Page 261

-112

Chapter 12

RADIATION DETECTORS

This c h a p t e r presents a d e s c r i p t i o n of d i f f i c u l t i e s with the d e t e c t o r s most frequently used in advanced nuclear spectrometry. Although a detector is d i r e c t l y , and sometimes inseparably, c o n n e c t e d to electronics, discussions on the behaviour and faults of electronic parts is avoided; for this the reader should refer to Chapters 5 and 11. F u r t h e r m o r e , a t t e n t i o n is d e v o t e d entirely to the high resolution d e t e c t o r s ; simple d e t e c t o r s , such as Geiger-Mueller or scintillation, are described in Chapter 11. A c c o r d i n g l y , the chapter s t a r t s with the h i g h - r e s o l u t i o n x-ray and gamma-ray detectors.

12. 1

HIGH RESOLUTION X-RAY AND GAMMA DETECTORS

Today, we find resolution detectors:

on the market the following

Coaxial pure germanium d e t e c t o r , p - t y p e Planar pure germanium d e t e c t o r Coaxial pure germanium d e t e c t o r , n-type Si(Li) d e t e c t o r Pure Si(Li) d e t e c t o r Surface barrier detector types of high for gamma x-rays x-rays and gamma x-rays x-rays alpha and beta

The Ge(Li) d e t e c t o r , which started the revolution in the field of high resolution spec troscopy, is not p r o d u c e d anymore by the respected manufacturer of semiconductor detectors.

12.1.1

Selection of a High Resolution D e t e c t o r

When ordering a germanium detector for gamma rays, one must spec ify : 1. 2. 3.

Resolution (an average d e t e c t o r will have resolution 1.8 keV). E f f i c i e n c y (varies from 8 to 40%). Type of cryostat, i.e. the size of the dewar, and the position of the dipstick.

NOTE ;

The price of a Ge detector steeply increases with the size (efficiency) and with improved resolution.

NOTE ;

A d e t e c t o r with a v e r t i c a l d i p s t i c k is the most common type. Other shapes of the dipstick are for special applications.

(279)

Page 262

-2-

Chapter 12

For a pure planar germanium d e t e c t o r , we must define 1. 2. 3. 4.

Thickness of the d e t e c t o r Area of the d e t e c t o r , in square mm R e s o l u t i o n , in keV Size of the dewar and type of the d i p s t i c k

| NOTE : Planar Ge d e t e c t o r s are used in measurement of x-rays | | w i t h energy above 20 keV. Due to intense escape | i p e a k s , they cause d i f f i c u l t i e s in i n t e r p r e t a t i o n of | | c o m p l e x spectra. | I________________________________________________________________I When buying a Si(Li) d e t e c t o r , we must determine: 1. 2. 3. 4.

Resolution of the d e t e c t o r , in keV (an average d e t e c t o r has 180 eV) A r e a , in square mm (the usual sizes are 30 and 80 sqmm) Type of c r y o s t a t and d i p s t i c k Type of p r e a m p l i f i e r (optical or resistive feedback)

12.1.2

Installation and Testing of a High Resolution Detector

The high r e s o l u t i o n solid s t a t e d e t e c t o r s are shipped together w i t h the dewar, but w i t h o u t l i q u i d n i t r o g e n , except if the customer specifies otherwise. A pure germanium d e t e c t o r and a modern Si(Li) d e t e c t o r can be stored for years at room temperature, without loosing their c h a r a c t e r i s t i c s . B e f o r e put into o p e r a t i o n , the dewar of such a d e t e c t o r must be filled with l i q u i d n i t r o g e n , and the d e t e c t o r cooled for a period of 4 to 6 hours. The manual of the d e t e c t o r s p e c i f i e s the minimum time for cooling, and it is a d v i s a b l e to adhere to the r e c o m m e n d a t i o n s of the m a n u f a c t u r e r s . With each solid state d e t e c t o r , the customer receives a certificate describing the p r o p e r t i e s of the d e t e c t o r , and the c o n d i t i o n s under which these c h a r a c t e r i s t i c s have been d e t e r m i n e d . An example of the c e r t i f i c a t e (in this case, for a Si(Li) d e t e c t o r ) is given in F i g . 12.1. When d e t e r m i n i n g the features of the d e t e c t o r , one must try to r e p r o d u c e as close as possible the measuring c o n d i t i o n s given in the c e r t i f i c a t e . A general observation is that the m a n u f a c t u r e r s are rather c o n s e r v a t i v e with the s p e c i f i c a t i o n s presented in the certificate. With careful adjustment of e l e c t r o n i c s and correct measuring p r o c e d u r e s , the value given by the m a n u f a c t u r e r can as a rule be easily achieved and surpassed. Before you start to write angry letters to the company that supplied the detector about d i s s a t i s f a c t o r y performance of their d e t e c t o r , be sure about your electronics s e t u p , grounding and overall conditions in your laboratory. For example, high humidity in the laboratory can easily deteriorate the resolution of a Ge

5000V+ as the bias voltage.

(280)

detector that needs

Page 263

-3-

Chapter

12

Section 6 DIÏTKCTOR Sl'KCIFICATIONS AND PKR FOR M A NCK D A T A (KKAK KNVKI.OPK) 6.1

SPECIFICATIONS

Serial Number b 86151

Tne purcluso specifications and LliereLore the warranted pcrfoninrtce of tlus detector are as fol

Model

___7229P-7500-1320

Rel. FJficiency __

13

Resolution

2.0 keV (B-.1M) keV (FVIN)

at 1.33 MeV keV (1-V.1M)

keV (FV.TM) Peak/Ccnipton at 122 keV

: 1

Cryostat Description

Vertical d i p s t i c k , type 7500

6.2 PHYSICAL/PERFORMANCE DATA

Date

Actual performance of this detector when tested is given below. Geometry Coaxial one open end, closed end facing window. Dianeter

48 nm

Length

35 nm

Active area facing window

18.1 5 an1 nm

Depletion Voltage

(+)4000

Vdc.

Recomrended Bias Voltage

(+)4500

Vdc.

Distance from window

ELECTRICAL CHARACTERISTICS

.01

Leakage Current at Reccmnended Bias

Preamplifier Test Point Voltage at Recoimended Bias nA - 2.3

Vdc.

RESOLUTION AND EFFICIENCY *

FWIM(keV)

,, ftf) GJ 1332 1.79 3.34

Peak/Comuton

42.4

Efficiencv(%)

13.0

Isotone Enercv(keV) [•WN(kcV)

r Co V

122 .836

:1

All measurements performed at 4 microseconds sliaping time

Fig.

12.1: The certificate for a Si(Li) x-ray d e t e c t o r

(281)

Page 264

C h a p t e r 12 12.1.3

-4G e r m a n i u m Gamma-Ray Detector

The International Elect r o t e c h n i c a l Commission published d e t a i l e d i n s t r u c t i o n s for t e s t i n g of s e m i c o n d u c t o r d e t e c t o r s . For gamma-ray d e t e c t o r s , the IEC Publication 656 (1979), "Test Procedures for H i g h - P u r i t y G e r m a n i u m D e t e c t o r s for X and Gamma R a d i a t i o n " can be ordered from the B u r e a u Central de la Commission E l e c t r o t e c h n i q u e I n t e r n a t i o n a l e , 1, rue de V a r e m b e , Genève, Suisse.

12.1.3.1

D e t e r m i n a t i o n of resolution

R e s o l u t i o n of a gamma-ray s e m i c o n d u c t o r d e t e c t o r is defined as the fu 1 1 - w i d t h - a t - h a ï f - m a x i m u m (FWHM) of the peak t h a t corresponds to the g a m m a - r a y s from Co-60 at the energy of 1.3MeV. The m e a s u r e m e n t is s i m p l e : place a s u i t a b l e Co-60 c a l i b r a t i o n source at a d i s t a n c e of 25 mm from the face of the d e t e c t o r . The c o u n t i n g r a t e of the d e t e c t o r should be at least 1000 counts/s. The r e s o l u t i o n of a detector depends on the counting rate; with a higher counting r a t e , t h e r e s o l u t i o n decreases. The commercial c o m p a n i e s m a k e t h e i r m e a s u r e m e n t s for the c e r t i f i c a t e at a counting r a t e very close to the nominal 1000 counts/s. Using a spectroscopy amplifier with carefully adjusted pole-zero c a n c e l l a t i o n and a good m u l t i c h a n n e l analyzer, c o l l e c t a s p e c t r u m of Co-60. In the c e n t r a l channel of the 1.3 MeV peak, at l e a s t 10,000 c o u n t s s h o u l d be c o l l e c t e d .

With a m o d e r n m u l t i c h a n n e l analyzer, the FWHM can be d e t e r m i n e d by the f i r m w a r e of the i n s t r u m e n t . O t h e r w i s e , you can do it a p p r o x i m a t e l y on the m o n i t o r of the M C A , or precisely by p l o t t i n g the p a r t of the s p e c t r u m w i t h the 1.3 MeV peak, and d e t e r m i n i n g t h e F W H M from t h e p l o t . Put all the c o n d i t i o n s of the measurement and the results in a log b o o k , t o g e t h e r w i t h the copy of the c e r t i f i c a t e .

12.1.3.2

Efficiency determination

E f f i c i e n c y is a very i m p o r t a n t p r o p e r t y of a high resolution gamma ray d e t e c t o r . The Ge d e t e c t o r e f f i c i e n c y is defined as the net area u n d e r the Co-60 peak at the energy of 1.3 M e V , c o m p a r e d to the net area u n d e r the same peak measured w i t h a 3x3 inch Nal d e t e c t o r . The e f f i c i e n c y is given in %. Place a Co-60 c a l i b r a t i o n source at a d i s t a n c e of 25 cm from the surface of the 3x3 inch Nal d e t e c t o r , take the s p e c t r u m and d e t e r m i n e the area under the 1.3 MeV peak. R e p e a t the measurement with the Ge d e t e c t o r under i d e n t i c a l g e o m e t r y . The r a t i o of the i n t e n s i t i e s g i v e s you the e f f i c i e n c y :

eff

(282)

=

I(Ge) I(Nal)

x

100

Page 265

-5-

C h a p t e r 12

Some o t h e r very i m p o r t a n t i n f o r m a t i o n t h a t tells us a b o u t the q u a l i t y of the d e t e c t o r , is the d e p e n d e n c e of the efficiency on the energy of the d e t e c t e d gamma rays. A n u m b e r of m e t h o d s have been designed for a c c u r a t e m e a s u r e m e n t of this p r o p e r t y , and computer programs are available for e v a l u a t i o n of m e a s u r e m e n t s . The simplest methods involve the use of a set of calibrated gamma sources. For d e t a i l s see "A Guide and I n s t r u c t i o n for D e t e r m i n i n g Gamma-Ray Emission R a t e s w i t h Germanium Detector Systems", by K. D e b e r t i n , May 1985. 12.1.4

Si(Li) and Planar Ge D e t e c t o r

They are for high r e s o l u t i o n x-ray s p e c t r o m e t r y . From the outside, there is only one essential observable difference, compared to the gamma-ray Ge d e t e c t o r s : on the top of the d e t e c t o r cap we find a b e r i l l i u m window. It r e p r e s e n t s a barrier that separates t h e vacuum inside the detector from the atmospheric pressure o u t s i d e , and does not permit light to f a l l on the detector. It does not stop x-rays a p p r e c i a b l y , e x c e p t if they are of very low e n e r g y , s a y , below 4 keV. A newly received x-ray detector should be t e s t e d for its performance i m m e d i a t e l y upon arrival. A f t e r cooling it w i t h the l i q u i d n i t r o g e n for several hours, the t e s t s can start. 12.1.4.1

Connections and i n i t i a l t e s t s

The x-ray solid state d e t e c t o r s use two t y p e s of p r e a m p l i f i e r : w i t h r e s i s t i v e f e e d b a c k , or w i t h o p t i c a l feedback. Most of the c o n t e m p o r a r y Si(Li) d e t e c t o r s use the o p t i c a l feedback because t h i s p e r m i t s t h e m to o b t a i n b e t t e r resolution. The one honourable e x c e p t i o n are the d e t e c t o r s p r o d u c e d by ORTEC, using r e s i s t i v e feedback.

Let us consider the case of a preamplifier with o p t i c a l feedback. A f t e r connecting the high voltage (SHV c o n n e c t o r s , usually a RG59 cable; observe the p o l a r i t y , in most cases a negative voltage is a p p l i e d ! ) , use a BNC cable to d i s p l a y the output (sometimes called E N E R G Y ) from the p r e a m p l i f i e r in an oscilloscope. Before the h i g h v o l t a g e is a p p l i e d , the o s c i l l o s c o p e (try to use the a m p l i f i c a t i o n on the v e r t i c a l scale t h a t gives 5 V / c m , and a t i m e c o n s t a n t of the sweep of 100ms) will show t h a t t h e r e is some n e g a t i v e v o l t a g e (about 1 2 V ) on the o u t p u t . This you can see if you s w i t c h the oscilloscope from DC to G R O U N D o p e r a t i o n . Slowly s t a r t to increase the v o l t a g e , say a b o u t 100 V per second. If you s t o p for a m o m e n t at 300 V, you will see on the oscilloscope the picture as shown in Fig. 12.2. A f t e r a few seconds, the s a w t o o t h shape of the signal w i l l become s t r e t c h e d out (Fig. 12.3), u n t i l you see only a s t r a i g h t line. N o w , the v o l t a g e can be increased to the p r e s c r i b e d v a l u e .

(283)

Page 266

C h a p t e r 12

-6-

Fig.

12.2: Scope shows sawtooth shapes when high v o l t a g e is applied to an o p t i c a l feedback p r e a m p l i f i e r

Fig.

12.3: In a short t ime, the shape changes to a low frequency saw

It is time to put a r a d i o a c t i v e calibration source on the d e t e c t o r . It is p r e f e r a b l e to use a Fe-59 source, e m i t t i n g x-rays of 5,9 keV. If you do not have a c a l i b r a t i o n source, use an e x c i t a t i o n source in a s u i t a b l e holder, and place a sample that contains mainly iron, in the position of the s a m p l e , on top of the holder. If the counting rate is h i g h , the sawtooth w i l l appear again. Increase the sweep frequency and the v e r t i c a l a m p l i f i c a t i o n of the oscilloscope, and you will observe the steps on the increasing line of the sweep, Fig 12.4.

(284)

Page 267

-7-

Fig.

12.4:

C h a p t e r 12

Shape of the p r e a m p l i f i e r o u t p u t w i t h x-rays coming into the d e t e c t o r

The d e t e c t o r and the p r e a m p l i f i e r are working.

12.1.4.2

Résolut ion

After connecting the p r e a m p l i f i e r to the main a m p l i f i e r , and t h i s to the m u l t i c h a n n e l analyzer, adjust the a m p l i f i c a t i o n of the main a m p l i f i e r so t h a t the 5.9 keV peak will be registered in channel 300 or there about. Use long shaping time (try to m a t c h the s p e c i f i c a t i o n on the c e r t i f i c a t e ! ) , say 16 m i c r o s e c o n d s , and the a m p l i f i c a t i o n a b o u t 300.

Make the energy c a l i b r a t i o n of the MCA using two c a l i b r a t i o n x-ray sources, or two d i f f e r e n t pure m e t a l s on top of the holder w i t h the e x c i t a t i o n source. One of the two should be the Fe-59 source. Collect sufficient counts in the peak to o b t a i n good s t a t i s t i c s . Determine the FWHM of the k-alpha (the largest) of the Fe-59 r a d i o i s o t o p e peaks. The FWHM value of the 5.9 keV peak, in eV, is the resolution of the d e t e c t o r , as given in the c e r t i f i c a t e . If no Fe-55 c a l i b r a t i o n source is available, you can use an e x c i t a t i o n source, and an iron sample. A l t h o u g h the resolution is determined at 5.9 keV (the Mn K-alpha line), the value obtained with the iron target (6.4keV for the K-alpha line) will be a good approxima t ion. 12.1.4.3

Efficiency

Determining the efficiency of the x-ray d e t e c t o r as the function of x-ray energy is a tedious and long procedure (see for example the IEC Publication 759, "Standard Test Procedures for Semicondict or X-Ray Energy Spectrometer". Around 10 keV, a Si(Li) detector efficiency is 100%, falling to zero at 2 keV, and decreasing less radically on the high energy side.

(285)

Page 268

Chapter 12 12.1.5

-8-

Maintenance of High Resolution D e t e c t o r s

In addition to all the established routines for the proper h a n d l i n g of nuclear i n s t r u m e n t s , some specific p r e c a u t i o n s should be a p p l i e d to the solid s t a t e d e t e c t o r s that o p e r a t e at low t e m p e r a tures .

All m o d e r n , high resolution solid state d e t e c t o r s can be stored at room t e m p e r a t u r e . In your l a b o r a t o r y , you might still have an old Ge(Li) d e t e c t o r , or a Si(Li), m a n u f a c t u r e d before 1982. These d e t e c t o r s have to be kept at a low t e m p e r a t u r e at all times.

How do we know when the level of the liquid n i t r o g e n in the dewar is too low? One can buy some level measuring devices (or, as an e l e c t r o n i c s m a n , you can design one yourself). More simple is to put the d e w a r w i t h the d e t e c t o r on a b a t h r o o m scale, and d e t e r m i n e how heavy the system is w i t h , and without liquid nitrogen. In many l a b o r a t o r i e s , the dewar is refilled by a rather dangerous p r o c e d u r e of taking the d e t e c t o r from the dewar, filling the l i q u i d nitrogen by p o u r i n g it in the dewar, and reinstalling the d e t e c t o r . If you are a very careful and rather strong m a n , this t e c h n i q u e works. But it is m u c h more a d v i s a b l e , p r a c t i c a l and safe to a c q u i r e a transport dewar w i t h a filling a t t a c h m e n t (see Fig. 12.5).

RELIEF/DRY NITROGEN PRESSURIZATION INLET

Fig. 12.5: The nitrogen transfer device The d e t e c t o r cap is connected to the dipstick with a flange, see Fig 12.6. An 0-ring takes care of vacuum tightness of the joint. Be careful not to pour liquid nitrogen on the flange. The cooling will make the rubbery material of the 0-ring b r i t t l e , and a vacuum leak will develop. Some manufacturers use metal rings; they to are safe as far as the cooling is concerned. but are d i f f i c u l t replace.

(286)

Page 269

Chapter 12

-9-

By m i s h a n d l i n g , liquid nitrogen can be poured on the flange. While this is bad, it is not necessarily a c a t a s t r o p h y . Tighten the screws h o l d i n g the b o t t o m and top flange t o g e t h e r ; you m i g h t be lucky, and the 0-ring can be squeezed enough to still hold the vac uum.

It is a recommended policy to keep the d e t e c t o r in l i q u i d n i t r o g e n all the time. The s h o r t a g e of liquid nitrogen m i g h t force you to let the d e t e c t o r warm up. If this h a p p e n s , let it warm up all the w a y , i.e. let it spend a day or two at room temperature before it is cooled again. Dflnn

Of lee

Copper

Aluminium

Fig.

12.6:

The inner life of a semiconductor d e t e c t o r . the flange with an 0-ring.

Note

High resolution detectors are sensitive to mechanical shocks. Be very careful when moving them. A special feature of these d e t e c t o r s is microfony. Mechanical vibrations or accustic signals will be t r a n s f e r r e d to the d e t e c t o r , and u n d e s i r a b l e noise will be produced. Different precautions can be taken to avoid these effects:

- place the dewar of the detector on a soft, rubbery mat; - put sound-absorbing m a t e r i a l around the d e t e c t o r c a p ;

- put some fine sand in the b o t t o m of the dewar, to a level that the end of the dipstick will be inserted in the sand; this prevents bubbles from forming on the surface of the d i p s t i c k end. With the Si(Li) d e t e c t o r s , the berillium window is the most sensitive part. It can be easily broken; high h u m i d i t y will cause

(287)

Page 270

C h a p t e r 12

-10- corrosion resulting in the appearance of m i c r o s c o p i c vacuum leaks. When not in use, the detector should be p r o t e c t e d with the p l a s t i c cap. If s t o r e d for a longer p e r i o d , the d e t e c t o r top should be covered w i t h a p l a s t i c s h e e t , with some s i l i c a gel to avoid h i g h humid it y. 12.1.6

T r o u b l e s h o o t i n g of Ge and Si(Li) D e t e c t o r s

The s e m i c o n d u c t o r h i g h - r e s o l u t i o n d e t e c t o r s are normally bought together with the p r e a m p l i f i e r . The p r o p e r t i e s of the d e t e c t o r (and w i t h them its p r i c e ) are d e f i n e d for the c o m b i n a t i o n de t e c t or + p r e a m p l ifier + dipst ick. With some detectors, the preamplifier can be ordered separately as a replacement, to be i n s t a l l e d locally.

There are several possible combinations and each s h o u l d be t r e a t e d in a s e p a r a t e way:

and c o n f i g u r a t i o n s ,

(i)

The ÜRTEC d e t e c t o r s are recognized by their shape: the p r e a m p l i f i e r is i n t e g r a t e d in the cylinder of the d e t e c t o r head, see Fig. 12.7. This s t r e a m l i n e d solution is p r a c t i c a l for some m e a s u r e m e n t s where the d e t e c t o r must be inserted in a measuring c h a m b e r , and the protruding p r e a m p l i f i e r is an o b s t a c l e . These d e t e c t o r s are somewhat more d i f f i c u l t to service and repair.

(ii)

A widely accepted configuration is shown in Fig. 12.8. H e r e , the p r e a m p l i f i e r is located in a separate box that is attached to the d i p s t i c k . Such a p r e a m p l i f i e r can contain the i n p u t s t a g e , w i t h the FET at room t e m p e r a t u r e : this solution is sometimes a p p l i e d to the gamma-ray detectors. In most cases, the p r o t r u d i n g box does not c o n t a i n the input stage; t h i s is somewhere inside the d e t e c t o r c a p , w i t h the FET at low t e m p e r a t u r e .

Suppose t h a t you have a gamma-ray detector, with a p r e a m p l i f i e r o u t s i d e the d e t e c t o r cap. Something seems to be wrong w i t h it. B a s i c a l l y , there are two possibe types of defects, mechanical and e l e c t r o n i c . The d e f e c t s and their symptoms are listed in Table 12.1

(288)