Page 1 of 8840A Digital Multimeter Instruction Manual

Manual · Fluke

8840A Digital Multimeter Instruction Manual

Manual180 pagesPDF 42 MB

Instruction manual for the Fluke 8840A 5-1/2 digit bench multimeter: operation, remote programming, calibration and service.

Download the PDF42 MB

Covers · 1 model

The models it covers

Summary

What is in it

The Fluke instruction manual for the 8840A Digital Multimeter, a 5-1/2 digit instrument for bench or systems use. It gives the full specifications, operating instructions, remote programming over the optional IEEE-488 interface, a measurement tutorial, theory of operation, maintenance with performance tests, calibration and troubleshooting, the list of replaceable parts, options and accessories, and schematic diagrams. It also covers the True RMS AC option. This copy is a scan, so some text may read imperfectly.

Its parts

  1. Introduction and Specifications
  2. Operating Instructions
  3. Remote Programming
  4. Measurement Tutorial
  5. Theory of Operation
  6. Maintenance
  7. List of Replaceable Parts
  8. Options and Accessories
  9. Schematic Diagrams

From the manual

Key facts from this manual

Some pages are scans or cannot be read; this says only what the readable text supports. Each line names its page in the PDF.

Instruction manual (PN 879304, December 1991, Rev. 3, 5/97) for the Fluke 8840A Digital Multimeter, a scanned copy with an OCR text layer. It gives the specifications, installation and front-panel operation, remote programming through the IEEE-488 Interface option, and service information. It also covers the True RMS AC option and the rack kits, probes and test leads offered for the meter.

Read for this summary: pages 1 to 3, 7 to 8, 12, 17 to 25, 27 and 31 of 180.

What it is
5-1/2 digit digital multimeter for bench or systems use, with a vacuum fluorescent display p. 17
Measurements
DC voltage, 2-wire and 4-wire resistance, DC current; AC voltage and AC current with Option -09 p. 17
DC voltage
Ranges 200 mV to 1000V; maximum input 1000V dc or peak ac on any range p. 18
Basic DC accuracy
0.005% for 1 year p. 17
Current
2000 mA range; 2A dc or rms ac maximum, protected by a front-panel 2A, 250V fuse and an internal 3A, 600V fuse p. 20, 21
Resistance
2-wire or 4-wire, ranges up to 20 MΩ p. 21, 22
True RMS AC (Option -09)
AC voltage ranges 200 mV to 700V; maximum input 700V rms or 1000V peak p. 19
IEEE-488 (Option -05)
Full programmability, external trigger input and sample complete output connectors, automated calibration p. 17
Power
100, 120, 220 or 240V ac +10% (250V ac maximum), switch selectable, 50, 60 or 400 Hz, 20 VA maximum p. 23
Power-line fuse
1/4A, 250V slow-blow for 90V to 132V; 1/8A, 250V slo-blo for 198V to 250V p. 25
Temperature
0 to 50°C operating, -40 to 70°C storage p. 23
Size and weight
8.9 cm x 21.6 cm x 37.1 cm (HxWxD); 3.4 kg (7.5 lb) net p. 23, 24
Rack mount kits
Y8834 single, Y8835 dual, Y8836 center p. 27
Probes and leads
TL70A, Y8134 and Y8140 test leads; 80T-150U and 80TK temperature probes; 85RF and 83RF RF probes; 80J-10 shunt; Y8100, Y8101, 80i-400, 80i-600 current probes; 80K-6, 80K-40 high voltage probes p. 12

What this manual walks you through

  • Installing the power-line fuse and setting the line voltage p. 25
  • Mounting the meter in a 19-inch rack p. 25, 27
  • Reading error messages and the overrange indication p. 27, 31
  • Running the diagnostic self-tests p. 31
  • Choosing autorange or manual range p. 31
  • Triggering readings in continuous or external trigger mode p. 31
  • Programming the meter remotely over the IEEE-488 bus p. 8

Good to know

  • Leave the inputs disconnected during the self-tests, or the meter may report errors. p. 31
  • For line voltages of 198V to 250V the line fuse must be changed to a 1/8A, 250V slo-blo fuse. p. 25
  • Use clamp-on probes when measuring circuits above 2 amps. p. 3
  • In fast reading rate with data sent over IEEE-488, the display must be blanked (command D1) for stated accuracy. p. 18

Questions this manual answers

What does ERROR 30 mean on a Fluke 8840A?

An AC function was selected on a meter without the True RMS AC Converter option (Option -09). p. 27, 31

What do 8840A error codes 01 to 29 mean?

They correspond to the self-tests. If they appear with the inputs disconnected, the manual says there may be a hardware problem. p. 31

What fuse does the 8840A power line take?

A 1/4A, 250V slow-blow fuse for 90V to 132V; for 198V to 250V, a 1/8A, 250V slo-blo fuse of 6.3mm x 32mm. p. 25

How fast can the 8840A take readings?

Up to 100 readings per second in the fast rate. p. 17, 22

The whole text · 18 parts

Read the manual here as text

Its text as extracted from the PDF, unedited, in parts; each page opens the PDF at that page.

  1. Pages 1 to 101-10
  2. Pages 11 to 2011-20
  3. Pages 21 to 3021-30
  4. Pages 31 to 4031-40
  5. Pages 41 to 5041-50
  6. Pages 51 to 6051-60
  7. Pages 61 to 7061-70
  8. Pages 71 to 8071-80
  9. Pages 81 to 9081-90
  10. Pages 91 to 10091-100
  11. Pages 101 to 110101-110
  12. Pages 111 to 120111-120
  13. Pages 121 to 130121-130
  14. Pages 131 to 140131-140
  15. Pages 141 to 150141-150
  16. Pages 151 to 160151-160
  17. Pages 161 to 170161-170
  18. Pages 171 to 180171-180

The file

Revision and size

RevisionPN 879304, December 1991 Rev. 3, 5/97
PublisherFluke Corporation
Pages180
PDF42 MB
TextThe pages are scans; their text was read by OCR, which may hold reading errors.