Page 251
(267)
Fig.
11.13
Main mechanical elements of a medical scanner
NUCLEAR DETECTOR
PATIENT
PM
DETECTOR AND RECORDER
Nal
HV AND SIGNAL CABLE
SUPPORTING DRIVEN BEAM (LINE-WISE MOTION) "Xx" DRIVE "Y" DRIVE |
RECORDER HEAD WITH & at we Se COLOR TYPEWRITER RIBBON CC. uaj --T AND TAPPER
™~ wT
LIMIT SWITCHES
SCINTIGRAM PAPER w, CONTROL BOX FOR
SCANNER CONTROLS DETECTOR POSITIONING
-27-
NUCLEAR CHANNEL
CONTROLS
CASTORS FOR f=q Semen THE SCANNER
Chapter ll
Page 252
Chapter 11
11.2.3
-28-
C o n t r o l s , Turning-On and Quick Checks
The nuclear channels are controlled by the ISOTOPE SELECTOR (see F i g . 11.14); this switch or p u s h - b u t t o n adjusts the energy and window s e t t i n g s o p t i m a l for the isotope used. If they are not p e r i o d i c a l l y checked for a c c u r a c y , poor p i c t u r e s m i g h t result. In most s y s t e m s , MANUAL ENERGY and WINDOW controls are also a v a i l a b l e . If the manual and the preset a d j u s t m e n t s give d i f f e r e n t r e s u l t s , it's time for r e a d j u s t m e n t and c a l i b r a t i o n .
ISOTOPE SELECTOR r
1 ! 1
^, -^
RECOMMENDED
S4-T 4<?
r".. .......
/^
ô COUNTS/SEC
. PRESET MODE A SELECTOR ^ MANUAL ENERGY
WINDOW
H. V.
©
<E Fig.
11.14
LOUDSPEAKER
^ ^ s . ON
1
I
1
1
1
OFF
1
1
COUNTING-RATE METER RANGE SELECTOR
Nuclear channel controls
The c o u n t i n g - r a t e m e t e r range a d j u s t m e n t is conventional. If a r a d i o a c t i v e source is introduced into the field of view of the detector and the corresponding p r e s e t or manual energy and window s e l e c t i o n was done, the d i s p l a y m e t e r should i n d i c a t e changing values as the source is moved under the d e t e c t o r .
Fig.
An e x a m p l e of the front panel of a scanner control is shown in 11.15.
1000 j^rif^fe^^^^ «fl£
GOOD E-t W 2 S W Q W D W O U bj o « u o; UhJK3 M >
SPEED
[cm/min]
X 0
COUNTS/cm* INFORMATION DENSITY
COLOR DISPLAY
1 1 1 1
»
STEP SIZE MANUAL CO POSITIONING
Fig.
(268)
11.15:
SCANNING
Scanner controls lOj ^-/ COLOR ADJUSTMENT
OFF MATMC nALNt >
Page 253
-29-
C h a p t e r 11
The a c t i v i t y-1o-colour servo should follow the movement of the display m e t e r as the source is moved under the d e t e c t o r .
The next test is concerned with the X-Y movements of the detector. First the LIMIT or LIMIT SWITCHes should be set for a short, few c e n t i m e t e r long line l e n g t h , with full-size step-wise notion and the scanning should be s t a r t e d with the START switch in the lowest SPEED range. The scanner should start m e a n d e r i n g within the preset borders. The speed should be adjusted for higher and higher speeds up to the m a x i m u m . Take care, because if the limit switch is b a d , the detector might try to run away, breaking p a r t s of the mechanical s y s t e m ; this is why the b o r d e r control checking is recommended on the lowest speed range. If the system is already moving, a small source, a so-called "phantom-organ", should be put under the detector and imaging should be a t t e m p t e d . Here one should follow the recommended s e t t i n g - u p p r o c e d u r e described in the manual. First p o s i t i o n the d e t e c t o r above the source and adjust the CALIBRATION control to get the display meter pointer to reach the red colour zone, representing the highest a c t i v i t y zone on the scintigram. The speed should be set to get the recommended 600 count s/square centimeter information density. An image of the source should result from this a c t i v i t y similar to those which were taken previously under r e f e r e n c e conditions.
The d e t e c t o r and source distance is critical in imaging; the best resolution could be obtained from the "focal plane" of the m u l t i h o l e d i a p h r a g m before the sein t i llator. The optimal imaging d i s t a n c e is given on the d i a p h r a g m . Each scanner has some means for the detector positioning. This function is used during the setting-up adjustment. By looking on the counting-rate meter or listening to a variable frequency loudspeaker sound, controlled by the r a d i a t i o n i n t e n s i t y , the operator should search for the area with maximum isotope concentration. This manual mode can be used to test the safety switches on the border of the maximum scannable area. If the detector is not stopped there a u t o m a t i c a l l y , the drive mechanism can be harmed. Fig.
A block 11.16.
11.2.4
11.2.4.1 schematic of a typical scanner can be found in
Troubleshooting and Stock Faults
Power supplies
In most systems separate low-voltage regulated power supplies are used for the nuclear channel and for the detector motors. A DC/DC converter feeds the scintillation detector, which receives power from the nuclear channels voltage regulator. The troubleshooting procedure is similar to all scintillation detector high-voltage power supplies.
(269)
Page 254
Chapter 11
-30-
£l EH OS CO i I.UQO !H H HM l -_|
Cn Q Z1 los z os Si
:o w o x| të Sa al
.u n « 2l
1 cn z
0! 8
§
OS U
X iJ 0.
S
0<
S cn
« M
S" ÖJ s 1|
§| o< <
>H
Ö EH
Z
O u n O
O
O. < a u
PS H o o u D
Fig. 11.16:
(270)
Block schematic of a typical scanner
Page 255
-31-
C h a p t e r 11
The r e g u l a t e d voltage s u p p l i e s of the X-Y m o t o r s are r a t h e r simple; their rated output is b e t w e e n 150-300 VA. They are d e s i g n e d to deliver ample power to s t a r t moving the d e t e c t o r by the s t e p p i n g motors. The speed r e g u l a t i o n is a c c o m p l i s h e d by c h a n g i n g the f r e q u e n c y of the steps. The r o t a t i o n d i r e c t i o n is changed by a l t e r i n g the phase r e l a t i o n s in the c o n t r o l coils of the m o t o r s . The high power s e m i - c o n d u c t o r s m i g h t fail more o f t e n if the v e n t i l l a t i o n of the h e a t s i n k s are obstructed or high voltage t r a n s i e n t s are present on the s u p p l y lines. This o f t e n h a p p e n s in h o s p i t a l s w h e r e the c a b l e s are h e a v i l y overloaded and the wire c o n n e c t i o n s are c o r r o d e d . An AC v o l t a g e s t a b i l i z e r can improve the s i t u a t i o n for the scanner. There is only one simple rule in power diode r e p l a c e m e n t , the symmetry should be conserved in the r e c t i f i e r , t h a t is if the same diode which failed is not available, both diodes should be replaced w i t h p r o p e r l y rated ones. 11.2.4.2
Scintillation detector
All t r o u b l e s h o o t i n g r e c o m m e n d a t i o n s given for scintillation d e t e c t o r s in this m a n u a l are valid for the d e t e c t o r s of the scanners as well. But t h e r e are certain stock f a u l t s which are not common in o t h e r a p p l i c a t i o n s .
D e t e c t o r cable. They are exposed to c y l i c a l b e n d i n g s and movements after accomplishment of each scanning l i n e ; this can amount to 1 U . O O U cycles/week and two m i l l i o n cycles/year. Most cables are worn out in 3-4 years. The cable d e t e r i o r a t i o n is slow; sometimes noises appear w i t h o u t any apparent reason, only in c e r t a i n areas of the s c i n t i g r a m even w i t h o u t any isotope. If such e f f e c t a p p e a r s , it is a sure sign t h a t the cable is d e t e r i o r a t i n g .
C o n t a m i n a t i o n of the d e t e c t o r . Under c l i n i c a l c o n d i t i o n s it happen that the d e t e c t o r surface is contaminated by the p a t i e n t . This can be p r e v e n t e d by enclosing the d e t e c t o r in an easily r e p l a c e a b l e p l a s t i c sheet or b a g . could
Scintillation detector stability. If the gain of the p ho t omul t i p lier changes during the i m a g i n g , it could happen t h a t a f r a c t i o n of the photopeak i n f o r m a t i o n w i l l be lost. On the scintigram it m i g h t look as if certain organ areas had lower m e t a b o l i c f u n c t i o n s . It is easy to test the system by a t t a c h i n g a small r a d i o a c t i v e source to the d e t e c t o r and keeping the scanner record the measured intensity in this fixed g e o m e t r y . A good scanner should not drift in one hour more than ten percent in recorded intensity.
11.2.4.3
P u l s e - h e i g h t analyzer
All recommendations given for p u l s e - h e i g h t analyzer t r o u b l e s h o o t i n g are v a l i d for the scanners as well; there is, h o w e v e r , one a d d i t i o n a l uncommon stock f a u l t .
(271)
Page 256
Chapter 11
-32-
Isotope selector d r i f t . Most scanners have certain preset channels for the most o f t e n used isotopes. The adjusting e l e m e n t s might change their value due to e n v i r o n m e n t a l e f f e c t s . It is important to check the accuracy of the preset channel positions periodically by comparing the manually a d j u s t e d p h o t o p e a k counting rates to the preset channel r e s u l t s . If they d i f f e r by more than 10%, readjustment is i n d i c a t e d .
11.2.4.4
Counting-rate meter
The c o u n t i n g - r a t e m e t e r s d r i v e the in t ens ity-to-colour code c o n v e r t e r in the display system. The linearity of the conversion is critical and since it can lead to false medical diagnosis, it has to be checked p e r i o d i c a l l y . The procedure is r a t h e r s i m p l e . One should measure the f r e q u e n c y of a pulse generator which drives the p u l s e - h e i g h t analyzer. A graph should be p l o t t e d on input frequency vs. d i s p l a y e d value r e l a t i o n . A c t i o n should be taken if the i n t e g r a l non-linearity exceeds + /- 5%. A pulse generator is an essential of c o u n t i n g - r a t e meter faults. 11.2.4.5 instrument in the d i a g n o s i s
Colour a d j u s t m e n t p o t e n t i o m e t e r
The most often used c o n t r o l e l e m e n t of the scanner is the p o t e n t i o m e t e r between the c o u n t i n g - r a t e m e t e r s o u t p u t and the input of the in t ens i ty-1o-colour c o n v e r t e r servo c i r c u i t . It must be adjusted so that the maximum a c t i v i t y area should appear in red colour on the s c i n t i g r a m . An early sign of wear of this p o t e n t i o m e t e r is if the colors are changing while the c o u n t - r a t e remains c o n s t a n t during t e s t s w i t h a source fixed to the d e t e c t o r .
11.2.4.6
In tensity-1o-colour code converter servo
The c i r c u i t diagram and m e c h a n i c a l c o n s t r u c t i o n of a t y p i c a l converter is shown in Fig. 1 1 . 1 7 . In its s i m p l e s t version the p o s i t i o n of the coloured typewriter r i b b o n under the t a p p e r is sensed by a p o t e n t i o m e t e r , m e c h a n i c a l l y a t t a c h e d to a DC m o t o r . The n o n - i n v e r t i n g input of the operational a m p l i f i e r is connected to the source of the i n t e n s i t y i n f o r m a t i o n , t h a t is, to the counting-rate meter output through the colour adjustment potentiometer. The inverting input is connected to the moving contact of the p o t e n t i o m e t e r , while one end is on ground and the other on a reference DC v o l t a g e level. The DC motor is c o n n e c t e d w i t h such p o l a r i t y to give n e g a t i v e f e e d b a c k , t h a t is, the v o l t a g e from the p o t e n t i o m e t e r compensates the input v o l t a g e and the motor r o t a t e s u n t i l the error signal is m i n i m i z e d . The most c r i t i c a l part is the p o t e n t i o m e t e r . They wear out after 10 m i l l i o n t u r n s , so their expected life is 3-4 years in a nuclear medicine unit with 5.000 imagings per year.
(272)
Page 257
-33-
Chapter 11
COUNTING RATE METER OUTPUT COLOR ADJUSTMENT
DC SERVOMOTOR
REFERENCE VOLTAGE
COLOR RIBBON POSITION TRANSDUCER
COLOR RIBBON CARRYING CAGE NUT ON SCREW rur
1
/
X, J ' J W/////JP Pj/j/rf
SCREW
1
1 l
1
! $ /-w \
® GEAR BOX
DC SERVO MOTOR
/ COLOR RIBBON POSITION TRANSDUCER
COLOR RIB
PRINTING EDGE
Fig.
11.17:
Circuit diagram and mechanical c o n s t r u c t i o n of a t y p i c a l converter
(273)
Page 258
C h a p t e r 11
-34-
Wear c o u l d be one reason for the component d e t e r i o r a t i o n , but under-used p o t e n t i o m e t e r s can d e v e l o p noises under tropical c o n d i t i o n s as well. However, they are r e d u c e d to a normal l e v e l a f t e r a few hours of o p e r a t i o n . 11.2.4.7
C o n t r o l logic
The movement of the d e t e c t o r is u n d e r the control of the scanners logic s y s t e m . Its main f u n c t i o n s are the sensing of the p o s i t i o n of the scanning area d e t e r m i n i n g b o a r d e r limit switches, the p o s i t i o n of the b o r d e r l i m i t p r o x i m i t y switches and the safety limit s w i t c h e s , and to c o n t r o l the s t e p p i n g m o t o r s a c c o r d i n g l y . Some s i t u a t i o n s and a c t i o n s :
s c a n n i n g line-wise border switch a c t i v a t e d : step-wise m o t i o n s t a r t s , a f t e r this new line scanning s t a r t s w i t h reversed d i r e c t i o n . step-wise border s w i t c h a c t i v a t e d : s c a n n i n g is f i n i s h e d , all motors are s t o p p e d , b o r d e r limit p r o x i m i t y switch a c t i v a t e d : m o t o r torque is r e d u c e d , breaking a c t i o n s t a r t s before scanning d i r e c t i o n is r e v e r s e d , this a c t i o n is i m p o r t a n t o t h e r w i s e the m e c h a n i c a l system is exposed to heavy strain. safety limit switch activated:
f u r t h e r m o t i o n in same d i r e c t i o n is i n h i b i t e d . Under the manual d e t e c t o r p o s i t i o n i n g m o d e , only the safety limit s w i t c h e s are a c t i v e in p r o t e c t i n g the m e c h a n i c a l system. 11.2.4.8
S t e p p i n g motor c o n t r o l c i r c u i t s
The block diagram of the circuit is illustrated in Fig. 1 1 . 1 8 . The step frequency selection is done by the SPEED control switch. In a phase s p l i t t e r , d o u b l e phase signals are g e n e r a t e d from the step frequency signals to feed the r o t a t i o n d i r e c t i o n c o n t r o l circuit together w i t h the line-wise border limit switch information.
Whenever a border limit switch is a c t i v a t e d , the b i - s t a b l e m u l t i v i b r a t o r , sensing it, changes the phase of the triggering pulse to the t r i a c which d r i v e s the r o t a t i o n d i r e c t i o n control coil of the motor .
The motor torque is always reduced to zero before the rotation "firing angle" under the d i r e c t i o n is reversed by m o d u l a t i n g the control of the limit p r o x i m i t y switch.
(274)
Page 259
-35-
C h a p t e r 11
The step-wise m o t i o n is under the c o n t r o l of a preset c o u n t e r , p r o g r a m m e d by the s t e p - l e n g t h s w i t c h . In manual mode all r o t a t i o n d i r e c t i o n orders are c o n t r o l l e d by the manual s w i t c h e s and the m o t o r s move only while they are pushed and only the s a f e t y logic can i n h i b i t them. All r e p a i r a c t i v i t i e s should start with the c h e c k i n g of the s a f e t y s w i t c h e s , because if t h e y are s h o r t e d , the m e c h a n i c a l system w i l l not stop on the b o r d e r s , and if the torque limiters are b a d , expensive d a m a g e s can be e x p e c t e d . For this reason, it is a good p r a c t i c e to put the d e t e c t o r into the c e n t e r of the scanning area d u r i n g the i n i t i a l p a r t of the testing. If t r i a c s should be r e p l a c e d , the rule is to change them in pairs. Use fuses to p r o t e c t them during first t r i a l s after rep l a c e m e n t . 11.2.4.9 The
A l i g n m e n t of the m e c h a n i c a l system d e t e c t o r s are rather heavy and the m e c h a n i c a l system o p e r a t e p r o p e r l y only if the beam can s u p p o r t i n g the heavy lead s h i e l d is a c c u r a t e l y l e v e l e d . The m e c h a n i c a l system wears out q u i c k l y w i t h o u t this p r e c a u t i o n . It is important to check the leveling of t h e scanner a f t e r t r a n s p o r t a t i o n . 11.2.5
Q u a l i t y C o n t r o l of the Repaired Instrument
Recommendations of the IAEA TECDOC-317 on " Q u a l i t y Control of
N u c l e a r M e d i c a l I n s t r u m e n t s " s h o u l d b e followed.
11.2.6
Preventive Maintenance
Recommendations given to the Nuclear L a b o r a t o r i e s should be f o l l o w e d , t h a t is: - instruments like d u s t ,
Medical
Pilot should be p r o t e c t e d from e n v i r o n m e n t a l effects moisture condensation, excessive line v o l t a g e f l u c t u a t i o n s , and t r a n s i e n t s ; - i n s t r u m e n t s should be c o v e r e d a f t e r u s e ; - rarely used i n s t r u m e n t s s h o u l d be turned on p e r i o d i c a l l y and m e c h a n i c a l l y o p e r a t e d to p r e v e n t r u s t i n g , and to dry o u t ; - l u b r i c a t i o n and safety c h e c k - u p s manua1 ;
should be a c c o r d i n g to the
- only t r a i n e d operators s h o u l d use the i n s t r u m e n t ; - a log book should be kept on use and r e p a i r s , reference tests and t h e i r r e s u l t s should be a t t a c h e d to the book.
(275)
Page 260
Chapter 12 RADIATION DETECTORS