3. Operation Controls and Indicators Operator controls and indicators are located on the front panel. Figure 8 shows the panel arrangement. Table 3 identifies the function of each element. PRIMARY DISPLAY SECONDARY DISPLAY
WHEN CAL, CONFIG, STATUS, OR DIAG
20-SEGMENT BARGRAPH FOR OUTPUT % OF SPAN WHEN IN MEASURE MODE
KEY IS PRESSED, TWO SEGMENTS ONLY APPEAR UNDER THE APPLICABLE LEGEND.
7.00 pH 25.0 C
CONFIGURATION KEY
CALIBRATION KEY
ENTER KEY
MEASURE KEY
HOLD KEY
UP, DOWN, LEFT, AND RIGHT ARROW KEYS
DIAGNOSTIC KEY STATUS KEY
Figure 8. Indicators and Controls Table 3. Indicators and Controls Control/Indicator Digital Bargraph Primary Display Secondary Display
CALibration Key
CONFIGuration Key
HOLD Key
MEASUREment Key
Function Displays percent-of-range indication. Displays measurement and some error messages. Displays the absolute value of the measurement signal in millivolts, the output in 4 to 20 mA, or the temperature, and error messages and configuration/calibration prompts. Places transmitter in Calibration mode. This enables you to calibrate the transmitter automatically to your buffer solutions or manually to values entered. Places transmitter in Configuration mode. This enables you to configure the value or status of each parameter required for your application. Places transmitter in Hold mode. This enables you to hold the output at a determined value while configuring or calibrating the transmitter or performing sensor maintenance. Places transmitter in Measurement (normal operation) mode.
11
Page 18
MI 611-211 – October 2005
3. Operation
Table 3. Indicators and Controls (Continued) Control/Indicator
Function
STATUS Key
Places transmitter in Status mode. This enables you to view the measurement and system parameters and thus assess the performance of the loop. DIAGnostics Key Places transmitter in Diagnostic mode. This enables you to further identify specific errors and possible solutions to an error flagged in measurement mode. Direction (Arrow) Keys Move you through the structure diagram and subsequent menus and values. They may be used to increment one position at a time, or to move through a menu or string of values by holding a key in the depressed position. ENTER Key Enters the menu item or value selected.
Error Messages The 870IT will display one of several messages when a problem occurs. These are summarized below: Flashing Display: Occurs when there are not enough significant digits on the display to display a number. For example, if the PRIMARY SCALE was configured as 99.99 and a measurement of 100 was received, the display would flash +99.99. CAL ERROR: Occurs when a limit condition was exceeded during an attempted calibration. CAL appears on the top line and ERROR on the bottom line of the display. XMTR ER: Occurs in Measure mode only to indicate a transmitter problem. The transmitter is still configurable but measurements are held at the failsafe value if set. PROBE ER: Occurs in Measure mode only when one of the following probe errors occurs: Glass, Lowslope, Coat, Leakage, Preamp, Aging, ATC Short, and ATC Open. For more information, see “Diagnostics” on page 15. Message appears on the bottom line of the display and flashes alternately with the bottom line measurement. The transmitter is still configurable. For probe errors of ATC Short or ATC Open, measurements are held at the temperature default value if set. CONFIG ER: Occurs in Measure mode only when one of the following configuration errors occurs: Calibration Required, 4-20 Range, Measurement Range, and Compensation Range. For more information, see “Diagnostics” on page 15 Message appears on the bottom line of the display and flashes alternately with the bottom line measurement. ER DATA: Occurs after you change a value and try to enter it with the ENTER key. Message appears on the bottom line of the display for 1 to 2 seconds and then automatically returns to the last legal value. The top line does not change from the previous edit prompt. Situations causing this message and appropriate responses are: 1. New value is out of range. 2. The slope of a chemical composition curve has changed sign. 3. The first value of a compensation point pair (TEMPx in TEMPCOMP or ABSx in CHEMCOMP) was entered that was not greater than the last one.
12
Page 19
3. Operation
MI 611-211 – October 2005
Passcode Access Two levels of passcode protection are available. The Minor level permits access to Calibration and Output Hold. The Major level permits changing parameters in Configuration as well as access to Calibration and Output Hold. Both are 4-digit passcodes. The passcodes may or may not be the same. They are set as part of configuring the instrument. Upon entering the Configuration, Calibration, or Hold mode by pressing the appropriate key, the display will read PASSCODE 0000. Upon entering the proper passcode and then pressing the ENTER key, the display will change to the next item in the appropriate structure. The initial factory default value for both Minor and Major passcodes is 0800.
Viewing and Changing Data In the Configuration, Calibration, Hold, and Status modes, you will need to move through a structure of parameters (using the arrow keys and ENTER key) to view and/or change the value or status of a particular parameter. Structure diagrams, located in Appendix A, will aid you in doing this. Note that the Up and Down arrow keys will scroll through a menu in either direction. Also, that the arrow keys may be used to increment one position at a time, or to move continuously through a menu or string of values by holding the key in the depressed position. The following example will show you how to use the keys in moving through the structure. The example used is that of configuring temperature features. For the example, assume Fahrenheit temperature units, an automatic temperature mode, a default value of 77 ° F, and a 3-wire, 1000 ohm RTD. In following the procedure, refer to Figure 9. 1. Use the Right arrow key to move from CONFIG TEMP to TEMP UNITS and then to UNITS CELSIUS. 2. Use the Down arrow key to move to UNITS FAHREN and press the ENTER key. This will enter your Fahrenheit selection and move the display to TEMP MODE. Every temperature entry in the transmitter will now be in Fahrenheit units. 3. Use the Right arrow key to move to MODE AUTO. Press the ENTER key. This will enter your Auto selection and move the display to DEFAULT 000.0. The tenths digit will be flashing. 4. Because you do not need to change the tenths digit, use the Left arrow key to move to the units digit. Use the Up arrow key to change this digit to 7. 5. Use the Left arrow key to move to the tens digit and the Up arrow key to change this digit to 7. You have now achieved your goal of setting the value to 77.0, so press the ENTER key. This will enter your value of 77 and move the display to TEMP TYPE. 6. Use the Right arrow key to move to TYPE RTD and press the ENTER key. This will enter your selection of RTD and move the display to RTD 2 WIRE. 7. Use the Down arrow key to move to RTD 3 WIRE and press the ENTER key. This will enter your selection of 3 Wire and move the display to 3 WIRE 100 OHM. 8. Use the Down arrow key to move to 1000 OHM and press the ENTER key. This will enter your selection of 1000 ohm and move the display back to CONFIG TEMP. This completes the example.
13
Page 20
MI 611-211 – October 2005
CONFIG TEMP
3. Operation
TEMP UNITS
UNITS CELSIUS
UNITS FAHREN
TEMP MODE
TEMP TYPE
MODE AUTO
DEFAULT
MODE MANUAL
MANUAL
TYPE RTD
RTD 2 WIRE
2 WIRE 100 OHM
2 WIRE 1000 OHM
RTD 3 WIRE
3 WIRE 100 OHM
3 WIRE 1000 OHM TYPE BALCO
Figure 9. Example of Structure Diagram
Measure Measure is the normal operating mode of your transmitter. The pH, ORP, or ISE measurement is shown on the top line of the display. Upon returning to Measure mode after changing many configuration parameters, the display may read CONFIG ER. This typically means that a calibration is required. The second line of your display shows the absolute value of the sensor signal in millivolts, the output in 4 to 20 mA, or the temperature, depending on your configuration. You can view one of the other two selections by using the arrow keys. The display will return to the configured selection when the TIMEOUT time is reached. Transmitter (XMTR ER) and sensor (PROBE ER) errors are also shown on your display while in the Measurement mode.
14
Page 21
3. Operation
MI 611-211 – October 2005
Diagnostics (Refer to “Structure Diagram 11” on page 48.) The Diagnostics mode enables you to perform checks on your system at any point in time. While you are in the Diagnostic mode, two segments of the bar graph will appear under the legend DIAG at the top of the front panel. To enter the Diagnostics mode, press the DIAG key on the front panel. If there are no errors, your display will read DIAGS OK. If there was an error flagged as XMTR ER, PROBE ER, or CONFIG ER in the Measurement mode, the Diagnostics mode will help you further identify the specific error and corrective action. Possible errors and the recommended corrective actions are listed in Table 4. Table 4. Diagnostic Error Messages Message
Description
Corrective Action
Transmitter Error Probe Errors GLASS Resistance of glass electrode is checked versus solution Replace electrode ground. Error is flagged if resistance is less than user set limit. LOWSLOPE Flags an error after a 2-point calibration in buffers Replace electrode or when the Nernst slope of the sensor becomes less than solution the user set limit. COAT Checks reference junction resistance against solution Clean electrode ground. Error is flagged if resistance is greater than user set limit. Sensor is not immersed in solution. Check that solution ground and reference junction are immersed in solution. LEAKAGE Checks solution leakage into the probe by checking Replace resistance between solution ground and the RTD. Error is flagged if resistance is anything but infinite. PREAMP Checks preamplifier by monitoring voltage coming out Replace of it. Error is flagged at voltage greater than 2 volts. AGING Flags an error after a 2-point calibration in pH buffers See Note below when the Nernst slope of the sensor has decreased sequentially five times. ATCSHORT Flags an error if the resistance of the temperature Replace temperature ATC OPEN compensator is greater or less than the expected compensator resistance of the device configured.
15
Page 22
MI 611-211 – October 2005
3. Operation
Table 4. Diagnostic Error Messages (Continued) Message
Description
Corrective Action
Configuration Errors CAL REQ Calibration required 4 - 20 RNG Measurement outside configured range MEAS RNG Measurement outside configured precision or scale COMP RNG Measurement outside compensation curve
Calibrate Reconfigure range or correct process error Reconfigure precision or correct process error Reconfigure compensation or correct process error
NOTE
Corrective action for AGING error is to check the value of the slope in the Status menu. The glass or ISE electrode may be reaching the end of its useful life if the SLOPE differs greatly from the theoretical value (80% is usual limit but application requirements should dictate lowest acceptable slope).
Status (Refer to “Structure Diagram 8” on page 45.) The Status mode enables you to view the measurement and system parameters and thus assess the performance of the loop. You may not edit them in this mode. For a string display of more than eight characters, use the Left and Right arrow keys to scroll the display. All measurements (outputs) continue while in the Status mode. To enter the Status mode, press the STATUS key. Move through the menu by using the Up and Down arrow keys. While you are in the Status mode, two segments of the bar graph will appear under the legend STATUS at the top of your front panel. The Status menu contains the following parameters. Table 5. Status Parameters Parameter TEMP SLOPE ABSOLUTE mA GLASSRES ASYMM REF RES ATC RES CAL NAME LAST CAL 16
Description Process Temperature Efficiency of sensor expressed as the deviation in percent of the slope of the most recent calibration compared to the standard (e.g., mV/pH/59.16 mV/pH) Uncompensated millivolt signal from sensor (a) Analog Output: Analog output value in mA Digital Output: Output shown on secondary display in mA. Indicates whether resistance of glass electrode is acceptable Asymmetry potential (mV difference from the expected theoretical isopotential pH voltage of zero) Resistance of reference electrode Resistance of temperature compensator Name of last calibrator Date of last calibration (mm/dd/yy)
Page 23
3. Operation
MI 611-211 – October 2005
Table 5. Status Parameters (Continued) Parameter TAG NUM TAG NAME LOCATION SONUMBER MS CODE S/W VER DEV NAME DEV TYPE
Description Tag number Tag name* Location of measurement (b) Sales order number Model code Software version Device name* Device type*
(a)See Table 12 for expected absolute mV readings. Absolute mV is a very useful sensor troubleshooting tool. (b)Appears only on devices coded 870ITPH-F.
Configuration Configuration is the process of enabling functional capability in the transmitter firmware for your application. This section will enable you to systematically determine and configure the value or status of each parameter required for your application. Many applications require only changing a few of the default values already entered. While you are in the Configuration mode, two segments of the bargraph will appear under the legend CONFIG at the top of your front panel. These segments will flash if you are also in a Hold mode. To enter the Configuration mode, press the CONFIG key. At the prompt, enter the four character Major passcode and press ENTER. If the passcode is correct, the display will flash UNLOCKED and then change to the first item in the configuration structure, CONFIG DISPLAY. You may then edit any of the parameters in the Configuration mode. If the passcode is incorrect, the display will flash LOCKED and then change to the first item in the configuration structure. You may view the parameters in this mode but will not have the ability to change them. In the UNLOCKED condition, the analog output defaults to 3.8 mA. If you do not want this to happen, turn the HOLD feature on first. ! CAUTION Always exit the Configuration mode by pressing the Measure key before power is turned off. This allows proper reinitialization of internal data base security checks.
Config Display (Refer to “Structure Diagram 1” on page 38.) This section of the structure enables you under DISPLAY PRIMARY to specify the PRIMARY SENSOR as pH, ISE, or ORP. This choice configures your analyzer as a pH, ORP or ion selective electrode transmitter. Only parameters pertinent to the specific measurement technology will appear in the configuration software after this initial choice is made. If pH, you can further specify the PRIMARY RESOLUTION as 0.1 or 0.01 pH. If ISE, you can further specify the PRIMARY SCALE as 9.999, 99.99, 999.9, or 9999 ppm. This choice specifies the largest value that may be displayed as well as the decimal placement in ppm.
17
Page 24
MI 611-211 – October 2005
3. Operation
NOTE
A change in the Primary Sensor, Resolution, or Scale requires a calibration. Next, under DISPLAY SECOND, you can configure the secondary display to show the absolute value of the measurement in millivolts, the output in 4 to 20 mA, or the temperature.
Config Output (Refer to “Structure Diagram 1” on page 38.) This section of the structure first asks you to specify the MODE, referring to the communications and measurement output. If your transmitter was specified with analog output only (870ITPHA), specify ANALOG. If your transmitter was specified with I/A Series communications (870ITPH-F), specify ANALOG if you want to have a 600 baud digital signal superimposed over an analog 4 to 20 mA signal, or specify DIGITAL if you want to have a 4800 baud digital signal only. If you configured the mode as ANALOG, you will be asked, under OUTPUT ANALOG, to specify the 4 to 20 mA output to represent the uncompensated (ABSOLUTE) measurement, the temperature compensated MEASUREment, or the TEMPerature. If you configured the mode as DIGITAL, the 4 to 20 mA output will be fixed at 12 mA. Next, under OUTPUT RANGE, enter the MINimum (4 mA) and MAXimum (20 mA) range values for the output just specified. Specify between +2000 and -2000 mV for ABSOLUTE, in ° C or ° F for TEMPerature, or in measurement units (pH for pH, mV for ORP, ppm for ISE) for MEASURE. To configure the output as reverse acting, enter higher value as MINimum and the smaller as MAXimum. NOTE
1. 1. The units for temperature, ° C or ° F, are specified in the section titled “Config Temp” on page 20. 2. 2. A change in OUTPUT ANALOG will cause the OUTPUT RANGE to default to 0.0 and thus new MIN and MAX values must be entered. Lastly, you can specify a fail-safe signal for all transmitter errors and for critical sensor diagnostic errors. Specify ON to specify a fixed output when in a failed condition. If ON is selected, further specify the DC mA output between 3.8 and 20.5 mA. The default value is 3.8 mA. If you want to have this signal outputted with ±0.5 mA pulse (as shown below) for greater visibility on a chart recorder or data display, specify ON as explained above and then go back and specify PULSE.
The OFF selection is not functional in the current versions of firmware. Any sensor diagnostic that has been set to ENABLE (see “Config Diags” on page 26) causes the output to go to the failsafe DC mA signal. If sensor diagnostics are enabled, it is important to set the fail-safe DC mA value to an acceptable safe output level. If you do not want the output to respond to some or all of the sensor diagnostics, then you must set the selected diagnostics to DISABLE in the Config Diags menu.