Browse
(440) 299-7778
Menu
(440) 299-7778
One-year parts and labor warranty30-day money-back guaranteeContact

The text of a manual · part 1 of 1

Instruction Manual R82000, T82000, P82000, P82000 FR (including ATEX & HF), pages 1 to 7

Pages 1 to 7 of 7

The original document is the PDF (419 KB). This is its text as extracted, unedited; tables and drawings read best in the PDF, and each page heading opens it at that page.

Filed in the catalog under the GIC Thermodynamics PT100

Page 1

INSTRUCTION MANUAL R82000 T82000 P82000 P82000 FR (including ATEX & HF)

S-Products B.V. Nijverheidscentrum 26, NL-2761JP Zevenhuizen, The Netherlands, Tel: +31 180 633455 , Fax: +31 180 633644 S-Products Inc. 35, Kings Highway East, Fairfield, CT 06825, USA, Tel: +1 203 331 9546, Fax: +1 203 335 2723 E-mail: info@s-products.com Internet: www.s-products.com

Document No. Instructie82.cdr - Issue 031105

Page 2

SPECIFICATIONS GENERAL INTRODUCTION: The 82000 series are highest quality free to scale 2-wire transmitters. The range is set with an external solder link on top of the transmitter. Can even be re-scaled in the field. Fine ZERO and SPAN adjustments using independent precision potentiometers. Electronics are completely sealed in rugged epoxy-coated metal housing. RTD (Pt-100) units are completely linearized and thermocouple units are voltage linear. The Potentiometer transmitter represents a 0...100% setting Off-the-Shelf availability for RTD (Pt-100) and thermocouples type K and J. Other Thermocouple inputs and an Intrinsically safe version is available.

Specifications at Vnom 24VDC; Tam b25°C and Span nom. RTD=100°C; T/C=250°C GENERAL 4-20 mA Output 12-36 VDC Power supply* ±0.02%/°C Zero Drift ±0.02%/°C Span drift ±0.002%/V Supply Voltage Effect Upscale >23 mA, limited to 40 mA Open Circuit Detection (Vbat-12V) / 20 mA Load Capability -20° to 80°C (-4° to 160°F) Ambient Temp. -20° to 100°C (-4 to 210°F) Storage Temp. <1% (typ.) Frequency Effect 43mm Dia. x 25mm H. (1.7"x1.0") Dimensions IP20 Degree of protection Pt100(alpha=0.00385) Zero Adjustment Span Adjustment

Model R82000 between -50° and +50°C (-60° and +120°F), 20 turn Potentiometer adjustment -50° to 800°C (90° to 1440°F) divided into 4 parts. Scaleable via solder-link on top of unit. Fine adjustment via 20-turn Potentiometer. Sensor Lead Resistance Effect 0.01°C/Ohm; max. 500 Ohm 0.1% Linearization Thermocouple Model T82000 Thermocouple K Model T82000-K Thermocouple J Model T82000-J approx. 50 °C (60°F) via 20-turn potentiometer Zero Adjustment Type K 150° to 1200°C (270° to 2160°F) divided in 3 parts. Span Adjustment Type J 150° to 600°C (270 to 1080°F) divided in 2 parts. 0.03°C/°C Cold Junction drift 10 K Ohm max. Lead Resistance Voltage Linear Output Potentiometer Model P82000 FR 1 kiloOhm Minimum potentiometer 50 kiloOhm Maximum potentiometer 0...100% = 4...20 mA Fixed Range Range Variable Resistor Model P82000 Factory set; specify Resistor Range Our Models R83000 and T83000 are the factory calibrated versions of R82000 and T82000. These part numbers are no longer in our program.

*NOTE: For ATEX approved equipment different power requirements may apply!

S-Products B.V. Nijverheidscentrum 26, NL-2761JP Zevenhuizen, The Netherlands, Tel: +31 180 633455 , Fax: +31 180 633644 S-Products Inc. 35, Kings Highway East, Fairfield, CT 06825, USA, Tel: +1 203 331 9546, Fax: +1 203 335 2723 E-mail: info@s-products.com Internet: www.s-products.com

Document No. Instructie82.cdr - Issue 031105

Page 3

ATEX GENERAL: The R/T82000 is available in an intrinsically safe version and meets the requirements of the ATEX directive 94/9/EC. MARKING: All ATEX approved R/T82000 units carry the following labels:

General product label

R/T82000 (HF) Year of production Serial No. i.e.:

S-Products BV R/T82000(HF) Nijverheidscentrum 26, NL-2761 JP

K E M A 0 3 AT E X 141 8 0344 II 1 G

X .

2 - WIRE TRANSMITTER MODEL 82000

INPUT RANGE

C05200

EEx ia IIC T4/T5/T6 See certificate for data Ui < 30V Ii <100mA Pi < 750mW

+

Certificate No.

KEMA 03ATEX 1418 X

R/T82000 (HF):

+

OUTPUT 4 - 20 mA

Parameters: Connect to a Zenerbarrier with the following maximum values: Uo = 30 V DC Io = 100 mA Po = 750 mW

R82000 HF & T82000 HF supply circuit KL1, KL2: UI = 30 V DC II = 100 mA PI = 750 mW LI = 55 µH CI = 58 nF input circuit KL3, KL4, KL5: Uo = 30 V DC Io = 100 mA Po = 750 mW Lo = 3,4 µH Co = 8 nF

R82000 & T82000 supply circuit KL1, KL2: UI = 30 V DC II = 100 mA PI = 750 mW LI = 22 µH CI = 15 nF input circuit KL3, KL4, KL5: Uo = 30 V DC Io = 100 mA Po = 750 mW Lo = 3,4 µH Co = 40 nF

+

DC POWER

Apparatus without further specification

0344 II 1 G EEx ia IIC T4 (Tamb -20...+80°C) T5 (Tamb -20...+80°C) T6 (Tamb -20...+50°C)

+

ZENER BARRiER

EX TRANSMiTTER SENSOR

-

-

NON-HAZARDOUS AREA

HAZARDOUS AREA

S-Products B.V. Nijverheidscentrum 26, NL-2761JP Zevenhuizen, The Netherlands, Tel: +31 180 633455 , Fax: +31 180 633644 S-Products Inc. 35, Kings Highway East, Fairfield, CT 06825, USA, Tel: +1 203 331 9546, Fax: +1 203 335 2723 E-mail: info@s-products.com Internet: www.s-products.com

Document No. Instructie82.cdr - Issue 031105

Page 4

NORMATIVE REFERENCE: The transmitters R82000 (HF) and T82000 (HF) have an EC Type examination certificate issued by KEMA and have been approved to the following standards:

EN 50014 EN 50020 EN 50284 SPECIAL CONDITIONS FOR SAFE USE: For electrical data and temperature data see EC-Type examination certificate

KEMA 03ATEX1418 X

INSTALLATION REQUIREMENTS - ATEX approved equipment is only to be connected to a Zenerbarrier. The Temperature Transmitter must be mounted in an enclosure providing a degree of ingress protection of at least IP20 per EN 60529. This enclosure must be in conformance with Clauses 4.3 and 4.4 of EN 50284, when the Temperature Transmitter is mounted in an area where the use of category 1 G apparatus is required. This enclosure must be in conformance with Clauses 7.3 and 8.1 of EN 50014, when the Temperature Transmitter is mounted in an area where the use of category 2 G apparatus is required. The mentioned Clauses describe how to avoid danger of ignition due to electrostatic charges. - All 82000 series transmitters have an ingress protection of IP20. This requires usage in a dry, clean and well controlled environment. Mount the transmitter in a (DAN) head as described below.. - Use twisted pair, shielded cable to connect the transmitter to its power supply in order to obtain the best immunity to Electromagnetic signals. On the power supply side connect the shield to the power supply’s earthing point, on the transmitter side connect the shield to the grounded head somewhere close to the transmitter. Make sure the transmitter is firmly connected to the head.

S-Products B.V. Nijverheidscentrum 26, NL-2761JP Zevenhuizen, The Netherlands, Tel: +31 180 633455 , Fax: +31 180 633644 S-Products Inc. 35, Kings Highway East, Fairfield, CT 06825, USA, Tel: +1 203 331 9546, Fax: +1 203 335 2723 E-mail: info@s-products.com Internet: www.s-products.com

Document No. Instructie82.cdr - Issue 031105

Page 5

CONNECTIONS

P82000 FR

P82000

POWER (24V DC) -

POWER (24V DC) -

+

KL2

KL2

KL1

ZERO

SPAN KL5

KL3

KL1

J1 J2 J3 J4

J1 J2 J3 J4

ZERO

+

SPAN KL5

KL3

KL4

KL4

20 mA at max. Ohm value

4 mA at min. Ohm value

Connect Potentiometer of min. 1 kiloOhm (Max. 50 kiloOhm)

Connect variable resistor

R/T82000 POWER (24V DC) -

+

KL2

KL1

J1 J2 J3 J4

ZERO

RANGE connect J1 J2 J3 J4

SPAN

50...100 150...300 150...300 100..200 300...600 300..600 200...400 600..1200 400...800

KL5

KL3 KL4

RTD + thermocouple table 1 [°C] Pt100 Thermocouple K J

R82000 -

T82000

S-Products B.V. Nijverheidscentrum 26, NL-2761JP Zevenhuizen, The Netherlands, Tel: +31 180 633455 , Fax: +31 180 633644 S-Products Inc. 35, Kings Highway East, Fairfield, CT 06825, USA, Tel: +1 203 331 9546, Fax: +1 203 335 2723 E-mail: info@s-products.com Internet: www.s-products.com

Document No. Instructie82.cdr - Issue 031105

Page 6

INTRODUCTION The series R/T82000 are highest quality free to scale 2-wire temperature transmitters. The P82000 FR is for connecting a Potentiometer. The range is set with an external solder link. The P82000 FR is Fixed Range. Fine adjustment of the range can be done by potentiometers located on the top of the transmitter With this transmitter the following ranges are covered: MODEL R82000 T82000-K T82000-J P82000 FR

INPUT RTD (Pt100) Thermocouple K Thermocouple J Potentiometer

ZERO -50...+50°C 0°C 0°C 0

SPAN 50.....800°C 150...1200°C 150.....600°C 100% or consult factory for other ranges or input sensors.

PROGRAMMING AND CALIBRATION INSTRUCTIONS

R82000 & T82000

In order to calibrate the transmitter you will need a welder and solder, 24V power supply with mA indicator (Model TL245 is recommended as a fast and simple calibration tool) and an input simulator (Pt100 or TC) 1.Select the desired Span from table 1. Please note if, for example, -50°...+150° is required the span is 200°, and not 150°. Link the two pads with solder. 2.Connect a power supply (24 V DC) and a digital milli amperemeter to the transmitter. 3.Connect an input simulator (Pt100 or Thermocouples with cold junction compensation) as shown in the connection diagram. 4.Simulate the required Zero and adjust the loop current to 4 mA with the "ZERO" potentiometer. 5.Simulate the maximum input value and adjust the loop current to 20 mA with the "SPAN" potentiometer. 6.Repeat step 4 and 5 until the correct reading is achieved. - + 7.Remove the input simulator and power supply. POWER Fill in the label with a permanent marker (fibrepoint). (24 VDC)

mA indicator

The transmitter is now ready to build in a connection head.

-

+ J1 J2 J3 J4

ZERO

SPAN

PROGRAMMING AND CALIBRATION INSTRUCTIONS P82000 FR In order to calibrate the transmitter you will need a 24V power supply with mA indicator (Model TL245 is recommended as a fast and simple calibration tool) and an input simulator (Potentiometer) Pt100

1.Connect any potentiometer with a value between 1 kÙ and 50 kÙ as shown in the connection diagram.. thermocouple 2. Connect a power supply (24 V DC) and a digital milli amperemeter to the transmitter. potentiometer 3.Simulate the required Zero and adjust the loop current to 4 mA with the "ZERO" potentiometer. 4.Simulate the maximum input value and adjust the loop current to 20 mA with the "SPAN" potentiometer. 5.Repeat step 3 and 4 until the correct reading is achieved. 6.Remove the input simulator and power supply. Fill in the label with a permanent marker (fibrepoint).

S-Products B.V. Nijverheidscentrum 26, NL-2761JP Zevenhuizen, The Netherlands, Tel: +31 180 633455 , Fax: +31 180 633644 S-Products Inc. 35, Kings Highway East, Fairfield, CT 06825, USA, Tel: +1 203 331 9546, Fax: +1 203 335 2723 E-mail: info@s-products.com Internet: www.s-products.com

Document No. Instructie82.cdr - Issue 031105

Page 7

DIMENSIONS

5 mm 0.2 "

33 mm 1.30 "

20 mm 0.79 "

43.3 mm 1.70 "

5 mm 0.2 "

43.3 mm 1.70 " 1

20 mm 0.79 "

6 2

5

4

3

MOUNTING in CONNECTION HEAD 1. Screw M4 2. Transmitter 3. Measuring insert plate 4. Ferrule 5. Spring 6. Notch

WARRANTY

Our transmitters have a 5 year warranty. Follow usage instruction in this manual. Do not try to open the transmitters. In case of malfunction return to your supplier.

S-Products B.V. Nijverheidscentrum 26, NL-2761JP Zevenhuizen, The Netherlands, Tel: +31 180 633455 , Fax: +31 180 633644 S-Products Inc. 35, Kings Highway East, Fairfield, CT 06825, USA, Tel: +1 203 331 9546, Fax: +1 203 335 2723 E-mail: info@s-products.com Internet: www.s-products.com

Document No. Instructie82.cdr - Issue 031105