Maintenance

TROUBLESHOOTING Table 6-19. Keyboard Wiring BUTTONS THAT BUTTONS THAT BUTTONS THAT SIGNAL PIN PRODUCE PRODUCE PRODUCE NAME INVERTED INVERTED INVERTED STROBE ZERO STROBE ONE STROBE TWO RLO U212-38 SRQ 2002/mV VDC RL1 U212-39 LOCAL 2 VAC RL2 U212-1 RATE 20 2 WIRE kQ RL3 U212-2 OFFSET 200 4 WIRE kQ RL4 U212-5 AUTO 2000 mA DC RL5 U212-6 TRIG 20M mA AC RL6 U212-7 EX TRIG Not used Not used directly since all of its inputs and outputs are available. Again, outputs of quad analog switches U804 and U808

Table 6-20. Analog Control Devices are not available and must be determined by analog means. DEVICE REF. DES. 6-60. Evaluating Static Signals Relay Butter U201

. Quad Comparator U305 Table 6-21 may be used to determine whether or not Quad Analog Switch U301 proper signals are reaching any particular analog control . device. It may also be used to quickly exercise all of the Quad Analog Switch U302 devices before beginning analog troubleshooting if it is Quad Analog Switch U303 still unclear as to which devices are suspect. A number of Quad Analog Switch U402 the inputs to these devices are static which makes them Quad Analog Switch U403 particularly easy to check. 8-Bit Latch U803 * For example, suppose quad comparator U305 appears not Quad Analog Switch U804 * to be working. Connect a scope to U305-11 and step the Quad Analog Switch U808 8840A through all functions and ranges in the following * Option -09 only sequence:

1. VDC: 200 mV, 2V, 20V, 200V, 1000V If the instrument is not in the slow reading rate, it gives the following pattern at U305-5: VAC: 200 mV, 2V, 20V, 200V, 700V

00000 00000 000000 000000 0 0 2 WIRE kQ: 2008, 2k, 20k, 200k, 2M, 20M

Next move to U305-7 and repeat. The pattern at U305-7 will be:

mA DC (one range only)

2.

3

4. 4 WIRE kQ: 200Q, 2k, 20k, 200k, 2M, 20M 00000 00000 000000 111111 0 0 6 mA AC (one range only)

6-61. Evaluating Dynamic Signals

The procedure for evaluating the dynamic signals is only slightly more involved. For example, consider U302-7 in Table 6-21. At the right end of that row the table says to sync on U302-3. The sync pulse is negative-going. Apply it to channel 1 of a dual-trace scope syncing on the leading (negative-going) edge. Observe the target pin (U302-7) on channel 2 of the scope. While stepping through the 24 ranges observe the state of the target pin exactly when the sync pulse goes from low to high. (See Figure 6-12.) (This procedure works best in the fast reading rate since the repetition rate of the sync pulse on U302-3 is greater.)

While doing this, observe the state of U305-11. As shown in Table 6-21, this 24-range sequence will produce the following pattern at U305-11:

11010 00000 111100 111100 0 0

Next move to U305-5 and repeat. The slow reading rate gives the following pattern at U305-5:

11111 00000 111000 111000 0 o

6-36

Page 142 read from a scan by OCR; it may hold reading errors

Le-9

(00¥> (Tay) (ZQ0) (Eq¥) (hav) (Gv) (700) (Zq¥)

é et st

é

S ét 9T

:

COMmmemOOO he OFM mm OD Ore Oem mOOOM Owe OCOM Onmew CGOMm bL€st we SOO OS Om OMe OO Or Comme COOMm eORmM OOOM cChMmm OO0OM

Orme CGOOrm ComrmeOoom oMwpmR ODOM CORR OD OM OR eOO OM Cee HOODOO,

OF re ODOM, Oe tem OO Ores ORF ODOM Or MEM OO Oh Oe mmeOOOm OM MMOD OR

CMe ODOM

A

“OTZeuNG ryuO05 Pazeuqrpesupn

472958 aus SINdiNO asauy T1Y at

Lo) ad at >) ores Omrn od ores or ho Td oad >) CMR S Sonne ores OMe oS

MrMOD RHO Orr oS oron~ onor~ ocsso roo wee OO ores Orr Ordo ooscs

Drom S

Che S

|

> us Subg

=2OrM oD Oro mOned Sh eh >) OM, -SHO oro ah Dood >) “OFS eOrD

On O eOrD HOMO ~OoMmd MOO oeror meOMmMD me She MOMS, OrnwdD OOH COroOrw ors ord

E- copn } uo aUAS

E-£0 uO DU

Bese

GN

Net et et

RUION

S¥d coEn

Ones orssd eee OO Oorso eee OO

Onhow ono opto Oho ono

Onoso Somos onoo ones Mnos HYhosd

Oonoco Onos obhoso Onbhos “boo Mbhbos

She Oo

0 ¢ Q t

6 Re £-TOEN uO Dug, “Oo Ju

She ee Ore ne Oooo tO mt pe, ooeo

KOO “Oooo "O90 HOS =o00

One pe ~ Or p be S ht pe et St ee “Ooo &Soo

Ot be ne DS ot et te DO bet et ee D ht ee eo HMO eros

“Sooo

~oscs ayy ae utd gap -Je9 344 4O 4845 aya Qe

') as

“astnd aufis % 40 afpa UTSTu

“a a UpyOOT pue ‘pazertpur tnd sn z-

T3e

Gutob - aa Hau a ug

Gur, une fig

Payxray) aq qsnw speu6ts asaya try

NGbAU

USESe Toen

OM mew SO bt at pe Ore mene dat ft Desk bet oe poe ee bo ee ee CD tt tO 6 bt OD ee OD ™- OO oornw Coomrn cScon~w-m Comm Corr Comm loler tT) oornrw oonrnw oor em lotos yy COMmpm ba 0 bed fee

249.895 ade STeubts ISaY4 tly oo os oc oso oo ou on ou oso oo oc oath 33 mein oo “Oo =O eo oo

OOTTET OOTTTY 00000 oOTrort

0

| an

UOT} 4uod

-dun4 utd

HIed "ON mo nw Be bom Uxexr aw

28 ys Lee eo Bo) bam UIE aw os UDBOID

| UDSDs OD ic Be tt We it a

T Gor I

31t60

2 $F out. 2f60 SNOILINI430

SF Tus $T uke

Uppead qsey ut Gutpeas @ors uf

6 a7ed aqeu

(WAT Paw JO MOTS UF YF ‘"a*T) aQe4 Guypeas qsey up you

See3S¢ 31607 josuOD Goyeuy ‘12-9 aiqe,

ONLLOOHSSTISNOUL aoueuawieyy

Page 143 read from a scan by OCR; it may hold reading errors

Maintenance TROUBLESHOOTING

Using this procedure, the following pattern should be seen: 00000 11111 000000 000000 0 1

Note that the last eight rows in Table 6-21 are actually outputs of U803. Therefore, observing those pins proves not only that the control signals are correct but also that U803 itself is functioning correctly.

6-62. DC Scaling Troubleshooting

Whenever there is a failure in the DC Scaling circuit, first check the power supply voltages for all active components. (Supply voltages and pin numbers are listed in Table 6-22.) A test of the bootstrap supplies for U306 is described later under this heading.

After checking the power supplies, use an oscilloscope to check the digital logic input pins of quad analog switches

4V, 2us SYNC PULSE OV _ v - toy LOGIC “1" AT 4 \ TARGET PIN Wu Soo

\

I

|

{ 4v - \

| LOGIC “0” AT i TARGET PIN ov, Lee om we = - i“ _ OBSERVE TARGET PIN AT THIS TIME

Figure 6-12. Typical Dynamic Control Signals

U301, U302 and U303. These should show digital signals with high 2+3V and low s+0.5V.

In the 20V range, any voltage applied to the HI INPUT terminal (relative to Reference Low) should be present at U306-3. If not, trace the voltage from the HI INPUT terminal to U306-3 to isolate the problem.

To check U306, select the VDC function and the 2V range. Measure the voltage at TP302 while applying first 1V and then OV (a short) across the HI and LO INPUT terminals. If 1V and then OV appear at TP302, U306 is probably OK. If not, the problem is in U306 or its bootstrap supplies (TP301 and TP303).

To check the bootstrap supplies, put the 8840A in the 20V range and measure the voltage at TP301, TP302, and TP303. TP301 should be 6.3V (nominal) above TP302, and TP303 should be 6.2V (nominal) below TP302. If the bootstrap supplies are operating correctly, measure the voltage at U306-3 and U306-6 for input voltages of +20V and -20V; if the voltage at U306-3 differs from U306-6, then U306 is bad.

To check the dc input path after U306, short the HI and LO INPUT terminals and read the display. If zero is displayed for ranges 13 and 15 but not for r1, 12, and r4, then the signal path including Q311 and U301B is suspect. To check Q311, apply a 1V dc input in the 2V range and check that the voltage at the drain and source of Q311 is 1V. If not, Q311 or its driver is bad. If 1V appears at U301-16, but not at the display, then U301 may be bad.

If zero is not displayed for r3 and 15 with the HI] and LO INPUT terminals shorted, then Z301 or U302D is probably bad.

6-63. Track/Hold Troubleshooting

If a problem is suspected in the Track/Hold (T/H) circuit, first check the power supply voltages of all active components. (See Table 6-22.)

Table 6-22. DC Scaling and Track/Hold Supply Voltages

PIN OR SUPPLY PIN OR SUPPLY DEVICE VOLTAGE DEVICE VOLTAGE U301-6 +5V U303-20 +7.5V U301-10 ov U303-11 -8.2V U301-20 +7.5V U304-4 -8.2V U301-11 -5V U304-7 +7.5V U302-6 +5V U305-3 +5V U302-10 OV U305-12 -5.5V (nom)" U302-20 +7.5V U307-4 -15V U302-11 -5V U307-7 +15V U303-6 +5V Q305,c +30V U303-10 OV Q306,c -30V

*With OV input.

6-38

Page 144 read from a scan by OCR; it may hold reading errors

Maintenance TROUBLESHOOTING

Next, check the T/H output waveform at TP103 with an oscilloscope. Set the 8840A to the VDC function and 2V de range, apply +1V de across the HI and LO INPUT terminals, and trigger the scope from the falling edge of line not-TR (TP201). The waveform should look like that in Figure 6-13. The circuit may be checked as follows:

1. Short the HI and LO INPUT terminals, and select the 2V dc range.

2. Short U301-14 to ground (Reference Low). The 8840A should read within 10 counts of zero. (The actual value is not as important as its stability.)

3. Connect U307-6 to U303-18, and monitor the voltage at TP103 using another multimeter. The 8840A should read about the same as the external multimeter, but with opposite sign.

If the 8840A fails step 2 but not step 3, then U303 is bad. If the 8840A fails both steps 2 and 3, then the fault is in the A/D Converter or the T/H Amplifier. To tell which, lift the end of R318 closest to the front panel (connected to U307-6) and apply an input of less than +2V to TP103. If the A/D Converter is OK, the 8840A will display the applied voltage with the opposite polarity. (For example, if you apply +1V, it should display -1V.) The readings may differ by a slight offset.

6-64. Ohms Current Source Troubleshooting

Malfunctions in the ohms functions can be caused by a fault in the Precision Voltage Reference, Ohms Current Source, or Ohms Protection. Malfunctions can also be caused by a fault in the DC Scaling circuit which loads the Ohms Current Source.

First check the power-supply levels (see schematic). Then check all digital logic input pins of the quad analog switches (U402 and U403). These should show digital signals with high 2+3V and low <+0.5V.

To determine whether the Ohms Current Source is being loaded down by the DC Scaling circuit, select the VDC function and connect a 10 kQ resistor between the collector of Q404 and ground (Reference Low). (Selecting the VDC function opens K401, and configures the Ohms Current Source in the 20 kQ range.) If the voltage across the 10 kQ resistor is 1V, then the Ohms Current Source is working (at least in the 20 kQ range), and the problem is probably due to a defect in the DC Scaling circuit.

To test whether the Ohms Current Source is actually being sourced out the HI and LO OUTPUT terminals, select the 20 k@ range and the 2-wire ohms function, connect a 10 kQ resistor across the HI and LO INPUT terminals, and measure the voltage across this resistor with another voltmeter. There should be a 1V drop across the resistor.

If the ohms functions do not work in any range, check the supplies at U401 (+15V), U404 (+30V and -5V), U402 (+15V, +5V, and OV), and U403 (+15V, +5V, and OV), and check the -7V reference at R416. Also, test the Ohms Protection circuitry as follows: Select the 20 kQ range and

2-wire ohms function, connect a 10 kQ resistor to the HI and LO INPUT terminals, and bypass the protection circuitry by connecting the emitter of Q402 to the junction of R410 and R309. If a reading of 10 kQ is displayed, the protection circuitry is defective. To isolate the problem, successively short each part of the protection circuitry that is in series with the Ohms Current Source, until the display reads 10 kQ.

If the ohms functions work in only certain ranges, suspect resistor network Z401 or analog switches U402 or U403. To test the analog switches, select a defective Tange and connect a short across the switches that are supposed to be closed in that range. If the Ohms Current Source then works, one of the analog switches is probably bad. If the range still doesn’t work, then 7401 is probably bad.

To test the first stage of the Ohms Current Source, short U402-19 to Reference Low through a 2 kQ resistor and check that the voltage across R401 is 7.0V (nominal) and that the voltage at U401-6 is -4V (nominal). If the voltages are correct, the first stage of the Ohms Current Source (U401 and Q401) is working. If not, suspect U401 or Q401. (Under no circumstances should U401-6 ever be positive.)

If the first stage of the Ohms Current Source is working, test the second stage as follows:

1. Select the 20 kQ range and apply a 10 kQ input.

2. Check that the voltage between TP403 and U402-16 is - +5V.

3. Measure the voltage at U404-6 with respect to ground. If the voltage at U404-6 is negative, U404 is bad. (Under no condition should U404-6 ever be negative.)

6-65. Precision Voltage Reference Troubleshooting

If there is a failure of the Precision Voltage Reference, check the power supply levels at U702. U702 requires two supplies, +15V and -15V, which must be within approximately +5% of their nominal value. Using an oscilloscope, check that the power supplies and op amp outputs (U702-1 and U702-7) are free from ripple and oscillations.

If the supplies are correct, check the output voltage levels at TP701 and TP702. The voltages should be +7.00000V +1000 ppm and -7.00000V +250 ppm. Also check that the reference amplifier output voltage (U702-1) is nominally +65V.

If the outputs are grossly out of tolerance (e.g., stuck at +15V or -15V), the most likely cause is a bad op amp (U702) or open resistor network (Z701 or Z702). If the outputs are slightly out of tolerance, the most likely cause is a defective or out-of-tolerance resistor in Z701 or Z702. Because Z701 is precisely matched with U701, Z701 and U701 must be replaced as a matched set.

Shorts between lands or runs can also cause small errors (10 ppm to several hundred ppm). Shorts between sense

6-39

Page 145 read from a scan by OCR; it may hold reading errors

Maintenance

TROUBLESHOOTING y BY TR2 HD2 adi f, - INPUT: +1V Ws aatirech VERT: ; _ 0.5V/DIV HD1 L = OV . pees HORIZ: 2 ms/DIV rg + RANGE: A/D INPUT SAMPLING |} | | TR2 HD2 INPUT: -1V es Set eel Se VERT: L : 0.5V/DIV 7 - HORIZ: ov a . : 2 ms/DIV T t 2v t TR2 ¢ Asmat Hb2 INPUT: OV * + 5 mV/DIV C OV Li. : + + HORIZ: 2 ms/DIV ' Em ee vol... ees peer Ses Oe 0 Ot | RANGE: L | $ f 200 mv “OFFSET VOLTAGE OF Q314 | t INPUT: + +100 mv os 53 0.5V/DIV + OV smn. . HORIZ: L 2 ms/DIV . EB: a + a RANGE: . HD 200 mV

ALL WAVEFORMS FOR SLOW READING RATE.

6-40

Figure 6-13. Typical Output Waveforms for Track/Hold Circuit {TP103)

Page 146 read from a scan by OCR; it may hold reading errors

Maintenance TROUBLESHOOTING and output lands can cause small errors that are not related to resistor networks. Load regulation problems can also be caused by shorts between sense and load lines.

In some rare cases, the op amps (U702A and U702B) could be out of spec, causing 2 small error. The maximum input offset voltage of the op amps used in the circuit is 3 mV.

6-66. A/D Converter Troubleshooting

If there is a failure of the A/D Converter, all power supply levels should be checked at the op amps (U102 and U103) and the A/D IC (U101). The A/D Converter has a total of seven supplies: +15V, -15V, +5V, +7.5V, -8.2V, +7.00000V, and -7.00000V. All supplies should be within 5% of their nominal values except for the +7.00000V and -7.00000V reference supplies, which should be within +1000 ppm and +250 ppm respectively. The bootstrap supplies (lines BS1 and BS2) should be +7V and -7V (+10%) referenced to the + input of the A/D amplifier (U103-3).

Troubleshooting the bootstrap supplies can often be made easier by putting the 8840A in EX TRIG (to stop the A/D Converter) and connecting the input of the A/D Converter (TP103) to INPUT LO (Reference Low on the schematic). The bootstrap supplies are then referenced to instrument common (Reference Low).

NOTE

For the following tests, set the 8840A to the VDC function and the 2V range, and trigger the oscilloscope from the falling edge of line not-IR (TP201).

If all supplies are correct, the next most useful trouble- Shooting tool is the A/D output waveform at TP101, which can be checked with an oscilloscope. The waveform should look like the one shown in Figure 6-14 when the input voltage is at 50% of the selected range. Various portions of the waveform correspond to different parts of the A/D cycle. By examining the waveform, problems in the A/D Converter can be isolated down to one or two components.

The waveform at the storage capacitors can often be used to locate leakage problems. The leakage can be due to contamination on the Main PCA or to defective switches in U101. Figure 6-15 shows the waveforms across storage capacitors C102 and C103 (U101-24 and U101-25, respectively) for a specific input.

The A/D Converter communicates with the In-Guard uC via the internal bus, which also goes to several other sections of the instrument. What looks like a problem in the A/D Converter may actually be caused by a problem in another section of the instrument which is loading down the bus data lines (U101-1, -2, -3, -38, -39, -40). A typical waveform at one of the data lines is shown in Figure 6-16. One of the data lines can be loaded down so that the In-Guard uC fails to recognize data sent over that line. If so, the amplitude of the signal of the bad line would be less than 3V peak-to-peak.

One technique of finding an overloaded or shorted data line is to remove the In-Guard wC and drive one data line at a time through a 1 kQ resistor. Measure the voltage drop across a length of the line. Normally the voltage drop across the line is zero volts (less than 5 uV). Voltage drops larger than 5 wV indicate a short. (The voltage drop is caused by excessive current flowing through the line.)

When troubleshooting the A/D Converter it may be desira-ble to determine what the reading is at the A/D Converter when the display is definitely incorrect. A digital problem between the A/D Converter and the In-Guard uC can cause erroneous or noisy readings or offsets. Similar prob-, | + | :

- : VERT 2V/DIV r; r 4 HORIZ .2 ms/DIV oe +

. f 3 ra tdi +1.0V INPUT r T + vid i ' t e J.J {

Figure 6-14. Output of A/D Amplifier (TP101)

6-41

Page 147 read from a scan by OCR; it may hold reading errors

Maintenance TROUBLESHOOTING

RAMP UP OR DOWN INDICATES LEAKAGE PROBLEM AT U101-24

RAMP UP OR DOWN INDICATES LEAKAGE PROBLEM AT U101-25 aan es os ores aa OS ee ee ees = ees Sears GPs ofA a + \ io + \ | | ; Tt | L pp eae VERT .2V/DIV - a oo errr HORIZ .2 ms/DIV a t +1.01V INPUT = i -, - _ a re | 4 a PIN 24 PIN 25 Figure 6-15. Waveforms at U101-24 and U101-25 lems may be caused by a failure of the Calibration Mem- CAUTION ory (U220) or by bad A/D calibration constants. (To check for bad A/D calibration constants, clear the calibration memory.)

Readings at the A/D Converter can be determined by interpreting the waveform at the DAC output (TP102). Waveforms at TP102 for several input levels are shown in Figure 6-17. The A/D reading can be calculated by knowing the weight of each bit and by weighting each nibble correctly. (The first nibble is weighted 1, the second 1/16, the third 1/162, the fourth 1/163, etc.) Figure 6-18 shows how to read the A/D output for an input of 0.66V by interpreting the waveform at TP102 using the first three nibbles.

Troubleshooting the A/D Converter for defective components can be simplified by setting the circuit in a quiescent state. This can be done by selecting EX TRIG, which causes all A/D activity to stop. The A/D Converter is then in the autozero configuration, and the offset of the amplifiers and the various levels in the bootstrap circuits can be easily measured with a voltmeter. Oscillations at the outputs of the amplifiers and other abnormal signals can easily be identified with an oscilloscope.

6-67. Power Supply Troubleshooting If the display does not light up, first check the following:

1. Is the instrument plugged in to an energized outlet providing alternating current at a frequency of 47 Hz to 440 Hz and a voltage within +10% of that selected by the rear panel line voltage selection switches?

Is the POWER switch ON (pushed in)?

Is the rear panel fuse blown?

6-42

For fire protection, use exact fuse replacement only.

If the rear panel fuse is blown, replace it with a 250V fuse of the proper rating for the line voltage selected. Use 1/4A slow-blow for 100V and 120V power-line voltage and 1/8A slow-blow for 200V and 240V power-line voltage. If the fuse keeps blowing, measure the resistances of the power transformer (T601) windings. They should be within 10% of the values shown on the schematic. If not, the transformer is probably shorted. Also inspect the area around the transformer POWER switch and power-cord connector to make sure there is nothing shorting out the traces. If the IEEE-488 Interface board is suspected of causing the problem, it can easily be unplugged. Check the crowbar circuit (CR615 and Q601). If either component is shorted or if there is a large amount of leakage around or through CR615, fuses will continue to blow.

If everything looks OK but the fuse keeps blowing, troubleshooting may be performed by powering the instrument through a variac, applying only enough line voltage to find the problem without blowing the fuse. NEVER USE A LARGER FUSE. To do so will only turn a smali problem into a big one.

Since power supply problems can produce symptoms in many different sections of the instrument, the first step in troubleshooting any problem should usually be a quick check of the power supplies. For each power supply (TP801-TP806), check the level with a voltmeter and check for ripple with an ac-coupled oscilloscope. The dc voltages should be within the limits given in Table 6-23.

Page 148 read from a scan by OCR; it may hold reading errors

Maintenance

TROUBLESHOOTING ~ t . VERT IV/DIV. ene nana ba ia Ba id Dai 2M Bnd BD HORIZ. .2 ms/DIV [ COCCPETy o2- - tr L Figure 6-16. Typical Bus Data Line Waveform f r + 0 100 - oe ne ae aeeede t 1 i Pel _ = = fants eee pes len} a | den! 4 j = * ~ = O%--- “I 0% - L_ f, .66V INPUT, 1V/DIV, .1 ms/DIV +.66V INPUT, 1V/DIV, .1 ms/DIV I | me ‘to a be, = + LS an = wt L. oo - 6 ‘euttins tacaes oes + a io r an = 9. - Mee: L i Ba ] f, .03V INPUT, 1V/DIV, .1 ms/DIV + .03V INPUT, 1V/DIV, .1 mv/DIV

Figure 6-17. Waveforms at TP102 for Several inputs on 2V DV Range

6-43

Page 149 read from a scan by OCR; it may hold reading errors

Maintenance TROUBLESHOOTING fo}sfol+fo|s N1 = -24+140+1/44+0+1/16 = -.6875

_NIBBLE 1 | , NIBBLE2 , | NIBBLES | r 7 I “| r ma bt bé bE rFrEE = FEE = rer = FrEooa B rFEgaao® rFek-poa |B) a) S$; S] S$} SI |S) 2/$iS/Sjs] [RIFIS{;S{SIs|

| |

[sfofofs||1| N2 = 0+0+0+1/4+1/8+1/16 =

A/D Reading (1st 3 Nibbles) = -.6875 + (1/16) (+.4375) + (1/16)? (0.0) = -.660156

1] 0[ of 0] o[o| N3 = 0+0+0+0+0+0 =

+.4375 0.0

Figure 6-18. Calculating the A/D Reading From TP102 Waveform

Table 6-23. Power Supply Voltages

LIMITS (in volts) TEST POINT MINIMUM MAXIMUM

+5V 4.75 §.25 +7.5V 7.00 7.87 +15V 14.25 15.75 +30V 28.45 31.55 8.2 8.61 -7.60 -15V -15.75 “14.25 -30V -31.55 -28.45

If a supply is too high, either its three-terminal regulator has failed or a fault elsewhere in the instrument has shorted two supplies together. After repairing such a problem, make certain that nothing else was damaged by the overvoltage.

If a supply is too low, there are a number of possible causes. First check the input to the affected regulator. If it is not at least 1V above the maximum output given in Table 6-23, the cause may be a bad transformer winding (check the resistance), open or shorted Tectifiers, a shorted filter capacitor, or a shorted regulator. The latter two failures will usually blow the line fuse.

All regulators incorporate current-limiting which allows them to shut down in the event of a load failure. Therefore if the power supply output is too low, the first step should be to determine if it is due to a high load caused by a failure elsewhere in the instrument. Frequently the faulty component can be found by using a multimeter with at least 5 digits resolution to check the supply pins of all components powered from that supply. Connect one lead of the voltmeter to the appropriate test point for the power

6-44 supply under test and use the other lead to probe the loads. Small voltage drops across the PCA traces can be detected in this way, and the fault isolated. If any component other than one of the regulators is too hot to touch, there is something wrong with it or with something connected to it.

The True RMS AC PCA, if installed, uses +5V and +15V. If there is a problem with one of those supplies, first disconnect the True RMS AC PCA. If the problem goes away, troubleshoot the True RMS AC PCA using the procedure given later in this section.

With most power supply problems, the output voltage is too low or too high. More subtle problems that may be encountered include high ripple or oscillation. If more than 10 mV of line-frequency ripple exists on one of the power supply outputs, it is usually caused by the input being too low, causing the regulator to drop out of regulation. It is also possible (but not likely) that the regulator itself is defective. High-frequency oscillation (frequently syn-chronized with the 1 Mhz or 8 Mhz clock) is usually the result of a bad regulator or output bypass capacitor. A fair amount of high-frequency noise is generally present on all the supplies, particularly +5V, and should not cause any concern unless the instrument behaves erratically or the Teading is noisy.

6-68. IEEE-488 interface Troubleshooting (Option -05)

6-69. SERVICE POSITION

To provide easy access to the IEEE-488 Interface PCA and the Main PCA, the IEEE-488 Interface PCA can be placed in the specially provided service position as follows:

1. Remove the case from the chassis according to the Case Disassembly procedure provided earlier in this section.

Page 150 read from a scan by OCR; it may hold reading errors

Maintenance TROUBLESHOOTING

2. Release the two nylon latches that hold the IEEE-488 CAUTION Interface PCA in place by pulling the latches upward.

3. Position the IEEE-488 Interface PCA vertically as Shown in Figure 6-19 and latch it in place be pressing the two nylon latches into the mounting supports specially provided on the chassis.

6-70. DIAGNOSTIC PROGRAM gram is running.

To facilitate troubleshooting, the IEEE-488 Interface pro- 1. Ensure the 8840A POWER switch is OFF.

vides a diagnostic program which places the instrument in

To avoid damage to the 8840A or other equipment, the 8840A must be discon- nected from all other IEEE-488 interface instruments while the diagnostic pro- known configurations. To initiate the diagnostic program, 2. Disconnect all cables from the rear panel IEEE-488 proceed as follows: connector.

Figure 6-19. Option -05 Service Position

Table 6-24. Diagnostic Modes

SWITCHES CONFIGURATION

A3 A2 Al

1 0 1 Static, odd-port bits = 1, even-port bits = 0

1 1 1 Static, odd-port bits = 0, even-port bits = 1 j x 0 Dynamic

0 x x ReadMWrite NOTES:

. “x” means switch setting does not matter.

. “Static” means the Out-Guard 1.C V/O port bits programmed as outputs are driven to a constant logic 1 or 0 level (as defined by switch A2).

“Dynamic” means the Out-Guard uC W/O port bits programmed as outputs are driven with a 610 Hz, 50% duty cycle square wave. All odd port bit numbers are 180 degrees out of phase with even port bit numbers.

“Read/Write” means that data is read from and written to the NEC7210 IEEE chip (U901) when DS (U901 -8)

is low. R/W (U901-7) determines wheth er the data is being read from or written to the NEC7120. The address bits are always 3 (0011) and the data bits are incremented each time.