Page 111
Maintenance PERFORMANCE TEST
5. Set the Current Source for zero mA and disconnect it from the 8840A.
6-8. AC Current Test (Option -09 Only)
The following procedure may be used to test the mA AC function:
1. Ensure the 8840A is on and warmed up for at least 1 hour.
2. Select the mA AC function.
Connect the AC Current Source to provide a current input to the 2A and LO INPUT terminals. If an ac current source is not available, the functionality of the 8840A can be checked at 10 mA by using a Fluke 5200A. set at 100V and connected to the 8840A 2A terminal through a 10 kQ, 2W, 1% resistor.
4. For each step in Table 6-7, set the AC Current Source for the indicated input and verify that the displayed reading is within the limits shown for each reading rate.
5. Set the AC Current Source to Standby and disconnect it from the 8840A.
6-9. CALIBRATION
CAUTION
To avoid uncalibrating the 8840A, never cycle power on or off while the CAL ENABLE switch is on.
NOTE
If U220 is replaced, perform the Erase Cali- bration Memory procedure (located later in this section) before attempting calibration. Failure to do so may result in an ERROR 29 on the 8840A front panel display.
The 8840A features closed-case calibration using known reference sources. The 8840A automatically prompts you for the required reference sources, measures them, calculates correction factors, and stores the correction factors in the nonvolatile calibration memory.
Closed-case calibration has many advantages. There are no parts to disassemble, no mechanical adjustments to make, and if the TEEE-488 Interface is installed, the 8840A can be calibrated by an automated instrumentation system.
The 8840A should normally be calibrated on a regular cycle, typically every 90 days or 1 year. The frequency of the calibration cycle depends on the accuracy specification you wish to maintain. The 8840A should also be calibrated if it fails the performance test or has undergone repair. To meet the specifications in Section 1, the 8840A should be calibrated with equipment meeting the minimum specifications given in Table 6-1.
The following paragraphs first present a basic calibration procedure. This is followed by a description of advanced features and special considerations, and by a description of remote calibration using the IEEE-488 Interface.
6-10. Basic Calibration Procedure
The basic calibration procedure consists of the following four parts. These parts must be performed in the order shown.
l. Initial Procedure. 2. A/D Calibration.
3. Offset and Gain Calibration for each function and range.
4. High-Frequency AC Calibration (True RMS AC option only).
Normally, it is recommended that the entire calibration procedure be performed. However, under some circumstances the earlier parts may be omitted. For example, if
Table 6-6. DC Current Test
DISPLAYED READING STEP INPUT SLOW MEDIUM FAST MINIMUM MAXIMUM MINIMUM | MAXIMUM MINIMUM MAXIMUM 1 0 mA -000.04 +000.04 -000.06 +000.06 -000.2 +000.2 2 1000 mA dc 999.56 1000.44 999.54 4000.46 999.4 1000.6 Table 6-7. AC Current Test STEP INPUT TEST LIMITS RANGE
NUMBER CURRENT FREQUENCY MINIMUM MAXIMUM 1 2000 mA 1900.00 mA 1 kHz 1890.40 1909.60 2 2000 mA 100.00 mA 1 kHz 97.60 102.40
6-6
Page 112
Maintenance CALIBRATION installing the True RMS AC option, it may be necessary only to perform Offset and Gain Calibration for the ac functions, followed by High-Frequency AC Calibration. But if the A/D Calibration is performed, it must be followed by a complete Offset and Gain Calibration for all functions and then by High-Frequency AC Calibration.
Some of the calibration calculations are complex and take the 8840A some time to execute. For example, when you store the zero input during the Offset and Gain Calibration for the VDC function, it takes around 22 seconds before the next prompt appears. (The 8840A automatically uses this input to calibrate the offset for all ranges.) While the 8840A is executing a calibration step, it ignores all of the front panel buttons and delays execution of all remote commands.
6-11. INITIAL PROCEDURE Always begin the calibration procedure as follows:
1. Allow the 8840A to stabilize in an environment with ambient temperature of 18°C to 28°C and relative humidity less than 75%.
2. Turn the 8840A on and allow it to warm up for at least 1 hour.
3. Enable the calibration mode by pressing the CAL ENABLE switch with a small screw-driver or other Suitable instrument. (The CAL ENABLE switch is located on the right side of the display and is normally covered by a calibration seal.)
When the calibration mode is enabled, the CAL annunciator lights up, and the 8840A displays the first prompt for the A/D Calibration procedure (Figure 6-2). To exit the calibration mode, press the CAL ENABLE switch again.
In the calibration mode, the front panel controls assume the functions described in Figure 6-3. Some of these functions are advanced features and are not Tequired for the basic calibration procedure. The display blanks briefly when a button is pressed.
The following functions are inappropriate during calibration, and are therefore unavailable:
* Offset * Autoranging
* External Trigger
Ss CAL rl v oC
Figure 6-2. First A/D Calibration Prompt
* Front Panel Trigger * Front panel SRO (Under local control)
* Diagnostic self-tests
6-12. A/D CALIBRATION
The A/D Calibration procedure calibrates the analog-to-digital converter for offset, gain and linearity. The 8840A automatically selects the A/D calibration procedure when the CAL ENABLE switch is first pressed. The procedure must be performed in its entirety, and may not be performed in part. If the A/D calibration is discontinued prior to completion, the last complete set of A/D calibration constants will be retained unchanged.
To perform A/D Calibration, proceed as follows:
1. Ensure the Initial Procedure has been completed. The 8840A then displays the prompt for the first reference source, zero volts (i.¢., a short).
2. Each time the 8840A prompts you for a reference source, apply the requested source to the HI and LO INPUT terminals, and press the STORE button. When STORE is pressed, the numeric display field blanks while the 8840A performs the necessary calculations. (Do not change the reference source while the display is blank.) The 8840A then displays the next prompt. For reference, all prompts are shown in Table 6-8.
NOTE
The 8840A automatically checks that the reference input is near the value prompted, and displays ERROR 41 if it exceeds a specific tolerance. (See Advanced Features and Special Considerations, later in this section.)
3. After the last input is stored, the 8840A begins taking readings in the 2V range of the VDC function. (The CAL annunciator remains on.) Verify the A/D calibration using the test points in Table 6-9. If you wish to repeat the A/D Calibration procedure, press the A/D button.
The A/D Calibration procedure is an iterative process. Each pass through the procedure uses the constants stored previously and improves them. Normally, one pass is adequate. However, if the calibration memory has been erased or replaced, or the A/D Converter has undergone tepair, the A/D Calibration procedure must be performed twice.
Because the A/D Calibration procedure is iterative, the instrument’s performance can be slightly enhanced by going through the procedure more than twice. However, this is not necessary to meet the published specifications.
6-13. OFFSET AND GAIN CALIBRATION
This procedure calibrates the instrament’s offsets and gains by applying a high and low input for every range of
6-7
Page 113
Maintenance
CALIBRATION VDC] }vaAc ka kn mA mA { Tac } {Ex Taig} {mre J = ~ 2wiRE| | «wine = ~ STORE asp RATE (er) Cex) (0) 200 2 20 200 2000 20 AUTO VAR IN HF AC ERASE fm un +/- DIGIT1 OIGIT 2 DIGIT 3 DIGIT « OGITs Ot é
Function Each button selects the Offset and Gain Calibration procedure for the respective function.
Buttons
Range These buttons have two functions. Normally, pressing a range button selects the Offset and Gain
Buttons Calibration procedure for that range in the presently selected function. When using the variable input feature, these buttons are used to increment the digits of the displayed prompt. The range buttons are ignored during the A/D Calibration procedure.
STORE Tells the 8840A that the input it is requesting is now present at the input terminals. Pressing the STORE button causes the 8840A to compute and store calibration constants.
A/D Selects the A/D Calibration procedure.
HF AC Selects the High-Frequency AC Calibration procedure.
ERASE Used to erase calibration constants from calibration memory. To complete an erasure of all or part of calibration memory, you must press another button within 1%seconds of pressing this button.
CAUTION If any part of calibration memory is erased, all or part of the calibration procedure must be performed.
RATE Changes the reading rate. Valid only when the 8840A is taking verification readings. Ignored at other times.
VAR IN Allows the use of reference sources that differ from the prompted values. After pressing VAR IN (variable input), the range buttons can be used to change the displayed prompt. The variable input feature is not available during the A/D Calibration procedure.
Figure 6-3. Calibration Functions Table 6-8. A/D Calibration Table 6-9. A/D Calibration Test STEP DISPLAYED PROMPT STEP INPUT ALLOWABLE ERROR
A .0 V DC (Short) A OV (short) +2 counts B - 03 V DC B -0.03V +2 counts Cc -1.01 V DC C +0.03V +2 counts D + 99 V DC D 0.660V +3 counts E + 51 V DC E +0.660V +3 counts F - 51 V DC F -1.970V +4 counts G - 26 V DC G +1.970V +4 counts H + 26 V DC
! + .135 V DC
J - 135 V DC
K - 0725 V DC
L + 0725 V DC
6-8
Page 114
Maintenance CALIBRATION each function. To save time, the 8840A uses each input for aS many ranges as possible.
A function is calibrated by pressing the corresponding function button. Once a function is selected, the 8840A automatically steps through each range of that function, prompting you for the necessary reference sources. (The prompts are shown in Table 6-10.) The 8840A does not automatically select another function after one function has been completely calibrated; therefore, the functions may be calibrated in any order.
To perform Offset and Gain Calibration, proceed as follows:
1. Ensure the A/D Calibration procedure has been completed.
2. Select the desired function by pressing the corresponding function button. The 8840A will display the first prompt for that function.
3. Each time the 8840A prompts you for a reference source, apply this source to the appropriate terminals, and press the STORE button. When STORE is pressed, the numeric display field blanks while the 8840A performs the necessary calculations. (Do not change the reference source while the display is blank.) The 8840A then displays the next prompt. For reference, all prompts are shown in Table 6-10.
NOTE
To use reference sources that differ from the prompted values, see Storing Variable Inputs later in this section.
4. After the last range is calibrated, the 8840A begins taking readings in the highest range so that you may verify its calibration. (The CAL annunciator remains on.) To verify the calibration for the other ranges, press the corresponding range button. (Pressing a function button begins the Offset and Gain Calibration procedure for that function.)
5. Repeat steps 2, 3, and 4 for the remaining functions. Note that both 2-wire and 4-wire ohms functions must be calibrated. (The VAC and mA AC functions require calibration only if the True RMS AC option is installed.)
NOTE
(True RMS AC option only.) The VAC and mA AC functions should normally be calibrated using reference sources at I kHz (+10%). For special applications, performance may be optimized at other frequencies. See Optimizing AC Calibration at Other Frequencies, later in this section.
6. When all functions have been calibrated, exit the calibration mode by pressing the CAL ENABLE switch and attach a calibration certification sticker over the CAL ENABLE switch. (If the True RMS AC option is installed, instead proceed to the High- Frequency AC Calibration procedure which follows.)
6-14. HIGH-FREQUENCY AC CALIBRATION
The High-Frequency AC Calibration procedure calibrates the response of the VAC function from 20 kHz to 100 kHz. If the True RMS AC option is not installed, selecting this procedure results in an error message.
Table 6-10. Offset and Gain Calibration
DISPLAYED PROMPT STEP 2 WIRE kQ voc vac! mA DC mA AC! 4 WIRE kQ A +00.0 mV DC 10.0 mV AC 0.00 Q 00.0 mA DC 100. mA AC (short) (short) (open) B +190.0 mV DC 100.0 mV AC 100.0 Q 1000. mA DC 1000.0 mA AC Cc +1.900 V DC 1.000 V AC 1.000 kQ D +19.00 V DC 10.00 V AC 10.00 ks2 E +190.0 V DC 100.0 V AC 100.0 kQ Steps C-G not applicable F +1000. V DC 500. V AC 1000. kQ for these functions. G Step G not applicabie 10.00 Ma for these functions. NOTE: 1. Inputs: should be at 1 kHz +/-10%. Performance may be enhanced for specific frequencies (see text).
6-9
Page 115
Maintenance CALIBRATION
The reference sources used in this procedure should normally be between 90 kHz and 100 kHz. 100 kHz (nominal) is recommended. For special applications, performance may be optimized at other frequencies. See Optimizing AC Calibration at Other Frequencies, later in this section.
To perform High-Frequency AC Calibration, proceed as follows:
1. Ensure Offset and Gain Calibration has been completed for the VAC function.
2. Select the High-Frequency AC Calibration procedure by pressing the HF AC button. The 8840A will display the first prompt (100 mV AC). The “U” in the display indicates the High-Frequency AC Calibration procedure has been selected.
3. Each time the 8840A prompts you for a reference amplitude, apply this amplitude to the HI and LO INPUT terminals, and press the STORE button. When STORE is pressed, the numeric display field blanks while the 8840A performs the necessary calculations. (Do not change the reference source while the display is blank.) The 8840A then displays the next prompt. For reference, all prompts are shown in Table 6-11.
NOTE
To use reference amplitudes that differ from the prompted values, see Storing Variable Inputs later in this section.
4. After the last range is calibrated, the 8840A begins taking readings in the highest range so that you may verify its calibration. To verify the calibration for the other ranges, push the corresponding range button. The CAL annunciator remains on.
5. The calibration procedure is now completed. Exit the calibration mode by pressing the CAL ENABLE Switch, and attach a calibration certification sticker over the CAL ENABLE switch.
6-15. Advanced Features and Special Considerations
The 8840A has several advanced calibration features which are not necessary for the basic calibration procedure, but which can make calibration easier. The following paragraphs describe these features and also discuss special considerations for optimizing the performance of the 8840A in special situations.
6-16. STORING VARIABLE INPUTS
As a convenience, the VAR IN (variable input) feature lets you calibrate the 8840A using reference source values which differ from the values prompted by the 8840A. For example, you may want to calibrate the 200 range using a reference resistor with a precisely known value of 99.875, rather than 100Q as prompted. This feature is not available during A/D Calibration.
6-10
Table 6-11. High-Frequency AC Calibration
STEP DISPLAYED PROMPT'2 A 100.0 mV AC B 1.000 V AC Cc 10.00 V AC D 100.0 V AC E 200.0 V AC NOTES:
1. The display also indicates "U" to show that HF AC cal is selected.
2.. Inputs should be between 90 kHz and 100 kHz. 100 kHz (nominal is recommended)
To use the variable input feature, proceed as follows:
1. When the 8840A prompts you for an input, press the VAR IN button. The blank digits will be replaced with zeroes. You can then increment each digit of the display by pressing the range buttons. The 200Q/mV button toggles the displayed sign.
2. Change the displayed prompt to correspond to the desired reference source by pressing the appropriate range buttons.
3. Connect the desired reference source to the appropriate input terminals of the 8840A.
4. Press the STORE button.
To meet the specifications over all ranges, the reference source for the high prompts must be between half and full scale. (The high prompts are those prompts that are between 50% and 100% of full scale.) The reference source for the low prompts must be equal to or greater than the prompted value, but not more than that value plus 4000 counts. (The low prompts are those prompts that are zero or 5% of full scale.)
For special applications, the 8840A can be calibrated at values outside the recommended range. This can enhance the performance at the calibration value. However, performance at other values may be degraded.
6-17. CALIBRATING INDIVIDUAL RANGES
During Offset and Gain and High-Frequency AC Calibration, it is possible to calibrate individually selected ranges. This feature does not apply to the mA DC and mA AC functions and is not permitted during A/D Calibration.
To calibrate an individual range, proceed as follows:
1. Select the desired calibration procedure by pressing the appropriate function button (or press the HF AC button if High-Frequency AC Calibration is desired).
Page 116
Maintenance CALIBRATION
2. Press the range button for the range to be calibrated. The 8840A then prompts for a low reference source for that range. (See Table 6-12.) (During High- Frequency AC Calibration, the 8840A prompts only for a high reference source. In this case, proceed to Step 4.)
3. Apply the requested reference source and press STORE. The display will blank briefly and then prompt for a high reference source. (See Table 6-12.)
4. Apply the requested reference source and press STORE. The display will blank briefly, and the 8840A will then begin taking readings in the selected Tange so that you may verify the calibration. The CAL annunciator remains on.
5. To continue, select another range. You may restart any of the calibration procedures by pressing the appropriate function button, the A/D button, or the HF AC button.
6-18. VERIFYING CALIBRATION
When you complete certain parts of the calibration procedure, the 8840A automatically begins taking readings so that you can verify the calibration is correct. (It is recommended that you do so.) The CAL annunciator remains lit. The 8840A continues to take readings until you select another calibration procedure or exit the calibration mode.
Specifically, the 8840A begins taking verification readings after:
Completing A/D Calibration.
2. Completing Offset and Gain or High-Frequency AC Calibration.
3. Completing the calibration of an individually selected range.
While the 8840A is taking verification readings, certain buttons are active or function differently:
1. If you just completed the Offset and Gain or High- , Frequency AC Calibration for an entire function (not just one range), the range buttons can be used to change ranges in order to verify all ranges were calibrated correctly.
2. If you just calibrated an individually selected range, pressing another range button begins the Offset and Gain Calibration procedure for the new range.
3. You can use the RATE button to verify the calibration at other reading rates.
CAUTION it is still possible to erase the calibra- tion memory while the 8840A is taking verification readings. .
6-19. ERASING CALIBRATION MEMORY
The 8840A allows you to erase some or all of the correction constants stored in calibration memory (U220). It is recommended that you erase the entire calibration memory before beginning calibration if the calibration memory is replaced or accidentally altered. The capability of erasing particular parts of the memory is mainly intended as a troubleshooting aid to the technician.
CAUTION
Once the calibration memory is erased, the 8840A must be recalibrated.
To erase all or part of the calibration memory, proceed as follows:
1. Press the front panel ERASE button. The display Should show the erase prompt “cl” (for “clear”). If you do not press another button within 1-1/2 seconds, the 8840A returns to its previous state.
2. To complete an erasure, press one of the following buttons within 1-1/2 seconds of pressing the ERASE button:
Table 6-12. Prompts When Calibrating Individual Ranges
PROCEDURE FUNCTION LOW PROMPT HIGH PROMPT Offset VDC Zero 95% of full scale! and Gain kQ Zero 50% of full scale Calibration > VAC 5% of full scale 50% of full scale
High-Frequency AC Calibration
(Not applicable)
(No low prompt)
50% of full scale?
NOTES:
1. Exception: The 1000V dc range has a high prompt of 1000V dc. 2. Exception: The 700V ac range has a low prompt of 100V ac and a high prompt of 500V ac. 3. Exception: The 700V ac range has prompt of 200V ac.
Page 117
Maintenance CALIBRATION a. STORE -- Erases the entire memory. A/D -- Erases the A/D Calibration constants c. Any function button -- Erases the Offset and Gain Calibration constants for all ranges of that function.
d. HF AC -- Erases the High-Frequency AC Calibration constants.
3. After an erasure is finished (a complete erasure takes about 3 seconds), the 8840A returns to one of the following states:
After complete erasure: Begins A/D Calibration. b. After A/D erasure: Begins A/D Calibration.
c. After Offset and Gain erasure: Begins Offset and Gain Calibration for erased function.
d. After High-Frequency AC erasure: Begins High- Frequency AC Calibration.
6-20. TOLERANCE CHECK
The 8840A automatically checks that the reference input is near the value prompted. This minimizes common errors such as applying a reference source with the wrong sign. If the reference input exceeds the tolerances shown in Table 6-13, the 8840A displays ERROR 41.
If ERROR 41 occurs, the most likely cause is that the reference input is incorrect (e.g., has the wrong polarity). If the input is in fact correct, refer to the Troubleshooting heading in this section.
6-21. AC CALIBRATION AT OTHER FREQUENCIES
For special applications where the 8840A is to be used to measure ac voltages or currents exclusively at a single frequency or narrow range of frequencies, accuracy may be enhanced at that frequency (or range of frequencies) by performing calibration according to the following procedure. Note that this may degrade the accuracy at frequencies significantly removed from the frequency of optimization.
Table 6-13. Tolerance Limits
CALIBRATION
1. A/D Calibration 2. Offset and Gain
TOLERANCE +244 counts from prompt
Calibration VDC, mA DC +488 counts from prompt Ohms +3002 counts from prompt VAC. mA AC +3002 counts from , prompt
3. HF AC Calibration +9999 counts from prompt
6-12
To optimize performance at a frequency less than 1 kHz, perform the offset and gain calibration procedure, above, using the frequency at which measurements will be made rather than 1 kHz. This technique may be used for both the VAC and mA AC functions. At the calibration frequency, the 8840A will yield accuracy closely approaching the specified mid-band performance.
To optimize performance at a frequency greater than 1 kHz, perform calibration as follows:
1. Perform the Offset and Gain Calibration procedure using inputs at 1 kHz.
2. Perform the High-Frequency AC Calibration procedure using inputs at the desired frequency of optimization rather than at 100 kHz. Skip step 5 in that procedure
3. Again perform the Offset and Gain Calibration procedure, this time using inputs at the desired frequency of optimization rather than at 1 kHz.
6-22. OPTIMIZING USE OF THE 5450A
If the Fluke 5450A Resistance Calibrator is used to calibrate the 2-wire ohms function, the following procedure is recommended to optimize the calibration of the lowest two ranges. (The 5450A has a 25 mQ “floor” which would otherwise result in 25 digits of error in the 200Q range of the 8840.) In this procedure, the 8840A is referred to as the unit under test (UUT).
1. Complete Offset and Gain Calibration for the UUT’s 4-wire ohms function. The UUT will then be taking verification readings.
2. Connect the UUT to the 5450A as shown in Figure 6-4,
3. Select the “SHORT” from the 5450A, and measure this value at the 5450A OUTPUT terminals using the UUT in 4-wire ohms. If in remote, take the average of four readings. (In remote calibration, the averaged value can be stored in the controller.) Record the value.
4. Select the “100Q” output from the 5450A, and measure this value as in step 3.
5. Find and record the numerical difference between the values calculated in steps 3 and 4. This value will be used as the variable input for calibrating the 200Q range in 2-wire ohms.
6. Repeat steps 4 and 5 using the “1 kQ” output from the 5450A; find and record the numerical difference between this and the “SHORT” measured in step 3. This value will be used as the variable input for calibrating the 2 kQ range in 2-wire ohms.
7. Press the UUT’s 2 WIRE kQ button. This selects the Offset and Gain calibration procedure for the 2-wire ohms function and prompts for zero input. Select the “SHORT” from the 5450A, and calibrate all the zeros by pressing STORE.
Page 118
Maintenance CALIBRATION
8. Select the “100” output from the 5450A and cali- 13. Using the same configuration shown in Figure 6-4, brate the high point for the 200Q range, entering the verify that the UUT measures the same value (within value computed in step 5 as a variable input. 1 digit) in 2-wire ohms (using the offset feature to
« *. . correct for 5450A floor error) as in 4-wire ohms. If
9. Select the “1 kQ” output from the 5450A and cali- the readings differ by more than 1 digit, reenable the brate the high point for the 2 kQ Tange, entering the calibration mode and repeat steps 2 through 8. value computed in step 6 as a variable input.
. _ 14. Cover the CAL ENABLE switch with a calibration
10. Calibrate the remaining ranges (steps D-G of Table certification sticker.
6-10) using the 5450A outputs.
11. Recalibrate the low point for each 2-wire ohms range 6-23. Remote Calibration using a shorting link (Pomona MDP-S-0 or equiva- If the IEEE-488 Interface is installed, the 8840A can be lent) across the UUT’s HI and LO INPUT terminals. calibrated under remote control. Remote calibration is very similar to local (front-panel controlled) calibration. Table
12. Exit the calibration mode by pressing the CAL 6-14 shows the remote commands which correspond to the ENABLE switch. front panel features.
INPUT HI SENSE HI OUTPUT HI SENSE HI
INPUT LO SENSELO OUTPUT LO SENSE LO, N° 8840A 5450A Figure 6-4. Optimizing Use of the 5450A Table 6-14. Commands Used During Remote Calibration FRONT PANEL CORRESPONDING COMMENTS MMAND FEATURE COMMA
Display G2 Loads the calibration prompt into the output buffer. Not valid when the
8840A is taking verification readings.
Function Fi through F6 In the calibration mode, these select the Offset and Gain Calibration
Buttons procedure for the corresponding function.
Range Ri through R6 In the calibration mode, these select the Offset and Gain Calibration procedure for the corresponding range in the presently selected function. (For entering variable inputs, see VAR IN below.)
STORE Co Tells the 8840A that the requested calibration input is valid. This command causes the 8840A to take readings, and compute and store calibration constants.
NOTE The CO and C3 commands can take up to 22 seconds to execute. You must determine when these commands are complete before sending more commands. See Timing Considerations in text.
AID Ci Selects the A/D Calibration procedure.
Page 119
Maintenance
CALIBRATION Table 6-14. Commands Used During Remote Calibration (cont) FRONT PANEL CORRESPONDING COMMENTS COMMAND FEATURE
HF AC C2 Selects the High-Frequency AC Calibration procedure.
ERASE C3 After receiving this command the display shows the erase mode prompt (cl). (The prompt is not loaded into the output buffer.) To complete the erasure you must then send C0. Sending any other command after the C3 command causes the 8840A to return to its previous state. There is no timeout as with the front panel ERASE button.
CAUTION The command string ‘C3 CO’ erases the entire calibration memory. A complete calibration must then be performed.
RATE SO through S2 Changes the reading rate while the 8840A is taking verification readings. Causes an error at any other time during calibration.
VAR IN N<value>P2 Enters <vaiue> as a variable input. (See Entering Variable Inputs, in text.) Causes an error if sent during A/D Calibration or when the 8840A is taking verification readings. You can check the command was successful by checking the error status, by sending the Get Calibration input command (G2), or by checking the Cal Step Complete bit in the serial poll register.
-- P3<string> Puts the <string> into calibration memory. The string may contain up to 16 ASCII characters, and can be recalled with the G3 command.
NOTE:
Other commands that may be used during calibration are: P1 (Put SRQ mask); the remaining Get commands; and X0
(Clear Error Register).
To facilitate remote calibration, there are some differences from local calibration:
1. In remote calibration, you can store a 16-character message in the calibration memory which can be read by the system controller. Possible uses include storing the calibration date, instrument ID, etc.
2. Although some buttons are ignored in local calibration (e.g., the AUTO button), the corresponding remote commands (e.g., RQ) load the output buffer with an error message.
3. The calibration memory is erased differently. (This is explained later.)
4. The rear panel SAMPLE COMPLETE signal acts slightly differently. During calibration, the SAMPLE COMPLETE signal is inactive. When the 8840A is taking verification readings, the SAMPLE COM- PLETE signal acts the same as in normal operation.
Note that a command may be valid in some parts of the calibration procedure but not in others. The Get Input Prompt (G2) command, for instance, is not valid when the 8840A is taking verification readings. The Rate (Sn) commands, for instance, are valid when the 8840A is taking verification readings, but they are not valid at any other
6-14 time during calibration. Table 6-15 shows when commands are invalid.
6-24. TIMING CONSIDERATIONS
The CO command, can take up to 22 seconds. If during this time the controller continues to send the 8840A more commands, the commands may fill up the 8840A’s input buffer. If this happens, errors will probably occur.
You can avoid this problem by knowing when these commands are completed. There are three ways to determine this:
1. Monitoring the Cal Step Complete bit in the serial poll status register. This status bit is set false every time the remote processor sends a command to the in guard processor. It is then set trme when the in guard processor completes the command and is ready to accept more. So you can send a command and loop on a check of the status, until the command is complete.
2. Setting the SRQ mask to generate an SRQ on Cal Step Complete. An SRQ is generated and the Cal Step Complete bit in the serial poll status response is set when a cal command is complete. This approach depends on capabilities of the controller being used.
3. Executing a delay in controller software after sending each command. (Not recommended.)
Page 120
Maintenance
CALIBRATION Table 6-15. Error Numbers Which Are Displayed When Commands Are Not Valid CALIBRATION MODE COMMAND NORMAL SELF TEST MODE A/D CAL OFFSET & HF AC VERIFICATION GAIN CAL CAL MODE Bn 60 52 52 52 52 co 51 60 54 C1 51 60 C2 51 60 C3 51 60 Dn 60 52 52 52 52 Fn 60 GO G1 G2 51 60 54 G3 60 G4 G5 G6 G7 N PO 60 52 52 52 52 P1 P2 51 60 56 54 . P38 51 60 RO 60 52 52 52 52 R1-6 60 52 R7 60 52 52 52 52 Sn 60 52 52 52 Tn 60 52 52 52 52 Wn x0 Yn 20 60 52 52 52 52 * 60 52 52 52 52 ? 60
Although not usually necessary, these methods can be used for other commands as well.
6-25. REMOTE ERASURE
The C3 command allows you to erase the entire calibration memory. The erasure is executed by sending the string “C3 CO” (equivalent to pressing ERASE and then STORE). Any command other than CO after C3 will abort the erasure. To facilitate remote calibration, the CO command does not timeout as does the front panel ERASE button. The selective erasure that is possible from the front panel is intended as a troubleshooting aid, and is not available over the IEEE-488 Interface.
Note that the erase command can take up to 3 seconds to execute. To prevent timeout problems with the controller, you must determine when the command is completed before continuing. Several methods are presented in Timing Considerations, above.