The text of a manual · part 12 of 28
Troubleshooting Nuclear Instruments, pages 111 to 120
Pages 111 to 120 of 273
The original document is the PDF (12 MB). This is its text as extracted, unedited; tables and drawings read best in the PDF, and each page heading opens it at that page.
Filed in the catalog under the Camac Output register Driver CAMAC
Page 111
-15-
Chapter 6
Troubleshooting considerations similar to the previous can be applied to instruments or parts of them based operational amplifiers.
ones upon
An instrument may be designed to use several operational amplifiers. The search for a fault in the instrument requires a fault analysis for the individual operational amplifier circuits, which can be introduced with a q u i t e general approach by discussing the connections as shown in Fig 6.10.
Fig.
6.10: Basic operational amplifier configuration
Suppose t h a t the o u t p u t voltage Vo, which should be close to zero if this circuit is intended to be a linear a m p l i f i e r , is found instead to be close to the positive saturation voltage (+11 to + 14V, depending on the type of operational amplifier e m p l o y e d , for a + 1 5 V bias). Assume also t h a t visual inspection has been accurately carried out on the circuit and that n o t h i n g has come to your a t t e n t i o n . Follow the instructions suggested in the flow chart presented in Fig. 6.11.
(119)
Page 112
Chapter 6
-16-
[MEASURE v0[
IresTOKCg.^"*"^*"1"^ S^-vca^ff*,--* -
~^^
.^y
V ilOHWM. <*1U» FUjf
1
,
Fig. 6.11: Flowchart for troubleshooting a basic operational amplifier configuration
(120)
Page 113
Chapter 6
-176.1.4
Signal Analysis
Have you finished the dc analysis of the i n d i v i d u a l parts your i n s t r u m e n t ? Have you made sure that everywhere c o n d i t i o n s meet the expected values?
If so, you can quickly i m p l e m e n t s its basic function.
check of in the instrument the dc whether your instrument
For instance:
If the i n s t r u m e n t is a p r e a m p l i f i e r or an a m p l i f i e r , does a signal appear at the o u t p u t when you a p p l y a signal at the i n p u t , and are the shape and the a m p l i t u d e of the o u t p u t signal reasonable? If the instrument is a single channel analyzer, does the logic o u t p u t signal appear at the o u t p u t when you apply an input position ing signal, and slowly move the threshold p o t e n t i o m e t e r u n t i l t h e input signal falls into the channel? If the i n s t r u m e n t is a t i m e - t o - a m p l i t u d e converter, does a ramp signal appear at the o u t p u t when you apply a s ignal at the START input and a delayed one at the STOP input?
d.
If the i n s t r u m e n t is a linear gate, is the signal applied at the analog input t r a n s m i t t e d almost u n a l t e r e d to the output during those intervals when it overlaps in time with the g a t i n g command a p p l i e d at the LOGIC INPUT? Some s i t u a t i o n s are summarized in F i g . 6.12. For each i n s t r u m e n t there is an a p p r o x i m a t e indication of the c o r r e c t and of a p o s s i b l e d e f e c t i v e o u t p u t signal. If the signal is a c c e p t a b l e , you can go on w i t h a d e t a i l e d analysis and calibration of the instrument. If the signal is strange, it is necessary to carry on w i t h a signal analysis of the i n d i v i d u a l stages.
To apply signal analysis, you must send a signal to the input and then follow its path t h r o u g h o u t the i n s t r u m e n t , making sure t h a t every c i r c u i t performs on the signal the correct function. Signal analysis will tell you some very i m p o r t a n t facts. 1.
The signal is present at the input of a certain circuit and is no longer present at the output. The circuit under test has an i n t e r r u p t i o n on the signal path.
2.
The signal appears at the output of an amplifier clipped on the t o p , although the input was in the correct range. The circuit dynamic range is out of the specified value.
3.
The signal passing through a linear amplifier appears at the output smeared by an oscillation whose frequency exceeds 100 MHz. You can almost be sure that in the circuit there is an ill-neutralized transistor.
(121)
Page 114
Chapter 6
-18-
E - » n mi ti*f( j - - i, rwM e.tMt*»>. fVJl U«M •».« Mrr>. UM ik*
WVUTUM »•-» «W«
TO IT r «•»« n
*»mr «« UNmniM
\ NÏMM to TW« «*»»e
^S.
CW^Vctlv« totM**A"*
' \-^^ir£ M
^1*0^ vrut i caff
Fig,
(122)
6.12:
Test set-up for some typical nuclear electronic instruments w i t h acceptable and d e f e c t i v e o u t p u t s ignals
Page 115
-19-
Chapter 6
The signal passing through a linear feedback amplifier presents at the o u t p u t a d a m p e d oscillatory behaviour. Most p r o b a b l y the f e e d b a c k loop is not s u f f i c i e n t l y compensated. A linear feedback a m p l i f i e r , whose function is that of a m p l i f y i n g and not that of s h a p i n g , slows down excessively the signal. Your feedback loop is over-compensated.
NOTE :
To carry on signal analysis, you need a rectangular pulse generator and a dual channel oscilloscope with at least 100 MHz bandwidth. It must be e q u i p p e d with two Ix probes and two lOx probes. Do not use the Ix probes if this is not necessary to achieve a high s e n s i t i v i t y of your scope. For an easier observation on the scope, use an e x t e r n a l trigger by sending to the oscilloscope input the advanced trigger signal from the generator. Set on the generator a 200ns delay. Make sure that no false contacts exist in the cables, connectors and probes. Check the ground wires of the probes; make sure that they are safely soldered to their alligator-clips. Add in series with each probe a 300 resistor to avoid spurious e f f e c t s due to probe c a p a c i t a n c e . H a v i n g done t h i s , send the signal from the generator o u t p u t to the input of the instrument under test. Use the two channels of the scope to m o n i t o r , for each part or elementary c i r c u i t in your i n s t r u m e n t , input and o u t p u t signals. Ground the probes to the ground conductors that are closest to the p o i n t s you are going to monitor.
Some ideas about how to proceed with the signal analysis will now be discussed. They will refer to p a r t i c u l a r l y simple s i t u a t i o n s where the instrument is made only of a m p l i f y i n g p a r t s , all working in the linear range and not requiring any logic command. More c o m p l i c a t e d s i t u a t i o n s will be considered afterwards with s p e c i f i c cases. The analysis referred to in the linear case is, however, e x t r e m e l y useful in i n t r o d u c i n g the basic principles. Suppose that following the signal p a t h , you arrive at a circuit i l l u s t r a t e d in Fig. 6.13, w i t h signal shapes as indicated.
(123)
Page 116
Chapter 6
-20-
J"
Fig.
6.13
AC-coupled operational amplifier with open signal path
If the circuit had previously undergone dc analysis and was found to operate correctly de-wise, then the present f a u l t , that is, open signal path between input and o u t p u t , may depend on an open C capacitor or on an interrupted connection in series with C on the printed-circuit board.
ATTENTION:
Such a fault cannot be analys is of DC cond it ions.
detected from the
Another frequent source of fault is a shunting capacitor which , because of a mistake made in a previous repair, is much larger than it should be. Such a situation is illustrated in Fig. appe as it it is 6.14. The signal actually does nott v"appear to the oouuttppuutt,, as short-circuited to ground by the lOOnF capacitor connected by mistake. The collée tor-to-ground capacitor of large value interrupts the signal path again. In this case DC analysis would not reveal the fault.
(124)
Page 117
Chapter 6
-21-
DUTPUT lOOpF NOMINAL lOOnF ACTUAL INPUT
-E
Fig.
6.14:
Transistor stage w i t h d e f e c t i v e o u t p u t signal
Consider now the following two-1ransis t or operational a m p l i f i e r (Fig. 6.15) and look at the o u t p u t signal (a). Suppose that the oscillation frequency exceeds 100 MHz. Such a limit is purely e m p i r i c a l , but it h e l p s you to guess w h e t h e r an oscillation has to be a t t r i b u t e d to a component or to the f e e d b a c k loop. The transistor T2 appears to be unneutralized. Most likely the oscillation is due to T2. To damp it, add a 100 ohm t r a n s i s t o r in at T2 , and as close as possible to series w i t h the c o l l e c t o r lead its can .
Fig.
6.15:
Two-transistor operational amplifier with signal oscillator
If the o u t p u t signal looks like that of case (b), with the frequency of the damped oscillation of 50 MHz or less, most likely
(125)
Page 118
-22-
Chapter 6 the o s c i l l a t i o n is d e t e r m i n e d by the feedback loop. Add then a small c o m p e n s a t i n g c a p a c i t o r , 5pF to 50pF, between the c o l l e c t o r of Tl and g r o u n d . The d a m p e d o s c i l l a t i o n will d i s a p p e a r . Look now at the a m p l i f i e r c i r c u i t of Fig. 6.16. Suppose you find t h a t its signal gain differs considerably from the nominal v a l u e of 10. Such a d e f e c t may d e p e n d on a change t h a t o c c u r r e d in one or b o t h r e s i s t o r s because of some kind of a c c i d e n t or on the fact t h a t for some reason the o p e r a t i o n a l a m p l i f i e r is completely out of s p e c i f i c a t i o n s as far as its signal gain is concerned. In t h i s case the r e s i s t o r s do not draw c u r r e n t in the standing state. Turn off p o w e r , d i s c o n n e c t t h e m from the c i r c u i t and measure their values w i t h an o h m m e t e r . If the measured values are close to the nominal ones, proceed to replace the o p e r a t i o n a l amplifier. The d e s c r i b e d f a u l t o b v i o u s l y e s c a p e d d c analysis.
NGMINAL VALUES
Fig. 6.16: O p e r a t i o n a l a m p l i f i e r with a gain of 10
Consider now the following a m p l i f i e r , whose nominal gain should be 10, but actually is found to be 1 (Fig. 6.17). The fault may d e p e n d on Rl o p e n - c i r c u i t e d or R2 s h o r t - c i r c u i t e d . Open and s h o r t c i r c u i t may be r e l a t e d to faults on the p r i n t e d circuit b o a r d . In this case visual i n s p e c t i o n aiming at d e t e c t i n g possible interruptions, accidentally disconnected c o m p o n e n t s , or stray jumpers has to be carried out f i r s t . If nothing comes out of this i n s p e c t i o n , measure the resistors Rl and R2 and change the d e f e c t i v e one(s). Remember that t h e r e is still the possibility of a gain in the operational a m p l i f i e r much below the nominal v a l u e , in which case it has to be replaced.
(126)
Page 119
Chapter 6
-23- lOnF
ÜUTPUT_]
NOMINAL VALUES
Fig. 6.17:
6.2
AC-connected operational amplifier with a gain of 10
TROUBLESHOOTING OF SPECIFIC EQUIPMENT
6.2.1
Troubleshooting of a Charge-Sensitive Preamplifier
Existing commercial charge-sensitive preamplifiers fall four c a t e g o r i e s .
(i)
Resistive feedback preamplifiers:
(ii)
Resistive feedback preamplifiers:
(iii)
O p t i c a l feedback preamplifiers:
into with input FET operating at room temperature with input FET cooled with input FET operating at room temperature (iv)
O p t i c a l feedback preamplifiers:
with input FET cooled Each category requires specific troubleshooting procedures. As an e x a m p l e , a preamplifier with resistive feedback and input FET o p e r a t i n g at room t e m p e r a t u r e is analyzed. 6.2.1.1
General considerations
A t y p i c a l block diagram of such a preamplifier (PA) is given in Fig. 6.18, where the shielding box, the connectors, the d e t e c t o r b i a s i n g and the test networks are also i n d i c a t e d .
(127)
Page 120
Chapter 6
-24-
DETECTOR BIAS
AMPLIFIER
DETECTDR
INPUT
POLE-ZERO
ADJUST NETWORK
OUTPUT TEST INPUT
Fig.
6.18: C h a r g e - s e n s i t i v e p r e a m p l i f i e r
The basic b l o c k s are the c h a r g e - s e n s i t i v e l o o p , the pole-zero a d j u s t m e n t n e t w o r k , and the o u t p u t amplifier. The c h a r g e - s e n s i t i v e loop d e t e r m i n e s , to a major e x t e n t , the noise p e r f o r m a n c e of the preamplifier. The f u n d a m e n t a l c o n c e p t s a b o u t these blocks can be found in document IAEA TECDOC-363, SELECTED TOPICS IN NUCLEAR E L E C T R O N I C S , pages 72 t h r o u g h 77, and in the d o c u m e n t IAEA TECDOC-309, N U C L E A R E L E C T R O N I C S L A B O R A T O R Y M A N U A L , p a g e s 77 t h r o u g h 83. The d r a w i n g of Fig. 6.16 also shows, i n s i d e the shielding box, the d e t e c t o r b i a s network consisting of (RICH) low-pass f i l t e r , and d e c o u p l i n g r e s i s t o r RB , the i s o l a t i n g c a p a c i t o r Ci, and the test network consisting of terminating resistor Ro and injection c a p a c i t a n c e Cinj. the
If you have such a t y p e of p r e a m p l i f i e r to t r o u b l e s h o o t , check first whether or not the o u t p u t lies w i t h i n its linear range, w i t h o u t even o p e n i n g the PA b o x . For such a purpose:
- c o n n e c t the PA o u t p u t to a scope; - put the scope on the position of h i g h e s t v e r t i c a l s e n s i t i v i t y , p o s s i b l y 2 to 5 m V / c m ; - use ac c o u p l i n g in the scope;
- t r i g g e r the scope on AUTOMATIC;
- turn on power on PA and look at the o u t p u t baseline. If the o u t p u t baseline looks to be smeared by noise (Fig. 6.19a), t h e n your PA is likely to be in a correct dc condition.
(128)