Page 81

Chapter 5 5.4

-10PLUG-IN NIM POWER SUPPLY

As already mentioned, there are NIM-plug-in power supply modules available to supply a NIM bin. Such a supply module is described below, and shown in Fig. 5.3. This module has some l i m i t a t i o n s :

- it is available only for + 1 2 V / 1 A , +24V/0.5A, - 1 2 / l A , -24V/0.5A; - its maximum power capability is 48W at 50C.

DO NOT plug in sue h a module in a NIM crate that already has a powe r supply in t he back .

1 NOTE: | 1

5.4.1

Circuit Description

The power transformer provides four separate low-voltage ac sources. These sources are full-wave-rectified, capacitorf i l t e r e d , and regulated by electronic series regulator circuits, which have the capability of r e g u l a t i o n and current l i m i t a t i o n w i t h foldback characteristics. Each supply voltage has its own reference source and the p o s s i b i l i t y to adjust the output v o l t a g e by a p o t e n t i o m e t e r .

All four supplies use the same active components; however, the voltage d i v i d e r s are tuned to the requirements of the o u t p u t voltage. Two d i f f e r e n t preregulators are used: one consists of the voltage doubler (Dl, D2, Cl, C2, na d zener diode D4) for the +12V and

+ 24V

-24V supply.

supply.

The second uses a 10V zener d i o d e for -12V

The reference voltages of and

6.95V -are derived from a reference diode (Ul, U3, U5, U7).

The regulators are of the type LAS1000 or equivalent. As a pass transistor a npn-Darlington transistor is used (type TIP100 or equivalent) . A reverse-current protection is also built in by the reversed biased diode from the type 1N5819 or equivalent. Following the above observations, troubleshoot the instrument. Nevertheless, useful.

the following it additional should be easy to comments might be

There are two shock-hazard locations to watch for: the wiring side of the input line cord connector block, and the two thermal switches (SI for thermal cutout, S2 for thermal warning) mounted against the heatsink. These locations are exposed to the ac-line voltage .

(88)

Page 82

-11-

Chapter 5

In Table 5.1, the t y p i c a l de-voltages (measured with respect to TP1 or ground.), are g i v e n , w h i c h should be very h e l p f u l for troubleshooting. The pin assignments for the NIM connector are presented in Table 5.2.

5 .5

MCA POWER SUPPLY

From the c i r c u i t diagram (Fig. 5.4), it is d i f f i c u l t to realize t h a t the -»-24V line is the d o m i n a n t supply v o l t a g e . From t h i s v o l t a g e all others are derived by using the 24V e i t h e r as r e f e r e n c e v o l t a g e or by s u p p l y i n g it to the o p e r a t i o n a l a m p l i f i e r s w h i c h are used to g e n e r a t e the v a r i o u s v o l t a g e s .

This 24V line is c o n t r o l l e d by a o v e r h e a t i n g sensor switch and a p p e a r s only if the t e m p e r a t u r e of the h e a t s i n k at the b a c k - p a n e l is below the c r i t i c a l value. In the 24V line r e g u l a t i o n c i r c u i t , there is a zener d i o d e . It is used to p r o v i d e the r e g u l a t e d 24V as the r e f e r e n c e v o l t a g e for the o p e r a t i o n a l a m p l i f i e r . This c i r c u i t r e d u c e s the influence of the line v a r i a t i o n s to the o u t p u t v o l t a g e . At all other supply v o l t a g e s , the r e g u l a t i o n is implemented in a c l a s s i c a l way using operational amplifier and current limitation with foldback characteristic. Only the 5V supply uses two parallel power t r a n s i s t o r s because a h i g h c u r r e n t has to be d e l i v e r e d .

In a d d i t i o n , it is necessary to p r o t e c t the d i g i t a l ICs from overvoltage: an o v e r v o l t a g e p r o t e c t i o n circuit is i n t r o d u c e d . After power is s w i t c h e d on, 5 LEDs indicate the following v o l t a g e s : + 5V, + 1 2 V , +24V, - 1 2 V , -24V. This is very c o n v e n i e n t because it is a l r e a d y an indication as to where and how to start with troubleshooting. If no LED is on, and the main fuse ( F l ü l ) is not b l o w n , the fault may be in the + 2 4 V supply line. Check the t e m p e r a t u r e s w i t c h ; if it is c l o s e d , then measure the reference v o l t a g e 018. N e x t measure the o u t p u t of the o p e r a t i o n a l a m p l i f i e r A3, w h i c h s h o u l d be a p p r o x i m a t e l y 1 5 V . The base v o l t a g e of Q2 compared to ground should be a l i t t l e higher than +25V. If n o t , check the v o l t a g e across Q3.

If this v o l t a g e is less t h a n 1 . 5 V , it indicates that there is an overload at this o u t p u t and current l i m i t a t i o n is a c t i v e . This is valid for all o t h e r o u t p u t v o l t a g e s , except for the +5V. There, it is not easy to d i s t i n g u i s h b e t w e e n an overload s i t u a t i o n and the presence of the a c t i v a t e d o v e r v o l t a g e protection. Due to the overvoltage protection, the thyristor Q19 is f i r e d and shorts the o u t p u t to g r o u n d . As long as the current l i m i t a t i o n works p r o p e r l y , there is only one way to reset the c u r r e n t l i m i t a t i o n : by switching the i n s t r u m e n t off and on again, after removing the load. Only in this way is it possible to determine whether current l i m i t a t i o n or o v e r v o l t a g e causes the action. If you suspect that the o v e r v o l t a g e p r o t e c t i o n has become a c t i v e , you should switch off the i n s t r u m e n t s , and connect it via a VARIAC to the m a i n s , as shown in Fig. 5.5. Increase slowly the voltage to the i n s t r u m e n t . In this way it is possible to d i s t i n g u i s h if there was a bad adjustment of the overvoltage p r o t e c t i o n or if a transient occurred during the s w i t c h i n g on of the ins t rumen t.

(89)

Page 83

Chapter 5

TABLE

5.1:

-12-

Typical üC-Voltages wit hou t load ) Nod«

Volt*»»

Nod«

VolUffi

U2 pin 8 3 5 7 2 1

1-36.5 + 2.5 0 0 + 2.5 +24.0 +23.9 T25.3

U6 pin 8 3 5 7 2 1

+ 9.5 - 9.5 -12.0 -12.0 - 9.5 0 - 0.2

US pin 8

7

-36.5 * 2.5 0 0

2 1

+ 2.5 +12.0

2 1 10 6

10 6

U4 pin 8 3 5

10

6

TABLE 5 . 2 :

Pin

(measured with respect

10 6 3 5 7

+11.8 -M3.5 assignment, for

+ 1.4 + 9.5 -21.5 -24.0 -24.0 -21.5 0 - 0.2 + 1.3

NIM modules

BIN/MODULE CONNECTOR PIN ASSIGNMENTS FOR AEC STANDARD NUCLEAR INSTRUMENT MODULES PER TID-20893 (Rev 4) (adopted by DOE) Pin 1

2 3 4

5 6 7

8

9 •10 •11 12 13 14 15 •16 •17 18 19 20

21 22

(90)

Function

-3 volts -3 volts Spare Bus Reserved Bus Coaxial Coaxial Coaxial 200 volts dc Spare -6 volts -6 volts Reserved Bus Spare Spare Reserved -12 volts -12 volts Spare Bus Reserved Bus Spare Spare Reserved

Pin

23 24 25 26 27 '28 '29 30 31 32 '33 "34 "35 "36 "37 38 39 40 '41 '42 G

Function

Reserved Reserved Reserved Spare Spare *24 volts -24 volts Spare Bus Spare Spare 117 volts ac (Hot) Power Return Ground Reset (Sealer) Gate Reset (Auxiliary) Coaxial Coaxial Coaxial 117 volts ac (Neut ) High Quality Ground Ground Guide Pm to

TP1

Page 84 read from a scan by OCR; it may hold reading errors

(91)

Fig.

5.4:

Low Voltage Power Supply of a MCA

=> sey Patame saves olay PREAMP | niav pmcane

«lal: a °f27=] a]

TT

= Jewlooe

REV (UPDATED

(rels<

Sy REND

Ltd See D

4] ry

RSI FivisidD ax Chk!

F pete

Reviseo 229

-20- Ht

REWSFO AR INOO be Bt

- 2a CRU

REVISED- ARN227 jer {5 ws oo fm

‘Is.

Tey t

L u

“TL 8 &

See ee -d

SEE ECN zemalut 5-21 61 REviSED ARN284 2 es 10 37-8) REVISED aR iz? 1-284 REMED AR NIS37 _ [reerlaw [ra-4? AR NSEG ase)

o2uay sol zr tes 33 iv See BEIiat | zP jiz-¥-38

PREAMP cao j MVPS CMD

T=T « alalelel « i

88 z r(x]

= il< jJo}ulul a] >] x] al ste o

L

*] ty

MOTES: I. THERE ARE WO 4 WAY TIES ON THIS SCHEMATIC

ZLUNLESS MOTED ALL RESISTORS AGE Yaw, 257, SUNLESS NOTED ALL CAMCITORS ARE I pi

4 UNLESS NOTED ALL OIODES ARE 1nN4003. $.@ INDICATES ANGOC RESISTORS.

6. Zed MmOICATES WIRE BOINT ON BOTH SIDES -"* OF Re BOARD.

“Ge . SAME HCATSINK,

Tun t mR 2000 302 30v

‘te % cs

(Lc Re FERN] t Qn ®, g 3 & a 2 At 6 zt eB LW &,

NO] TYPE LC all uate Al] Leer AST LP PaL anal uret Ast une AG L822 ATI LM 7a:

Ad? eke

LAST COMPONENT ACF WO

RESISTORS R72 CAPRLITORS «= C23 Oro0E os TRANSISTORS 921 WR RES we TEST POINT) = TPS Ke ay WIPE PONT wre Leos LEO5.

SCHEMATIC LVPS CANBERRA

POODLL 4200

-13-

Chapter §

Page 85

Chapter 5

-14- variac ac-line

MCA line input

Fig 5.5:

MCA

Using a variac to i d e n t i f y a fault

The arrangement as shown in Fig 5.5 can be used also for t e s t i n g in the case when the fuse (F101) is blown i m m e d i a t e l y when the instrument is switched on. This i n d i c a t e s t h a t there is a short c i r c u i t somewhere in the power s u p p l y . Before you s t a r t to look. into the c i r c u i t in d e t a i l , remove the whole load (take out all b o a r d s , disconnect the display unit). Slowly increase the output voltage of the v a r i a c and observe the c u r r e n t w i t h an a m m e t e r (see Fig 5.5). When the a m m e t e r i n d i c a t e s roughly 60-80% of the nominal fuse c u r r e n t , stop increasing the v o l t a g e . Now measure the voltage at the o u p u t of the variac. Here you will already have some i n d i c a t i o n as to where the short circuit might appear.

(i)

Less than 4% of the ac line voltage i n d i c a t e s d e f e c t ive line f i l t e r network, or a f a u l t y transformer.

(ii)

A b o u t 4% of line voltage can i n d i c a t e a bad transformer or a bad rectifier. S e p a r a t i o n of t r a n s f o r m e r and r e c t i f i e r diodes or b r i d g e s has to be m a d e to d e f i n e the d e f e c t i v e p a r t of the s u p p l y .

(iii)

6-8% of line v o l t a g e i n d i c a t e s a s h o r t c i r c u i t either in the rectifier p a r t or a defective filter c a p a c i t o r ; by s e p a r a t i o n of these two p a r t s of the power s u p p l y , it is p o s s i b l e t o d e t e c t t h e d e f e c t i v e component.

(iv)

8-10% of the line v o l t a g e the pas s-1ransis t or , and working p r o p e r l y , t h e r e f o r e i n d i c a t e s a short c i r c u i t a f t e r the c u r r e n t limitation is not all o u t p u t v o l t a g e s have to be m e a s u r e d t o locate t h e d e f e c t i v e c o m p o n e n t . The s e m i c o n d u c t o r s must be checked in the c o n v e n t i o n a l way. In measuring base-emitter voltage (around 0.7V), c o l l e c t o r - e m i t t e r v o l t a g e m u s t not be less t h a n 0.7V otherwise the transistor would o p e r a t e in the s a t u r a t i o n r e g i o n and would no longer be linear. If these v o l t a g e s are c o m p l e t e l y d i f f e r e n t , as previously m e n t i o n e d , remove the t r a n s i s t o r s and check it again. To check the o p e r a t i o n a l a m p l i f i e r , its two input v o l t a g e s should have the same value. If t h e y do n o t , compare them w i t h the o u t p u t . If t h e r e is a discrepancy, remove this operational amplifier.

(92)

Page 86

-15-

Chapter 5

After replacing the d e f e c t i v e c o m p o n e n t s , check the power supply again. Increase the o u t p u t voltage of the v a r i a c to the nominal mains v o l t a g e , but also measure the ac current used by the power supply. This current should not be more than 20% of the fuse current since there is no load. Readjust all supply voltages according to the circuit diagram or service manual.

A f t e r this p r o c e d u r e , the o u t p u t voltage of the +5V line follows when the o u t p u t v o l t a g e of the +24V line is increased. Simultaneously measure both o u t p u t v o l t a g e s . When the +5V line reaches around 5.6V, the o v e r v o l t a g e p r o t e c t i o n should be activated. In this case, decrease the o u t p u t v o l t a g e of the +24V line to the nominal value, switch off the instrument to reset the memorized o v e r v o l t a g e , and switch the instrument on again. Then measure the o u t p u t voltages to be sure that everything is p r o p e r l y set. If the overvoltage p r o t e c t i o n did not t r i g g e r , try to detect the d e f e c t i v e components (D13, Q20, Q190, or the passive components around them). The c o m p o n e n t s 1C A5, Q 1 5 , and Q16 are there for power reset, which is necessary for the microprocessor reset. 1C A6 generates a fixed clock f r e q u e n c y .

The current limitation can now be checked. For this p u r p o s e , an ammeter in series with a variable power resistor has to be connected to the o u t p u t , and simultaneously the o u t p u t v o l t a g e has to be measured. Increase the current by changing the resistor. When exceeding 100% of the nominal current, the o u t p u t v o l t a g e should decrease. After removing the load, the nominal o u t p u t voltage should come back again. A f t e r c o m p l e t i n g this p r o c e d u r e , you can i n s t a l l all removed boards and the display unit. Make sure the MCA is now working. If not see Chapter 9, MULTICHANNEL ANALYZER.

5.6

DETECTOR BIAS SUPPLIES

In this section, bias supplies for^detec t ors are described. can d i s t i n g u i s h between several types of such supplies. (i)

High voltage supplies which are designed for d e t e c t o r s must be very stable in scintillation respect to o u t p u t v o l t a g e , and relatively powerful (up to l-2mA at a b o u t 2000V); and r i p p l e is not so c r i t i c a l .

(ii)

We noise

High v o l t a g e s u p p l i e s w h i c h are designed for solid s t a t e d e t e c t o r s should have low noise and r i p p l e , but are not r e q u i r e d to be highly stable with r e s p e c t to o u t p u t v o l t a g e . The current r e q u i r e d for d e t e c t o r s such as Si(Li), p u r e Ge , or Ge(Li), is in the order of less than lOOnA. Surface barrier d e t e c t o r s sometimes require a higher current of to 2uA (at relatively low voltage of about 100V).

up

(93)

Page 87

Chapter 5

-16-

For troubleshooting, a high-voltage instrument is recommended. This is

NOTE:

- either a digital v o l t m e t e r with lOMohm input resistance using a HV-probe with built-in v o l t a g e divider 1000:1 or 100:1, or - s t a t i c voltmeter

5.6.1

Example 1; ORTEC Model 459

Fig. 5.6 shows a solid state d e t e c t o r . This

HV-supply circuit diagram of a HV-bias supply for a can deliver up to lOOuA and the circuit is simple to understand.

NOTE ;

NEVER measure with a digital voltmeter in the HV-path unless you use a HV-probe with a voltage divider 1000:1 (such a high voltage probe is available from Fluke) or 100:1. Through such a probe the input resistance of a d i g i t a l voltmeter is 10000 MOhms. A n o t h e r possibility is to use a s t a t i c v o l t m e t e r , but pay a t t e n t i o n to HV isolation problems.

A HV-bias supply is usually built around a DC/DC c o n v e r t e r . An oscillator, push-pull c i r c u i t r y , a transformer and a regulation c i r c u i t are necessary on the primary s i d e ; on the secondary s i d e , the AC is r e c t i f i e d and f i l t e r e d or a v o l t a g e m u l t i p l i e r may be used . Before you start troubleshooting, make sure t h a t the working area where the HV supply will be tested is free of all conducting material, such as wire ends, e t c . , to prevent HV-shock hazards.

5.6.1.1

Troubleshooting

The loop .

specific HV-bias supply described here has no feedback

Be sure that all required input supply voltages are provided. Using an oscilloscope, check that the oscillator (Ql, Q2) is working. Frequency can be changed wit h potent iome ter R42. If n o t , check Ql and Q2. The next step is to inspect the operation by measuring with the voltmeter the emit ter-vol tage of transistor Q7 Increasing the while changing the front panel potent iometer R23 setting of the potentiometer should induce the voltage change of If the emitter-voltage of transistor Q7 (from -24V up to -t-lOV)

(94)

Page 88 read from a scan by OCR; it may hold reading errors

(95)

F

:

ig.

5.6

HV-bias supply (ORTEC Model 459)

Of Le.

“INaie3 | _ 22, o +24V¥ a RI et AeK x Lye

MPSOf31

MPS6431

R6 R7 R8 1&0 270 270

| RS mT 4 aw

BA [2 LED!

i Hy ON (ee) C2 x Iz Bett) PIPS@SSi iInatSS a3

MPS6°524 ot x .

03

IN?74A

| ar? | R24 MFCESS

Qs MyE3055 ki?

+ $2465 2 - ry) ed 180 REmMcrTEe vom RIB

24K 66 T ce 1+

LCie

MIGH VOLTAGE Adu £28 ~ - 5Kv

+ { w Ww

Ol R23 shes Sev |

1T

Rss} oe MOSESSI

20K T 93 = CRP)

loutpur| Y A! YOLTALE!

24k 6 . MJE3055 24a -24qv | 10K _2

LAST| NOT NO pced

Rso leas Co

HIGH VOLTAGE cs ee $2B (-) o1 01

7) L ¢ [ee TI M2278A cis ci? CIs 2a 0

04 06 D8 \ o1 05 07 09 x

202

1 Rx . oo VOLTAGE 27K IN41S3, ran R37

2 2 x a ze a Pp , a,

NOTES. (UNLESS OTHERWISE a SPEC IFIFD) I. RESISTORS MARKED ® ARE ik, Vow MF 2 CAPACITORS ARE IN wf, 3 04-017 ARE INQ44I, 4 RESISTCRS AREI/aw, 5% CC. 5 CAPACITORS MARKED & ARE MICA CIPPED 6. RESISTORS MARKEDAMARE 1%,'/4w, MF c=

é

2: t fs Fiziatol

S¢ei@

Reb*” Rese? R208? | at) 909K >

- MH CORMPORATEO

MIDLAND ROAD OAK RIDGE TENNESSEE 378630

MOLE

PEL or KY eT

-l7-

Chapter 5

Page 89

Chapter 5

-18- this is not o b s e r v e d , remove the load by unsoldering capacitor C13. Check again.

If there is no voltage at the o u t p u t , check the current l i m i t i n g c i r c u i t r y which consists of zener diode D3 and transistor Q9; it works like a shutdown device by driving the transistor Q10 into saturation. When e v e r y t h i n g seems normal, check w i t h the oscilloscope the waveform of the signal which appears at collector of transistors Q5 , Q6. If there is no rectangular pulse visible (the amplitude of the pulse should be d e p e n d e n t on the e m i t t e r - v o l t a g e of transistor Q7 ) check transistors Q3 to Q6 or measure with an ohmmeter if the transformer winding has an i n t e r r u p t i o n . When the t r a n s f o r m e r has a shorted w i n d i n g , this a p p e a r s as an overload, and c u r r e n t l i m i t a t i o n should take place. In t h i s case, transformer Tl has to be replaced. A f t e r inserting capacitor C13 and you observe t h a t an overload appears a g a i n , a d e f e c t i v e component must either be in the filter capaci t or' C27 to C29 or in the voltage m u l t i p l i e r . Remove one a f t e r a n o t h e r c a p a c i t o r s C27, C28 or C29, and check again. If no d e f e c t i v e c o m p o n e n t was found, the next step is the inspection of the voltage multiplier. When t r o u b l e s h o o t i n g a v o l t a g e m u l t i p l i e r , work carefully! Measure the voltage of each m u l t i p l i e r stage e i t h e r w i t h a HV-probe or with a s t a t i c - v o l t m e t e r . If a stage has the same voltage level as the previous one, it is d e f e c t i v e (either the diode or the capacitor). Replace the d e f e c t i v e component and check again. When e v e r y t h i n g seems to be working properly, you can start to recalibrate the instrument.

Set the front panel dial to 5.00, which should correspond to 5000V, and measure the o u t p u t voltage. W i t h two p o t e n t i o m e t e r s R22 and R 1 5 , the voltage can be adjusted: R22 is responsible for the s e t t i n g of the 5000V level, R15 is used to adjust the low voltage region. Check the linearity of the dial s e t t i n g by measuring the o u t p u t v o l t a g e ; do this slowly so that the capacitors can follow by increasing t h e i r charge. W i t h p o t e n t i o m e t e r R42 you can change the oscillator f r e q u e n c y , which is necessary to minimize the current consumption of the + 24V and -24V supply line. Most of the faults in the earlier version of this HV-supply were the consequence of burning out of the polarity indication lamps. N o w a d a y s , these lamps are replaced by LEDs. Also, there were sometimes problems with the tantalum capacitors due to the spikes w h i c h occur from switching an inductor on and o f f . 5.6.2

Example 2; Canberra Model 3102

Another type of HV-bias supply is the rather complicated one from Canberra, Model 3102. This circuit has a feedback circuitry to stabilize the HV-bias voltage. Canberra even uses special ICs, which are not very common on the market. Fig. 5.7 shows the circuit of the above-mentioned bias supply. It is controlled and regulated by a feedback loop and the ratings are up to 2mA at 2000V.

(96)

Page 90

-19-

Chapter 5

NOTE : This is an example of a case when you cannot trust the circuit diagram. There are some changes in the c i r c u i t , which are not shown in the circuit d i a g r a m , e.g. oscillator swing is from -10V to 0V and due to the slow rate of the used operational amplifier it has a trapezoidal waveform. The feedback resistor network R107, R i l l is connected to -12V and not to ground. The positive supply of the operational amplifier is connected to ground and not as it is shown in the c i r c u i t diagram to +12V. Therefore, the oscillator operates without the module being switched on. Also, the component number IC104 does not correspond to the actual lay-out, which is in this case IC101. There might be other inconsistencies.

5.6.2.1

Circuit description

ATTENTION:

Circuit operation is described according to the diagram, and does not include the modifications later introduced by the manufacturer.

Zener diode CR101 and operational amplifier IC104A generate the -9V reference voltage. The oscillator is built around IC104B. The driver stage, built around IC105, is coupled through C108 and C107 to the power transistors Q101 and Q102 of the push-pull stage. The d r i v e r stage is controlled via capacitor C104 by the oscillator and de-controlled via the amplifier bias current input by operational a m p l i f i e r IC102B. This operational amplifier is itself controlled by the voltage setting operational amplifier IC103A. The output to the voltage monitor is taken from amplifier IC101. The over-voltage detection is performed by a summing amplifier (1C 103B, 1/2 of CA3240) and operates in the same way as the shut-down function of the inhibit gate. 5.6.2.2

NOTE:

Troubleshoot ing

Use the HV-probe to measure the output voltage. Such a probe should have a built-in 1000:1 or 100:1 voltage divider. Be careful when measuring high voltage.

(97)