8840A Digital Multimeter Instruction Manual, pages 21 to 30

Pages 21 to 30 of 180

The original document is the PDF (42 MB). This is its text as extracted, unedited; tables and drawings read best in the PDF, and each page heading opens it at that page.

For the Fluke 8840A

Page 21 read from a scan by OCR; it may hold reading errors

Introduction and Specifications SPECIFICATIONS

Table 1-1. Specifications (cont)

MEDIUM AND FAST READING RATES ... In medium rate, add 50 counts to number of counts. In fast reading rate, for sinewave inputs >1000 counts and frequencies >100 Hz, the accuracy is +(0.4% of reading + 30 counts).

NONSINUSOIDAL INPUTS ............... For nonsinusoidal inputs 210,000 counts with frequency components <100 kHz, add the following % of reading to the accuracy specifications.

FUNDAMENTAL CREST FACTOR

FREQUENCY 1.0 TO 1.5 1.5 TO 2.0 2.0 TO 3.0 45 Hz to 5 kHz 0.05 0:15 0.3 20 Hz to 45 Hz 0.2 0.7 415

Operating Characteristics

TEMPERATURE COEFFICIENT .......... Less than 0.1 x accuracy specification per °C from 0°C to 18°C and 28°C to 50°C. MAXIMUM INPUT ..................00005 2A de or rms ac. Protected with 2A, 250V fuse accessible at front panel, and internal 3A, 600V fuse. BURDEN VOLTAGE ..................00. 1V de or rms ac typical at full scale. RESISTANCE Input Characteristics ee eee ee oS “RESOLUTION cee RANGE FULL SCALE CURRENT 5% DIGITS 5% DIGITS 4% DIGITS* THROUGH UNKNOWN 2000 199.9990 1mo 10 ma 1mA 2 kQ 1.99999 kQ 10 mQ 100mQ - _e 1mA | 20 kQ 19.9999 kQ 100 mQ 1Q 100 yA 200 kN 199.999 kQ 10 102 10 pA 2000 kQ 1999.99 kQ 10Q 10002 - 5 wA 20 MQ 19.9999 MQ 1002 1kQ 0.5 pA *4% digits at the fastest reading rate. Accuracy NORMAL (S) READING RATE ...... Jo eeee +(% of Reading + Number of Counts)“ RANGE 24 HOUR 23+1°C2 90 DAY 23+5°C 1 YEAR 23+5°C 20003 0.004 +35 0.011 + 45 0.014 + 45 2kQ 0.0028. + 2 0.01 +3 0.013. + 3 20 kQ 0.0028 +2 0.01 +3 0.013 + 3 200 -kQ 0.0028 + 2 0.01 +3 . 0.013 +3 2000 kQ 0.023 +3 0.027 +3 0.028 + 3 20 MQ 0.023 +3 0.043 + 4 01044 + 4

‘Within one hour of zero, using offset.control. ? Relative to calibration standard. * Applies to 4-wire ohms only.

“When in fast reading rate with internal trigger and transmitting data out of the IEEE-488 interface, the 8840A _ display must be blanked (command D1) to ensure stated accuracy.

* When offset control is not used, the number of counts are 5,7, and 9 for 24 hour, 90 day, and 1 year respectively.

Page 22 read from a scan by OCR; it may hold reading errors

Introduction and Specifications

SPECIFICATIONS

Table 1-1. Specifications (cont)

Accuracy, cont

INPUT PROTECTION

READING RATES READING RATES

MEDIUM AND FAST READING RATES ..

MEASUREMENT CONFIGURATION OPEN CIRCUIT VOLTAGE

WITH INTERNAL TRIGGER

Operating Characteristics TEMPERATURE COEFFICIENT

Pee errr wcrerece

Pee ere rmercorseere weecercece

. in medium rate; add 2 counts to the number of counts forthe 2000 tnrough 200 kQ ranges and 3 counts for the 2000 kQ and 20 MO ranges. in fast reading rate, use3 counts for the number of counts for the 200Q range, and 2 counts for ail other ranges.

Less than 0.1 x accuracy specification per °C from 0°C to 18°C and 28°C to 50°C. 2-wire or 4-wire.

Less than 6.5V on the 2002 through 200 kQ ranges. Less than 13V on the 2000 kQ and 20 MQ ranges. To 300V rms.

(readings per second).

i

POWER LINE FREQUENCY* RATE 50 HZ 60 HZ 400 HZ cS) 2.08 2.5 2.38 M 16.7 20 19.0 F 100 100 100

*Sensed automatically at power-up.

AUTOMATIC SETTLING TIME DELAY

Time in milliseconds from single trigger to start of A/D conversion, Autorange off.

READING RATE NUMBER OF COUNTS FUNCTION RANGE FROM FINAL VALUE" S M F vpc 200 mv 342 61 9 5 2V-1000V 342 17 9 5 VAC All 551 | 551 | 551 30 (Note 2) mA DC 2000 mA 342 17 9 5 mA AC 2000 mA 551 | 551 | 551 30 (Note 2) Ohms 2002 394 | 105 17 5 2 322 17 13 5 20 kA 342 17 13 5 200 kQ 141 | 121 21 5 2000 kQ 141 | 101 81 10 20 MQ 1020 | 964 | 723 30

1. Difference between first reading and final value for an in-range step change coincident with trigger. 2. For slow reading rate. 50 counts for medium rate; 10 counts for fast rate.

Page 23 read from a scan by OCR; it may hold reading errors

Introduction and Specifications SPECIFICATIONS

Tabie 1-1. Specitications (cont)

EXTERNAL TRIGGER TIMING CHARACTERISTICS

The following diagram shows the nominal timing for the various processes which take place between an external trigger and data sent out on the IEEE-488 interface. Delays will vary if a second trigger comes before the data handshake is _ complete. REAR BNC TRIGGER J ar een te - AUTOMATIC SETTLING TIME DELAY’ A/D CONVERSION? ‘ SAMPLE COMPLETE® a DATA HANDSHAKE* t 12ms - 3 he

NOTES: 1. Time from single trigger to start of A/D conversion. (See “Automatic Setting Time Delay” on previous page.) If the delay is disabled by using the T3.or T4 command, then the delay is 1 ms +150 ps. When the 8840A is triggered with an IEEE-488 command (GET or ?), the automatic setting time delay begins after the trigger command has been processed and recognized.

2. A/D conversion time is dependent on the reading and power-line frequency:

RATE A/D CONVERSION TIME (ms) 50 Hz 60 Hz 400 Hz S) 472 395 414 M 52 45 47 F 7 7 7 3. Sample Complete is a 2.5 us pulse which indicates that the analog input may be changed for the next | reading. 4. When talking to a fast controller. GENERAL COMMON MODE VOLTAGE ........cccccceseeeeee 1000V dc or peak ac, or 700V rms. ac from any input to earth. TEMPERATURE RANGE.. 0 to 50°C operating, -40 to 70°C storage. HUMIDITY RANGE ........... 80% RH from 0 to 35°C, 70% to 50°C. WARMUP TIME................. 1 hour to rated specifications. POWER.........ceeccesesseteceees 100, 120, 220, or 240V ac +10% (250V ac maximum), switch selectable at rear panel, 50, 60, or 400 Hz, automatically sensed at power-up. 20 VA maximum. VIBRATION .........eeeeeeeee Meets requirements of MIL-T-28800C for Type Ill, Class 3, Style E equipment. ELECTROMAGNETIC COMPATIBILITY .............. ¢ Specifications apply when used in an environment with field Strengths <1 V/m. * For field strengths up to 3 V/m multiply floor adder by 12 for VDC and Resistance and 100 for DC Current. e VAC and AC Current have no adders up.to 3 V/m. PROTECTION...........0000. ANSI-C39.5 and IEC 348,.Class |. SIZE (HxWxD)... 8.9 cm x 21.6 cm x 37.1 cm (3.47 in x 8.5 in x 14.4 in).

Page 24 read from a scan by OCR; it may hold reading errors

introduction and Specifications SPECIFICATIONS Table 1-1. Specifications (cont) WEIGHT .......0....sceecesseeeeees Net, 3.4 kg (7.5 Ib); shipping, 5.0 kg (11 Ib). INCLUDED...............0..cccee Line cord, test leads, Instruction/Service Manual, IEEE-488 Quick Reference Guide,

(Option -05 only), and instrument performance record.

IEEE-488 INTERFACE FUNCTION ...........ccccceseeeeees Option allows complete control and data output capability, and supports the following interface function subsets: SH1, AH1, T5, L4, SR1, RL1, DC1, DT1, E1, PPO, and Co. St 1) 21.6 em (8.5") 37.1cem .. (14.4") 00 ' 9 Grr) Ke UT ea) | |. 25.4m | - (10.0") | i, - | 89 cm (3.47")

Figure 1-1. External Dimensions

Page 25 read from a scan by OCR; it may hold reading errors

2-1. INTRODUCTION

This section proyides instructions for installing and operating the 8840A. Refer to Section 4 for measurement considerations.

NOTE

Do not allow the meter to be used if it is damaged or its safety is impaired.

2-2. INSTALLATION

2-3. Installing the Power-Line Fuse

WARNING

FOR POWER-LINE VOLTAGES OF 198V TO 250V, THE POWER-LINE FUSE MUST BE REPLACED WITH A 1/8A, 250V SLO-BLO FUSE FOR FIRE PRO- TECTION. TO AVOID ELECTRIC SHOCK, REMOVE THE POWER CORD BEFORE REPLACING THE EXTERNAL LINE FUSE.

The 8840A has a rear-panel power-line fuse in series with the power supply. A 1/4A, 250V slow-blow fuse is installed in the factory for operation from 90V to 132V. For operation with power-line voltages of 198V to 250V, the fuse must be replaced with a 1/8A, 250V slo-blo fuse.

To replace the power-line fuse, first remove the power cord. Then turn the rear-panel fuse cover 1/4-turn counterclockwise with a screwdriver.

For power-line voltages of 198V to 250V, use only a 1/4 x 1 1/4 (6.3mm x 32mm) fuse with at least a 100A breaking capacity.

Section 2 Operating Instructions

2-4. Connecting to Line Power

WARNING

TO AVOID SHOCK HAZARD, CONNECT THE INSTRUMENT POWER CORD TO A POWER RECEPTACLE WITH EARTH GROUND. TO AVOID INSTRUMENT DAMAGE, CHECK THAT THE REAR PANEL LINE VOLTAGE SELECTION SWITCHES ARE SET TO THE POWER- LINE VOLTAGE IN YOUR AREA.

The 8840A can be configured to accept line power of 100, 120, 220, or 240V ac (+10%, 250V maximum) at 50, 60, or 400 Hz. The voltage must be selected by setting the rear panel LINE SET switches as shown in Figure 2-1) The 8840A automatically senses the power-line frequency at power-up, so that no adjustment for frequency is necessary.

2-5. Adjusting the Handle

The handle provides two viewing angles for bench-top use. To adjust its position, pull the ends out to a hard stop (about 1/4 inch on each side) and rotate it to one of the four stop positions shown in Figure 2-2. To remove the handle, adjust it to the vertical stop position and pull the ends all the way out.

2-6. Rack Mounting. Kits ©

You can mount the 8840A in a standard 19-inch rack panel using the accessory rack mounting kits shown in Figure 2-3. To install the Single Rack Mount Kit, remove the handle and handle mounting plates, and attach the rack ears with the screws provided (Figure 2-4). The Dual Rack Mount Kit is installed similarly. (Both kits include mounting instructions.) The rear feet may be rotated 180 “pee to Clear a marrow rack space.

| 2-1

Page 26 read from a scan by OCR; it may hold reading errors

Operating Instructions OPERATING FEATURES

FOR THIS LINE USE THIS.SWITCH

VOLTAGE: SETTING:

90V to 109V LINE SET

109V to 132V LINE

SET

FOR THIS LINE USE THIS SWITCH

VOLTAGE: SETTING:

198V to 229V “LINE SET

229V to 250V LINE SET

Figure 2-1. Line Voltage Selection Settings

PULL ENDS OUTWARD TO ROTATE

1. Viewing position

=

2. Alternate viewing position

3. Carrying position

4. Removal position (to remove, pull ends out)

= i.

Figure 2-2. Adjusting the Handle

2-7. OPERATING FEATURES

2-8. Power-Up Features

When the 8840A is turned on, all display segments light up for about 2 seconds while the instrument performs an internal self-test of its digital circuitry. The 8840A then assumes the following configuration:

* VDC function

° Autorange, starting in the 1000V range

* Slow reading rate

* Continuous, internal trigger

* OFFSET off

* Local (front panel) control

While all display segments are lit during the power-up self-test, you can freeze the display by pressing the SRO button. All display segments will then remain lit until you press any button.

2-2

2-9. Front and Rear Panel Features

Front panel features are explained in Figure 2-5. Rear panel features are explained in Figure 2-6.

The alternate functions embossed below the front panel Tange buttons and the special feature buttons are enabled by the CAL ENABLE switch. These functions are for use only when calibrating the instrument. See the Maintenance section for further explanation.

CAUTION

To avoid accidentally uncalibrating the 8840A, do not press the CAL ENABLE switch unless calibrating the instru- ment. Never cycle power on or off while the CAL ENABLE switch is on.

Note that the VAC and mA AC functions are available only with the True RMS AC option. If this option is

Page 27 read from a scan by OCR; it may hold reading errors

Operating Instructions OPERATING FEATURES, poo0p0g. Ses} (Y* pooooaay?P

Single Rack Mount Kit | (Accessory Y8834)

Center Rack Mount Kit (Accessory Y8836) also available.

€

90 90 ” ook ; = DOOG0G Ss2 = Q@O0000 ess id BOSSOBOSE= (C= paccnanz2s ||. Dual Rack Mount Kit (Accessory Y8835)

(aa

Figure 2-3. Rack-Mount Kits

REMOVE HANDLE MOUNTING

ATTACH RACK EARS

Figure 2-4. Installing the Single Rack Mount Kit absent, pressing the VAC and mA AC function buttons causes the 8840A to briefly display an error message (ERROR 30).

2-10. Display Features

The 8840A features a vacuum fluorescent display with a numeric field and annunciators. The annunciators are explained in Figure 2-5.

2-11. Error Messages

If the 8840A detects an operator error or an internal failure, it displays an error message for about 2-1/2 seconds and then resumes normal operation. During this time, the front panel buttons are ignored. The error message consists-of the ERROR annunciator and a two-digit error code. (See Figure 2-7.) Error codes are explained in Table 2-1.

while the mA DC or mA AC function is selected, ERR' 31 is displayed. In this case the error message is display: until you return the switch to the FRONT position or select another function.

If the FRONT/REAR switch is set to the REAR ERROR R d

2-12. Overrange Indication |

An input is overrange if it exceeds the full scale of the selected range. In most ranges, the 8840A indicates an input is overrange by lighting the OVER annunciator and showing a “1” on the display. (See Figure 2-8.) The sign, the position of the decimal point, and the other annunciators are not affected. |

As a safety feature, the 8840A treats the 1000V. de and

700V ac ranges differently. In these ranges, the 8840A |

23

|:

Page 28 read from a scan by OCR; it may hold reading errors

2-4

Figure 2-5. Front Panel Features

FUNCTION BUTTONS:

voc] DC Volts

VAC] AC Volts*

2-Wire Ohms

HIGH and LO SENSE Terminals for 4-Wire Ohms Only

Hl and LO INPUT Terminals

2A INPUT Terminal (Houses 2A fuse)

ee

4-Wire Ohms

DC Current

Ea AC Current*

Display

FRONT/REAR switch selects either front or rear inputs.

| 200 mV or 2002

Calibration Functions _ i 2V or 2 kN (embossed) ;

*Available with True RMS AC only. Gives error message if option not installed.

RANGE BUTTONS

~~

1000V dc, 700V ac, 2 MQ ms or 2000 mA

CAL ENABLE switch enables calibration mode. (CAUTION! See text.)

SPECIAL FEATURES

POWER switch turns 8840A on or-off.-Also initiates power-up self-test and resets instrument to:

VDC function Autorange

Slow reading rate Continuous trigger OFFSET off

OPERATING FEATURES

Operating Instructions

Local (front panel) control

Page 29 read from a scan by OCR; it may hold reading errors

Figure 2-5. Front Panel Features (cont)

TRIG triggers a new reading.

Enabled in external trigger mode.

EXT TRIG toggles between internal and external trigger modes.

RATE cycles between slow, medium and fast reading rates. Automatically selects the optimum filter for each - reading rate.

SRQ generates a service request over the IEEE- 488 bus if enabled by the SRQ mask (IEEE-488 Interface option only). When pressed for 3 seconds, SRQ initiates diagnostic self-tests. NOTE: Leave inputs disconnected during self- tests or the 8840A may indicate errors.

If the 8840A is in remote, LOCAL returns it OFFSET stores the displayed reading as an to local control. If the 88404 is in local, the subsequent readings in the function its bus address for two and one half presently selected. Readings in the other seconds. Ignore if the |EEE-488 Interface is functions remain unaffected. Pressing not installed. OFFSET again cancels the offset, or stores a new offset if in a different function. Error Condition Overrange Calibration Mode Enabled

Reading Rate: Slow, Medium, Autorange On and Fast. Blinks off when a ading is triggered. ____. feading 's trig - Offset on External Trigger

(AUTO OFFSET Mode Enable, TRIG SMF OVER ERROR CAL AuTo OFFSET

Self-Test Enabled, - - - - - }- - _ TEST | chi ED rl j re a5 mA

REMOTE Function and Units IEEE-488 Interface | Usain Tf, ct iJ iJ -t. v1 l. x yi. 4 WIRE Annunciators a g |

Annunciators

OPERATING FEATURES

Operating Instructions

Page 30 read from a scan by OCR; it may hold reading errors

Figure 2-6. Rear Panel Features

SAMPLE COMPLETE output

TTL-level Normally high IEEE-488 Interface pulsed low when samples for Bus Connector* a reading are completed*

TALK ONLY and IEEE-488 ADDRESS Selection Switches*

OPERATING FEATURES

Operating Instructions

EX TRIG input* TTL-level, falling-edge triggered (internally selectable

HIGH and LO SENSE Terminals for 4-Wire Ohms

Only

Serial Number

HI and LO INPUT Rear feet rotat Line Voltage Terminals for rack mounting Selection Switches Power-Line Cord Connector

Power-Line Fuse al \

Rear Panel Insert

*Available with lEEE-488 Interface only. Otherwise, the upper portion of the rear panel is covered with an insert as shown at right.

CAUTION: The rear panel insert is attached from inside the case. Refer to Section 8 for instructions on removing it.