Chapter 2 ORGANIZATION OF THE LABORATORY
Page 12
-12
ORGANIZATION OF THE LABORATORY
2. 1
GENERAL REMARKS
Chapter 2
The contents of this chapter are limited to some general observations and conclusions on the organization and operation of an electronics laboratory that serves nuclear intrumentat ion and related equipment. The same recommendation applies as well for all donated instruments. It should be noted that there are several IAEA publications dealing with the topic (IAEA TECDOC-309, Nuclear Electronics Laboratory Manual, and TECDOC-363, Selected Topics in Nuclear Electronics). The optimal set-up of a service laboratory depends, to a large extent, on the social, economic and policy situation in the country. As an example, consider the specific regulations in a country, referring to the financing and control of material support to the laboratories. Therefore, it is d i f f i c u l t to present a detailed assessment of the best approach for establishing and operating such a laboratory. Below, only some points that seem to apply to all countries and situations are summarized.
2.2
PHYSICAL E N V I R O N M E N T
A summary of the main items of recommended facilities required for a service laboratory is given here. Only a reminder on the essential features to be considered when a laboratory is created are presented. Further details are given in the following chapters.
1.
The laboratory should have a minimal useful area of 10-12 square meters for each employee. If Computer Aided Engineering (CAE) facilities are envisaged, anair-conditioned computer room and a room for work stations should be made available.
2.
For countries in tropical regions, the laboratory should be air-conditioned, At least a part of the laboratory should be equipped to have a dry area, i.e a room with reduced humidity.
3.
An adequate storage room, for special has to be kept for emergency return available.
4.
The electronics laboratory should have a good, reliable mains supply, properly arranged electrical power distribution, a n d , preferably, a good dedicated grounding system.
5.
The laboratory should have good overall precise work on electronics instruments places should have additional lights.
6.
Depending on the size of the laboratory, and its activities, add it ional room space with reduced h u m i d i t y for storage of electronic instruments and components, as well as the proper packing material that s h i p m e n t , should be illumination. For individual working
(13)
Page 13
Chapter 2
-2- laboratories for arranged.
production of printed c i r c u i t s , should be
2.3
ADMINISTRATION OF A NUCLEAR INSTRUMENTATION LABORATORY
2.3.1
Staff
The p r o p e r staffing of the maintenance and service of a nuclear e l e c t r o n i c s laboratory is obviously the most i m p o r t a n t part of good m a n a g e m e n t . At the start of such a laboratory, it is advisable to make a complete inventory of all the instruments in the i n s t i t u t e or i n s t i t u t i o n s that are e x p e c t e d to be served by the laboratory. According to the volume and complexity of the i n s t r u m e n t a t i o n , the r e q u i r e d skills should be d e t e r m i n e d , and on this basis, the persons w i t h s u i t a b l e professional profiles recruited. It is highly a d v i s a b l e to recruit for the laboratory such staff m e m b e r s who can cover a wide spectrum of instruments. N e v e r t h e l e s s , it will be necessary, with the increasing demands placed on the work of the l a b o r a t o r y , to specialize some of the s t a f f m e m b e r s in p a r t i c u l a r a s p e c t s of nuclear i n s t r u m e n t a t i o n . Some of the topics of such expertise are: radiation detectors, analog e l e c t r o n i c s , d i g i t a l e l e c t r o n i c s , i n t e r f a c i n g , c o m p u t e r s a n d electromechanical apparatus.
Note:
the recent trends in e l e c t r o n i c s tend to over-emphasize d i g i t a l e l e c t r o n i c s and c o m p u t e r software d e v e l o p m e n t . P a r t i c u l a r l y in the nuclear f i e l d , there is a c r i t i c a l need for persons with knowledge and experience in high quality analog e l e c t r o n i c s , and this should not be neglected. F u r t h e r m o r e , it should be kept in mind t h a t many i n s t r u m e n t s in the nuclear l a b o r a t o r i e s are not strictly nuclear. Electronic b a l a n c e s , sample c h a n g e r s , o p t i c a l s p e c t r o m e t e r s , d i f f r a c t orne t ers, and ph-meters are research tools that also need maintenance and service, and the e l e c t r o n i c s laboratory should be in a position to offer i t.
The management of the laboratory should design a s u i t a b l e scheme to evaluate and promote the activities of the staff. S p e c i a l attention should be given to the fact t h a t continuous training of the s t a f f is required to keep up w i t h the e x t r e m e l y fast progress in the world of electronics and c o m p u t e r science. The IAEA is developing a maintenance; necessarily computer-based management scheme such a system can be developed computer-supported, and can for p r e v e n t i v e locally, not also include repair and servicing aspects. 2.4
INSTRUMENTATION AND ELECTRONICS COMPONENTS
Chapter 3 provides detailed information on the type and amount of testing instruments and tools required for normal operation of a service laboratory, at d i f f e r e n t levels. It should be emphasized that the testing instruments in a laboratory should be regularly checked and calibrated for proper operation; such controls should include connectors and cables.
(14)
Page 14
-3-
Chapter 2
F u r t h e r m o r e , a c e r t i f i e d recalibrat ion of instruments used as m e a s u r e m e n t s t a n d a r d , at a national i n s t i t u t i o n , should be planned. The acquisition of any instrument, p r e f e r a b l y w i t h two sets of operator and service manuals even at a d d i t i o n a l expense, should be accompanied w i t h the provision to order some spare parts and components. One set of operator and service manuals should stay at the location of each instrument while the other set should be kept in the t e c h n i c a l library of the service laboratory. A good rule can be t h a t 1 1/2 % of the value of the i n s t r u m e n t s should be invested in spare p a r t s , at the time when the purchase is m a d e . In the following years, each instrument will need b e t w e e n 1 and 3 % for replacement p a r t s , d e p e n d i n g on the c o m p l e x i t y and design of the i n s t r u m e n t . Each laboratory should have a basic supply of e l e c t r o n i c c o m p o n e n t s . A list is presented in C h a p t e r 3, Section 3.4; this is considered to be a minimal set of components t h a t have to be on stock in the l a b o r a t o r y and need to be u p d a t e d regularly. It is considered absolutely mandatory that a nuclear e l e c t r o n i c s laboratory have access to some local and foreign p e t t y cash, for rapid and u n b u r e a u c r a t i c a c q u i s i t i o n of p a r t s and c o m p o n e n t s t h a t are or are not available on the home market. This is e s s e n t i a l for r a p i d turn-around of r e p a i r s ; it is not t o l e r a b l e t h a t the s t a f f m u s t wait m o n t h s for the a p p r o p r i a t e a p p r o v a l to buy a minor electronic p a r t , and the e x p e r i m e n t e r cannot use his ins t rumen t.
It should be pointed out here that there is a tendency more and more often in modern e l e c t r o n i c i n s t r u m e n t a t i o n to make use of h y b r i d analog, c u s t o m i z e d d i g i t a l c i r c u i t s , EPROMs, PALs and other programmable c h i p s , to achieve higher packing d e n s i t i e s , b e t t e r overall performance, and to reduce production cost. It will not be p o s s i b l e , even in a very w e l l - e q u i p p e d l a b o r a t o r y , to have all these s p e c i a l spare parts available in a stockroom. Fortunately, the f a i l u r e r a t e of such components is low. These spare p a r t s have to be ordered from the manufacturer of the instrument or its r e p r e s e n t a t i v e and are not available from the s e m i c o n d u c t o r manufacturer. Even so, one has to realize the fact that the on-site service of highly sophisticated nuclear instrumentation may often have to limit itself to the board level just by swapping boards. Board test e q u i p m e n t , which is required to enable repair on the chip level, will only be affordable to the instrument manufacturer in some strategically located service centers worldwide. One may complain about this situation, but it also offers a considerable advantage by reducing the number of boards in a t y p i c a l nuclear instrument, and consequently, the failure rate. Many manufacturers are now able to use a functionally p a r t i t i o n e d approach to break down a design to the board level, which allows them to p r o v i d e diagnostic routines and facilities to the customer for easy fault location to the board level.
(15)
Page 15
Chapter 2
-it-
Special emphasis should therefore be given to this fact when ordering e q u i p m e n t if such diagnostic facilities are available. In such a case, r e t u r n of an instrument to the m a n u f a c t u r e r may never be necessary and the p o s s i b i l i t y of shipping damages when returning a board to the m a n u f a c t u r e r largely vanishes. Of specific importance for a service laboratory is the a v a i l a b i l i t y of s u i t a b l e extension boards and cables. If they cannot be p u r c h a s e d , they must be p r o d u c e d in the laboratory. A n o t h e r essential a c t i v i t y of a properly organized l a b o r a t o r y , is the good o r g a n i z a t i o n of the t e c h n i c a l library. This should include: - Originals of all the operator manuals
- Originals of all available service manuals, circuit digrams, p a r t s l i s t s , and t r o u b l e s h o o t i n g i n f o r m a t i o n ; - A set of d a t a books on electronic c o m p o n e n t s ; as a m i n i m u m , it is recommended to have a set of U.A.T.A. Books (D.A.T.A., Inc. P.O. Box 26875, San Diego, California 92126, USA), possibly c o m p l e m e n t e d w i t h some p u b l i c a t i o n s of i n d i v i d u a l p r o d u c e r s of electronic c o m p o n e n t s ; - a p p l i c a t i o n notes referring instrument manufacturers;
to nuclear electronics from
- catalogues. 2 .5
ORGANIZATION OF THE WORK IN AN ELECTRONICS LABORATORY
F r e q u e n t l y , a nuclear electronics l a b o r a t o r y combines its a c t i v i t i e s in service w i t h some d e v e l o p m e n t work in order to give its s t a f f a chance to keep up with the r a p i d d e v e l o p m e n t in nuclear electronics. In such cases, the basic rule should be:
| The repair and s e r v i c i n g of i n s t r u m e n t s has p r i o r i t y to any | | o t h e r a c t i v i t y of the l a b o r a t o r y . | I_______________________________________________________________I The electronics laboratory should be involved in a research instrument from the moment of its delivery. A staff member should assist in the u n p a c k i n g , i n s t a l l a t i o n , and initial testing of every newly acquired instrument. For those instruments that cannot be r e p a i r e d l o c a l l y , and where there is a p o s s i b i l i t y for them to be sent for r e p a i r , special s h i p m e n t m a t e r i a l should be stored.
(16)
Page 16
-5-
Chapter 2
In l a b o r a t o r i e s and i n s t i t u t e s in d e v e l o p i n g c o u n t r i e s , we f r e q u e n t l y observe the "wooden box e f f e c t " : the delivered e q u i p m e n t is k e p t in b o x e s , s o m e t i m e s for years. This is not t o l e r a b l e , and the electronics laboratory can contribute to the action: upon d e l i v e r y , each i n s t r u m e n t should be i m m e d i a t e l y u n p a c k e d , i n s p e c t e d for p o s s i b l e damage (and all claims should be s u b m i t t e d as soon as possible, otherwise the w a r r a n t y m i g h t be lost), installed and t e s t e d . A d o c u m e n t should be p r e p a r e d s p e c i f y i n g the measured p r o p e r t i e s of the i n s t r u m e n t ; l a t e r , this w i l l p e r m i t a comparison on the instrument's p e r f o r m a n c e .
For every i n s t r u m e n t , a logbook and repair list should be o p e n e d , at the time of its arrival. All s u b s e q u e n t a c t i o n s , be it for p r e v e n t i v e maintenance or repair of the i n s t r u m e n t , should be registered in this book. The head of the workshop should set a good example on the u t i l i z a t i o n of the logbooks, otherwise it will not be used by the personnel. It is highly recommended to start weekly repair case discussions and to use the logbook during this session.
It is a d v i s a b l e for the repair of an instrument to be organized in the following way: a copy of the c i r c u i t diagram should be m a d e , and the values obtained in measuring the q u a n t i t i e s , such as DC v o l t a g e and signal s h a p e , should be registered. Such an "updated" diagram should be stored in the logbook. If no diagrams are a v a i l a b l e , a careful record should be maintained of the m e a s u r e d values at selected points of the circuit.
The staff of nuclear electronics laboratories should learn from their mistakes. The staff member of the e l e c t r o n i c s laboratory who is e x p e c t e d to r e p a i r a faulty instrument should make inquiries on how the fault has developed. This might require an a d e q u a t e a p p r o a c h , some c a r e f u l , not i n q u i s i t o r y i n v e s t i g a t i o n , and should p r e f e r a b l y be made by a senior person. The findings should be recorded in the logbook. P r e v e n t i v e maintenance is d e s c r i b e d in Chapter 13.
(17)
Page 17
Chapter 3 TOOLS, INSTRUMENTS, ACCESSORIES, COMPONENTS, SKILLS
Page 18
-13
Chapter 3
TOOLS, INSTRUMENTS. ACCESSORIES, COMPONENTS. SKILLS Tools are needed for proper m a i n t e n a n c e of instruments.
The number of tools, however, is very large, even if only a c e r t a i n category of service work, let us say t r o u b l e s h o o t i n g , of nuclear instruments has to be performed. The following compilation of tools shall give an idea of what is needed at a work bench or in the laboratory. The discussions on the a p p r o p r i a t e tools and instruments will distinguish between d i f f e r e n t levels of troubleshooting. A c c o r d i n g l y , the following list of tools will be d i v i d e d into t h r e e groups. Group A:
necessary as a minimum to solve simpler troubleshooting and maintenance tasks, e.g. cleaning, replacing of simple components, yes or no tests.
Group B for advanced repair work, replacement or repair of m u l t i p l e step switches, e.g.
in addition to
A, more complex components like hybrid circuits connectors moving coil i n s t r u m e n t s , e t c .
Group C:
for s o p h i s t i c a t e d repair work, e . g - calibration, v e r i f i c a t i o n of m a n u f a c t u r e r s p é c i f i c a t i o n , m o d i f i c a t i o n and new developments (or prototypes).
In the following list, you will also find a Group F: s p e c i a l l y for field service. This group will strongly overlap the other groups of tools but is meant for troubleshooting and maintenance away from the work bench. For this purpose, pre-packed tool bags are a v a i l a b l e but m o s t l y they do not fully satisfy the needs.
Either some items are missing or some of them will never be used. Therefore, it seems b e t t e r to put tools used at the work bench into a bag for field service.
(21)
Page 19
Chapter 3
-2-
3. 1
LIST OF TOOLS (MATERIALS)
3.1.1 3.1.2 3.1.3 3.1.4 3.1.5
Room Work Benches Chairs Bench lights Trolley table
ABC ABC ABC ABC ABC
3.1.6 3.1.7 3.1.8 3.1.y 3.1.10
Shelf Cupboard Storage cabinet Storage boxes Cabinet
ABC ABC ABC ABC ABC
3.1.11 3.1.12 3.1.13 3.1.14 3.1.15
Screwdrivers Allen keys Pliers Diagonal cutting nippers Jewellers snips
ABC BC ABC ABC
ABC
F F F F F
3.1.16 3.1.17
Knives Spanners
ABC ABC
F F
3.1.18 3.1.19 3.1.20
Soldering units Tin (different types) Desoldering units
ABC ABC ABC
F F F
3.1.21 3.1.22 3.1.23 3.1.24 3.1.25
Desoldering tapes (different types) Calipers Steel ruler Measuring tapes Punches
ABC BC BC ABC ABC
F F F
3.1.26 3.1.27
Hammer Drills
ABC ABC
F F
3.1.28 3.1.29 3.1.30
Trepanning cutters Correcut drills Drills for printed circuit boards
BC C C
F
3.1.31 3.1.32 3.1.33 3.1.34 3.1.35
Hand drilling machines El. drilling machine (with stand) Stand-alone drilling machine Dental drills Dental drilling machine
ABC BC C C C
F
3.1.36 3.1.37 3.1.38 3.1.39 3.1.40
Thread taps Die nuts (threading dies) Files Vices Clamps
C C ABC BC C
F
3.1.41 3.1.42 3.1.43 3.1.44 3.1.45
Tweezers Inspection mirrors Magnifier mirrors, lenses Brushes (cleaning) Sprays (liquids for different purposes)
ABC ABC BC ABC ABC
F F F F F
(22)
F
Page 20
-33.1
Chapter 3
List of tools (materials) (cont'd.)
3.1.46 3.1.47 3.1.48 3.1.49 3.1.50
Tapes Saws Saws (special) Tool bags Tool kits
ABC ABC C
3.1.51 3.1.52 3.1.53 3.1.54 3.1.55
Storage boxes Cable crimping tools Wrapping tools Blower and vacuum cleaner Screw punches
BC BC C ABC C
3.1.56 3.1.57
Tools for surface mounted devices Glue (instant action type like Loctite or similar) Winding machine (small size with turn counter)
C (F) ABC F
3.1.58
Detailed descriptions of presented in the following pages.
F F
F F the most important
F F F tools are
(23)