Page 8
Chapter 1
Introduction Medtronic is the global leader in medical technology. With deep roots in the treatment of heart disease, Medtronic now provides a wide range of products and therapies. Medtronic Subcutaneous Diagnostics and Monitoring (SQDM) is a recently established organization residing in the Medtronic Arnhem facility, where also the Medtronic Vitatron organization is located (in the sequel we will depict these two organizations by Vitatron). Vitatron focuses on subcutaneous (under the skin) implantable medical monitoring devices, used for diagnostic data collection during a number of years. Before the switch to these implantable monitoring devices, pacemakers were developed at Vitatron. This project is carried out at the firmware development department. This department is responsible for designing, implementing and testing all firmware located in the pacemakers and implantable monitors based on requirements delivered by the system department. A considerable amount of the development effort consists of testing. Besides the firmware development department, the Research&Development organization consists of a software group, basically developing Object-Oriented, PC-based software and a hardware development group developing the chip-set used in the pacemakers and implantable monitors. Basically Vitatron works from idea until production preparation for their products. The firmware forms the intelligent core of the devices that are developed at Vitatron, i.e. it controls the behavior of the device hardware. In a pacemaker the firmware controls the hardware such that an adequate heart rate is maintained in the patient’s heart, which is necessary either because the heart’s native pacemaker is insufficiently fast, or there is a block in the heart’s electrical conduction system. An adequate heart rate is required to ensure that the heart pumps enough blood into the body, according to the patient’s activity level. To accomplish this adequate heart rate, the pacemaker is connected to the chambers of the heart by electrodes (pacemaker leads). Via these leads, electrical stimuli (paces) can be delivered to force a contraction of the heart’s muscle cells. The pacemaker also retrieves information from the heart via these leads; contractions of the heart are measured and result in input signals on the pacemaker leads. Together with hardware notifications that indicate a delivered pace, these are the main inputs for the pacemaker firmware, which calculates whether the current heart rate is appropriate on every input signal. These calculations always result in commanding the hardware to deliver a pace to the heart on a certain moment in time. The planned pace will ensure a heart contraction and
7
Page 9
prevents that the heart stops pumping blood into the body if it does not contract spontaneously anymore. Because errors in Vitatron’s medical devices could potentially harm humans, testing of the firmware that controls these devices is extremely important. Vitatron spends a lot of effort in testing and verification of their firmware designs and implementations. Currently this is mainly done by reviewing and functional testing, but the coverage of these techniques is not optimal. Therefore it could be the case that certain errors are not discovered. Combining the current test methods with formal methods is perhaps a solution to test the firmware designs more extensively. But at this moment, Vitatron is unaware of the possibilities of formal methods in their environment. Formal methods are based on solid mathematical principles and increase understanding of systems, increase clarity of descriptions and help solve problems and remove errors. The use of formal methods for software and hardware designs is motivated by the expectation that performing appropriate mathematical analyses can contribute to the reliability and robustness of these designs. However, the high cost of using formal methods means that they are usually only used in the development of safety-critical systems. Model checking is one of these formal methods that can be used to verify whether a model of a system design satisfies its requirements. Models of hardware designs, software designs and also communication protocols can be verified by means of model checking. The technique is called model checking because it is used to check whether a system requirement holds in a formal model that represents the verified system design. The great advantage of design verification by using model checking is the completeness of the verification; after validation of a requirement on a formal model of the system design, it is absolutely sure that this model contains no behavior that violates the requirement. Of course such a validation is only useful if the verified model is a valid representation of the system design. To make model checking feasible, simplifications must be applied when a system design is translated to a formal model. Without these simplifications the model verification generally needs too much time and space resources. Because of this need for simplifications, it generally appears to be very hard to create a formal model that is a valid representation of the system design. Furthermore inventing the right requirements to the system design and correctly formulating them is often not straightforward. This makes clear that the application of formal model checking to Vitatron’s firmware designs is certainly not straightforward. Therefore this project investigates the feasibility of model checking in the Vitatron environment. For this investigation, the firmware design of Vitatron’s DA+ pacemaker is used as a case study. We investigate how formal models can be developed that represent (parts of) the firmware design, which requirements to the firmware design (parts) are suitable candidates to be formally verified, and whether the verification of these requirements on the formal models is feasible. In these investigations, we consider two different model checking tools, i.e. mCRL2 and Uppaal. Developing a formal model that represents a system design, means that the behavior and the responsibilities of all components in the system design and the communications between them are made explicit. As described above, the challenge in formally modeling a system design consists of inventing the right simplifications without letting the formal model deviate too much from the meant behavior in the system design. In our investigation of the 8
Page 10
development of formal models representing the pacemaker firmware design, we also address this point of inventing the right simplifications. The final goal of these investigations is providing recommendations to Vitatron, concerning the applicability of model checking as a minimum. To achieve this final goal, we try to verify requirements on (parts of) the firmware design by means of model checking in both mCRL2 and Uppaal. The report is divided into two parts. The first part consists of chapters 2, 3, and 4 and describes the background of the project. Chapter 2 introduces the model checking technique and the used model checking tools, chapter 3 describes some important aspects of the physiology of the human heart, and chapter 4 describes the architecture of the DA+ pacemaker and the design of the pacemaker firmware. The second part consists of chapter 5, 6, and 7 and describes the investigation that has been carried out during this project. Chapter 5 describes the translation of the pacemaker firmware design to formal models, and chapter 6 describes how model checking is applied on these formal models and which results we have obtained. Chapter 7 describes the recommendations to Vitatron about how they could apply model checking on their firmware designs in the future. Finally, chapter 8 presents our main conclusions and gives a summary of our recommendations from chapter 7. Note that the formal models that are produced in this project are not included as appendices to this report for confidentiality reasons. This report is accompanied by a CD on which all relevant formal models can be found. Furthermore, in this version of the report some parts of the text are removed or changed for confidentiality reasons. These changes all concern pacemaker details which are not crucial for understanding the results of the investigations.
9