Page 31
Operating instructions _ OPERATING | EATURES
ERROR a7 af Ls Error 30: AC function selected without True RMS AC Converter option.
Figure 2-7. Typical Error Message indicates when the input exceeds the input overload limit of 1000V dc or 700V ac, respectively, by lighting the OVER annunciator and flashing the display. Readings are still displayed.
2-13. Diagnostic Self-Tests
The 8840A features diagnostic self-tests which check both the digital and analog Circuitry in the instrument. The self-tests consist of 21 analog tests followed by the in-guard program memory, calibration memory, and display self-tests. To initiate the self-tests, press the SRO button for 3 seconds. The instrument can be stopped in any of the test configurations by pressing the SRO button while the test number is displayed. Press any button to continue the tests.
During the test, the TEST annunciator lights, and the numeric field displays the number of each analog test as it is performed. Then all display segments light up while the instrument performs the in-guard program memory, calibration memory, and display self-tests. The 8840A then retumms to the power-up configuration. The self-tests are described in greater detail in the Maintenance section.
NOTE
The inputs must be left disconnected while the self-tests are performed or the 8840A may indicate that errors are present.
If the 8840A detects an error, it displays an error message for about 2-1/2 seconds. (Error codes 01 through 29 correspond to the self-tests.) If self-test errors are displayed even when the input terminals are disconnected, there may be a hardware problem in your 8840A. In that event, refer to the Maintenance section or contact your local Fluke representative.
2-14. Ranging
Measurement ranges can be selected using either autorange (by pressing the AUTO button) or manual range (by pressing another range button). The 8840A displays explicit units in every range, so that the display may be read directly.
2-15. AUTORANGE
In autorange, the 8840A goes to a higher range when the input exceeds full scale (199999 counts), and goes to a lower range when the input falls below 9% of full scale (18000 counts). While the instrument changes range, the numeric field on the display is blanked until a new reading is completed. However, the decimal point and units annunciators always indicate what range the instrument is in.
Pressing the AUTO button when the instrument is already in autorange toggles the 8840A from autorange to! manual range. This causes the instrument to remain locked in the present range.
2-16. MANUAL RANGE.
In manual range, the 8840A remains fixed in the selected range until you select another range or press AUTO. If you select a range which is not valid for the present function, or select a function which is not valid for the | present range, the 8840A selects the nearest valid range. For example, if the 8840A is in the VDC function and you press the 20 MQ button, the 8840A selects the 1000V Tange. i
The range buttons have no effect in the mA DC and mA AC functions, since all measurements in these functions are made in the 2000 mA range.
2-17. Triggering
Triggering causes the 8840A to execute a measurement cycle and display the result. During each measurement cycle, the instrument samples the input a numberof ines and then averages the samples to compute a reading. The number of samples averaged for each reading lepends upon the reading rate. |
Each time a reading is triggered, the rate annunciator (S, M, or F) blinks off. In the fast reading rate, the F annunciator flashes so rapidly it appears to be almost constant.
How the 8840A is triggered depends on whether the continuous trigger mode or external trigger mode is Selected. Pressing the EX TRIG (external trigger) button toggles the 8840A between the two modes.
2-18. CONTINUOUS TRIGGER MODE
In the continuous trigger mode, readings are triggered bya continuous, internal trigger. The rate of the trigger is set by the RATE button.
2-19. EXTERNAL TRIGGER MODE
In the external trigger mode, readings are triggered by pressing the TRIG button. If the IEEE-488 Interface! option is installed, readings can also be triggered by remote commands or by using the rear panel external trigger (EXT TRIG) connector. (See the Options and Accessories section.) |
Page 32
a
Operating Instructions OPERATING FEATURES
Table 2-1. Error Codes
ERROR
ERROR
NOTE: See the Maintenance section for a detailed description of self-tests.
CODE MEANING CODE MEANING ANALOG SELF-TEST ERRORS 1 200. VAC, Zero 40 Computed calibration constant out of nay Be 2 700 VAC, Zero (Previous cal may be wrong or there may 3 mA AC, Zero a hardware problem.) 4 mA DC, Zero 5 200 VDC, Zero 41 Calibration input out of acceptable 1: 6 1000 VDC, Zero Check that input is correct. (Previous cal 7 1000 VDC + 20 MQ may be wrong or there may be a hardware 8 20 VDG + 20 MQ problem.) 9 20 VDC + 2000 kQ 10 .2 VDC + 2000 kQ 42 Calibration memory write error. (Probably a 11 2002, Overrange hardware probiem.) 12 2 kQ, Overrange 13 20 kQ, Overrange 50 Guard crossing error detected by in-Guard 14 200 kQ, Overrange uc. 15 1000 VDC + X10 T/H + 20 MQ 51 Calibration command not valid unless cali- 16 200 VDC + 200 kQ bration mode is enabied. 17 200 VDC + 20 kQ 18 200 VDC + 2 kQ 52 Command not valid at this time. 19 200 VDC, Filter On . 20 200 VDC + 2 kQ, Filter Off 53 invalid calibration value in Put command. 21 200 VDC, Filter Off (Example: Sending a negative value during ac calibration.) | DIGITAL SELF-TEST ERRORS 25 In-Guard uC Internal RAM 26 Display RAM 54 Command not valid in calibration verification. 27 In-Guard uC Internal Program Memory 28 External Program Memory 56 Variable inputs not allowed during A/D call 29 Calibration Memory bration. Use prompted value. OPERATION ERRORS 30 AC functions available only. with True RMS 60 Device-dependent commands not valid dur. AC option. ing self-tests 31 mA AC or mA DC function selected while 71 Syntax error in device-dependent command REAR inputs selected. String. 72 Guard crossing error detected by Out-Guard uc. 32 OFFSET selected with reading unavailable 73 _Guard crossing error detected at power on or overrange. or CAL ENABLE switch on at power on. 77 IEEE-488 Interface _self-test_error.
Page 33
Operating Instructions OPERATING FEATURES
OVER
£ v 0¢ s .
! I.
OVER: Input 22V dc in 2V range.
Figure 2-8. Overrange Indication
In the external trigger mode, pressing any front panel button blanks the numeric field on the display until a new measurement is triggered. This ensures that all readings correspond to the instrument configuration indicated by the display annunciators. The blanking also occurs in the continuous trigger mode, but usually isn’t noticed because new measurements are triggered automatically.
The TRIG button does not trigger readings in the continuous trigger mode. However, it does blank the last reading . to acknowledge a button was pressed.
2-20. Reading Rates and Noise Rejection
The RATE button allows you to optimize either measure- _ Ment speed or noise rejection. The 8840A uses both analog and digital filtering.to allow measurements in the présence of unwanted environmental noise (especially line-related noise). However, since filtering introduces a delay in Tesponse to a change in the input signal, there is an inherent trade-off between noise rejection and measurement speed.
The instrument has three Teading rates: slow (S) and medium (M), with a 5-1/2 digit display, and fast (F), with a 4-1/2 digit display. To provide optimum combinations of measurement speed and noise rejection, the RATE button allows control of both the internal trigger rate and the degree of filtering. The same degree of filtering is used in both the continuous and external trigger modes.
In the continuous trigger mode, the actual number of readings displayed per second for each reading rate is determined by the line-power frequency. At power-up, the 8840A senses the line-power frequency and adjusts the analog-to-digital converter timing characteristics for optimum normal-mode noise rejection. The resulting reading rates are shown in the specifications in Section 1.
2-21. Automatic Settling Time Delay.
When the external trigger mode is selected, the 8840A automatically inserts a delay after receiving a trigger signal, but before starting the first input sample. The delay is just long enough so that the reading will be correct (within a specified number of counts of the final value) even if the trigger signal occurs as the input makes a step change between zero and full scale (10,000 counts and full scale in the ac functions). For resistance measurements, the delay also allows for settling when the step change is made from an open circuit to an in-range value.
The length of the delay depends on the Tange, function, and reading rate, as shown in the specifications in Section 1, The delay is enabled only in the external trigger mode. It can be turned off with a remote command over the IEEE-488 interface bus to accommodate special timing considerations.
2-22. External Trigger Input (Option -05 Only) The rear panel EXT TRIG input is a TTL-level input which can be used to trigger measurements when’ the 8840A is in the external trigger mode. A measurement is triggered on the falling edge of the input. Since the EXT TRIG input is pulled high internally, it can also be controlled by a normally open switch. A measurement is triggered when the switch is closed.
For special applications using the IEEE-488 Interface, the automatic setting time delay can be disabled using remote commands. (See Section 3.) Refer to Section 1 for timing details.
The polarity of the EXT TRIG input can be reversed by changing internal jumpers. Refer to the Maintenance section for instructions.
2-23. Sample Complete Output (Option -05 Only) The SAMPLE COMPLETE output indicates when analog input sampling for a reading is completed. The output is a TTL-level signal which is pulsed low for approximately 2.5 us when the input-sampling portion of the A/D conversion is completed. The signal is useful for interfacing with other equipment when the 8840A is used in external trigger mode in an instrumentation system. For example, the SAMPLE COMPLETE output could be used to advance a scanner to the next channel..
2-24. MAKING MEASUREMENTS
2-25. Input Overload Protection Limits
WARNING
TO AVOID SHOCK HAZARD AND/OR INSTRUMENT DAMAGE, DO NOT APPLY INPUT POTENTIALS THAT EXCEED THE INPUT OVERLOAD LIMITS SHOWN IN TABLE 2-2.
The 8840A is protected against input overloads up to the limits shown in Table 2-2. Exceeding these limits may damage the instrument and/or pose a shock hazard.
. 2-9
Page 34
Operating Instructions MAKING. MEASUREMENTS
2-26. Measuring Voltage and Resistance
To measure voltage or resistance, select the desired function and connect the test leads as shown in Figure 2-9. Resistance can be measured in either the 2-wire or 4-wire configuration.
2-27. Measuring Current
To measure current, select the desired function and connect the test leads as follows:
1. Turn off power in the circuit to be measured (see Figure 2-10).
2. Break the circuit (preferably on the ground side to minimize the common mode voltage), and place the 8840A in series at that point.
3. Turn on power in the circuit, and read the display.
4. Turn off power in the circuit, and disconnect the 8840A.
2-28. Current Fuse Protection
The 2A input terminal is protected from overloads by a 2A, 250V fuse which is accessible from the front panel, and by an internal 3A, 600V fuse. If either fuse blows, the 8840A will respond as though the input were zero.
WARNING
TO AVOID ELECTRIC SHOCK, REMOVE THE TEST LEADS BEFORE REPLACING THE FRONT PANEL FUSE.
To replace the front panel fuse, first remove the test leads. Then press in the lip of the 2A input terminal slightly and rotate it 1/4-turn counterclockwise. Spring tension will force the fuse and fuse holder out of the front panel. The internal 3A fuse should be replaced only by qualified service personnel.
2-29. Offset Measurements
WARNING
WHEN THE OFFSET FEATURE IS IN USE, DISPLAYED READINGS ARE RELATIVE AND MAY NOT INDICATE THE PRESENCE OF DANGEROUS POTENTIALS AT THE INPUT CONNEC- TORS OR TEST LEADS. USE CAUTION TO AVOID ELECTRIC SHOCK OR INSTRUMENT DAMAGE.
The OFFSET feature allows you to store a reading as a relative reference value. When the OFFSET button is pressed, the 8840A stores the present reading and displays subsequent measurements as the difference between the measured value and the stored reading. The OFFSET annunciator is lit whenever. an offset is in use.
The OFFSET feature may be used in all functions. Since the display represents a numeric difference, it always has a sign, even in the resistance and ac functions.
The offset can be cancelled by pressing the OFFSET button again, in which case the OFFSET annunciator disappears from the display. The. offset can also be! cancelled by storing an offset in another function. If a reading is overrange or unavailable when the OFFSET. button is pressed, the 8840A indicates ERROR 32 and does not store the offset.
If you change functions while an offset is stored, the OFFSET annunciator disappears and the offset temporarily disappears. However, when you retum to the original function, the offset is restored (and the OFFSET annunciator reappears) unless a new offset was established in another function.
Note that the input overload limits are not changed by the use of the offset feature. However, the display flashes if the 8840A is in the 1000V dc or 700V ac ranges and the input exceeds 1000V dc or 700V ac, respectively.
Table 2-2. input Overload Limits
All Functions
FUNCTION CONNECTORS MAXIMUM INPUT VDC , INPUT HI and LO: 1000V de mA.DC 2A INPUT and INPUT LO: 2000 mA 2 WIRE/4 WIRE kQ INPUT HI and LO: 300V rms SENSE HI and LO: 300V rms _
VAC INPUT HI and LO: 700V rms, 1000V peak, or 2 x 107 V-Hz (whichever is less) mA AC 2A INPUT and INPUT LO: 2000 mA rms
Any terminal to earth:
1000V dc or peak ac
2-10
Page 35
Operating Inetructions MAKING MEASUREMENTS voc} | vac ko - ma) foe's] sanna ne = D m os mo ‘Ooo cal
Figure 2-10. Measuring Current
While an offset is enabled, the 8840A indicates an overrange condition if either of the following conditions occur:
° The input signal is overrange * The calculated reading is overrange
For example, suppose the instrument is in the 20V range of the VDC function and you store an offset of +15V. The maximum positive voltage reading that can be displayed without overranging is +4.9999V, which is actually a +19.9999V input signal. The maximum negative voltage reading that can be displayed without overranging is kn SENSE terminals for -19.9999V, which is actually a -4.9999V input si al. You 4 WiRE 4-wire resistance can measure a greater range of voltages by selecting a measurement only. higher range.
When in autorange, the 8840A selects the range appro-
= priate for the input signal, regardless of any stored offset.
rm «ko If, for example, a +10V offset is stored, and a +1V input is
Lt 4 WIRE applied, the 8840A will autorange to the 2V e and
, display an overrange condition since it cannot di: May -9V for ae wes on the 2V range. Manual range control could be jused to oo; oo lock the 8840A into the 20V range in this case. OO re] |
Applications of the offset feature include correcting for test lead resistance in 2-wire resistance measurements, nulling offset currents or voltages, measuring voltage devi-ations, and matching resistors.
2-30. EXTERNAL CLEANING
Vl WARNING
TO AVOID ELECTRIC SHOCK OR INSTRUMENT DAMAGE, NEVER GET WATER INSIDE THE CASE. TO AVOID INSTRUMENT DAMAGE, NEVER APPLY SOLVENTS TO THE INSTRUMENT. |
} ; } Should the 8840A case require cleaning, wipe the instrv- Figure 2-9. Measuring Voltage and Resistance ment with a cloth that is lightly dampened with water or a mild detergent. solution.
Page 36
Operating Instructions EXTERNAL CLEANING
2-12
Page 37
NOTE
This section contains programming instruc- tions for use with the IEEE-488 Interface (Option -05). For installation instructions, refer to the Options and Accessories section
3-1. INTRODUCTION
The IEEE-488 Interface tums the 8840A into a fully programmable instrument for use with the IEEE Standard 488-1978 interface bus (IEEE-488 bus). With the IEEE-488 Interface, the 8840A can become part of an automated instrumentation system. The 8840A can. be under complete, interactive control from a remote bus controller; or it can be set to the talk-only mode, connected to a data logger or printer, and dedicated to a single task.
This manual assumes you know the basics of the TEEE-488 interface bus. For an introduction to the bus, request Fluke Application Bulletin AB-36, “IEEE Standard
488-1978 Digital Interface for Programmable Instrumentation.”
3-2. CAPABILITIES
The IEEE-488 Interface provides remote control of all front panel controls except for the POWER, CAL
ENABLE, and FRONT/REAR switches. Other features include:
* A simple and predictable command set ¢ Fast measurement throughput
* Full talk/listen capability, including talk-only operation
* Full serial poll capability, with bit-maskable SRO
* Full remote/local capability, including local lockout
Section 3 Remote Programming val
* EXTERNAL TRIGGER and SAMPLE COMPLETE connectors
* Remote calibration
¢, Programmable trigger sources, including two bus trig-i | gers
*, Informative output suffix’ (suppressible) * Selectable output terminators
The 8840A supports the following interface function subsets: SH1, AH1, T5, L4, SR1, RL1, DC1, DT1, E1, PPO, and Co. |
3-3. BUS SET-UP PROCEDURE
To set up the 8840A on the IEEE-488 bus, proceed as follows:
1. Tum the 8840A POWER switch OFF and set the 8840A IEEE-488 address using the rear panel JEEE-488 address switches shown in Figure 3-1.
2, With the 8840A POWER switch OFF, plug the
~~
IEEE-488 cable into the 8840A rear panel IEEE-488 connector.
3. Switch on . the 8840A. Whenever the 8840A is in the local state, the IEEE-488 address can be displayed on the front panel by pressing the LOCAL button. |
3-4. AN OVERVIEW OF REMOTE OPERATION
An overview of remote operation is presented in the block diagram in Figure 3-2. Each block represents a register, buffer, etc., contained in the 8840A. The status registers in the center column indicate the instrument’s status, including its function, range, reading rate, etc. The input buffer receives data from the IEEE-488 bus. The output buffer receives data from the blocks to its left, and sends data on to the IEEE-488 bus.
Page 38
Remote Programming AN OVERVIEW. OF REMOTE OPERATION
“A
ADDRESS
30 SoPHRETE
AS. AZ Al - TALK ONLY ADDRESS|TALK A5 A4 A3 A2 Al ADDRESS! TALK AS AG A3 A2 A1 ADDRESS| TALK A5 A4.A3 A2) Al ONLY ONLY ONLY 00 0 0 0.0 0 °0 11 0 0 4 01 1 22 0 1°01 14° 07 01 0000 0 1 12 0 0 1 4.0 0 23 0 1 0°14 4)-1 02 000 0 1 0 13 0 0 14 14 0 1 24 0 1 14 +0 0.0 03 000 01 1 14 0 0.1. .-4 #1.:0 25 o 1 1 °0 =0).1 04 000 1 0 0 15 0 0 1 1 1 4 26 0 1.1 0 14 OF 05 00071 0 1 16 0 1°00 0.0 27 01 14 0 14/1 06 00014 1 0 17 0 1:0 0 0-1 28 0 141-1 +4 +#0/°0 07 0 00°14 1 1 18 o 1.0 01 0 29 0114 47-0) 1 08 001 0 0 0 19 o 14 0 0-1 1 30 o 1 14 4 4} 0 09 0 0 1 0 0 1 20 0 4.01 0 0 31 ' Not allowed 10 001 °0 1 «0 21 Oo 1°06047~-0 1 TALK 1X X XX! x il ONLY {X= setting does not matter|
Figure 3-1. IEEE-488 Address Selection
Information is transferred between blocks by device-dependent commands. Each command is shown next. to an arrowhead which indicates the resulting information trans-fer. For example, Put command PO takes a number from, the input buffer and stores it in the primary status registers. Likewise, Get command GO gets the content of the primary status registers and copies it into the output buffer.
3-5. A NOTE ABOUT EXAMPLES
In the examples in this manual, device-dependent commands are shown enclosed within quotation marks, as they would be entered in Fluke BASIC. For clarity, the commands are also separated by spaces. However, the spaces are are not necessary and may be omitted.
Example Explanation
“* F3R1 S112” This example is equivalent to “*F3R1S1T2” or “*,F3,R1,S1,T2”.
Using the Fluke 1722A Instrument Controller, these commands might be written into.a BASIC program as shown in Figure 3-3. Examples using other controllers are given at the end of this section.
Examples of 8840A output data show the terminators CR and LF. The terminator EOI is not shown because it is a uniline message. However, the terminators CR, LF, and EOI are all selectable using the Write commands.
For reference, the ASCII. and IEEE Std 488-1978 bus codes are shown at the back of this section.
3-2
3-6. DEVICE-DEPENDENT COMMAND SET
Device-dependent commands are the heart of 8840A remote control. They tell the 8840A how and when to make measurements, when to. put data on the bus, when to make service requests, etc. Commands which correspond directly to the front panel controls or display are shown in Figure 3-4. The complete set’ of device-dependent. commands is listed in Figure 3-5. The commands may be entered using either upper- or lower-case letters. See table 6-15 for conditions under which certain commands a re not valid.
Device-dependent commands. are device-dependent jmes-sages. For the 8840A to receive them, they must be sent over the IEEE-488 bus when the 8840A is in remote and has been addressed as a listener.
| The. following paragraphs describe the device-dependent commands in alphabetical order. Special characters ( and ?) are described last.
3-7. Bn (Offset Commands)
The Offset commands duplicate the function of the front panel OFFSET button. When the 8840A receives the Bi command, the 8840A stores the present reading as an offset for the present function. The BO command cancels the offset. As with front panel operation, only one:o: fset is allowed at a time.
The offset status (not the offset value). can be read jusing the GS command. The 8840A defaults to BO on both power-up and on any device-clear command (*, D L, or SDC).
Page 39
Remote Progral ming
DEVICE-DEPENDENT COM
ND SET
P FUNCTION feet Ri- R6 RANGE PO GO INSTRUMENT. y <>, CONFIGURATION READING Se RATE ‘ TRIGGER Ta MODE ‘ \ F,R,S,T.B,Y,W FRONT/REAR |g X0,P3,C:.".Dal INPUT INPUTS ( - - oourreR RO) O. | AUTORANGE R7 | TAT, AB STATUS ON/OFF 3 N(VALUE) 4 OFFSET a, NUMERIC ON/OFF PO,P1,P2 L ENTRY - REGISTER SUFFIX Yo ENABLE/DISABLE G6 YW STATUS < yo, OUTPUT Wa TERMINATORS . ‘| GO-G7 DATA DISPLAY Dn, > MEASUREMENT ON/OFF | MEASURE- CIRCUITRY _| MENT DATA OUTPUT ° TGET TIO, >! BUFFER (3 ° ERROR Ph MESSAGES , | ERROR xe STATUS a USER P3 G3 SERIAL MESSAGE POLL REGISTER MASK ( P2 VARIABLE gh? CAL VALUE CALIBRATION CALIBRATION G2 STATUS PROMPT , CALIBRATION |.C, | FUNCTION [*
IEEE-488 INTERFACE BUS
Figure 3-2. Remote Operation Block Diagram
Page 40
Remote Programming DEVICE-DEPENDENT COMMAND SET
Here is a typical command string as it might be sent from a Fluke 1722A instrument Controller. The string configures the 8840A and triggers.a reading. The PRINT command automatically sends terminators (CR, LF, and/or EOI) to the 8840A at the end of the command string.
lEEE-488 bus address................32
Resets the 8840A to the power-up configuration
Pec ercesecrccccrece
Selects the 2 WIRE kQ function.......... Dob: oS 7
Selects the 2000 range vcecececccsceseaceet Selects the medium reading rate... ..sececeeee!
Selects the external trigger mode : (Rear panel trigger disabled) ...............2
Triggers a reading
Cee e eo eee ee esa aesrerevrenseses
Figure 3-3. Typical Command String
3-8. Cn (Calibration Commands)
CAUTION
The command string “C3 CO” erases the entire calibration memory. A com- plete calibration must then be per- formed.
The Calibration commands allow the 8840A to be calibrated under remote control. Commands CO, C1, and C2 duplicate the front panel calibration functions STORE, A/D, and HF AC, respectively. For a complete description of remote calibration, see the Maintenance section of this manual.
For the 8840A to accept these commands, the 8840A must be in the calibration mode (enabled by pressing the front panel CAL ENABLE switch). Otherwise, the commands generate an error message. ,
3-9. Dn (Display Commands)
The Display commands allow the user to biank the numeric field in the 8840A front panel display. The DO command causes the display to operate normally, and is the default on power-up and upon any device-clear command (*, DCL, or SDC).
The D1 command blanks the numeric field in the display. The annunciators remain active, and all of the annunciators still flash if the input exceeds 1000V dc or 700V ac in the Tespective ranges. The D1 command is. used for best performance when high IEEE-488 Interface Data rates are required.
3-4
3-10. Fn (Function Commands) |
| The function commands duplicate the front panel function buttons. The 8840A defaults to F1 on power-up and on any device-clear command (*, DCL, or SDC). If FO is sent to the 8840A, it is internally converted to F1. The function setting can be read using the GO command. |
As with the front panel commands, selecting F5 or F6 automatically selects the 2000|mA range (R5). Also, if the instrument is in a resistance function (F3 or F4) and in R6, selecting any other function automatically selects RS.
Example Explanation | “F3” Selects 2 WIRE kQ function; it does not affect any other settings. | : | “* FS” Selects mA DC function and/ 2000 mA range. Resets all other settings to default.
3-11. Get Commands |
The Get commands “get” information from the 8840A for the controller. Each Get command loads the output buffer with an output string in the format shown in Figure 3-6. Status data (the output from Get commands GO, G1, G3, G4, G5, G6, G7, and G8) is interpreted as shown in| Table 3-1. The Get commands should not be confused with the interface message GET (Group Execute Trigger). |
The output data from some Get commands starts with a leading 1 or 10. This prevents the controller from suppressing leading zeroes and gives a uniform four-character length to all instrument configuration data (the data from Get commands G0, G4, G5, G6, and G7). po