The text of a manual · part 8 of 18

8840A Digital Multimeter Instruction Manual, pages 71 to 80

Pages 71 to 80 of 180

The original document is the PDF (42 MB). This is its text as extracted, unedited; tables and drawings read best in the PDF, and each page heading opens it at that page.

For the Fluke 8840A

Page 71 read from a scan by OCR; it may hold reading errors

Remote Programming EXAMPLE PROGRAMS

/* The following application program is written in C for the IBM PC-AT. The National Instruments Model AT-GPIB board provides the interface between the PC and the Fluke 8840A DMM. The program assumes that the configuration program IBCONF has been run to initialize the interface board with the device name I18840A assigned to the GPIBO board.

This program illustrates one possible use of the serial poll register. In this case it is merely looking for data available. The function of the program is to display on the screen the lowest resistance value measured on the input terminals of the 8840A using the 2-wire ohms function in autorange, The range and function commands are programmed using the Put Instrument Configuration (PO) command.

Microsoft C Version 6.0 on IBM PC-AT */ /* Link this program with appropriate mcib*.obj. : */

#include <stdio.h>

#include “decl.h"

#include <string.h> #include <math.h> char rd{512]; /* read data buffer */ int brd0; /* interface board number: */ int dmm; /* device number */ char spr; /* serial poll response byte */ float r,s;

char rd_string[11],ss[11];

main( )

{ brdO = ibfind( "GPIBO" ); /* initialize interface board */ dmm = ibfind( "I8840A" ); /* initialize 8840A */ ibcir( dmm ); /* clear device */ ibwrt( dmm, "N3001PO0Y1?", 10); /* write functions to instrument */ ibrd(dmm,rd,16); /* get data */ strnepy(rd_string,rd,11); r=atof(rd_string); /* convert from string to floating point */ labell: printf("%sS (2-wire) LOWEST READING\r\n",rd); :

do { ibwrt(dmm,"?",1); /* trigger ‘the device */ do ibrsp(dmm,spr); /* serial poll the device - */ while (spr & 0x40 != 0x40); _/* until data available */ ibrd(dmm,rd,16); /* get next data */ s=atof(rd); /* convert to floating point */ printf("\r");

} while (s>=r); /* throw away data if not lowest */ r=s; | goto label1; /* print new low */

3-36 Figure 3-14. Example Programs: Using the IBM PC° (cont).

Page 72 read from a scan by OCR; it may hold reading errors

Remote Progiemniing EXAMPLE ‘PROGRAMS

/*

/*

The following application program is written in C for the IBM PC-AT. The National Instruments

Model AT-GPIB board provides the interface between the PC and the Fluke 8840A DMM. The program assumes that the configuration program IBCONF has been run to initialize the interface board with the device name 18840A assigned to the GPIBO board.

This program illustrates a method of recording any errors produced by the 8840A selftest function.

It should be noted that: 1. If more than one digital error occurs, only the last one will be reported.

2. The response to a Get Instrument Config. (GO) command during selftest is "Sxxx". 3. The response to a Get Error Status (G7) command with no errors present is "1000".

Microsoft C Version 6.0 on IBM PC-AT Link this program with appropriate mcib*.obj. #include <stdio.h>

#include "decl.h" #include <string.h> char rd[512}; /* read data buffer int brd0; /* interface board number int dmm; /* device number char selftest_active[1}; /* selftest active flag char errcode[4]; /* error code buffer main( ) { brdO = ibfind("GPIBO"); /* initialize interface board dmm = ibfind(18840A"); /* initialize 8840A ibclr(dmm); /* clear device | ibwrt(dmm,"Z0",2); /* start 8840A self test do{ /* do while self test active ibwrt(dmm,"G7",2); /* get error status ibrd(dmm,rd,16); strncpy(erreode,rd,4); if (stremp(errcode,"1000")!=0) /* test for error. {ibwrt(dmm,"X0",2); printf("Error %s occurred\r\n",errcode); ibwrt(dmm,"G0",2); /* get instrument configuration ibrd(dmm,rd,16); strncpy(selftest_active,rd,1); } while (stremp (selftest_active,"9")==0); _/* check for self test active ibwrt(dmm,"G7",2); /* get error status ibrd(dmm,rd,16); sumncpy(errcode,rd,4); if (srcmp(errcode,"1000")!=0) /* test for last digital error printf("Error %s occurred\r\n",errcode); | printf("\r\nSelftest complete\r\n’); }

*/

*/

*/ */ */ */ */

*/ */ */ */ */ */

*/

*/

*/ */

*/

Figure 3-14. Example Programs: Using the IBM PC® (cont)

3-37

Page 73 read from a scan by OCR; it may hold reading errors

Remote Programming EXAMPLE. PROGRAMS ____Table 3-4. ASCIIEEE Std 488-1978 Bus Codes “Asti BINARY | DEV. | MESSAGE | | BINARY A DECIMAL | OCTAL| HEX ASCII DEV.| MESSAGE CHAR. 7654 3210 | NO. | ATW=TRUE cuan, [DECMAL) OCTAL | HEX) oo arg | wo.| ATA. TRUE NUL 0 000 | 00 0000 (a 64 | 100 | 40 | 01000000 | 0 | MTA SQH 1 001 | 01 } 0000 0001 GTL A (65 | 101 | 41 | 01000001 | 1 | MTA STX 2 | 002 | 02 | 00000010 B | 66 | 102 | 42 | 01000010 | 2 | MTA ETX 3 {| 003 | 03} 00000011 - C | 67 | 103 | 43 | 01000011 | 3 | MTA RL EOT 4 | 004 } 04 | 0000-0100 soc | z D 68 | 104 } 44 | 01000100 | 4 | MTA ENO 5 | 005 | 05 | 0000 0101 pec | = E 69 | 105 | 45 | 01000101 | 5 | MTA ACK 6 | 006 | 06 | 0000 0110 = F 70 | 106 | 46 | 01000110 | 6 | MTA BELL 7 | 007 | 07 | 00000111 8 6 71 | 107 | 47-| 01000111 | 7] MTA BS 8 | 010 | 08 | 0000 1000 GeT | & H | 72 | 110 | 48 | 01001000 | 8 | MTA| HT 9 | 011 | 09 | 0000 1001 TCT | @ 73 | 111.1 49 | 01001001 | 9 | MTA LF 10 | 012 | 0A | 0000 1010 = J ‘74 | 112 | 48 | 0100-1010 | 10 | MTA VT 11 013 | 0B | 0000 1011 e K 75 |. 113 | 4B | 0100 1011, | 11 | MTA FF 12 | 014 | OC | 0000 1100 L ‘76 | 114 | 4C [0100 1100 | 12 | MTA! & CR 13 | 015 | OD | 0000 1101 M | 77 | 115 | 4D | 01001101 | 13 | MTA| & Sw) 14 | 016 | OE | 0000 1110 N | 78 | 116 | 4€ | 01001110 | 14 | MTA] &@ S! 15 | 017 | OF | 00001111 O | 79 | 117 | 4F | 01001111 | 15 | MTAL = oa DLE 16 | 020 | 10 | 0001 0000 P 80 | 120 | 50 | 01010000 | 16 | mTA| © DCt 17} 021 |.11 | 0001 0001 LLO Q 81 | 121] 5t | 01010001 | 17 | MTA] = Dc2 18 | 022 | 12 | 0001 0010 R 82 | 122 | 52 | 01010010 1 18 | MTA] = DC3 19 | 023-|. 13 } 0001 0011 » S | 83 | 123 | 53 | 01010011 | 19 | MTA OC4 20 | 024 | 14 | 0001 0100 ocL | & T 84 | 124 | 54 | 01010100 | 20 | MTA NAK 21 | 025 | 15 | 00010101 ppeu| = u | 8 | 125 |. 55 | 0101 0101 | 21 | MTA SYN 22 | 026 | 16 | 00010110 = v1 86 | 126 | 56 | 01010110 | 22 | MTA ETB 23 027 | 17 | 0001 0111 8 Ww ‘87 | 127-4 57 | 01010111 | 23 | MTA CAN 24 | 030 | 18 | 0001 1000 SPE | = x | 88 | 130 | 58 | 0101 1000 | 24 | MTA EM 25 | 031 | 19 | 0001 1001 spo | 2 Y 89 | 131 | 59 | 0101 1001 | 25 | MTA SUB 26 | 032 | 1A | 0001 1010 ¥ Z 9 | 132 | 5A | 0101 1010 | 26 | MTA ESC 27 | 033 | 18 | 0001 1011 2 { 91 | 133 | 5B] 0101 1011 | 27 | MTA. FS 28 | 034 | 1C | 0001 1100 \ 92 | 134 | SC | 0101 1100 | 28 | MTA GS 29 | 035 | 4D-} 0001 1101 } 93 | 135 | 5D | 0101 1101. | 29 | MTA RS 30} 036 | 1€ | 0001 1110 A | 94 | 136 | SE | 01011110 | 30 | MTA [US 31 | 037 | tf | 00071111 =| 95 | 137_ | SF] 0101 1111 UNT SPACE| 32] 040 | 20 | 00100000 | 0 |MLA ‘96 | 140 | 60 | 01100000-].0 | MSA ' 3 041 21. | 0010:0001 1 |MLA a 97 141 61..]:0110 0001 | 1 MSA " 34 | 042 | 22 | 00100010 | 2 | MLA D 98 | 142 | 62 | 01100010 | 2 | MSA # 35__| 043 | 23 | 00100011 | 3 | MLA C 99 | 143 { 63 | 01100011 | 3 | MSA $ 36 | 044 | 24 | 00100100 | 4 | MLA d | 100 | 144 | 64 | 01100100 | 4 | MSA % 37 | 045 | 25 | 00100101} 5 | MLA e | 101 | 145 | 65 | 01100101 | 5 | MSA. & 38 | 046 | 26 | 00100110 | 6 | MLA f 102 | 146 | 66 | 01100110 | 6 | MSA] 39 | 047 | 27 | 00100111 | 7 {MLA g | 103-| 147 | 67 | 01100111 | 7. | MSA ( 40 | 050 | 28 | 00101000 | 8 [MLA h | 104 | 150 | 68 | 01101000 |8 | MSA } 41 | 051 | 29 | 0010 1001 | 9 | MLA 105 | 151 | 69 | 0110 1001 | 9 |MSA ° 42 052 | 2A | 0010 1010} 10..|.MLA J 406 152. | 6A | 0110°1010. | 10 | MSA + 43 053 | 28 | 00101011 | 11 | MLA k 107 | 153. | 6B | 0110 1011, | 11. | MSA 44 | 054 | 2C | 00101100 | 12 {MLA ] 108 | 154 | 6C | 0110 1100 | 12 | MSA] @ - 45 | 055 | 20 | 00101101 | 13 |MLA| @ m | 109 | 155 | 6D | 01101101 | 13 | MSA | & 46 056. | 2E | 00101110 | 14 | MLA) & n 110 186 | 6E | 01101110 | 14 | MSA we ! 47 | O67 | 2F | 00101111 | 15 | MLA} w o | 41 | 157 | GF | O1t01711 | 15 | MSA] S i=) * nd 0 48 | 060 | 30 | 00110000 | 16 |MLA/ & p | 412 | 160 | 70 | 01110000 | 16 | MSA] & 1 49 061 31. | 00110001 | 17 | MLA = q 113 161 77 | Q111°0001 | 17 | MSA | = 2 50 | 062 | 32 | 00110010 | 18 |MLA| = r 14} 162 | 72 | 01110010 | 18 | MSA} 3 51 | 063 | 33 | 00170011 | 19 {MLA| 5 s 115 | 163_| 73 | 0117 0011 | 19 | MSA] B 4 52 | 064 | 34 | 00110100 | 20 | MLA 1 116 | 164 | 74 | 01110100 | 20 [Msa] # 5 53 | 065 | 35 | 00110101 | 21 | MLA vu | 17 | 165 | 75 | 0111 0101 | 21 | MSA 6 54 | 066 | 36 | 00110110 | 22 | MLA v | 198 | 166 | 76 | 01110110 | 22 | MSA 7 55 | 067 | 37 | 00110111 | 23 {MLA} w | 419 | 167 | 77 | 01110111 | 23 | MSA 8 56 | 070 | 38 | 0011 1000 | 24 | MLA x | 120 | 170 | 78 {0111 1000 | 24 | MSA g 57 | 071 | 39 | 0011 1001 | 25 |MLA y | 421 | a7t | 79 | 0111 1001 | 25 | MSA 38 072 | 3A 1.0011 1010 | 26 | MLA Zz 122 172 | 7A 40111 1010 4-26 -) MSA 59 073 | 3B.| 0011 1011] 27 | MLA { 23 173. | 7B | 0111 1011 4.27 -|-MSA . 60 | 074 | 3C | 00111100 | 28 [MLA 24 | 174 | 7C | 0111 1100 | 28 | MSA 61 | 075 | 3D | 00111101 | 29 | MLA } 25 | 175 | 7D | 0111 1101 | 29 | MSA , 62 | 076 | 3E | 00111110 | 30 | MLA 26 | 176 | 7E | 01111110 | 30 | MSA ? 63 O77 | SF { OOM AIK UNL 27 177 : 7F { QUIT 1144 3-38 .

Page 74 read from a scan by OCR; it may hold reading errors

4-1. INTRODUCTION

This section discusses considerations and techniques to help you use the 8840A effectively. Among other things, this section discusses sources of error which are an inherent part of the measurement process and which occur for all multimeters. By understanding why and when these efrors occur, and by knowing how and when to correct for them, you can make accurate measurements with confi-dence.

This section also discusses the relative benefits of 2-wire and 4-wire ohms, describes special considerations for making ac measurements, and presents some unusual applications--for example, using the test current in the 2-wire ohms function as a troubleshooting tool in itself.

4-2. DC VOLTAGE MEASUREMENT

When measuring dc voltages in high-impedance circuits, there are two possible sources of error to consider: circuit loading and input bias current.

4-3. Circuit Loading Error

Whenever a voltmeter is connected to a circuit, the voltmeter’s internal resistance changes the voltage of the circuit under test. The resulting error is called circuit loading error. The error is negligible as long as the resistance of the circuit under test (the source impedance) is small compared to the input impedance of the meter. As the source impedance approaches the input impedance of the voltmeter, the error can be considerable. The percentage of error can be calculated using the formula in Figure 4-1.

The input impedance of the 8840A is 10 MQ in the 200V and 1000V ranges, and is greater than 10,000: MQ in the 200 mV, 2V, and 20V ranges.. Therefore, for the 8840A, circuit loading error is less than 0.01% as long as the source impedance is less than 1 MQ in the.200 mV, 2V, and 20V ranges, and less than 1 kQ in the 200V and 1000V ranges. The exceptionally high input impedance on the 20V de range allows high-accuracy teadings in CMOS and high-impedance analog circuitry.

Section 4 Measurement Tutor al

NOTE

Input protection circuitry can reduce the inpu, impedance to as low as 100 kQ when the inal is overrange. This may also occur momentaril when the instrument autoranges to a highel range. |

4-4. Input Bias Current Error |

Input. bias current error occurs because.a voltmeter’ si input bias current always changes the voltage of the circuit under test. However, the error is significant only when measuring voltages in circuits with very high urce impedance. The error can be measured as shown in Figure 4-2.

With the 88404, it is easy to correct for this error using the OFFSET button:

2. Connect the 8840A INPUT terminals to a which matches the source impedance of the. ci be tested.

esistor

Select the VDC function and the desired me uit to

3. Allow the displayed reading to settle. 4. Press the OFFSET button.

5. Remove the resistor.

6.

Proceed with the desired measurement. Example:

Measure a 1.5V source with 1 MQ source impedance, correcting for input bias current.

1. Connect a 1 MQ resistor between the INPUT’ HI and INPUT LO terminals.

2. Select the VDC function and the 2V range. Allow: the display to settle.

4. Press OFFSET. (This zeroes the input bias current error.)

Page 75 read from a scan by OCR; it may hold reading errors

Measurement Tutorial DC VOLTAGE MEASUREMENT

100 x__Rs

CIRCUIT LOADING ERROR. IN % = Rs + Ri where Rs = Source impedance Ri = 8840A input impedance _ (Ri =>10,000 MQ in 200 mV, 2V, and 20V ranges) (Ri = 10 MQ in 200V and 1000V ranges)

EXAMPLE

When measuring the voltage across the 10 kQ leg of 290 kN over 10 kQ voltage divider, the circuit loading error is less than 0.1% in the upper ranges, and less than 0.0001% in the lower ranges: :

90kQ - SOURCE “== 8840A VOLTAGE “= $10 kQ 3 Ri ¢ . \ Rg = (90. kQ) x (10 kQ) _ gh, 90 kQ +10 kQ Error in the 200V and 1000V ranges = 100 x KOs 0.09% 9kQ+ 10 MQ ° 9 kO

Error in the 200 mV, 2V and 20V ranges = 100 x

= 0.00009%

9kQ + 10,000 MQ

Figure 4-1. Circuit Loading Error Calculation

5. Remove the i MQ resistor.

6. Measure the voltage of the circuit under test.

Note that this procedure does not correct for circuit load- _ ing error. Also note that if input bias current error is not corrected for, it may be added to the circuit loading error.

4-5. RESISTANCE MEASUREMENT

The 8840A allows you to measure resistance in both 2-wire and 4-wire configurations. Each has its benefits.

4-6. 2-Wire Ohms

Two-Wire ohms measurements are simple tc set up and yield good results for most measurement conditions. Measurements are made as shown in Figure 4-3. An internal current source (the “ohms current source”) passes a known test current (Itest) through the resistance being tested (Runknown). The 8840A measures the voltage drop across Runknown, calculates Runknown using Ohm’s law (Runknown = Vtest/itest), and displays the result.

4-2

The test current and full-scale voltage for each resistance range are shown in Table 4-1. Since the HI INPUT test lead is positive with respect to the LO INPUT lead, these test leads ate not interchangeable when a semiconductor device is being measured. |

4-7. Correcting for Test Lead Resistance in |

2-Wire Ohms . |

In 2-wire ohms, the resistance of the test leads can introduce error when measuring low resistances. Typical test leads may add as much as 0.5Q to 2-wire ohms readings.

With the 8840A, it is easy to correct for this error using the OFFSET button: 1. Select the 2-wire ohms function.

2. Touch the test leads together. The 8840A should indicate the resistance of the test leads.”

Page 76 read from a scan by OCR; it may hold reading errors

Measurement Tutorial RESISTANCE MEASUREMENT

R SOURCE

8840A

TO MEASURE INPUT BIAS CURRENT ERROR:

OFFSET.

the HI INPUT and LO_INPUT terminals.

4. Allow the circuit to settle.

1. Select the VDC function and the desired range.

2. Eliminate any offset voltages by shorting the Hl INPUT and LO INPUT terminals and then pressing .

3. Select aresistor which matches the source impedance (Rsource) of the circuitto be tested, and connect it to

5. Record the displayed voltage. This is the input bias current error (Veraon).

The input bias current error may be calculated as a percentage as follows:

G = : FRROR (IN %) = TOCTAGE MEASUREMENT * 100%

The input bias current itself (Igias) May be calculated as follows:

ERROR

~ * 'SOURCE

Igias =

__ Verror

Figure 4-2. Measuring Input Bias Current Error

3. With the test leads still touching, press. the OFFSET button. The 8840A should read 0Q.

4-8. 4-Wire Ohms

Four-Wire ohms measurements, provide the highest accuracy for low resistance measurements. The 4-wire configuration automatically corrects for both test lead resistance and contact resistance. Contact resistance (the resistance between the test probe tips and the circuit being tested) is unpredictable, and therefore cannot be reliably corrected with a fixed offset.

Four-Wire ohms measurements are especially important when using long test leads. In a typical automated test system, for example, the test leads could be connected through four or five switching relays, each with 2Q of Tesistance!

The 8840A makes 4-wire ohms measurements as shown in

Figure 4-4. The HI and LO INPUT leads apply a known, internal current source to the unknown resistance, just as ete epee er eng me aren emp ei ene tee ag ig th oe on apne samme eg in 2-wire ohms. (See Table 4-1.) However, the voltage drop across the unknown resistance is measured with the SENSE leads rather than the INPUT leads. Since the current flow in the SENSE leads is negligible, the error caused by the voltage drop across the leads is also negligible. :

NOTE

In the 2 MQ and 20 MQ ranges of 4-wire ohms, the voltage across the unknown resis- tance is' sensed between the HI SENSE and LO INPUT terminals. Accuracy is not affected as long as the resistance of the LO INPUT lead is less than 10Q in the 2 MQ range, and less than 1002 in the 20 MQ range.

4-9. Applications of the Ohms Functions

The 2-wire and 4-wire ohms functions can be used for a variety of purposes in addition to measuring resistance, as the following applications show.

4-3

Page 77 read from a scan by OCR; it may hold reading errors

Measurement Tutorial RESISTANCE. MEASUREMENT ~

8840A | ( OHMS CURRENT SOURCE (| e i VOLTAGE SENSE (m)- {M) Jf HI INPUT LO INPUT R unknown

Figure 4-3. 2-Wire Ohms Measurement

Table 4-1. Ohms Test Current 1. TEST FULL SCALE RANGE CURRENT VOLTAGE 2. 20022 1 mA 0.2V 2 kQ 1 mA 2.0V 20 ko2 100 pA 2.0V 200 kQ 10 pA 2.0V 2000 kQ 5 yA 10.0V 20 MQ 500 nA 10.0V

4-10.. TESTING DIODES The 2-wire ohms function can also be used to test diodes.

1. Select the 2-wire ohms function and the 2 kQ range.

2. Measure the resistance. of the diode. If the diode is good, when forward-biased it will measure about 0.6 kQ to 0.7 kQ for silicon (0.25 kQ to 0.3 kQ for germanium), and when reverse-biased it will cause the 8840A to indicate overrange. (The forward-biased reading depends upon the range used.)

The 2 kQ range is used because its 1 mA test current provides a typical operating point, and its 2V full-scale voltage is sufficient to turn on most diodes (even two 5. diodes in. series).

4-11.

TESTING ELECTROLYTIC CAPACITORS

The 2-wire ohms function can also give a rough test of an electrolytic capacitor’s. leakage and dielectric absorption.

This test works well for capacitors 0.5 uF and larger.

4-4

Select the 2-wite ohms function, the 2 kQ range, and the medium reading rate.

Connect the test leads to the capacitor (with the INPUT HI lead to the + lead and the INPUT LO lead to the - lead). The 8840A will try to charge it to the open-circuit voltage of the 2 kQ range (about 6V).

Disconnect the + test lead.

To test for leakage, select the VDC function and the 20V Tange (leave the 8840A in the medium reading rate), and measure the voltage that was stored on the capacitor during step 2.

a. If the capacitor is good, the voltage across the capacitor will be about 6V, and will be relatively stable. | b. - If the capacitor is leaky, the voltage across the capacitor will be much less than 6V, and the voltage will be decreasing. The rate of change depends on how leaky the capacitor is. | c. With some electrolytic capacitors, the . reading will increase. This usually indicates the capacitor is defective.

To test the capacitor’s dielectric. absorption, briefly

_ short the capacitor’s leads together and then measure the voltage across the capacitor.

a... If the dielectric is good (i.e., has low dielectric absorption), the voltage across the capacitor will be nearly zero volts.

Page 78 read from a scan by OCR; it may hold reading errors

Measurement Tutorial RESISTANCE MEASUREMENT

HI INPUT

OHMS

CURRENT t) V rest R unknown SOURCE

LO INPUT

8840A

H! SENSE

DC VOLTAGE SENSING CIRCUITRY o ")

LO SENSE

OHMS CURRENT SOURCE -)

Figure 4-4. 4-Wire Ohms Measurement b. If the dielectric is poor (i-e., has high dielectric absorption), the voltage across the capacitor will be significantly above zero.

4-12. A PRECISION CURRENT SOURCE

The ohms current source (the internal current source used in the ohms functions) makes a useful troubleshooting tool in itself. It has excellent linearity and temperature stability. Its compliance voltage is typically 5V in the lower four ohms ranges, and 12V in the upper two ohms tanges. The inputs are protected against accidental applications of voltage up to 300V rms.

To use the ohms current source, connect the test leads to the HI and LO INPUTS, and select either the 2-wire or

4-wire ohms function. Press the range buttons to select.any of the current levels shown in Table 4-1.

The ohms current source can be used to troubleshoot circuits by injecting current into selected nodes, orcing the circuits to be in a specific test state. For example, the ohms current source can be used to set or modify the bias of amplifier circuits. The current level can bec] ged simply by changing range. | The ohms current source can also be used to test mA or uA panel meters. The accuracy of the current source is more than enough to verify panel meters, whose accuracy is typically 1% to 5%. To test an analog panel) meter, simply connect the current source across the meter! move-

45

Page 79 read from a scan by OCR; it may hold reading errors

Measurement Tutorial RESISTANCE MEASUREMENT ment (as though measuring its resistance). A 1 mA meter should show full scale when the ohms function is set on the 2 kQ range. The same technique also works with digital panel meters.

4-13. DC CURRENT MEASUREMENT

To get the best accuracy using the mA DC function, it is important to understand the concept of burden voitage error.

When a meter is placed in series with a circuit to measure current, error can be caused by the small voltage drop across the meter (in this case, across the protective fuses and current shunt). This voltage drop is called the burden voltage, and it is highest for full-scale measurements. The full-scale burden voltage for the 8840A is typically less than 1V.

The burden voltage can present a significant error if the current source being measured is ‘unregulated (i.e., not a true current source) and if the resistance of the fuse and shunt is a significant part of the source resistance. If burden voltage does present a significant error, the percentage of error can be calculated and corrected for using the formulas in Figure 4-5.

4-14. REDUCING THERMAL VOLTAGES

When making very low-level dc measurements, thermal voltages can present an additional source of error. Thermal voltages are the thermovoltaic potentials generated at the junction between dissimilar metals. Thermal voltages typically occur at binding posts and can be greater than 10 pV.

Thermal voltages can also cause problems in the low ohms ranges. Some low-value resistors are constructed with dissimilar metals. Just handling such resistors can cause thermal voltages large enough to introduce measurement errors.

The effect of thermal voltages can be reduced by using the following techniques:

1. Use tight connections.

2. Use clean connections (especially free of grease and dirt).

3. Use similar metals for connections wherever possible (€.g.,. copper-to-copper, gold-to-gold, etc.).

4. Use caution when handling the circuit under test. Wait for the circuit to reach thermal equilibrium.

(Thermal voltages are generated only where there is a temperature gradient.)

4-15. AC VOLTAGE AND CURRENT MEASUREMENT

When making precise measurements of ac voltage and current, there are several considerations in addition to

4-6

4-16. True RMS Measurement

|

: | those discussed under dc voltage and current measurement. These include the concepts of rms conversion, crest factor, bandwidth, and zero-input error.

The True RMS AC Option measures the true rms value of ac voltages and currents. In physical terms, the rms |(root-mean-square) value of a waveform is the equivalent dc value that causes the same amount of heat to be dissipated ina resistor. True rms measurement greatly simplifies the analysis of complex ac. signals. Since the rms value jis the de equivalent of the original waveform, it provides a reliable basis for comparing dissimilar waveforms.

By contrast, many meters in use today use average-responding ac converters rather than true rms converters. The scale factor in these meters is adjusted so that they display the rms value for harmonic-free sinusoids. However, ifa signal i is not sinusoidal, average-responding meters do not display correct rms _ readings. | The 8840A actually derives the rms value using . alog computation. This means that the 8840A readings repre- Sent true rms values not only for harmonic-free sinusoids, but also for mixed frequencies, modulated signals, square waves, sawtooths, random noise, rectangular pulses with 10% duty cycle, etc.

4-17. Waveform Comparison

Figure 4-6 illustrates the relationship between ac and de components for common waveforms, and compares readings for true rms meters and average-responding meters. For example, consider the first waveform, a 1.41421V, (zero-to-peak) sine wave. Both the 8840A aad rms-calibrated average-responding meters display the correct rms reading of 1.00000V (the dc component equals 0). However, consider the 2V (peak-to-peak) square wave. Both types of meter correctly measure the dc component (OV), but only the 8840A correctly measures the ac|component (1.00000V). The average-responding meter measures 1.110V, which amounts to an 11% error.

Since average-responding meters have been in use for so long, you may have accumulated test or reference data based on them. The conversion factors in Figure 4-6 should help you convert between the two measurement methods.

4-18. Crest Factor

Crest factors are useful for expressing the ability of an instrument to measure a variety of waveforms accurately. The crest factor of a waveform is the ratio of its) peak voltage to its rms voltage. (For waveforms where the positive and negative half-cycles have different peak voltages, the more extreme peak is used in computing the crest factor.) Crest factors start at 1.0 for square waves (for which the peak voltage equals the rms voltage) and

Page 80 read from a scan by OCR; it may hold reading errors

Measurement Tutorial AG VOLTAGE AND CURRENT MEASUREMENT

LO INPUT

Es = Source voltage (measured) E» = Burden voitage (measured) R.= Load + source resistance

Rv = 8840A internal resistance

E ERROR IN mA = (Displayed current) X, _

Ep. ERROR IN PERCENT = = X 100 Ss

Es- Ep

EXAMPLE:

Correct current = 1460 mA + 40 mA = 1500 mA

NOTE: MEASURING BURDEN VOLTAGE measure voltage at the 2A terminal.

Displayed current = 1460 mA = 15V (measured with 8840A in VDC function) E,= = 0.4V (measured with 8840A as described below)

Error in percent = ov X 100 = 2.67% 15V Error in mA = (1460 mA) X

To get the correct current, add the error in mA to the displayed current:

The 8840A allows you to measure burden voltage (Es) directly. Leaving the 2A and LO INPUT leads in place for current measurement, select the VDC function. Then, connect a third lead to the HI INPUT terminal and with it

0.4V, - - - = 40mA 15V - 0.4V

Figure 4-5. Burden Voltage Error Calculation increase for more “pointed” waveforms as shown in Figure 4-7,

The 8840A has a full-scale crest factor limit of 3.0 for the 20V and 700V ranges, and 6.0 for the other ranges. For full-scale input signals with a crest factor above these limits, dynamic range limitations can begin to cause large errors. However, as Figure 4-7 shows, signals with a crest factor above 3.0 are unusual.

If you don’t know the crest factor of a particular waveform but wish to know if it falls within the crest factor limit of the 8840A, measure the signal with both the 8840A and an ac-coupled oscilloscope. If the rms reading on the 8840A is 1/3 or more of the waveform’s zero-to-peak voltage, the crest factor is 3.0 or less.

4-19. AC-Coupied AC Measurements

Input signals are ac-coupled in the ac functions. One of the advantages of ac coupling is that ac measurements can be

' made on power supply outputs,- phone lines, etc. Ripple measurements, for instance, cannot be made with dc coupling. Remember, however, that when the 8840A measures signals with the ac functions, the reading on the display

4-7