Page 141

Chapter 7 7.2.4

-6The O u t p u t Circuitry

The output pulses from this unit are intended to actuate the inputs of coincidence plug-in units or of time to a m p l i t u d e converters. The output circuitry should be able to drive terminated 50fl cable; discrete circuits are used here for this purpose. You should check if the positive-going pulse at the output connector J2 complies with TTL standards. A c t u a l l y , from the circuit you should e x p e c t logic 0 to be less than 0,2V for currents as large as 50mA (Q7 and Q9 are s a t u r a t e d , Q8 is cut off), and logic 1 to be higher than 2.5V for c u r r e n t s larger than 50mA (Q7 and Q9 are cut o f f , Q8 acts as an emitter-follower). The circuit for the negative-going NIM-pulse should sink a current of roughly 17mA; if this current flows through a 50-Ohm resistor, a signal of -0,8V develops; this signal is rather narrow due to the d i f f e r e n t i a t i n g time constant of a b o u t 20ns at the circuit's input. If the discrete o u t p u t circuits are suspected of malfunctioning, the t e c h n i q u e s given in Chapter 6 for the t r o u b l e s h o o t i n g of transistor c i r c u i t s may be followed. 7.2.5

Constant Fraction Timing Discriminator

Another common m e t h o d of obtaining accurate time marks, with a small a m p l i t u d e d e p e n d e n t time walk, is the constant fraction method. A c i r c u i t example may be taken from Canberra's Model 2035A constant fraction timing SCA, see Fig. 7.4. The plug-in unit differs from the one previously discussed e s s e n t i a l l y in the All input c i r c u i t s .

To check the circuit observe the input pulse at the e m i t t e r of Q13; at test point TP6, which is the negative input of A l l , the pulse occurs at the same time than at the emitter of Q13 and with approximately half amplitude. At test point T P 7 , which is the positive input of A l l , the pulse is delayed by 0.5, 2 or lOus according to the delay line s e l e c t e d , and its a m p l i t u d e is reduced to about 1/3. The All timing comparator responds to the voltage difference between these two pulses, changing s t a t e when the difference crosses zero. As b e f o r e , it may be seen that RV10 is a walk adjustment p o t e n t i o m e t e r . Note that when no input pulse is p r e s e n t , the o u t p u t of All (pin 5) is at logic 1.

(150)

Page 142 read from a scan by OCR; it may hold reading errors

mya a ol ae if lanwe 2

€ [awnuwe ord Se

Of arne fel REA pie lwop 2a “Mi; - + [eo TNote 10 Apoe® Feb ch jaro’ my | FL REV ED - ELA a {8

G [Ragas TRY ww al |

Me---5 Cup arar! TL OEE Cmaer PE oar] TC. 7 - aww ea a - peaD a- M0 “3- maD Pr) a. mrs Me - ate Al nm park aa - rEMs ave eres 4 a0 ~ DaTED aH ~ Ctl Met we feeg owner LAST MOEEPORE Camel wien INTEQEATED Cominrs att

337008 CBM + OCS mvaots Vrarne rot TEST Ons

- a7 ow -o" -@r “ras

(ist)

; “a COR ETTORE - 6M Ane -/ 6 tan mg aT A LOMSARS -t vO ee Sokted ear Ser CWE - ONT ar $9 Mak AK + err TRIMS TOR arr . WAKE POLES “e

TERE ABE NO RAM WA TAS ON TMS BEETS MVENC ATES isla BOT OX Bret ML FT AKLD RNIMIORS At Pi Derd LITA sarc ES AE Pens PB LORD 29 Amt OCHS @ NOW ORES MA ON FAM CORA ef PON AE? A A AOS (Tarn Fl atl Conenanenl (Ora rage Ty tw OrCa CS BEAL Mn, COMPONENT (RATION we MVR RR ESS t PECATES TVET LD Ate gh dol PS en mmo FOSSA = B: 60 | = be G

{ aajadeug

Page 143

Chapter 8 SCALERS, TIMERS, RATEMETERS

Page 144

-1-

Chapter 8

SCALERS. TIMERS. RATEMETERS 8. 1

TYPICAL STRUCTURES OF SCALERS. TIMERS AND RATEMETERS

To be able to service and repair sealers, timers and digital ratemeters, it is essential to correctly identify the d i f f e r e n t blocks in which the instrument may be logically d i v i d e d , and to u n d e r s t a n d their interactions. To help in this procedure, a short d e s c r i p t i o n of typical s t r u c t u r e s is given below. A detailed d e s c r i p t i o n of examples of commercially available instruments completes this c h a p t e r . 8.1.1

Introduct ion

The general block diagram of a t y p i c a l counter count c a p a b i l i t i e s is shown in Fig. 8.1.

INPUT

GATE

COUNTING STAGES with preset

DISPLAY

PRESET CIRCUITRY

Fig.

8.1: Block diagram of a typical counter

The signal to be counted can, in general, be easily followed through its path in the counter, and d e f e c t i v e ICs can be located without too much d i f f i c u l t y . C i r c u i t r y associated with preset count and gating is generally more d i f f i c u l t to troubleshoot because of interactions among several c o n t r o l signals. As an example of faults which are hard to f i n d , we may quote the spurious counts due to bad filtering by a defective capacitor.

8.1.2

Input circuits

Most input circuits accept NIM logic signals, positive and/or negative. Some counters have discriminat or-type inputs, allowing analog signals of small amplitude to be counted.

One must check whether the input circuit passes the signals for counting in a proper way. In p a r t i c u l a r , this involves checking for t r i g g e r i n g at the a p p r o p r i a t e levels, and for m u l t i p l e triggering due to cable reflections or other causes. NIM positive signals are frequently received at a Ik-Ohm impedance; significant reflections are absent provided the cables are not too long (and the signal is not too fast). NIM n e g a t i v e signals must be received

(155)

Page 145

Chapter 8

-2- in an i m p e d a n c e t h a t closely m a t c h e s the cable. F r e q u e n t l y , this impedance is m a d e up of a resistor in series with the input impedance of a common base transistor; a significant impedance m i s m a t c h may i n d i c a t e a d e f e c t i v e transistor.

D i s c r i m i n a t o r - t y p e input circuits most f r e q u e n t l y use the 710 or o t h e r fast d i s c r i m i n a t o r . Good power s u p p l y d e c o u p l i n g near the d i s c r i m i n a t o r , and a not too small hysteresis, are needed to avoid multiple triggering. 8.1.3

O u t p u t Circuits

Output circuits are prone to be damaged by improper connections. Most f r e q u e n t l y , however, they are just unable to cope w i t h the demands put on t h e m . For e x a m p l e , a TTL g a t e , or a CMOS d r i v e r , have a driving c a p a b i l i t y that can easily be e x c e e d e d . To d r i v e a t e r m i n a t e d 50-Ohm cable, one needs a transistorized o u t p u t stage or a special integrated circuit. A check should be made to verify whether legal logic levels are b e i n g delivered. O u t p u t c o n n e c t i o n to p r i n t e r s or e q u i p m e n t buses is eIsewhere.

8.1.4 addressed

C o u n t i n g Gates

Counting gates are open/closed manually or by signal-levels and/or pulses. Debouncing of manual switches is m a n d a t o r y ; the corresponding c i r c u i t r y is easily checked and the same is true for the latches a s s o c i a t e d w i t h pulse control. These b i s t a b l e elements are easily t r i g g e r e d by very short pulses; this may be a cause of t r o u b l e if d e c o u p l i n g is i n e f f e c t i v e , or if g l i t c h e s are allowed to occur and are not f i l t e r e d out. 8.1.5

The Counting Stages

The counter generally provides several counting decades whose o u t p u t is a v a i l a b l e in BCD format. N o r m a l l y , only the first (faster) d e c a d e , or decades d i r e c t l y associated with some types of preset c o n t r o l , are of the synchronous type. F r e q u e n t l y , ICs w i t h two d e c a d e s are u s e d , but LSI c i r c u i t s w i t h four or six d e c a d e s are already p r e s e n t in various i n s t r u m e n t s . Preset count is f r e q u e n t l y limited to the c o m b i n a t i o n p q x l O r where p.q and r represent decimal numbers. Generally it involves e i t h e r a comparison of the BCD-coded numbers p and q with the contents of the s e a l e r , or the zero d e t e c t i o n in a count-down process of two decades p r e v i o u s l y loaded with contents p and q. Troubles may easily arise from g l i t c h e s or spikes in the power lines.

(156)

Page 146

Chapter 8

-38.1.6

Display Circuits Seven-segment digits, implemented in LED or LCD arrays, generally used in displays.

In LED are typical, one from the decades to the BCD-to-7 segment another from the decoder to the are displays two multiplexed buses digits.

In decoder,

LCD displays, all digits should be permanently a c t i v a t e d ; circuits including a latch and a converter for each digit are generally used.

4-bit

Scanning c i r c u i t s are typically made out of a binary counter and a decoder (for example, an LS93 and an LS138) whose o u t p u t lines successively activate the open collector or tri-state circuits that connect the various decades to the b u s , and select the appropriate digit. Some LSI c i r c u i t s , like the 7301, already include the scanning circuits.

8.1.7 six decade

Time Generators

In timers or counter/timers, time marks are derived from the mains (typically at 1/100 or 1/120 second period) or, most frequently, from a crystal oscillator. These generally oscillate at a f r e q u e n c y of several MHz, which is suitably d i v i d e d to give time marks either in a seconds or in a minutes scale. In preset type count er/1imers, the preset count can be associated with either the counting events or the time marks.

Rat erne t er s

8.1.8

A typical d i g i t a l ratemeter has a structure similar to that of a preset count er/1imer.

preset

The integration period corresponds to a t i m e ; the displayed count, corresponding to the total count of the previous period, is u p d a t e d once per period.

Analog ratemeters typically shape the input pulses in a we 11-de f ined form, and integrate them to have a voltage to the incoming rate in either a linear or in a log proport ional scale.

8.2

TROUBLESHOOTING

The first step in troubleshooting is to establish diagram of the blocks. This the block instrument and to identify the faulty block or is generally easy for an absolute (no signal)

failure, but may be rather tricky for the faults related to d e g r a d e d performance (spurious counts, erroneous preset counting, etc . ) . 8.2.1 the

Checks at the Component Level

First check whether power arrives at the appropriate pin of 1C. The t r u t h table of gates must be seen to be followed, otherwise the component is defective. A check should be made to verify if the voltage levels are well w i t h i n the legal l i m i t s for

(157)

Page 147

Chapter 8

-4- the i n p u t s and the o u t p u t s ; t h i s a p p l i e s to a^l 1 d i g i t a l c i r c u i t s . In latches and f l i p - f l o p s , check if the Q and Q o u t p u t s are in fact c o m p l e m e n t a r y ; in general, check the t r u t h table of the component. If the c o m p o n e n t is supposed to b e h a v e in a dynamic w a y , like a LSI counter with m u l t i p l e x e d o u t p u t , verify whether the scanning and r e l a t e d signals are being sent o u t .

Analog c o m p a r a t o r s and t r a n s i s t o r s are sometimes found, e s p e c i a l l y in input and output circuitry. Measurement of the v o l t a g e levels at the input and o u t p u t will generally show whether a c o m p a r a t o r is working p r o p e r l y . Transistors are easily checked as discussed elsewhere; f r e q u e n t l y they will be either on or o f f , and the v o l t a g e levels they give to the following c i r c u i t s should be seen to comply with the appropriate legal levels under the load c o n d i t i o n s in w h i c h they are e x p e c t e d to o p e r a t e . D i s p l a y s may easily give short trouble.

F r e q u e n t l y this is due to c i r c u i t s or m i s e o n n e c t ions in LED d i s p l a y s , and to improper contact in the c o n d u c t i n g p a t h (usually at the rubber/glass junction) of LCD d i s p l a y s . These are also prone to m a l f u n c t i o n if excess h u m i d i t y p r o v i d e s leaky p a t h s for the segment v o l t a g e s . Faulty c a p a c i t o r s may be responsible for a t o t a l failure if they are s h o r t - c i r c u i t e d , or just to d e g r a d e p e r f o r m a n c e of the i n s t r u m e n t if they do not a d e q u a t e l y by-pass spikes or g l i t c h e s to ground (a much more d i f f i c u l t s i t u a t i o n to t roubleshoot ) . 8.2.2

Checks at the Block Level

The input block must be seen to give a single pulse for each pulse r e c e i v e d , even if the pulse has a reasonable amount of noise; a pulse generator will h e l p to d e t e r m i n e the t r i g g e r i n g threshold. If the input is supposed to work on analog signals, m u l t i p l e signals; a sine triggering is more likely to occur w i t h slower to ground by a d i o d e , is wave, with the n e g a t i v e half wave c l i p p e d a suitable test signal. If m u l t i p l e t r i g g e r i n g occurs (with the t h r e s h o l d well above noise level), the h y s t e r e s i s circuit of the c o m p a r a t o r must be checked.

Pulse p a i r r e s o l u t i o n is d e t e r m i n e d e i t h e r by the input block or by the first counting circuit. If otherwise unavailable, a simple test instrument can be b u i l t around two 74LS221 i n t e g r a t e d monostables and an OR g a t e , as shown in Fig. 8.2.

100nS

500 nS *••», »»

Fig. 8.2:

(158)

1 llOOnS, - Hr

Double pulse generator with adjustable separation between pulses

Page 148

Chapter 8

-5-

The f u n c t i o n i n g of the counting circuits can be easily checked if they consist of individually accessible decades or binary counters. When one is dealing with LSI c o u n t e r s , where the output lines are t i m e - s h a r e d by several d e c a d e s , the functioning of each d e c a d e may be observed by externally t r i g g e r i n g a scope with the scanning clock signal. Preset count circuitry may be checked starting from the storage of preset i n f o r m a t i o n , followed by observation of the c o u n t - d o w n , or by examination of the comparator c i r c u i t s that test when preset count is r e a c h e d , according to the approach used in the i n s t r u m e n t to implement the preset function.

8.3

TROUBLESHOOTING EXAMPLE:

TIMER/ SCALER, CANBERRA. MODEL 2070

A block diagram of this t i m e r / c o u n t e r is shown in Fig. 8.3; some s e m i c o n d u c t o r components (not all) are i n d i c a t e d to help in identifying the blocks.

INPUT EVENT GATE Q6,Q9,Q10,A56C

GATE CONTROL

PRESET COUNTER

EVENT/TIMEMUX A45-46.A52 « - - > D20, D21

A42-43,A51 A31.A41, A6C à i.

TIME BASE TIME GATE A33-36, A15C A46b

Fig. 8.3.1

8.3:

START/ STOP 4, - SINGLE/RECYCLE *, - - - - - - - - - - . 4, - - - - - - - A15a,A14,A3

A24.A41

Block diagram of Canberra 2070 timer/counter

Input and Event Gate

The dc voltage at the input is approximately 0V for it is by Q9 whose base is at about 0,7V, as set by the diode connected transistor Q6. NIM positive logic pulses (actually, any p o s i t i v e pulse larger than about 0,7V) saturate Q10 and cut off Q9; NIM negat ive pulses do not a f f e c t Q10, but draw enough current t h r o u g h Q9 for a logic 0 voltage to be d e v e l o p e d at the input of NAND g a t e A56c. Input pulses should appear at pin 8 of A56 if pin 9 is at logical 1; this level should appear if PRESET COUNTS mode is chosen (if n o t , check A 1 5 c for logic 0 at pin 10 and A52 for logic 1 a t pin 2). imposed

8.3.2

Time Generator and Time Mark Gate

Check the output of the 12 MHz generator and the divide by 12 1C (A33); the o u t p u t of A33 is t r a n s m i t t e d to the ck input (pin 5)

of A35 if gate A 1 3 d is opened;

this should occur if the external

GATE input is left u n c o n n e c t e d , and should stop if this input is g r o u n d e d . The 1 MHz o u t p u t of A33 is d i v i d e d by 10 by A35, giving

(159)

Page 149

Chapter 8

-6- time marks spaced by 0,01s; A34 provides a division by 60, producing time marks separated by 0,01 min. A36 a and b, together w i t h A46c, form a 2-to-l m u l t i p l e x e r allowing the selection of the time scale by a switch. In TP8, a time mark wave should be seen of either 0.01 min or 0.01s (in the PRESET 0.01s or COUNTS position). A46b is the time base g a t e ; the control input of this NOR gate (pin 6) is c o n n e c t e d to the same line as the control input of the NAND event gate (pin 10 of A56), thus assuring that one, but only one, of these gates is always open. 8.3.3

Gate C o n t r o l and Event/Time Multiplexer

The e x t e r n a l GATE signal actuates through NOR gate A45c. If the gate is closed (logic 1 at pin 8), no time pulse will pass; again, the o u t p u t (pin 10) will force m u l t i p l e x i n g d i o d e D20 into c o n d u c t i o n , therefore fixing the o u t p u t of the MUX (the common p o i n t of D20 and D21) to logic 0, and thus also disabling the counting of event pulses. If jumper A is moved to the B position, the e x t e r n a l gate signal cannot disable counting.

The external gate also actuates on the time generator by closing A 1 3 d (provided jumper 5 is in the appropriate position); t h u s , time c o u n t i n g can be controlled in Ips intervals. Note t h a t A45c is a c t u a t e d through A45b, being open when the output of A45b is at logic 1 ; this requires pin 5 of A45b to be at logic 0 - a c o n d i t i o n that is satisfied as discussed below. 8.3.4

S t a r t / S t o p and Enable C i r c u i t r y

The s t a r t signal, from the START input or from the manual s w i t c h , sets f l i p - f l o p A3. While the flip-flop is s e t , counting is e n a b l e d , e x c e p t for an initial 10;is dead time period set by the A24 monostable to reset the A51 six decade counter. To allow a reset to follow q u i c k l y , the set signal should be narrow; this is accomplished in the manual starting mode by keeping pin 2 of A13 only m o m e n t a r i l y at logic 0- only while R15 charges up C5. Through the output of

A 1 3 a , the start signal triggers the lOus

A24 monostable to clear the A51 c o u n t e r and to enable the parallel load of the A42 and A43 down counters. N o t e that a lOps dead time is thus i n t r o d u c e d in the c o u n t i n g time.

The o u t p u t of the A3 f l i p - f l o p is ANDed with the preset count signal (output of A6c) in A 1 4 d , and wired-OR through this gate with the e x t e r n a l ENABLE signal to control counting gate A45c through A45b. Note that the ENABLE port may be used as an o u t p u t driven by open collector gate A l 4 d . The s t a r t signal line at the o u t p u t of A14d is also used to keep the A35 and A34 time generator counters reset while the line is at logic 1 (counting not started); simultaneously, it also keeps the parallel load of the down counters enabled.

(160)

Page 150

-7-

Chapter 8

A2e and d form a falling edge triggered monostable that starts a counting cycle t h r o u g h A13a if D2 is moved to the Z position. This is necessary if one wants to allow counting to s t a r t at the falling edge of the external ENABLE signal. The stop signal, either from the manual switch or the preset count o u t p u t , resets the A3 flip-flop t h r o u g h the A15a g a t e ; t h u s counting is disabled. When pin 11 of A14 goes to logic 1, the 10s monostable implemented in one-half of A24 is a c t i v a t e d if the RECYCLE mode is s e l e c t e d . A f t e r the 10s d e l a y , the 10 us A24 monostable is t r i g g e r e d to clear the A51 sealer and to parallel load the A42 and A43 c o u n t e r s , while the A3 flip-flop is set through A46d and A 1 3 a . This keeps the counter recycling w i t h a 10s interval from the end of one counting period to the beginning of another. 8.3.5

The C o u n t i n g and P r e s e t C i r c u i t

The p r e s e t count value takes the form p q x l O r where r is set to a value in the range 0 to 6, and pq is a t w o - d i g i t number taking values from 0 to 99. This information is loaded in the down c o u n t e r s A42 (units) and A43 (tens). The power r is used to control the o u t p u t of the six d e c a d e c o u n t e r A 5 1 ; if r = 0, the o u t p u t has the same number of pulses as the i n p u t ; in general, the o u t p u t c o r r e s p o n d s to a division by 10 r. Pulses for the down counters come from the o u t p u t of the d e c a d e d i v i d e r A51 t h r o u g h g a t e A 4 1 d . This g a t e is enabled by the output of A 4 1 b while the p r e s e t count is not r e a c h e d . When the preset value is reached, the output of A41 becomes 1 and t h r o u g h A28d and A6 stops the counting process. C a p a c i t o r C3 f i l t e r s out e v e n t u a l n e g a t i v e spikes that may t r i g g e r the following c i r c u i t s . Gate A28d may be closed by A 3 1 , allowing the counter to be open i n d e f i n i t e l y ; A 3 1 , an 8-input NOR g a t e , closes A28d if the preset value to the down c o u n t e r s is zero.

8.4.

TROUBLSHOOTING EXAMPLE; C A N B E R R A MODEL 2071

LCD DISPLAY OF DUAL COUNTER/TIMER,

Here we just refer to the d i s p l a y section of this i n s t r u m e n t , the block diagram (channel A) of which is presented in Fig. 8.4. The 6-digit display can e i t h e r show the c o n t e n t s of decades D5 to DO, or of d e c a d e s D7 to D2, a c c o r d i n g to w h e t h e r the c o n t e n t s of the most s i g n i f i c a n t decades (D7, D6) is zero or non-zero; this is i m p l e m e n t e d by the scaling block. The c o n t e n t s of the various decades is transferred t h r o u g h a bus to the l a t c h - d e c o d e r ICs associated w i t h each d i g i t ; the latches are successively enabled by the strobe signals from A8, a counter that moves w i t h the same clock (A32) t h a t a c t u a t e s the scanning c i r c u i t s included in the LSI counter A54. The scan reset input of the LSI (pin 1) is used to assure the synchronism necessary for each digit to display the corresponding decade. The scan reset input signal can be made synchronous w i t h the signal strobe 1, in which case the display

(161)