A two-page specification sheet for the Wedgewood Model 622 conductivity monitor, which measures the conductivity and temperature of process liquids. It lists ranges, outputs, power, size and an order code table.
- Type
- Conductivity monitor for the conductivity and temperature of process liquids p. 1
- Ranges
- Scale A: 0-10µS/cm to 0-100mS/cm; Scale B: 0-20µS/cm to 0-200mS/cm full scale p. 1
- Accuracy
- +/- 2 % of full scale p. 1
- Outputs
- 4-20 mA (400 ohm load) and 0-2 Vdc (10k ohm load), for conductivity and temperature p. 1
- Power
- 115, 230VAC, 50-60Hz, 6VA; optional 24Vdc, 5W p. 1
- Operating environment
- 0-55 °C; 0-90 % relative humidity p. 1
- Inputs
- Dual channel resistive input; four-electrode conductivity sensor; Pt1000 RTD (alpha 0.00375) p. 1
- Temperature compensation
- Programmed curves for NaCl and KCl p. 1
- Range selection
- Two conductivity scale ranges, selectable remotely p. 1
- Size
- Type 4 DIN plug-in unit, 3U x 14HP x 160mm deep (2.8" W x 5.1" H x 6.7" D) p. 1
- Application
- Chromatography in the biopharmaceutical industry p. 1
What this sheet walks you through
Questions this sheet answers
What does the Wedgewood Model 622 measure?
The conductivity and temperature of process liquids. p. 1
What power does the Wedgewood Model 622 need?
115 or 230 VAC at 50-60 Hz, 6 VA, or optionally 24 Vdc at 5 W. p. 1
What outputs does the Model 622 conductivity monitor have?
4-20 mA into a 400 ohm load and 0-2 Vdc into a 10k ohm load, for both conductivity and temperature. p. 1
What temperature sensor does the Model 622 accept?
A Pt1000 RTD with an alpha constant of 0.00375. p. 1
The file
Size and text
Every page · 2
Read it here
Each page as it is printed, and its words beside it, so a phone can search and copy them.
Page1of 2
Open in the PDF
The text of page 1, from the PDF itself
Model 622 Conductivity Monitor The Model 622 conductivity monitor is used with either the Model B607 or the Model B605 pH/conductivity sensor to measure the conductivity and temperature of process liquids
Choice of two conductivity scale ranges with remote range selection capability Current and voltage outputs for conductivity and temperature Programmed temperature compensation curves for NaCl and KCl Four-electrode conductivity sensor input for enhanced measurement range Accepts Pt1000 RTD for improved accuracy (alpha constant = 0.00375) Ideally suited for chromatography applications in the biopharmaceutical industry Type 4 DIN plug in unit 3U x 14HP x 160mm deep (2.8” W x 5.1” H x 6.7” D) conserves space and simplifies mounting
Model 622 Signal inputs Range
Accuracy Signal outputs Power Operating environment
Specifications Dual channel resistive input for conductivity or temperature Scale A. 0-10µS/cm, 0-100µS/cm, 0-1mS/ cm, 0-10mS/cm, 0-100mS/cm Full Scale Scale B. 0-20µS/cm, 0-200µS/cm, 0-2mS/ cm, 0-20mS/cm, 0-200mS/cm Full Scale +/- 2 % of full scale 2, 4-20 mA, 400 ohm load for cond. and temp. 0-2 Vdc, 10k ohm load for cond. and temp. 115, 230VAC, 50-60Hz, 6VA optional 24Vdc, 5W Temperature 0-55 °C; relative humidity 0-90 %
Page2of 2
Open in the PDF
The text of page 2, from the PDF itself
Dimensions Diagram2.65 [67]2.79 [71]
Expanding The Science of Analysis
4.42 [112]5.06 [129]4.80 [122]75 6. 72] [1
Wedgewood Technology INCORPORTATED
2.19 [56]Order Code622 A B D
Power Supply 115 Vac 230 Vac 24 Vdc 6 7
Flowcell Kynar flowcell (BT721/BT724 sensors) Stainless steel flowcell (BT728 sensor) TU-
Transmitter Range 0 - 200 mS/cm 0 - 100 mS/cm -T BQ -N
Cable Type To BT721/724 inline cond./temp. sensor To B725 cond./temp. tank probe sensor None F N
Cable Length Feet None (BT728 sensor) Enter 3 digit cable code from choices below: 10 ft: 010 15 ft: 015 25 ft: 025 For no cable, enter ---
Wedgewood Analytical, Inc. 4123 East La Palma Avenue, Suite 200 Anaheim, CA 92807 Toll Free: 1-800-835-5474 Direct: 1-714-577-5600 Fax: 1-714-577-5688 www.WedgewoodAnalytical.com