Browse
(440) 299-7778
Menu
(440) 299-7778
One-year parts and labor warranty30-day money-back guaranteeContact

The text of a manual · part 2 of 8

870ITPH Intelligent Electrochemical Transmitter for pH, ORP, and Ion Selective Electrode Measurements, pages 5 to 12

Pages 5 to 12 of 62: Figures; Tables; 1. Introduction

The original document is the PDF (605 KB). This is its text as extracted, unedited; tables and drawings read best in the PDF, and each page heading opens it at that page.

For the Foxboro 870ITPH

Page 5

Figures 1 2 3 4 5 6 7 8 9

Wiring ......................................................................................................................... Supply Power vs. Load ................................................................................................. Data Label .................................................................................................................... Panel Mounting ........................................................................................................... Surface Mounting ........................................................................................................ Pipe Mounting ............................................................................................................. Wiring ......................................................................................................................... Indicators and Controls ................................................................................................ Example of Structure Diagram .....................................................................................

2 5 6 7 8 8 10 11 14 v

Page 6

Tables 1 2 3 4 5 6 7 8 9 10 11 12 vi

Product Safety Specifications ........................................................................................ Recommended Conduit and Fittings ........................................................................... Indicators and Controls ................................................................................................ Diagnostic Error Messages ........................................................................................... Status Parameters ......................................................................................................... Invensys Foxboro Sensor Data ..................................................................................... American Buffer Tables ................................................................................................ NIST Buffer Tables ...................................................................................................... European Buffer Tables ................................................................................................ List of Characters ......................................................................................................... Sensor Diagnostics ....................................................................................................... Millivolt Supply Formulas ............................................................................................

5 9 11 15 16 20 23 23 25 26 29 33

Page 7

1. Introduction The 870ITPH Intelligent Transmitter converts pH, ORP, or Ion Selective Electrode measurements to a 4 to 20 mA and/or digital output signal depending on the model selected and the configuration implemented. The analog only output version (Model 870ITPH-A) supports configuration, calibration, and diagnostic troubleshooting locally at the front panel. The transmitter has one 4 to 20 mA dc output which can be software configured (connected) to temperature compensated measurement, uncompensated (absolute) measurement, or temperature. The digital output version (Model 870ITPH-F) supports configuration, calibration, and diagnostic troubleshooting locally at the front panel, via a PC-based configurator anywhere on the signal leads, or from any I/A Series workstation via an Intelligent Transmitter Fieldbus Module. The digital output version supports either digital only communication or a hybrid 4 to 20 mA dc/digital communication as configured. When operating in either of these modes, the transmitter continuously outputs three digital measurement values: temperature compensated measurement, uncompensated (absolute) measurement, and temperature. In the 4 to 20 mA mode, you may assign the output to any one of the three measurement parameters.

Quick Start The purpose of this section is to: ♦

Help you to wire your transmitter

♦

Familiarize you with the instrument configuration as received from the factory

♦

Assist you in verifying that your transmitter is in calibration

♦

Explain normal operation in Measure mode.

Wiring Your 870IT is a 2-wire transmitter. As such, it transmits output over the same two wires through which it receives power. If your transmitter was specified with analog communications (870ITPH-A), you must supply its power source. If it was specified with digital communications (870ITPH-F), power can be provided via the I/A Series Fieldbus Module. Connect this power/output signal to the power terminals and the signals from your measurement sensor to the appropriate terminals near the top of the case as shown in Figure 1. ! WARNING Wiring installation must comply with any existing local regulations. ! WARNING A safety agency certified energy barrier must be used with this apparatus to meet agency certification requirements. See MI 611-206 (FM/CSA) or MI 611-208 (ATEX) for details.

1

Page 8

5 5A 6

7

8 G

NOT USED

DIAGNOSTIC

REFERENCE ELECTRODE

REF. ELEC. SCREEN (SHLD)

SOLUTION EARTH (GROUND)

MEAS. SCREEN (SHIELD)

2 2A 3 3A 4

SENSOR POWER, +

1

SENSOR POWER, -

4 - 20 mA CAL

+

RTD 3-WIRE

RTD MEASUREMENT RTD RETURN

–

MEASUREMENT ELECTRODE

1. Introduction

CONFIGURATOR

PC-BASED

MI 611-211 – October 2005

INPUT POWER

CASE EARTH (GROUND)

Figure 1. Wiring

Factory Default Passcode The factory default passcode for both minor and major passcodes is 0800.

Checking Factory Configuration Following the structure diagrams in Appendix A and using the arrow keys, you can step through the product structure as configured in the factory. The factory default values are listed in Appendix B. There is space provided to make notations in the last column of the table that comprises this appendix.

Verifying Valid Measurements Your transmitter was calibrated at the factory. Therefore, you should not have to calibrate it. However, if you wish to check your calibration, place your sensor in a solution of known value and, using the Down arrow key, check that the temperature on the second line of the display agrees with the actual temperature of the solution. Then check the value on the top line of your display. If the value does not agree with the known value of the solution, recalibrate your transmitter. The simplest method of doing this is to perform a single point manual calibration using the solution approach. See “Manual Calibration” on page 31.

Basic Operation in Measure Mode Measure is the normal operating mode of your transmitter. The pH, ORP, or ISE measurement is shown on the top line of the display. The second line of your display shows the absolute value of the sensor signal in millivolts, the output in 4 to 20 mA, or the temperature, depending on your

2

Page 9

1. Introduction

MI 611-211 – October 2005 configuration. You can view one of the other two selections by using the Up and Down arrow keys. The display will return to the configured selection when the configured TIMEOUT time is reached. Configuration errors and transmitter and sensor errors are also shown on your display while in the Measurement mode.

Looking for More Information? For more detailed information, refer to the following sections of this manual: ♦

For installation information, refer to “Installation” on page 7.

♦

For detailed explanation of the controls and indicators, refer to “Controls and Indicators” on page 11.

♦

For detailed configuration instructions, refer to “Configuration” on page 17.

♦

For detailed Calibration instructions, refer to “Calibration” on page 30.

For dimensional information, refer to DP 611-163. General information on this and other electrochemical products are available on line at www.foxboro.com (pull down Products menu and select Echem Analytical). If you need additional help, please call 1-866-746-6477 in the U.S.A. or contact your local Invensys Foxboro representative.

Reference Documents Document(a) DP 611-163 MI 020-495 MI 020-501 MI 020-504 MI 611-206 MI 611-208(b) PL 611-153

Document Description Dimensional Print - 870IT Intelligent Transmitters PC20 Intelligent Field Device Tool PC50 Intelligent Field Device Tool PC50 Intelligent Field Device Tool (FoxCom Communication) FM and CSA Intrinsic Safety Connection Diagrams and Nonincendive Circuits Electrochemical Products Safety Information (for European installations) Parts List - 870IT Style B Intelligent Transmitters

(a) Documents are available on line at www.foxboro.com (pull down Products menu and select Documentation). (b) Available in many languages on line at www.foxboro.com/us/eng/products/ instrumentation/downloads/ecanalyticalsafety/ecanalyticalsafety.htm.

3

Page 10

MI 611-211 – October 2005

1. Introduction

Specifications Sensors: Virtually any glass or antimony pH, ISE, or noble metal ORP sensor including the following: PH10 DolpHin™ Series pH Sensor ORP10 DolpHin™ Series ORP Sensor 871A pH/ORP Sensor (see Note) 871PH pH/ORP/ISE Sensor (see Note) EP459A Fluoride Sensor EP460 pH/ORP Sensor EP462 pH Sensor EP466 pH Sensor NOTE

871A-2, 871PH-1, and 871PH-2 Sensors, although compatible with 870ITPH Transmitters, do not supply full sensor diagnostic capability. Maximum Distance Between Sensor and Transmitter: With preamplified sensor: 152 m (500 ft) With non-preamplified sensor that utilizes a driven shield (PH10-.N, 871A-1, -3; 871PH-5, -6): 15.2 m (50 ft) Temperature Compensation Inputs: 100 ohm platinum RTD, 2- and 3-wire 1000 ohm platinum RTD, 2- and 3-wire Balco 3000 ohm RTD Ambient Temperature: -25 to +55° C Relative Humidity: 0 to 95% noncondensing Electromagnetic Compatibility (EMC): The Model 870ITPH Intelligent Transmitter complies with the requirements of the European EMC Directive 89/336/EEC when the sensor cable and I/O cables are enclosed in rigid metal conduit. (See Table 2.) Power Supply Requirements (for 4 to 20 mA output): 12.5 to 42 V dc (see Figure 2) Load Limitations (for 4 to 20 mA output): 0 to 1450 ohms (see Figure 2)

4

Page 11

1. Introduction

MI 611-211 – October 2005

1450 1400

TYPICAL SUPPLY VOLTAGE AND LOAD LIMITS

1400

OUTPUT LOAD, Ω

1300 1200

V dc

Load (ohms)

1100

24 30 32

250 and 565 250 and 860 250 and 960

1000 900

NOTE The transmitter will function with an output load less than 250 ohms provided that a PC-based configurator is not connected to it. Connecting a PC-based configurator while operating in this area may cause disturbances and/or communication problems.

800 700 600 500 OPERATING OPERATING AREA AREA

400 300 250 200 100 0

SEE NOTE 0

20 12.5 30 40 42 SUPPLY VOLTAGE, V dc

Figure 2. Supply Power vs. Load Product Safety: The 870IT transmitters have been designed to meet electrical safety descriptions listed in Table 1. For detailed information, or status of testing laboratory approvals/certifications, contact Invensys Foxboro. Table 1. Product Safety Specifications Testing Laboratory, Type of Protection, and Area Classification

Electrical Safety Design Application Conditions Code

ATEX intrinsically safe for II 1 G, EEx ia IIC hazardous Temperature Class T4 - T6 locations. ATEX Type n energy limited for II 3 G EEx nL IIC Temperature Class T4 - T6 hazardous locations CSA intrinsically safe for Class I, II, and III, Groups A, Connect per MI 611-206. B, C, D, E, F, and G, Division 1 hazardous locations. Temperature Class T4. CSA suitable for use in Class I, II, and III, Groups A, B, Temperature Class T4. C, D, F, and G, Division 2 hazardous locations. FM intrinsically safe for Class I, II, and III, Groups A, B, Connect per MI 611-206. C, D, E, F, and G, Division 1 hazardous locations. Temperature Class T4. FM nonincendive for Class I, II, and III, Groups A, B, Temperature Class T4. C, D, F, and G, Division 2 hazardous locations.

AAA ANN CAA CNZ FAA FNZ

5

Page 12

MI 611-211 – October 2005

1. Introduction

RoHS/WEEE Compliance Statement This product is exempt from the European Directive 2002/95/EC on the restriction of the use of certain hazardous substances in electrical and electronic equipment (RoHS), as provided by Article 2 of that Directive in conjunction with the Product Category #9: “Monitoring and Control Instruments.” This product complies with the European Directive 2002/96/EC on Waste Electrical and Electronic Equipment (WEEE) and is marked accordingly. At end of product life users should contact Invensys Foxboro for return authorization and shipment instructions.

Identification A data label is fastened to the left side surface of the enclosure. This label provides the model number and other information pertinent to your particular transmitter.

MODEL AND STYLE

870ITPH-FWFAA MODEL

ELECTRICAL CLASSIFICATION CODE SALES ORDER / SERIAL NUMBER (IF APPLICABLE)

CERT SPEC REF NO. ORIGIN SUPPLY POWER FUSE0 - 14 PH CALIB FoxCom CONFIG CD ALARM OUTPUT CUST DATA

PLANT OF MANUFACTURE, DATE, SERIAL NO. SUPPLY VOLTAGE POWER CONSUMPTION MEASUREMENT RANGE COMMUNICATIONS OUTPUT USER INFORMATION

FOXBORO MA USA

Figure 3. Data Label

6