulse trains and the switching of equipment (incubators, deep freezers, r e f r i g e r a t o r s , floor polishers, workmen with drills and other machines) in own or neighbouring laboratories.
(74)
away, and If yes:
check whether pulses appear as i n t e r m i t t a n t trains
Page 69
-7-
Chapter 4
4.
Check whether pulses are coincident with the fluorescent lamps.
5.
Or with starting or running parking.
6.
Unproperly filtered AC.
flickering of motorbikes and cars on an outside
Other origins of spurious counts may be: 7.
E l e c t r o s t a t i c discharges, especially in very dry laboratories.
8.
HV sparking due to dust or humidity or parts starting to fail.
9.
N e a r b y switching thyristors or triacs.
10.
When there are nearby radio or TV receivers, check whether the instrument is properly grounded. Try different points, and check connections and coaxial cables. •
11.
E l e c t r i c a l discharge humid ity.
12.
Loose cable c o n t a c t s . ics in PM tube.
4.9 on surface of components, due to high
Cable-movement-caused noise, raicrophon-
THE SPREAD IN THE NUMBER OF COUNTS IS LARGER THAN MAY BE EXPECTED ACCORDING TO STATISTICAL FLUCTUATIONS WHEN M E A S U R E M E N T S OF THE SAME SOURCE IN FIXED GEOMETRICAL POSITIONS ARE DONE EACH DAY
Possible causes may be:
1.
An irreproducible source positioning.
2.
The use of different source holders or test tubes.
3.
Changes in the absorbing medium:
4.
The timer uses (the f l u c t u a t i n g ) AC frequency as time base.
5.
The setting of the HV is irreproducible or the HV is unstable.
6.
I r r e p r o d u c i b l e settings of the amplifier and/or SCA.
7.
Large temperature ins t rumen t .
8.
Spurious counts.
9.
Large AC fluctuations.
10.
Unstable LV power supplies.
11.
Bad earthing.
dust or dirt in factories.
changes, or too short warming-up times of
(75)
Page 70
Chapter 4 4.10
-8-
LOSS OF SPECTRUM RESOLUTION
Possible origins are: 1.
Disturbances in the HV supply:
2.
Detector colouration or a broken crystal.
3.
A bad optical joint.
4.
Unstable LV power supplies.
5.
An unstable preamplifier or main amplifier.
6.
Light leaks into the PMT or detector.
7.
The deterioration of the photo-cathode of the PMT.
8.
An oscillating preamplifier.
9.
Bad earth connections.
10.
Mechanical scanners ) .
(76)
vibrations of ripple and fluctuations.
dynodes in
PMT
(rectilinear