For the Fluke 8840A

Page 101 read from a scan by OCR; it may hold reading errors

Theory of Operation POWER. SUPPLY

Secondary T601-14,15,16 supplies the vacuum fluorescent display filament with 4.5V ac. The center tap is connected to the in-guard +5V supply in order to correctly bias the display. An isolated secondary supplies 16V ac to the power supply on the TEEE-488 Interface.

Zener diode CR615 and SCR Q601 comprise a protective crowbar circuit. If the line voltage exceeds the nominal value by approximately 30 percent or more, CR615 con-ducts, tuming on Q601, shorting out the power transformer secondary and blowing the line fuse. In normal operation, _ these components have ‘no effect.

5-37. IEEE-488 INTERFACE (OPTION -05)

The IEEE-488 Interface has five major parts, as shown in the. block diagram in Figure 5-16. All components are contained in a single printed circuit assembly (PCA). Reference designations are numbered in the 900 series.

5-38. Out-Guard Microcomputer

The Out-Guard Microcomputer (uC) (U901) communicates with the IEEE-488 talker/listerner IC (U911) and the In-Guard pC (U202).

The Out-Guard pC is similar to the In-Guard Z8 uC except that it contains 8K bytes of ROM and 236 butes of

RAM. For further description of the Z8 uC, refer to the heading “In-Guard Microcomputer” above.

5-39. ‘Guard Crossing

The guard crossing circuit permits serial asynchronous communication between U901 and U202 while isol ting the two electrically. One-half of the guard crossing circuit is contained’ on the Main PCA; the other half is on the IEEE-488 Interface PCA. Operation of the guard crossing circuit is described in an earlier heading.

5-40. Bus Interface Circuitry

The IEEE-488 bus protocol is handled by the wPD7210 IEEE-488 talker/listener IC (U911). It is controlled by U901 as a memory mapped peripheral through an 8-bit data bus.

Bus transceivers U912 and U913 buffer U911 from the IEEE-488 bus. They provide the bus with the r equited output drive capability and receiver impedance.

|

5-41. Signal Conditioning | The SAMPLE COMPLETE and EXT TRIG signals 1903 and J904) are. conditioned by U909. Diodes CR903, CR904, CR905, and CR906 and resistors R917 and ‘R918 provide protection from excessive voltages. Jumpers E902 and E903 allow selection of the polarity of the EXT TRIG

‘a _ IEEE-488 INTERFACE OPTION ~ - ~ | 4 } | IN-GUARD ', > | uc | “GUARD | U202 ;CROSSING IEEE-4s8 | | TALKER/ TRANSFORMER OUT-GUARD IC U911 488 T601 § - - - py «= POWER fg AND BUS BUS | SUPPLY Us01 TRANS- CEIVERS . U912 | AND | SAMPLE , U913 COMPLETE™ SIGNAL ‘ | EXTERNAL CONDITIONING TRIGGER 71>) > : | iz | NU

Figure 5-16. IEEE-488 Interface Block Diagram

Page 102 read from a scan by OCR; it may hold reading errors

Theory of Operation IEEE-488 INTERFACE (OPTION -05)- signal. (A polarity selection procedure is given in the Maintenance section.) The 8840A is configured in the factory so that it is triggered on the falling edge of the EXT TRIG signal.

5-42. IEEE-488 Interface Power Supply

The IEEE-488 Interface power supply circuit provides the TEEE-488 Interface PCA with +5V. The circuit consists of rectifying diodes CR908 and CR909, filter capacitor C910, and voltage regulator VR901. Power comes from transformer T605 on the Main PCA. U908 and associated circuitry resets the Out-Guard uC at power-up and following power-line voltage dropouts.

5-43, TRUE RMS AC (OPTION -09)

The True RMS AC circuit (Figure 5-17) performs two primary functions. First, it scales ac input voltages and ac current sense voltages to a range of OV to 2V ac mms.

-. Second, it converts the scaled ac voltages to an equivalent de voltage which is then directed to the A/D Converter via the Track/Hold Amplifier. The True RMS AC circuit is trimmed for flat high-frequency response using a variable filter which is set by the High-Frequency AC Calibration procedure.

The following paragraphs describe how these functions are . performed. Components are laid out-on’a single printed - circuit assembly (PCA). Component reference designators are numbered in the 800 series.

5-44. VAC Scaling

AC voltage inputs are ‘directed from the HI INPUT terminal to the True RMS AC PCA through protection resistor R309 on the Main PCA. In this way, voltage transients greater than 1560V are clamped by MOVs (RV301, RV402, RV403, and RV404) as in the VDC function. With the VAC function selected, K801 is closed. The input voltage is thus applied to C801, which blocks dc inputs.

U807 and resistor network Z801 provide selectable attenu-ation and 1 MQ input impedance. In the upper two ranges, K802 is closed and Q806 is off, providing a gain of -1/500. In the lower three ranges, K802 is open and Q806 is on, shorting Z801-4 to ground; this configuration provides a gain of -1/5. CR801 and CR802 provide protection by clamping the inverting input of U807 to approximately +0.6V. Q805 shifts logic levels to control Q806.

U806A, U806B, and a voltage divider (R804 and R805) provide gain which is selected for each range by the analog switches in U804. The configuration for each range is shown in Figure 5-17. (In this figure, the CMOS analog switches are represented by mechanical switches.) When U806A is not used, its non-inverting input is grounded by Q804. When U806B is not used, its non-inverting input is connected to the CURRENT SENSE line.

5-45. mA AC Scaling

The mA AC function uses the same current shunt and protection network which is used for de current. In the mA AC function, Q802 switches the CURRENT SENSE line to the non-inverting input of U806B, which provides a gain of 10.

5-46. Frequency Response Trimming The frequency response is trimmed by software calibration

~ using a digitally controlled one-pole low-pass filter (R832 and a combination of C826, C827, C828, and C829). The analog switches in U808 configure the four capacitors to select one of 16 possible RC constants. The input of the digitally controlled filter is buffered by voltage follower U801A. The individual gain stages are also provided with fixed frequency compensation.

5-47. True RMS AC-to-DC Conversion

U801B buffers the input to rms converter U802. U802 computes the rms value of the scaled input voltage as shown in Figure 5-18. Rather than explicitly squaring and averaging the input, U802 uses an implicit method in

R804, R80S uso? Iw VAC: 13, 15 VARIABLE RMS a FILTER CONVERTER = AC INPUT VAC: 12.14 i VOLTAGE f U806A, o FILTERS Lo» TRACK U806B \ ; (+500 FOR r4, 15) x5 2. \ CIRCUI \ VAC: rt CURRENT > | or mA AC SENSE

Figure 5-17. True RMS AC Option Block Diagram

Page 103 read from a scan by OCR; it may hold reading errors

Theory of Operation TRUE RMS AC (OPTION -09)

which feedback is used to perform an equivalent analog computation.

The filter averages the divider output signal. This filter consists of U809A, C813, R815, and the internal 25 kQ resistor and op amp between pins 8 and 9 of U802. The output is further filtered by a three-pole post-filter com-prised of U809B and associated resistors and capacitors. This output is then switched into the Track/Hold Amplifier of the dc front end via U302 pins 15 and 14. The Track/Hold Amplifier is set up for unity gain on all. ac Tanges.

FILTER Vout

Figure 5-18. True RMS AC-to-DC Converter

5-20

Page 104 read from a scan by OCR; it may hold reading errors

A Message From Fluke Corporation a Static awareness

Some semiconductors and custom IC's can be damaged by electrostatic discharge during handling. This notice explains how you can minimize the chances of destroying such devices by:

1. Knowing that there is a problem.

2. Leaning the guidelines for handling them.

3. Using the procedures, packaging, and bench techniques that are recommended.

The following practices should be followed to minimize damage to S.S. (static sensitive) devices.

al 3. DISCHARGE PERSONAL STATIC BEFORE HANDLING DEVICES. USE A HIGH RESIS-

1. MINIMIZE HANDLING TANCE GROUNDING WRIST STRAP.

2. KEEP PARTS IN ORIGINAL CONTAINERS UNTIL READY FOR USE. 4, HANDLE S.S. DEVICES BY THE BODY.

Page 105 read from a scan by OCR; it may hold reading errors

f

5. USE STATIC SHIELDING CONTAINERS FOR HANDLING AND TRANSPORT.

6. DO NOT SLIDE S.S. DEVICES OVER ANY SURFACE.

7. AVOID PLASTIC, VINYL AND STYROFOAM® IN WORK AREA.

PORTIONS REPRINTED WITH PERMISSION FROM TEKTRONIX INC. AND GENERAL DYNAMICS, POMONA DIV.

® Dow Chemical

8. WHEN REMOVING PLUG-IN ASSEMBLIES

HANDLE ONLY BY NON-CONDUCTIVE EDGES AND NEVER TOUCH OPEN EDGE CONNECTOR EXCEPT AT STATIC-FREE WORK STATION. PLACING SHORTING STRIPS ON EDGE CONNECTOR HELPS PROTECT INSTALLED S.S. DEVICES.

9. HANDLE S.S. DEVICES ONLY ATA STATIC-FREE WORK STATION.

10. ONLY ANTI-STATIC TYPE SOLDER- SUCKERS SHOULD BE USED.

11. ONLY GROUNDED-TIP SOLDERING IRONS SHOULD BE USED.

9/93

Page 106 read from a scan by OCR; it may hold reading errors

WARNING

THESE SERVICE INSTRUCTIONS ARE FOR USE BY QUALIFIED PERSONNEL ONLY. TO AVOID ELECTRIC SHOCK, DO NOT PERFORM ANY PROCEDURES IN THIS SECTION UNLESS YOU ARE QUALIFIED TO DO SO.

6-1. INTRODUCTION

This section presents maintenance information for the 8840A. The section includes a performance test, a calibration procedure, troubleshooting information, and other general service information.

Test equipment recommended for the performance test and calibration procedure is listed in Table 6-1. If the recommended equipment is not available, equipment that meets the indicated minimum specifications may be substituted.

6-2. PERFORMANCE TEST

This test compares the performance of the 8840A with the specifications given in Section 1. The test is recommended as an acceptance test when the instrument is first received, and as a verification test after performing the calibration procedure. If the instrument does not meet the performance test, calibration or repair is needed.

To ensure optimum performance, the test must be performed at an ambient temperature of 18°C to 28°C, with a relative humidity of less than 75%. Also, the 8840A should be allowed to warm up for one hour prior to beginning any test other than the self-test.

6-3. Diagnostic Self-Tests

The diagnostic self-tests check the analog and digital circuitry in the 8840A. There are 21 analog tests followed by in-guard program memory, calibration memory, and display tests. Out-guard program memory is tested when self-test is initiated by a remote command. Microcomputer RAM tests are done only at powerup.

Section 6 Maintenance

NOTE

The inputs must be left open-circuited while the self-tests are performed. Otherwise, the 8840A may indicate errors are present. Errors may also be caused by inductive or capacitive pick-up from long test leads.

If the FRONT/REAR switch is in the REAR position, the 8840A skips tests 3 and 4. Also, if Option -09 is not installed, the 8840A skips tests 1, 2, and 3.

To initiate the self-tests, press the SRQ button for 3 seconds. The TEST annunciator will then light up, and the 8840A will run through the analog tests in sequence. Each test number is displayed for about 1 second. The instrument can be stopped in any of the test configurations by pressing the SRQ button while the test number is displayed. Pressing any button continues the tests.

After the last analog test is performed, all display segments light up while the instrument performs the in-guard program memory, calibration memory, and display tests. The instrument then assumes the power-up configuration: VDC, autorange, slow reading rate, offset off, local control.

If the 8840A detects an error during one of the tests, it displays the ERROR annunciator and the test number for about 2-1/2 seconds, and then proceeds to the next test. The test number thus becomes an error code. (Error codes are listed in Table 2-1, Section 2.)

Passing all diagnostic self-tests does not necessarily mean the 8840A is 100% functional. The test, for example, cannot check the accuracy of the analog circuitry. If one or more errors are displayed, the 8840A probably requires service.

6-4. DC Voltage Test

The following procedure may be used to verify the accuracy of the VDC function:

6-1

Page 107 read from a scan by OCR; it may hold reading errors

Maintenance PERFORMANCE TEST

Tabie 6-1. Recommended Test Equipment

INSTRUMENT TYPE

MINIMUM SPECIFICATIONS

RECOMMENDED MODEL

DC Calibrator

PREFERRED:

Voltage Range: 0-1000V dc Voitage Accuracy: 10 ppm Absolute Linearity: +1.0 ppm

Fluke 5700A or Fluke 5440A

ALTERNATIVE:

(Must be used with Kelvin-

Varley Voltage Divider)

Voltage Range: 0-1000V dc

Voltage Accuracy: 20 pprn + 20 ppm of range

Fluke 343A

Kelvin-Varley Voltage Divider: Ratio Range: 0-1.0 Absolute Linearity: =1 ppm of input at dial setting

Fluke 720A

Resistor Calibrator

Resistance Accuracy: 0.005%

Fluke 5700A or Fluke 5450A, ES] DB62

DC Current Source

Accuracy: +0.025%

Fluke 5700A or Fluke 5100B

Oscilloscope

General purpose, 60 MHz, with 10 MQ probe

Philips 3055 or 3355

Digital Multimeter

Voltage Accuracy:

0.01% in V de

1.0% for 1V in V ac @ 100 kHz

Input Impedance:

10 MQ or greater in V dc;

1 MQ or greater in parallel with <100 pF in V ac

Fluke 8840A (with Option -09)

AC Calibrator

Fiuke 5700A and Fluke 5725A

Minimum Required Accuracy (By Range) F en Frequency 4, 10, 100 mv" 1, 10, 100V2 1000v2 Range

20 Hz - 30 Hz -1 + .005 1+ .005 12 + .005 30 Hz - 20 kHz 02 + 10 02 + .002 .04 + .004 20 kHz - 50 kHz 05 + 20 05 + .005 08 + .005 50 kHz - 100 kHz 05 + 20 05 + .005 -1 + .01 1. +(% of setting + pV) 2. +/-(% of setting + % of range)

AC Current Source

Fluke 5700A or Fluke 5100B

Frequency Minimum Required Accuracy Range (All Ranges)

30 Hz - 1 kHz +(.07% + 1 mA)

1 kHz - 5 kHz +(.07% + 1 mA) X frequency in kHz Shorting Bar Resistance: <1.5 mQ Pomona MDP-S-0

Construction: Soldered (not rivetted)

6-Inch Jumper, E-Z-Hook 204-6W-S or equivalent Optional Test | 9010A, 9005A or Micro-System Troubleshooter; 9000A-8048 Interface Pod; Equipment

6-2

Page 108 read from a scan by OCR; it may hold reading errors

Maintenance Test PERFORMANCE TEST

Ensure the 8840A is on and has warmed up for at least 1 hour.

Select the VDC function.

Connect the DC Calibrator (see Table 6-1) to provide a voltage input to the HI and LO INPUT terminals. If using the Fluke 343A and the Kelvin-Varley Voltage Divider instead of the Fluke 5440A, connect the test leads as shown in Figure 6-1.

For each step in Table 6-2, select the indicated range, set the DC Calibrator for the specified input, and verify that the displayed reading is within the limits shown for each reading rate. (For step A, connect a short across the HI and LO INPUT terminals and press OFFSET. The measurement in step C should be relative to this offset.)

Set the DC Calibrator to input negative voltages, and repeat steps C through G of Table 6-2.

With the unit is the 2V range, check the A/D linearity by setting the DC Calibrator for each step in Table 6- 9, while verifying the display reading is within the limit shown. Set the DC Calibrator for zero volts and disconnect if from the 8840A.

6-5. AC Voltage Test (Option -09 Only)

The following procedure may be used to verify the accuracy of the VAC function:

Ensure the 8840A is on and warmed up for at least | hour.

Select the VAC function and the slow (S) reading rate.

Connect the AC Calibrator to provide a voltage input to the HI and LO INPUT terminals.

(Low- and Mid-Frequency Test.) For each step in Table 6-3, select the indicated range, set the AC Calibrator for the specified input, and verify that the displayed reading is within the limits shown for each reading rate.

NOTE

This procedure tests the extremes of each range. You may shorten the procedure by testing only the “quick test points” indicated in Table 6-3 with asterisks.

(High-Frequency Test.) for each step in Table 6-4, select the indicated range, set the AC Calibrator for the specified input, and verify that the displayed reading is within the limits shown for each reading rate.

NOTE

This procedure tests the extremes of each range. You may shorten the procedure by testing only the “quick test points” indicated in Table 6-4 with asterisks.

DC CALIBRATOR SENSE (FLUKE 343A) OUTPUT ort REVERSE THESE LEADS FOR NEGATIVE INPUT { - VOLTAGE TO 8840A KELVIN-VARLEY VOLTAGE DIVIDER (FLUKE 720A) 0,, | INPUT ore L OO00 OOO fotexe) 9QOO= CONNECTIONS ARE SHOWN INPU FOR POSITIVE INPUT VOLTAGES OO TO 8840A NPS) oO oO oo ‘o}m) om NOTE: THE KELVIN-VARLEY VOLTAGE DIVIDER IS UO rol

NOT REQUIRED FOR INPUT VOLTAGES ABOVE 2.0V DC

Figure 6-1. Connections for Kelvin-Varley Voltage Divider

6-3

Page 109 read from a scan by OCR; it may hold reading errors

Maintenance PERFORMANCE TEST

Table 6-2. DC Voltage Test

DISPLAYED READING INPUT STEP | RANGE (V dc) SLOW MEDIUM FAST? MINIMUM MAXIMUM MINIMUM MAXIMUM MINIMUM MAXIMUM A' | 200 mV {OV (short)} -00.003 +00.003 -00.005 +00.005 -00.02 +00.02 B 2V, 20V,!0V (short)| -3 counts +3 counts -5 counts +5 counts -2 counts +2 counts 200V, 1000V C* | 200 my | 100 mv +99.989 +100.014 +99.987 +100.013 +99.97 +100.03 D OV 1V +.99993 +1.00007 +.99991 +1,00009 +.9997 +1.0003 E 20V 10V +9.9992 +1.00008 +9.9990 +10.0010 +9.997 +10.003 F 200V 100V +99.992 +100.008 +99.990 +100.010 +99.97 +100.03 G 1000v. | 1000V +999,92 +1000.08 +999.90 +1000.10 +999.7 +1000.3 NOTES: 1. Relative to high-quality short stored using OFFSET feature. 2. When in fast reading rate with intemal trigger and transmitting data out of the IEEE-488 interface, the 8840A display must be blanked (command D1) to ensure stated accuracy.

6. Set the AC calibrator to standby and disconnect it 3.

from the 8840A.

6-6. Resistance Test

The following procedure may be used to verify the accuracy of the 2-wire and 4-wire ohms functions.

1. Ensure the 8840A is on and has warmed up for at least 1 hour.

2. Connect the Resistance Calibrator to the 8840A for

For each step in Table 6-5, select the indicated range, set the Resistance Calibrator for the specified nominal input, and proceed as follows:

a. Test the 4-wire ohms function:

1. Select the 4-wire ohms function.

2. Verify that the displayed reading is within the limits shown for each reading rate.

b. Test the 2-wire ohms function:

1. Select the 2-wire ohms function. (The

4-wire ohms. SENSE test leads need not be disconnected.) Table 6-3. Low- and Mid-Frequency AC Voltage Test STEP INPUT ERROR TEST LIMITS (IN VOLTS) RANGE IN NUMBER VOLTAGE FREQUENCY COUNTS MINIMUM MAXIMUM 1 av 0.01000V 100 Hz 201 0.00799 0.01201 2 av 0.10000V 100 Hz 114 0.09886 0.10114 3 av 0.30000V 100 Hz 142 0.29858 0.30142 4 2Vv 1.00000V 100 Hz 240 0.99760 1.00240 5" a 1.90000V 100 Hz 366 1.89634 1.90366 6 2Vv 0.10000V 20 Hz 220 0.09780 0.10220 7 2Vv 1.90000V 20 Hz 2380 1.87620 1.92380 8 2vV 0.10000V 45 Hz 135 0.09865 0.10135 9 2vV 1.90000V 45 Hz 765 1.89235 1.90765 10° 200 mV 0.001000V 100 Hz 201 0.000799 0.001201 1 200 mV 0.190000V 20 kHz 366 0.189634 0.190366

“Quick test points.

6-4

Page 110 read from a scan by OCR; it may hold reading errors

Maintenance

PERFORMANCE TEST

Table 6-4. High-Frequency AC Voltage Test

STEP INPUT ERROR TEST LIMITS (IN VOLTS) RANGE IN NUMBER VOLTAGE FREQUENCY COUNTS MINIMUM MAXIMUM 1 200 mV 0.010000V 50 kHz 169 0.009831 0.010169 2 200 mV 0.010000V 100 kHz 350 0.009650 0.010350 3 2Vv 0.10000V 100 kHz 350 0.09650 0.10350 4 20V 1.0000V 100 kHz 350 0.9650 1.0350 5 200V 10.000V 100 kHz 350 9.650 10.350 6" 700V 100.00V 100 kHz 350 96.50 103.50 7 20V 19.0000V 100 kHz 1250 18.8750 19.1250 8 200 mV 0.190000V 100 kHz 1250 0.188750 0.191250 “Quick test points. 2. Zero the test lead resistance by pressing the 1. Ensure the 8840A is on and has warmed up for at

6-7. DC Current Test

OFFSET switch while shorting the HI and LO INPUT test leads together.

2. Verify that the displayed reading is within the limits shown for each reading rate. 3. 4.

The following procedure may be used to test the mA DC least 1 hour.

Select the mA DC function.

Connect the Current Source to the 2A and LO INPUT terminals.

For each step in Table 6-6, set the Current Source for the indicated input and verify that the displayed read-function: ing is within the limits shown for each reading rate. Table 6-5. Resistance Test INPUT ERROR FROM INPUT STEP RANGE (Nominal) (in Counts)

SLOW MEDIUM FAST24 1 2002 02 (short) +4 +0 +3 2 20028 1002 215 +17 #4 3 2 kQ OQ (short) +3 +5 +2 4 2 kQ 1 kQ 213 +15 +3 5 20 kQ O02 (short) +3 +5 +2 6 20 kQ 10 kQ 213 #15 +3 7 200 kQ OQ (short) x3 25 +2 8 200 k22 100 kQ +13 +15 +3 9 2000 kQ 022 (shor) +3 +6 22 10 2000 kQ 1 MQ +30 +33 +5 11 20 MQ 02 (short) +4 +6 22° 12 20 MQ 10 MQ +47 +50 +8

NOTES:

1. Using Offset Control.

2. 4-1/2 Digit Counts.

3. Applies to 4-Wire Ohms Only. 4

When in fast reading rate with internal trigger and transmitting data out of the IEEE-488 interface, the 8840A display must be blanked (Command D1) to ensure stated accuracy.

6-5