Page 1 of Feasibility of formal model checking in the Vitatron environment

Manual · Vitatron

Feasibility of formal model checking in the Vitatron environment

Manual91 pagesPDF 6.4 MB

Master's thesis on applying formal model checking (mCRL2 and Uppaal) to the firmware design of Vitatron's DA+ pacemaker.

What this file really is

The old site listed this file as a manual for the Vitatron EE244. This is not an equipment manual for the Vitatron EE244 but a 2007 university master's thesis on formal model checking of pacemaker firmware.

Download the PDF6.4 MB

Filed with · 1 model

Also filed with these models on the old site

Summary

What is in it

A master's thesis by J.E. Wiggelinkhuizen from Technische Universiteit Eindhoven, carried out at Medtronic SQDM/Vitatron in Arnhem. It investigates whether formal model checking can verify pacemaker firmware designs, using Vitatron's DA+ pacemaker as a case study. It explains model checking and the mCRL2 and Uppaal tools, the human heart and arrhythmias, the pacemaker architecture, formal models of the firmware, verification results, and recommendations to Vitatron.

Its parts

  1. Introduction
  2. Model checking
  3. The human heart
  4. Pacemaker architecture
  5. Formal modeling of pacemaker firmware
  6. Model checking approaches and results
  7. Recommendations to Vitatron
  8. Conclusions

From the manual

Key facts from this manual

Each line names its page in the PDF.

This is the public version of a December 2007 master's thesis from Technische Universiteit Eindhoven, carried out at Medtronic SQDM/Vitatron in Arnhem, on whether formal model checking can verify pacemaker firmware designs. It is not a user manual and does not name the EE244; its case study is the firmware design of the Vitatron DA+ pacemaker. It describes the pacemaker and firmware architecture, formal models in mCRL2 and Uppaal, the verification results and recommendations.

Read for this summary: pages 1 to 8, 41 to 44 and 86 to 88 of 91.

Document type
Master's thesis, public version, December 2007, Technische Universiteit Eindhoven p. 1
Case study
Firmware design of the Vitatron DA+ pacemaker p. 2
Tools compared
mCRL2 and Uppaal model checking tools p. 2
Pacemaker parts
Can, hybrid (digital and analog hardware with firmware in ROM), battery, antenna for the programmer, lead connector p. 41
Pulse generator
Interface between leads and firmware: delivers paces on firmware request and reports sense and pace events, with atrial and ventricular timers p. 42
Firmware design
Event-driven: an interrupt service routine handles each external event, which is then queued in the operating system p. 42
Specification method
Hatley and Pirbhai structured analysis method p. 43
Telemetry
Antenna and telemetry hardware let an external programmer program and read firmware parameters p. 41
Heart model result
Verified for low to average heart rates; high rates made the state space explode p. 3
Design part result
Verifying the part that handles one specific arrhythmia found a known deadlock p. 3
Tool advice
mCRL2 recommended over Uppaal for firmware designs p. 3
Conclusion
Not confirmed as an independent verification technique; possibly valuable as a supplementary test technique p. 86

What this manual walks you through

  • The parts of an implantable pulse generator and the firmware environment p. 41, 42
  • The Hatley and Pirbhai requirements model used for the firmware design p. 43, 44
  • Summary of the model checking approaches and results p. 2, 3
  • Recommendations for specifying and checking future firmware designs p. 87, 88
  • Glossary of pacemaker and modeling terms p. 6, 7

Good to know

  • This public version has some text removed for confidentiality reasons. p. 42
  • The firmware models were analyzed for only one particular setting of the programmable parameters. p. 86

Questions this manual answers

Which Vitatron pacemaker does this report study?

Its case study is the firmware design of the Vitatron DA+ pacemaker. p. 2

Which model checking tools were compared for Vitatron firmware?

mCRL2 and Uppaal; the report recommends mCRL2 because of limited results with Uppaal. p. 3

What is the pulse generator in a Vitatron pacemaker?

In the DA+ pacemaker it is the hardware module between the pacemaker leads and the firmware. p. 6

Did model checking find an error in the pacemaker firmware design?

Checking the design part that handles one specific arrhythmia found a deadlock Vitatron already knew about, rather soon. p. 3

Contents

Chapters and sections

Each line opens the PDF at its page.

  1. Summary2
  2. Glossary6
  3. Chapter 1Introduction8
  4. Chapter 2Model checking11
  5. Chapter 3The human heart33
  6. Chapter 4Pacemaker architecture41
  7. Chapter 5Formal modeling of pacemakerfirmware49
  8. Chapter 6Model checking approaches andresults61
  9. Chapter 7Recommendations to Vitatron77
  10. Chapter 8Conclusions86
  11. Bibliography90

From the PDF’s own bookmarks.

The whole text · 14 parts

Read the manual here as text

Its text as extracted from the PDF, unedited, in parts; each page opens the PDF at that page.

  1. Summary1-3
  2. Pages 4 to 74-7
  3. Chapter 1Introduction8-10
  4. Chapter 2Model checking, pages 11 to 1811-18
  5. Chapter 2Model checking, pages 19 to 2619-26
  6. Chapter 2Model checking, pages 27 to 3227-32
  7. Chapter 3The human heart33-40
  8. Chapter 4Pacemaker architecture41-48
  9. Chapter 5Formal modeling of pacemakerfirmware, pages 49 to 5449-54
  10. Chapter 5Formal modeling of pacemakerfirmware, pages 55 to 6055-60
  11. Chapter 6Model checking approaches andresults, pages 61 to 6861-68
  12. Chapter 6Model checking approaches andresults, pages 69 to 7669-76
  13. Chapter 7Recommendations to Vitatron77-85
  14. Pages 86 to 9186-91

The file

Revision and size

RevisionMaster's Thesis, public version, December 2007
PublisherTechnische Universiteit Eindhoven, Department of Mathematics and Computer Science
Pages91
PDF6.4 MB
TextText from the PDF itself.