The text of a manual · part 16 of 18

8840A Digital Multimeter Instruction Manual, pages 151 to 160

Pages 151 to 160 of 180

The original document is the PDF (42 MB). This is its text as extracted, unedited; tables and drawings read best in the PDF, and each page heading opens it at that page.

For the Fluke 8840A

Page 151 read from a scan by OCR; it may hold reading errors

Maintenance TROUBLESHOOTING

3. Short TP903 to TP905.

4. Power up the 8840A. The 8840A should display ERROR 50. To exit the troubleshooting mode, open the jumper and cycle the POWER switch from off to on.

Once the diagnostic program is started, Tear-panel TEEE-488 address switches A3, A2, and Al can be used to select one of four diagnostic modes, as shown in Table 6-24. In this table, Configuration indicates which Out- Guard uC I/O port bits are programmed as outputs and driven with a signal, as shown in Table 6-25.

6-71. True RMS AC Troubleshooting (Option -09)

6-72. SERVICE POSITION

To provide easy access to the True RMS AC PCA and the Main PCA, the True RMS AC PCA can be placed in the specially provided service position as follows:

1. Remove the case from the chassis using the Case Disassembly procedure provided earlier in this sec-

2. Release the four nylon latches that hold the Tue RMS AC PCA in place by pulling the latches upward. (See Figure 809-1E in Section 8.)

3. Disconnect the red ac input lead from both the True RMS AC PCA and the Main PCA.

4. Position the Tue RMS AC PCA vertically as shown in Figure 6-20 and latch it in place by pressing the bottom two nylon latches into the specially provided mounting supports on the chassis.

5. Connect the Main PCA ac take-off point (stud connector W301) to the True RMS AC PCA input (the stud connector labeled AC IN) with a 6-inch jumper (E-Z- Hook 204-6W-S or equivalent).

6-73. MAJOR PROBLEMS

The signal flow through the Tme RMS AC option is straightforward, with no feedback paths between individual stages. This simplifies troubleshooting and often makes it possible to isolate a single defective stage without tion. removing the instrument cover. Table 6-25. /O Port Configurations CONFIGURATION CONFIGURATION PORT BIT Static PORT BIT Static Read/Write . Read/Write . yw Dynamic Dynamic PO-0 OUT address P1-0 OUT data PO-1 OUT address Pi-1 OUT data PO-2 OUT address P1-2 OUT data PO-3 OUT address P1-3 OUT data PO-4 IN IN P1-4 OUT data PO-5 IN IN P1-5 OUT data PO-6 IN IN P1-6 OUT data PO-7 IN IN P1-7 OUT data P2-0 IN IN P3-0 IN IN P2-1 IN IN P3-1 IN IN P2-2 IN IN P3-2 IN IN P2-3 IN IN P3-3 IN IN P2-4 IN IN P3-4 IN IN P2-5 IN IN P3-5 IN IN P2-6 IN IN P3-6 clock clock P2-7 IN IN P3-7 serial serial NOTES: ° Due to external hardware conflicts, the following bits are NEVER driven by the Out-Guard uC in ANY diagnostic mode: PO-4,5,6,7; P2 (all bits); P3-1,2,3,4. : P3-6 is the 4 MHz clock for the NEC7210 IEEE chip (U901). \ . P3-7 is programmed as the serial output, and constantly transmits hex 55 every 820 us at 62,500 baud in all four diagnostic modes. This causes the front panel error message.

6-46

Page 152 read from a scan by OCR; it may hold reading errors

Maintenance TROUBLESHOOTING

Figure 6-20. Option -09 Service Position

Test the mid-frequency response of the VAC function around 1 kHz. If an accurate reading can be obtained on at least one range, the rms converter (U802) is working properly. Test the high-frequency response around 100 kHz. If, after calibration, an accurate reading can be obtained on at least one range, the digitally controlled filter (U801, U808, R832, and C826-829) is OK. If some Tanges are good and others are bad, the defective stage may be isolated using Table 6-26. If this table is used, the bad ranges must correspond exactly to the ranges listed in the first column and all other ranges must be good.

Most ac troubleshooting can be performed with the shields removed. To remove both shields, unscrew the Phillips screw on the back of the True RMS AC PCA. The only time it should be necessary to work on the PCA with the shields in place is when there is subtle high-frequency (>20 kHz) or low-level (<10 mV) error. In that case, the PCA should be left in its operating position, and the test points probed from the foil side of the PCA. Test points are labeled on both sides to facilitate such troubleshooting.

If no ranges work, the signal should be traced from input to output. At any point where the signal disappears, the preceding stage should be searched thoroughly. To trace the signal, lock the instrument into one range (200 mV is usually a good choice) and apply the appropriate voltage shown in Table 6-27 to the HI and LO INPUT terminals.

Table 6-26. Isolating a Defective AC Stage

DEFECTIVE RANGES DEFECTIVE STAGE 200 mV, 2000 mA U806B

20V, 700V R804, R805

2V, 200V U806A

200V, 700V Input (Q806, K802, 2801) 2V, 20V, 200V input (Q806, K802, 2801)

The input voltage should appear unchanged at pin 2801-1, and should appear at TP801 and TP802 as shown in Table 6-27. If no ranges work, it is likely that the rest of the scaling circuitry (U806B) is functional. .

If the signal at the input to U801A (pin 5) is incorrect, U804 may be defective, or the switch codes may be wrong. If the latter problem is suspected, refer to Table 6-28 and test the control lines to U804 (U804-1,8,9,16). If a logic error is found, it may be due to excessive loading or a faulty data latch (U803), or other cabling or main-board digital problems. High-frequency oscillation problems are usually caused by switches being on when they should be off, resulting in positive feedback loops being closed around portions of the scaling circuitry.

If the signal at TP802 is incorrect, but U801-5 is OK, the digitally controlled filter section (U801A and U808) is probably defective.

If the signal at TP802 is correct but the output signal (TP803) is incorrect, the rms converter is probably the source of the problem. Problems with U802 generally show up as an identical number of counts displayed in all Tanges OF as an overrange in all ranges. First isolate faults in the buffer amplifier (U802A) by ascertaining that the signal at U802-13 is the same as that at TP802 when each is observed with an ac-coupled oscilloscope, and that the dc offset at U802-13 is less than 4 mV. It is possible that a component in the loop filter (U809A and associated passive components) or the post-filter (U809B and associated passive components) is defective. The dc voltage at U802-6 should be the same as that at TP803 for frequencies above 500 Hz, and should be equal to the rms value of the input signal.

6-74. MORE OBSCURE PROBLEMS

Slow settling time or excessive jitter for low-frequency inputs is caused by rms converter loop errors. The cause

6-47

Page 153 read from a scan by OCR; it may hold reading errors

Maintenance

TROUBLESHOOTING

Table 6-27. AC Signal Tracing

RANGE VOLTAGE VOLTAGE | VOLTAGE (i kHa) | AT TPeot | aT TPa02 200 mV 100 mv | 20 mv tv ov 1V 200 mv (| 1 Vv 20V 10V eV 1V 200V 10V 20 mV 100 mV 700V 100V 200 mV 100 mv may be a fault in the rms converter or loop filter.

If the output voltage is stuck at the supply rails, the cause is probably a fault in the rms converter. A less common cause is op amp oscillation; this can be checked with a scope at TP802.

If one or more ranges are functional but cannot be calibrated at high frequencies, then either the digitally controlled filter (U801B, R832, and C826-C829) is defective, or a defective component elsewhere in the circuit has rendered the response out of calibration range. (The high-frequency calibration is designed to cover the range of error expected due to op amp variations, input dividers, PCA tolerances, shielding, etc.). A sweep generator is useful in troubleshooting difficult frequency response problems.

The calibration control lines to U808 are set by software to Store a high-frequency correction factor for each range. A state table cannot be given for these signals, but common sense will indicate if they are reasonable. For example, if all lines are at logic 0 for all ranges, something is probably wrong. Suspect U803, U808, or digital hardware on the Main PCA. If the control signals do indeed change with range, U808 or some part of the digitally controlled filter (U801B, R832, and C826-C829) may be defective. Even with the worst possible error in the high-frequency calibra-

Table 6-28. Truth Table for U804 and K2

PIN OR DEVICE

RANGE

U804-1 U804-8 | U804-9 | U804-16 /K2 2000 mA 1 0 1 4 (0) 200 mV 1 0 0 1 0 2vV 4 1 1 0 0 20V 0 1 1 1 0 200V q 1 1 0 1 700V 0 1 1 1 1 NOTE:

For U804, logic 0 = switch on. Logic 1 is >2.4V; logic 0 is <0.8V.

6-48 tion code, the reading should be within 10% of the correct value at frequencies up to 100 kHz. If the error is larger, there are analog problems.

It is safe to force one control line at a time high (+5V) or low (OV) to test the individual switches in U808. (The on resistance of switches in U808 should be less than 500Q; the off resistance should be greater than 10 MQ.) Forcing the control lines high or low should cause the reading to change when the voltages in Table 6-27 are applied to the input terminals at 100 kHz. If only certain ranges cannot be calibrated, refer to Table 6-26 to find the suspected stage.

6-75. Guard Crossing Troubleshooting

To troubleshoot the Guard Crossing circuit, place the 8840A in the In-Guard Troubleshooting Mode as described under Digital Controller Troubleshooting, earlier in this section. This causes the In-Guard uC to send a test pattern to the IEEE-488 Interface PCA via one-half of the Guard Crossing circuit. You should be able to observe the waveforms shown in Figure 6-22.

To troubleshoot the other half of the circuit, run the diagnostic program as described under IEEE-488 Interface Troubleshooting earlier in this section. The program causes the IEEE-488 Interface PCA to send the same test pattern to the In-Guard C. You should be able to observe the waveforms shown in Figure 6-21 at the corresponding points in this half of the circuit.

6-76. INTERNAL CLEANING

CAUTION

Failures due to electrostatic discharge can be caused by improper handling of the PCAs and by the use of a vacuum cleaner with static-inducing brushes. To prevent damage by electrostatic dis- charge, observe the precautions described on the Static Awareness sheet in front of this section.

If visual inspection of the instrament shows excessive dirt build-up in the instrument, clean the appropriate section using clean, oil-free, low-pressure air (less than 20 psi). If necessary, remove the option PCAs first.

6-77. Cleaning Printed Circuit Assemblies

If conditions warrant, individual printed circuit assemblies (PCAs) can be cleaned with water-based commercial cleaning systems such as dishwashers. If such systems are used, observe the following precautions:

1. Remove all shield covers (applies to the True RMS AC PCA) and socketed ICs.

2. Use Reagent Grade 2 or better water (de-ionized or distilled water) for the final rinse in geographic areas with exceptionally hard water. During the final rinse,

Page 154 read from a scan by OCR; it may hold reading errors

Maintenance INTERNAL CLEANING

Spray or run the water so that the surface is thoroughly covered to remove all ionized material.

3. Thoroughly dry all PCAs using one of the following itil [" methods: U221-11 = = _ a. Preferably, the PCA should be dried in a low- m7 temperature drying chamber or infrared drying U221-10

Tack with a temperature range of 49°C to 72°C (120°F to 160°F).

L lo

U221-8 b. If neither a drying chamber nor a drying rack is available, air dry the PCA at ambient room tem- J perature for at least two days.

A satisfactory cleaning method consists of holding the 5V PCAs under hot running water until they are clean. Follow | this wash with a final rinse. (See consideration 2, above.) U901-5

6-78. Cleaning After Soldering, 16 us |

CAUTION 112 us - >}

T.M.C. Cleaner and similar products can can attack the nylon latches and other 0.2V plastic pieces. U908-7 a a [ “" }SEEN ON, OV TIEEE-ag8

If a PCA has been soldered, it should first be cleaned with _ oovieon SPRAYON T.M.C Cleaner™ (rosin flux remover) or Ug08-5 | | | | | | i ov equivalent. The PCA should then be cleaned with water as described above. NOTE:-

@ THESE WAVEFORMS SEEN IN IN-GUARD TROUBLESHOOTING MODE.

@ REPETITION RATE = 500 us. INTERRUPTIONS EVERY 1.5 SEC.

Figure 6-21. Guard Crossing Test Waveforms

Trademark of E.E. DuPont De Nemours & Co.

6-49

Page 155 read from a scan by OCR; it may hold reading errors

Maintenance INTERNAL CLEANING

6-50

Page 156 read from a scan by OCR; it may hold reading errors

Section 7 List of Replaceable Parts

TABLE OF CONTENTS

ASSEMBLY NAME

DRAWING TABLE FIGURE NO. NO. PAGE NO. PAGE Digital Multimeter Assembly 8840A T&B 7-1 7-3 7-1 7-4 Al Main PCA 8840A-7601 7-2 7-8 7-2 711 A2 Display PCA 8840A-7602 7-3 7-12 7-3. 7-12

7-1

Page 157 read from a scan by OCR; it may hold reading errors

8840A INSTRUCTION MANUAL

7-1. INTRODUCTION

This section contains an illustrated list of replaceable parts for the 8840A. Parts are listed by assembly; alphabetized by reference designator. Each assembly is accompanied by an illustration showing the location of each part and its reference designator. The parts lists give the following information:

Reference designator

An indication if the part is subject to damage by static discharge

Description

Fluke stock number

Total quantity

Any special notes (1.e., factory-selected part)

CAUTION

A * symbol indicates a device that may be damaged by static discharge.

7-2. HOW TO OBTAIN PARTS

Electrical components may be ordered directly from the manufacturer by using the manufacturers part number, or from the Fluke Corporation and its authorized representatives by using the part number under the heading FLUKE STOCK NO. In the U-S., order directly from the Fluke Parts Dept. by calling 1-800-526-4731. Parts price information is available from the Fluke Corporation or its representatives. Prices are also available in a Fluke Replacement Parts Catalog which is available on request.

In the event that the part ordered has been replaced by a new or improved part, the replacement will be accompanied by an explanatory note and installation instructions, if necessary.

To ensure prompt delivery of the correct part, include the following information when you place an order:

¢ Part number and revision level of the pca containing the part.

* Reference designator e Fluke stock number e Description (as given under the DESCRIPTION heading)

e Quantity e Instrument Model, Serial Number, and Firmware

Numbers

7-3. MANUAL STATUS INFORMATION

The Manual Status Information table that precedes the parts list defines the assembly revision levels that are documented in the manual. Revision levels are printed on the component side of each pca.

7-4. NEWER INSTRUMENTS

Changes and improvements made to the instrument are identified by incrementing the revision letter marked on the affected pca. These changes are documented on a manual supplement which, when applicable, is included with the manual.

7-5. SERVICE CENTERS

To locate an authorized service center, call Fluke using any of the phone numbers listed below, or visit us on the World Wide Web: www.fluke.com

1-800-443-5853 in U.S.A and Canada 31 40 267 8200 in Europe 1-425-356-5500 from other countries oy

Ya? NOTE This instrument may contain a Nickel-Cadmium battery. Do not mix with the solid waste stream. Spent batteries should be disposed of by a qualified recycler or hazardous materials handler. Contact your authorized Fluke service center for recycling information.

A\ WARNING

THIS INSTRUMENT CONTAINS TWO FUSIBLE RESISTORS (PN 474080 AND 733915). TO ENSURE SAFETY, USE EXACT REPLACEMENT ONLY.

MANUAL STATUS INFORMATION

REF OR OPTION NO. ASSEMBLY NAME FLUKE PART NO. REVISION LEVEL Al Main PCA 728857 BY A2 Display PCA 728873, 05 IEEE-488 Interface PCA 879267 D o9 True RMS AC PCA 728899

7-2

Page 158 read from a scan by OCR; it may hold reading errors

Maintenance Test

PERFORMANCE TEST Table 7-1. 8840A Digital Multimeter Ref. Des. Description PN Qty. Notes Al MAIN PCA 728857 1 A2 DISPLAY PCA 728873 1 F301 FUSE,.406,1.375,3A,600V,FAST 475004 1 F302 FUSE, .25X1.0,2A,250V,FAST 376582 1 F601 FUSE, .25X1.25,0.25A,250V,SLOW 166306 1 F601 FUSE,.25X1.25,0.125A,250V,SLOW 166488 1 H8 FASTENER,PLUNGER,NYL,.271 682906 6 H9 FASTENER, GROMMET,POLYCARB,.271 682898 6 H10 NUT,EXT LOCK,STL,6-32 152819 3 H11 SCREW,PH,P,LOCK,STL,6-32,.250 152140 1 H13 SCREW,TH,P,SS,4-40,.187 721118 2 H14 SCREW,FIH,P,LOCK,STL,8-32,.500 721134 2 H15 SCREW,PH,P,THD FORM,STL,4-20,.500 721449 4 H17 SCREW,FHU,P,LOCK,SS,6-32,.250 320093 1 H18 SCREW,FH,P,LOCK,STL,8-32,.375 114116 2 H19 SCREW,PH,P,LOCK,SS,6-32,.312 424713 2 H20 SCREW,PH,P SEMS,STL,4-40,.375 281196 1 H24 SPACER 735407 1 H25 SPACER 735415 1 H27 SCREW,FH,P,LOCK,STL,6-32,.750 114504 2 MP1 SHIELD MAIN BOTTOM 728998 1 MP2 BRACKET, TRANSFORMER 765263 1 MP3 CASE, METAL 656389 1 MP4 FRONT PANEL 656397 1 MP5 PUSHROD, POWER 656413 1 MP6 PUSH ROD RAI 656488 1 MP7 PUSH ROD CAL ENABLE SWITCH 684142 1 MP8 MOUNTING PLATE, HANDLE, LEFT, MED PEW 729004 1 MP9 MOUNTING PLATE, HANDLE,RIGHT, MED PEW 660472 1 MP10 HANDLE, PAINTED DARK UMBER 810127 1 MP11 BEZEL,REAR 660480 1 MP2 SPACER, MATRIX 684126 1 MP13 CASE FOOT,BLACK 824433 2 MP14 CHASSIS ASSEMBLY 684134 1 MP15 TERMINAL BLOCK, RAI 656504 1 MP16 SEAL,CALIBRATION 735274 1 MP17 SUPPORT, PWB 683987 14 MP18 LENS,FRONT PANEL 656405 1 MP19 DECAL,FRONT 685081 1 MP20 DECAL, REAR 698431 1 MP21 INSERT, REAR PANEL 684191 1 MP22 IEEE CABLE GUIDE 728956 1 MP23 DECAL CSA 525527 1 MP25 SPACER,RND,NYL,.166,.250 772889 2 MP26 SHIELD, SUPPORT 735308 4 MP27 HLDR PART,FUSE,CAP, 1/4X1-1/4 460238 1 MP32 FUSE HOLDER/CAP ASSEMBLY 516039 1 MP57 NAMEPLATE, SERIAL -REAR PANEL- 472795 1 MP58 LABEL,CE MARK, BLACK 600707 1 S1 SWITCH,CONDUCTIVE ELASTOMERIG 680769 1 S601 LINE VOLTAGE SELECTION SWITCH 735258 1 TM1 INSTRUCTION MANUAL. 879304 1 TM2 GETTING STARTED MANUAL 879291 1 1; TMS QUICK REFERENCE GUIDE 879296 1 Ww4 CORD,LINE,R/A 5-15/IEC,3-18AWG 707018 1 W7 TEST LEAD ASSY, TL70A 855820 1

7-3

Page 159 read from a scan by OCR; it may hold reading errors

yee

~ - ~ MP9

MP10

MP 13 rve 2 Pic

8840A T&B (1 of 4)

7-4

' Figure 7-1. 8840A Digital Multimeter

Page 160 read from a scan by OCR; it may hold reading errors

List of Replaceable Parts 7

KMS AC

(OPTIONAL)

‘MP7

IE

B840A- 09

Tey

BB40A iEEE-488 INTERFACE {OP TIONAL)

F301, 4 FI x oa ag SH Qa, = oO = soa S & =

OETAIL T HD SCALE

8840A T&B (2 of 4)

Figure 7-1. 8840A Digital Multimeter (cont) 75