The text of a manual · part 28 of 28
Troubleshooting Nuclear Instruments, pages 271 to 273
Pages 271 to 273 of 273
The original document is the PDF (12 MB). This is its text as extracted, unedited; tables and drawings read best in the PDF, and each page heading opens it at that page.
Filed in the catalog under the Camac Output register Driver CAMAC
Page 271
-11-
Fig.
12.7: Streamlined d e t e c t o r p r e a m p l i f i e r setup
Chapter 12
Fig.
12.8
Preampl i f ier outside
TABLE 1 2 . 1 ; Defects and how they are recognized
(a)
(b)
Mechanical defects: - the vacuum inside the d e t e c t o r cryostat is bad moisture or ice accumulated at the neck of the cryotat
- the detector is broken no signal at the o u t p u t when gamma source is close
- the wires inside the detector heads are broken no signal with source or with pulser
Electronic faults - input FET is defective no o u t p u t signal, with test signal on the input
- d e t e r i o r a t e d resolution either damaged d e t e c t o r or moisture accumulation in the p r e a m p l i f i e r
- bad efficiency (x-ray)
dirty Be window or detector crystal surface
(289)
Page 272
Chapter 12
-12-
There are some actions that you can implement to repair a defective or improperly operating detector. But there are a number of faults when the detector must be returned to the factory for repair. Let us analyze some typical cases.
(i)
When touching the detector c a p , it feels very cold; moisture or ice a c c u m u l a t e at the point closest to the dewar. This is a sure sign that the vacuum inside the detector cryostat is bad. If you have a good vacuum laboratory, with a pumping system and a He leak d e t e c t o r , you can find the leaking spot, and repair it. Then you pump the cryostat for several hours, at the same time heating the bottom part of the cryostat, and the protruding dipstick to 80 or 100 degrees C. Without the proper vacuum e q u i p m e n t , the d e t e c t o r should be returned to the manufacturer's service laboratory.
ATTENTION : Do not apply high voltage to a detector which shows signs of a bad vacuum inside the cryostat. Some detectors (but not all) have a t e m p e r a t u r e sensitive switch inside the cryostst that prevents the a p p l i c a t i o n of high voltage to a d e t e c t o r that is not sufficiently cooled.
(ii)
The most obvious loss of vacuum happens if the Be window on an x-ray detector breaks. Immediately switch off high v o l t a g e , cover the d e t e c t o r with a t i g h t plastic cap, and remove the detector with cryostat from the dewar. If you are fast, the d e t e c t o r crystal itself can be saved, and the costs of repair will be only for mounting a new Be window. W i t h o u t vacuum but with high voltage on, the d e t e c t o r will heat up, the Li will d r i f t out of the d e t e c t o r , and you will have to buy a new one.
(iii)
If the detector is properly cooled, and yet there is no signal at the p r e a m p l i f i e r o u t p u t , there are several possible reasons: - you forgot detector ;
to put a calibration source in front of the the high voltage is not p r o v i d e d ; check the opérât ion the bias supply and the connections of the low voltage power supply (usually +/- 24 V) is not in order; check the o u t p u t s of the preamplifier connector (mounted on the main amplifier backplane) and the connecting leads;
the preamplifier is defective; closely inspect the printed board in the preamplifier and all connections;
(290)
Page 273
-13-
Chapter 12
- the inaccessible parts of the detector (the crystal itself or cooled FET) are f a u l t y ; in this case, without a well e q u i p p e d vacuum laboratory, you can do very little. (iv)
The d e t e c t o r resolution has d e t e r i o r a t e d . The most frequent reason is the h u m i d i t y and the condensed water in the p r e a m p l i f i e r . Use a hairdryer to remove moisture. A second possible cause is dirt on some components of the p r e a m p l i f i e r , or on the connector pins protruding from the cooled part of the cryostat. Clean t h e m , preferably with m e t h a n e , but at least w i t h a dry and clean cotton buds. Pay special a t t e n t i o n to the high voltage filter in the p r e a m p l i f i e r , usually made of two highohmic resistors and two high voltage capacitors.
The test input on the preamplifier box is connected to the d e t e c t o r crystal via the high voltage supply lead. Using a pulser, or a pulse g e n e r a t o r , you can inject the signals through the d e t e c t o r diode to the FET. If there are signals on the o u t p u t of the p r e a m p l i f i e r , you can conclude that the FET is working, and that the high voltage lead to the detector crystal is not interrupted. If the x- or gamma-rays are still not d e t e c t a b l e , you can conclude, w i t h a high p r o b a b i l i t y , that the detector itself is faulty .
NOTE: Modern high resolution semiconductor d e t e c t o r s are good and reliable instruments. If operated with proper care, they will last for many years. Even the old d e t e c t o r s can be trusted; fifteen-year-old Ge(Li) detectors that s t i l l operate with full original specifications are not unusual. In most cases, it is a human error that kills a detector. Handle these detectors with care, and you will not need the above troubleshooting advice.
(291)