The text of a manual · part 13 of 18
8840A Digital Multimeter Instruction Manual, pages 121 to 130
Pages 121 to 130 of 180
The original document is the PDF (42 MB). This is its text as extracted, unedited; tables and drawings read best in the PDF, and each page heading opens it at that page.
For the Fluke 8840A
Page 121
Maintenance CALIBRATION
NOTE
When erasing calibration memory, it is good practice to send the commands C3 and CO in the same command string. Sending C3 by itself could lead to accidentally erasing calibration memory, since the C3 command does not time out as does the ERASE button.
6-26. EXAMPLE CALIBRATION PROGRAM
An example A/D calibration program is shown in Figure 6-5. The program is written in Fluke BASIC for the Fluke 1722A Instrument Controller. It uses the Fluke 5440A Direct Voltage Calibrator to perform and then verify. the A/D Calibration procedure. In this program, the 8840A is at bus address 1, and the 5440A is at bus address 7.
6-27. DISASSEMBLY PROCEDURE
WARNING
TO AVOID ELECTRIC SHOCK, REMOVE THE POWER CORD AND TEST LEADS BEFORE DISASSEMBLING THE INSTRU- MENT. OPENING COVERS MAY EXPOSE LIVE PARTS.
CAUTION
To avoid contaminating the printed cir- cuit assemblies (PCAs), handle the PCAs by their edges. Do not ‘handle the areas of the PCAs that are not solder masked unless absolutely necessary. These areas must be cleaned if con-. taminated.
4/D Calibration is correct. Controller and the Fluke bus address 1;
10 y! TA 2 OQ.
R & STEP 1°55 REA »W(A) NEXT A INIT PORT 0 !
90. REMOTE. @1
100 !G0TO 290 !
PRINT "PUT 8840 IN CAL MODE,
INPUT Z$
FOR M=1 TO 2 PRINT @7, “SOUT 0.0; OPER" WAIT 1000 , PRINT @1, "Ci" FOR N=O TO 11 PRINT @1, "G2" INPUT @1,C IFC > 1000 GOTO 180 MASS": “STEP”, “INPUT ™
"SOUT"; C 250 PRINT @1, “co "
V6) W(6) 0.0, 4=0 TO D VCA)
STEP 1
NEXT N
NEXT M “PRINT "VERIFICATION" WAIT 3000 : FOR T=0 TO 6 STEP 1 PRINT @7,."SOUT"; V(T) WAIT 7500 INPUT @1,R RS = NUMSCR, "SH. #88eH" ) IE MID(R$. 2,1)-=- "0" THEN RS. = IF LEFT(R$.1) = " “ THEN RS PRINT R$,
END
03, 2,-0.03, 2, 0.66, 3, =0, £6, 3, 1. LOOP TO READ IN VERIFICATION VALUES.
PROVIDES MEANS To DO RIFICATION ONLY. THEN HIT CRETURN>.” ! ! STOP PROGRAM OF PASSRE URN? Is HIT.
Vm meme
: : ! aa CSPL(1%) AND 8%) <> 8% GOTO 260
LE “+"+RIGHT (RS,
' LOOP F
This program performs the A/D Calibration procedure and then verifies the The program uses the Fluke 172D3A Instrument
S440A Direct and the S440A is at bus
Voltage Calibrator.
The S840A is at address 7.
97; 4,-1.97, 4
INITIALIZE SYSTEM.
!_ INSTRUCTION NOTE.
NO. OF PASSE SET 5440A OUTPUT TO 0.0
PUT 8840A_ IN. A/D CAL. CALIBRATION LOOP.
GET 8840A CAL PROMPTS. CLEARS POSSIBLE ERROR MESSAGE IN BUFFER. LABELS
DISPLAY PASS, STEP, INPUT VALUE. SET OUTPUT OF 5440A. SETTLING TIME FOR 5440A.
STORE CAL.
! TEST 8840A FOR CAL STEP COMPLETE.
VERIFICATION LOOP.
SET 5440A_GUTPUT.
SETTLING TIME FOR 5440a.
SET READING FROM 8840a,
THIS AND NEXT 3 LINES FORMATS DISPLAY. FT(RS, L)+"_"+RIGHT(RS, 3)
+2) PRINTS Di LAY
: SPLAY. LF ABS(R-V(T) € SWC T) #1E-S+1E-6 THEN PRINT "PASS" ELSE PRINT "FAIL"
Figure 6-5. Example A/D Calibration Program
6-16
Page 122
| _ Maintenance _ DISASSEMBLY PROCEDURE -
The following paragraphs present a disassembly procedure for the 8840A. The procedure should be performed in the order presented: Remove the case first, and. then. remove Option -09 True RMS AC PCA, Option -05 IEEE-488 Interface PCA, the Main PCA, and the front panel. For reference, see the final assembly drawing in Section 7.
6-28. Case Removal
1. Remove the grounding screw from the bottom of the case. Remove two rear bezel mounting screws. (See Figure 6-6A.)
2. While holding the front panel, slide the case and rear bezel off the chassis (See Figure 6-6B). (At this point, the rear bezel is not secured to the case.)
6-29. True RMS AC PCA Removal (Option -09)
The True RMS AC PCA should be removed by reversing the last three steps in Figure 809-1 (see Section 8).
1. Release the True RMS AC PCA from the chassis by pulling the four plastic latches upward (Figure 809-1E).
2.. Raise the True RMS AC PCA slightly, and disconnect the red lead from the connector (J301) located on the Main PCA (Figure 809-1C).
3. Disconnect the ribbon cable from the Main PCA by releasing the ribbon.connector latches (push outward; see Figure 809-1D) and pulling the ribbon. cable directly outward from the connectors (Figure 809-1C).
"4. Lift the True RMS AC PCA out of the chassis.
6-30. IEEE-488 Interface PCA Removal (Option -05 Only) The IEEE-488 Interface PCA should be removed by reversing the last four steps in Figure 805-1 (see Section 8):
1. Remove the two jack screws and washers from the rear panel JTEEE-488 .connector (Figure 805-15).
2. Release the IEEE-488 Interface PCA from the chassis by pulling the two plastic latches upward (Figure 805-1H).
3. Raise the forward edge of the IEEE-488 Interface
_ PCA slightly, pull the PCA towards the front panel
(guiding the IEEE-488 jack and BNC connectors out of the rear panel), and lift the PCA out of the chassis (Figure 805-1G):
4. Disconnect the ribbon cable from the IEEE-488 Interface PCA by releasing the connector latches (push outward; see Figure 805-1F) and pulling the ribbon cable out from the connector (Figure 805-1E).
6-31. Main PCA Removal
1. Disconnect the leads from the four front panel input . terminals and the four rear panel input terminals by unplugging them. (Refer to Figure 6-6C.)
2. Remove the cable harness from the two cable clamps.
on the side of the instrument chassis. (Figure 6-6D.) Lift ‘the cable harness clear of the sidewall ‘cable guide.
10.
11. 12.
the 2A input terminal slightly and rotating it 1/4-tum
~ SET) PCA as follows (Figure 6-6G): a. Remove the upper screw that holds the LINE ber
Remove the front panel fuse by pressing in the lip of counterclockwise (Figure 6-6E).
(Disregard this step if the IEEE-488 Interface | was installed.) Disconnect the ribbon cable from the rear panel insert by pushing outward on the snap tab on either side of the ribbon cable connector.
Disconnect the two ribbon cables from the Display PCA by pulling the two plastic pull tabs direct! outward from the Display PCA.
Remove the two mounting screws on either side of the rear panel power receptacle.
Disconnect the green power supply ground lead from the rear panel mounting stud. (The stud is located near the rear panel power receptacle. See Figure 6-6F.)
Remove the Line Voltage Selection Switch (LINE
PCA to the upper rear panel standoff.
b. Unplug the ribbon cable from the Main PCA|and lift out the LINE SET PCA. |
| Remove the push rod for the CAL ENABLE switch as follows (Figuré 6-6H): |
| a. While supporting the white plunger of the CAL ___ ENABLE switch with a finger, pop the pushrod ~ Off the switch plunger by pulling the push | rod’ directly upward.
| b. Rotate the push rod 90 degrees toward the center of the instrument.
c. Pull the push rod toward the rear panel jand remove it.
Remove the FRONT/REAR switch push rod as follows (Figure 6-61):
a. Insert a blade-type screw driver in the slot visible on the top of the FRONT/REAR switch push|rod at the junction of the push rod and the switch. | b. Twist the screwdriver slightly to release the push rod from the switch shaft, then pull [the
FRONT/REAR switch push rod out through| the front panel.
Place the chassis on its side.
Remove the POWER switch push rod as follows. (Figure 6-65): | a. Insert a blade-type screwdriver in the slot visible. on the top of the POWER switch push rod at| the " Junction of the push rod and the switch. | b. Twist the screwdriver slightly to release the push - tod from the switch shaft, then pull the rear of the POWER switch push rod out through the bottom of the chassis. | c. Lift the push rod out and toward the rear panel, ~~ and remove it, a
Page 123
Maintenance DISASSEMBLY. PROCEDURE : ; ead g- - - - MOUNTING SCREW (1 OF 2) REAR PANEL : A. - PS eréuniome SCREW REAR BEZEL i His, ORANGE BROWN 'B. CHASSIS
6-18
Figure 6-6. 8840A Disassembly
Page 124
6-19
Figure 6-6. 8840A Disassembly (cont)
DISASSEMBLY PROCEDURE
Maintenance
Page 125
Maintenance DISASSEMBLY PROCEDURE
Figure 6-6. 8840A Disassembly (cont) 6-20
Page 126
Maintenance
DISASSEMBLY PROCEDURE
Figure 6-6. 8840A Disassembly (cont)
6-21
Page 127
Maintenance DISASSEMBLY PROCEDURE
13.
14.
15.
16.
17.
18.
Remove the two screws fastened to the transformer bracket, then remove the bracket. ‘
Remove the Main Shield as follows (Figure 6-6k):
a. Remove the screw that fastens the Main Shield to the Main PCA.
b. Grasp the Main Shield supports on one side of the instrument and pull the supports toward the center of the chassis, bowing the Main Shield. Remove the main shield.
Release the six plastic latches that hold the Main PCA to the chassis by pulling the latches upward (Figure 6-61).
Lift the front end of the Main PCA upward about 3 inches.
Free the white lead from the 2A INPUT tower as follows (Figure 6-6m):
a. Guide the wire, spring and fuse contact toward the front panel.
b. Thread the spring and fuse contact through the hole in the front end of the tower.
Slide the Main PCA forwards until it is free of the chassis.
6-32. Front Panel Disassembly
1.
Holding the chassis vertically (with the front panel downward), remove the mounting screws from the four corners of the Display PCA (Figure 6-7A).
Holding the chassis vertically (now standing the instrument on the rear panel), pull the front panel off the chassis and set it aside (Figure 6-7B).
Remove the the display assembly (Display PCA, spacer matrix, and keypad) from the chassis as follows (see Figure 6-7C):
a. Release the two plastic tabs on the front of the chassis.
b. Let the bottom edge of the display assembly swing toward the rear of the instrument.
c. Pull the Display PCA toward the bottom of the chassis.
Separate the spacer matrix from the Display PCA by releasing the two pairs of plastic snap tabs on the back of the Display PCA (Figure 6-7D).
Remove the keypad from the spacer matrix.
CAUTION
The vacuum fluorescent display should not be removed from the Display PCA; these are supplied as one part.
Remove the display window from the front panel as follows:
a. Slide the window upward (away from the buttons) about 1/32 inch (Figure 6-8A).
6-22 b. Push the window directly outward from the front panel (Figure 6-8B).
6-33. REASSEMBLY PROCEDURE To reassemble the instrument, proceed as follows:
1.
Assemble the front panel assembly by reversing the front panel disassembly procedure (Figure 6-7). (It is easiest to lay the keypad on the Display PCA before installing the spacer matrix on the Display PCA.)
CAUTION
The four Display PCA mounting screws are self-tapping. To avoid damaging the threads, ensure the screws are threaded properly before tightening. Do not overtighten them.
Turn the chassis upside down.
Install the Main PCA through the bottom of the chassis as follows:
NOTE
When installing the Main PCA, guide the rear ribbon cable around the shield connected to the rear panel so that the cable is next to the side of the chassis. Make certain that the cable is not pinched between the shield and the Main PCA.
a. Slide the Main PCA toward the rear panel, and position the power connector and fuse to fit through their respective openings in the rear panel.
b. Reinstall the white lead in the 2A INPUT tower and reinstall the 2A fuse in the front panel.
c. Make sure the six plastic latch heads are extended. Lower the Main PCA into position on the chassis and guide the six plastic latches into the circuit board supports on the chassis. Press the latch heads to lock the board in the chassis. Refer to Figure 6-6L.
Install one side of the Main Shield; bow it to install the the other side, and secure it to the Main PCA with the retaining screw.
Place the transformer bracket back into position and fasten down with the two screws.
Replace the push rod for the POWER switch as follows:
a. Insert the button end of the push rod into the rear of the front panel.
b. While supporting the opposite side of the plunger with your finger, snap the other end of the push rod onto the POWER switch plunger. Refer to Figure 6-6].
Page 128
6-23
Figure 6-7. Front Panel Disassembly
DISPLAY PCA
DISPLAY PCA (REAR VIEW)
SPACER MATRIX
MOUNTING SCREW (1 OF 4)
DISPLAY PCA ay
FRONT PANEL
- REAR PANEL
REASSEMBLY PROCEDURE
Maintenance
Page 129
Maintenance REASSEMBLY PROCEDURE
FRONT PANEL
DISPLAY WINDOW
10.
11.
12.
Figure 6-8. Removing the Display Window
Set the chassis right side up.
Insert the FRONT/REAR switch push rod through the front panel and snap it into place.
Reinstall the CAL ENABLE switch push rod by inserting the cylindrical end of the push rod into the rear of the front panel, then snapping it onto the CAL ENABLE switch plunger.
CAUTION
Make certain that the CAL ENABLE switch shaft is in the out (disabled) position after the CAL ENABLE push rod is installed. If the 8840A is switched on with the CAL ENABLE switch in the enabled position, the 8840A may require recalibration.
Position the slot in the lower edge of the Line Voltage Selection Switch PCA in the slot on the lower rear panel standoff. Secure the top of the Line Voltage Selection Switch PCA to the upper standoff using the single mounting screw, and plug the ribbon cable into the Main PCA.
Connect the power supply ground lead to the rear panel mounting stud. (The stud is located near the rear panel power receptacle as shown in Figure 6-6F.)
WARNING
TO AVOID ELECTRIC SHOCK, ENSURE THAT THE POWER SUPPLY GROUND LEAD IS FIRMLY ATTACHED TO THE REAR PANEL MOUNTING STUD.
Attach the two mounting screws on either side of the Tear panel power receptacle.
6-24
13.
14.
15.
16.
17.
18.
19. 20. 21.
6-34.
Connect the two ribbon cables to the Display PCA to the connectors. Push the cables straight in to avoid damage.
Reinstall the harness in the sidewall cable guide, and secure the harness to the chassis with the cable clamps.
Connect the leads to the four front panel input terminals according to the color codes marked on the rear side of the Display PCA.
Connect the leads to the four rear panel input terminals following the color codes as shown in Figure 6-6B.
(Option -05 only) Install the IEEE-488 Interface PCA according to the instructions in Section 8.
(Option -09 only) Install the True RMS AC PCA according to the instructions in Section 8.
Slide the case and rear bezel onto the chassis. Install the two rear panel mounting screws.
Install the case grounding screw in the bottom of the case.
WARNING
TO AVOID ELECTRIC SHOCK, ENSURE THAT THE GROUNDING SCREW IS FIRMLY ATTACHED TO THE CASE BOTTOM.
INTERNAL FUSE REPLACEMENT
CAUTION
For fire protection, use exact fuse replacements only.
The 8840A has an internal 3A 600V slow-blow fuse (F301) in series with the 2A input terminal. To replace this fuse, remove the case according to the disassembly instructions. The fuse is held in fuse clips on the Main PCA. Do not use makeshift fuses or short-circuit the fuse holder.
Page 130
Maintenance EXTERNAL TRIGGER POLARITY SELECTION (Option -05 Only)
6-35. EXTERNAL TRIGGER POLARITY SELECTION (Option -05 Only)
The EXT TRIG input is factory-configured with negative polarity (trigger on falling-edge). This polarity is set by jumper E902 on the IEEE-488 Interface PCA. To select positive polarity (trigger on rising-edge), remove jumper E902 and add jumper E903.
6-36. TROUBLESHOOTING
The 8840A is designed to be easily maintained and repaired. Both the analog and digital circuits have built-in diagnostic self-tests and troubleshooting modes to facilitate troubleshooting and repair. The instrument’s circuits allow troubleshooting and repair with basic electronic troubleshooting equipment such as a multimeter and oscilloscope. The troubleshooting mode in the digital controller Circuitry generates special test signals to allow troubleshooting and repair without a special test signal generator or complex logic analyzer. Using the information in this section, a technician should be able to troubleshoot and repair the 8840A very efficiently.
6-37. Initial Troubleshooting Procedure
WARNING
TO AVOID INJURY OR EQUIPMENT DAMAGE, USE EXACT REPLACEMENT PARTS FOR ALL PROTECTION COM- PONENTS.
When a problem occurs in the 8840A, first verify the problem is actually in the instrument. If the problem occurs when the instrument is in a system, check to see if the same problem exists when under local control. Watch the display as the instrument is turned on to see if any of the digital self-test error codes appear indicating a digital failure. If the malfunction does not involve the True RMS AC or IEEE-488 options, remove the option(s) from the instrument before proceeding.
If the display lights up, perform the self-test by pressing the SRQ button for 3 seconds. (Remember, the input terminals must be disconnected from the test leads during the self-tests. Otherwise, the 8840A may indicate errors are present.) The test numbers will appear consecutively. “ERROR” will appear on the display if a test should fail. The 8840A can be held in each of the test configurations by momentarily pressing the SRQ button. (Press any button to continue the tests.) With the description of the self-tests given below, it may be possible to isolate the failure. For reference, the states of various switches and logic lines are shown in Table 6-16 for each function, Tange, and reading rate.
If only one or a few failures appear in the self-tests, the problem is usually in the DC Scaling circuit. By carefully analyzing which failure(s) occurred, the fault can be located to within a few components. (Table 6-17 shows which components are exercised by each of the analog tests.) The heading DC Scaling Troubleshooting provides detailed instructions on locating and repairing DC Scaling circuit problems. However, before troubleshooting the DC Scaling circuit, all of the power supply levels should be measured to verify they are within the limits specified in Table 6-23 under Power Supply Troubleshooting, later in this section.
Some failures will cause many self-tests to fail. If this occurs, the fault is usually in the Track/Hold circuit, the A/D Converter, the Digital Controller circuit, or the Power Supply. Again, measure all of the power supply levels according to the limits specified in Table 6-23. The next Step is to isolate the problem to a specific section.
If the self-tests display a large number of errors or if readings are noisy and/or in error, the problem is usually in the A/D Converter or Track/Hold circuit. (A large aumber of errors can also be caused by a problem in the Ohms Current Source.) To isolate the problem, connect a jumper between TP103 and Reference Low (TP306, or the L-shaped shield around U202). The display should typically read less than approximately 35 counts (i.e., +.000XX where XX is less than 35) on the 2V dc range. If a good reading can be obtained (less than approximately 35 counts), the A/D Converter and Precision Voltage Reference circuits are most likely good. A more conclu-sive test can be made by connecting a low-impedance dc source between Reference Low and TP103 with an output voltage between -2.0V and +2.0V. The reading on the display will be of opposite polarity to the voltage applied to TP103. (Disconnecting one end of R318 will usually make it possible to display readings within 0.1% to 0.5% of the actual input.) After it has been determined that the A/D Converter or the Track/Hold circuit is not functioning properly, proceed to the corresponding heading for detailed troubleshooting instructions and guidelines.
A failure in the instrument may cause the 8840A to display random patterns or nothing at all. Usually, analog circuit failures do not cause the display to go blank or display random patterns. The best place to start troubleshooting a “dead” instrument or an instrument with a non-functional display is to check the power supply with a voltmeter for proper levels and to use an oscilloscope to check the supplies for oscillations. If all of the supplies are working correctly, check the clock for the In-Guard uC at U202-2. The signal should be an 8 MHz sine wave approximately 3.5V peak-to-peak. Then check the 1 MHz output of the A/D IC (U101) at U212-3. (If not present, check at the A/D IC at U101-14.) The signal should be a 1 MHz square wave approximately 5V peak-to-peak. The 8 MHz sine wave is generated by the clock circuit of the In-Guard uC, and the 1 MHz signal is the 8 MHz signal divided by a counter in the A/D IC. If the clock signals are correct, proceed to the heading Digital Controller Troubleshooting, below, for detailed troubleshooting instructions.
6-25