870ITPH Intelligent Electrochemical Transmitter for pH, ORP, and Ion Selective Electrode Measurements: 2. Installation
Pages 13 to 16 of 62, from 870ITPH Intelligent Electrochemical Transmitter for pH, ORP, and Ion Selective Electrode Measurements
The original document is the PDF (605 KB). This is its text as extracted, unedited; tables and drawings read best in the PDF, and each page heading opens it at that page.
For the Foxboro 870ITPH
Page 13
2. Installation Unpacking 1. Remove the transmitter from the shipping container and check for visible damage. 2. Save the container until you determine that no shipping damage has occurred. a. If no damage is observed, proceed to the Mounting procedure. b. If transmitter has been damaged, notify the carrier immediately and request an inspection report. Obtain a signed copy of the report from the carrier and contact the Customer Satisfaction Center at 1-866-746-6477 in the U.S.A., or contact your local Invensys Foxboro representative.
Mounting Panel Mounting Panel mounting of the transmitter is shown in Figure 4. For panel cutout size, space requirements, and other dimensional data, see DP 611-163.
PANEL TRANSMITTER BRACKET
NUT WASHER
Figure 4. Panel Mounting
Surface Mounting Surface mounting of the transmitter is shown in Figure 5. Fasten mounting bracket to transmitter using screws, washers, and lockwashers provided. Fasten mounting plate to surface using appropriate hardware. For space requirements and other dimensional data, see DP 611-163.
7
Page 14
MI 611-211 – October 2005
2. Installation
LOCKWASHER TRANSMITTER BRACKET
SURFACE
CAP SCREW
WASHER
Figure 5. Surface Mounting
Pipe Mounting Pipe mounting of the transmitter is shown in Figure 6. Securely fasten mounting bracket to a DN50 or 2-in pipe as shown. For space requirements and other dimensional data, see DP 611-163. BRACKET
LOCKWASHER PIPE WASHER CAP SCREW
TRANSMITTER NUT
U-BOLT
LOCKWASHER
Figure 6. Pipe Mounting
Wiring Your 870IT is a 2-wire transmitter. As such, it transmits output over the same two wires through which it receives power. If your transmitter was specified with analog communications (870ITPH-A), you must supply its power source. If it was specified with digital communications (870ITPH-F), power can be provided via the I/A Series Fieldbus Module. Connect this power/output signal and the signal from your measurement sensor as follows: 1. Open the hinged case door. 2. Remove the plastic shipping caps from the openings on the bottom of the case and replace with proper fittings. You will be running wires through the left and right openings and capping the middle opening.
8
Page 15
2. Installation
MI 611-211 – October 2005
! WARNING Wiring installation must comply with any existing local regulations. ! WARNING A safety agency certified energy barrier must be used with this apparatus to meet agency certification requirements. See MI 611-206 (FM/CSA) or MI 611-208 (ATEX) for details. ! CAUTION 1. Failure to adequately seal open ports or secure the cover will jeopardize NEMA 4X, CSA Enclosure 4X, or IEC IP65 and will void instrument warranty. 2. To maintain a rating of NEMA 4X, CSA Enclosure 4X, or IEC IP65, wiring methods, conduit, and fittings appropriate to the ratings must be used. Table 2 identifies recommended parts. ! CAUTION To mimimize RFI/EMI effect, use grounded metal conduit on sensor cable and input power leads.
Table 2. Recommended Conduit and Fittings Material
Conduit
Rigid Metal Semi-Rigid Plastic
1/2 inch Electrical Trade Size T&B #LTC 050
Semi-Rigid Plastic Metal Core Flexible Plastic
Anaconda Type HV, 1/2 inch T&B #EFC 050
Fitting T&B* #370 T&B #LT 50P or T&B #5362 T&B #LT 50P or T&B #5362 T&B #LT 50P or T&B #5362
*T&B = Thomas & Betts Corp., 1001 Frontier Road, Bridgewater, NJ 08807
3. Bring the wires from your 2-wire input power in the left opening and connect to the input power terminals shown in Figure 7. Bring the sensor cable in the right opening and connect by number to the appropriate terminals near the top of the case per Figure 7. 4. For non-Foxboro sensors with coax cable, connect RTD to terminals 1 and 2, measurement electrode to terminal 3, solution ground to terminal 4, and reference electrode to terminal 5. Solution ground must be wired to a metallic material in contact with the process solution. If not, jumper terminals 4 and 5. 5. For non-Foxboro sensors with triax cable, connect RTD to terminals 1 and 2, measurement electrode to terminal 3, measurement screen (shield) to terminal 3A, solution ground to terminal 4, and reference electrode to terminal 5. Solution ground
9
Page 16
MI 611-211 – October 2005
2. Installation must be wired to a metallic material in contact with the process solution. If not, jumper terminals 4 and 5. 6. If you are not wiring a temperature measuring device, short terminals 1, 2, and 2A to minimize noise. POWER IFD MODEM
SUPPLY (24 V dc) + –
RS-232
IBM COMPATIBLE
OPTIONAL
CABLE
PC
PRINTER
TRANSMITTER
REFERENCE ELECTRODE
7
8 G
CASE EARTH (GROUND)
Figure 7. Wiring
10
NOT USED
5 5A 6
INPUT POWER
NOTES: 1. IF NOT WIRING A TEMPERATURE MEASURING DEVICE, SHORT TERMINALS 1, 2, AND 2A. 2. TERMINAL 2A IS NOT USED WHEN YOUR SENSOR CONTAINS A 2-WIRE TEMPERATURE ELEMENT.
REF. ELEC. SCREEN (SHLD)
DIAGNOSTIC
2 2A 3 3A 4
SENSOR POWER, +
1
SENSOR POWER, -
4 - 20 mA CAL
+
SOLUTION EARTH (GROUND)
–
MEAS. SCREEN (SHIELD)
CONFIGURATOR
RTD 3-WIRE
PC-BASED
RTD MEASUREMENT RTD RETURN
250Ω
MEASUREMENT ELECTRODE
CABLE