Page 25
3. Operation
MI 611-211 – October 2005
Config Electrod (Refer to “Structure Diagram 2” on page 39.) NOTE
This parameter will not appear if ORP was selected as PRIMARY DISPLAY. This section of the structure asks you to specify the electrode: ♦
If measuring pH: Your choices are GLASS, ANTIMONY, or OTHER. GLASS sets the isopotential point at 7 pH. ANTIMONY sets it at 1 pH. OTHER allows the isopotential point to be set manually.
♦
If measuring ISE: You must specify the ion polarity. A positive ion selective measurement gets more positive (mV-wise) as the concentration increases (e.g., Na+, K+). A negative ion selective measurement gets more negative (mV-wise) as the concentration increases (e.g., F–, S–2). For fluoride measurements, this parameter should be set to negative. NOTE
A change in the electrode requires a calibration.
Config Tempcomp (Refer to “Structure Diagram 2” on page 39.) NOTE
This parameter will not appear if ORP was selected as PRIMARY DISPLAY. This section of the structure asks you to specify the temperature compensation if you are measuring pH or ISE. If measuring pH, specify STANDARD, AMMONIA, or CUSTOM. STANDARD assumes a glass or antimony electrode and applies the Nernst equation. The nominal temperature compensation range is -5 to +105 ° C (23 to 221 ° F). AMMONIA compensates for traces of ammonia in addition to the standard compensation. CUSTOM allows you to specify particular solution characteristics into the transmitter’s memory. When using this compensation, all measurements will be corrected back to a reference temperature measurement value. Therefore, you are asked to specify a reference temperature and a number of points to plot a compensation curve. The reference temperature (REF TEMP) should be within the nominal temperature compensation range. The number of points (NUM PTS) can be specified from 2 to 21 points and has a default value of 2. Each point specifies a value at a certain temperature. The temperatures (TEMPx) should be within the nominal temperature compensation range. The temperature units are specified in the next section. The values (VALx) can be specified from -2.00 to +16.00 pH and have a default of 0.00. If ISE was chosen as your PRIMARY DISPLAY, you must specify STANDARD or CUSTOM. CUSTOM causes you to specify a reference temperature and a number of points to plot a compensation curve. The number of points (NUM PTS) can be specified from 2 to 21 points and has a default of 2. Each point specifies a value at a certain temperature. The temperature units are specified in the next section. The values (VALx) can be specified from 0 to 2000 ppm and have a default of 0.
19
Page 26
MI 611-211 – October 2005
3. Operation
For CUSTOM compensation, the temperature values must be entered in increasing order and the slope of the curve cannot change direction. Deviation from this requirement will cause an ER DATA message and force a resequence through the point pairs.
Config Temp (Refer to “Structure Diagram 3” on page 40.) This section of the structure asks you to specify the temperature features. First, specify the TEMP UNITS as CELSIUS or FAHRENheit. All references to temperature will appear in the specified units within the transmitter software. Next, specify the TEMP MODE as AUTO or MANUAL. In the AUTO mode, the temperature follows the RTD. In MANUAL mode it is set at a fixed value. At this point you can set a DEFAULT value for the AUTO mode (in case the RTD fails) at the temperature at which you expect the process to be operating; or you can set a temperature for the MANUAL mode. Last, specify the TEMP TYPE as standard RTD or BALCO. If you specify the TYPE as RTD, specify 2-WIRE or 3-WIRE and specify the resistance as 100 OHM or 1000 OHM. Follow Table 6 for Invensys Foxboro sensors. Table 6. Invensys Foxboro Sensor Data Sensor PH10-..1 ORP10-..1 PH10-..2 ORP10-..2 PH10-..3 ORP10-..3 PH10-..4 ORP10-..4 871PH-1 871PH-2 871PH-3 871PH-4 871PH-5 871PH-6 871A-1 871A-2 871A-3 871A-4 EP459A EP460 EP462 EP466
20
Type 2-wire
Resistance 100 Ω
3-wire
1000 Ω
2-wire
100 Ω
3-wire
1000 Ω
2-wire 2-wire 3-wire 3-wire 2-wire 2-wire 2-wire 2-wire 2-wire 3-wire 2-wire 3-wire 3-wire 3-wire
100 Ω 100 Ω 1000 Ω 1000 Ω 1000 Ω 1000 Ω 100 Ω 100 Ω 1000 Ω 1000 Ω 100 Ω 100 Ω 100 Ω 100 Ω
Page 27
3. Operation
MI 611-211 – October 2005
Config Damping (Refer to “Structure Diagram 3” on page 40.) This section of the structure asks you to specify the damping response time (90% of step). Choices are 1, 5, 10, 20, 40, and 120 seconds. Damping applies only to measurement data, not to temperature data.
Config Chemcomp (Refer to “Structure Diagram 3” on page 40.) If measuring ISE, this section of the structure asks you to specify the chemical compensation (CHEMCOMP) as STANDARD or CUSTOM. This is similar to what you did in “Config Tempcomp” on page 19. CUSTOM causes you to specify a number of points (NUM PTS) to plot a compensation curve. The number of points can be specified from 2 to 21 and has a default of 2 points. Each point specifies an absolute value (ABSx) and the value you want it to be (NEWx). The values can be specified from 0 to 2000 ppm and have a default of 0. The decimal point location for these values was determined as PRIMARY SCALE in CONFIG DISPLAY. The slope of the curve cannot change direction or be zero between any two points. Deviation from this requirement will cause an ER DATA message.
Config Instrmnt (Refer to “Structure Diagram 4” on page 41.) NOTE
This section is not displayed unless the correct passcode had been entered for Configuration. INSTRMNT mA CAL
This subsection allows you to tune the 4 mA and 20 mA values. The procedure is as follows: 1. Connect a digital voltmeter to the to the calibration terminals (see Figure 7). 2. With INSTRMNT mA CAL displayed, press ENTER. The display will change to mA CAL 4 mA. Press ENTER again. 3. Calibration is performed by having your voltmeter read 100 mV. Do this by selecting one of the seven icons from the second line of your display by using the Left and Right arrow keys. When you have highlighted the icon you want, press ENTER. The icon meanings (shown below) are (from left to right) large increase, medium increase, small increase, accept and exit, small decrease, medium decrease, large decrease.
4. Repeat Step 3 until your meter reads 100 mV. Then highlight the center icon and press ENTER. The display will go to mA CAL 20 mA. Press ENTER. 5. Repeat the procedure shown in Steps 3 and 4 for 20 mA. The voltage reading should be 500 mV.
21
Page 28
MI 611-211 – October 2005
3. Operation
INSTRMNT STAB IND
When performing a calibration, the transmitter checks for stability in absolute measurement and temperature before accepting a change. This section enables you to configure the TIME allowed to attain stability and the amount of fluctuation (VAR) allowed. The value of VAR corresponds to the least significant digit of the measurement in the PRIMARY DISPLAY you configured in “Config Display” on page 17. For example, if you configured the PRIMARY RESOLUTION as 0.01 pH, then the value of VAR would be from 0.01 to 0.09 pH expressed as from 1 to 9. Both TIME and VAR are configured individually in STAB IND MEASurement and STAB IND TEMPerature. First, configure STAB IND MEAS as ON or OFF. If you configure this as ON, you need to specify the TIME between 5 and 60 seconds (in 5-second increments) and the VARiable between 1 and 9. A longer time period and a smaller measurement value assures more stability during calibration. The default TIME is 5 seconds and the default VARiable is 9. Thus, if the primary resolution was 0.01 and the default time and variable are used, the values can change at a rate of 0.09 pH over 5 seconds. Then repeat the procedure for STAB IND TEMP. If you configure this as ON, you need to specify the TIME between 5 and 60 seconds (in 5-second increments) and the VARiable between 1 and 9 ( ° C or ° F). The default TIME is 5 seconds and the default VARiable is 9.
Config Timeout (Refer to “Structure Diagram 4” on page 41.) This section of the structure asks you to specify the time in which the instrument will bring you back to the on-line/Measurement mode from another on-line mode (Status, Diagnostics, or the view only part of Configuration) when no keyboard input has occurred. In Measure mode, TIMEOUT also limits the time to view a secondary display other than that configured in CONFIG DISPLAY. The time may be specified between 0 and 999 seconds. A value of zero (0) disables the time-out function. The default time is 600 (10 minutes).
Config Buffers (Refer to “Structure Diagram 5” on page 42.) NOTE
This parameter will not appear if ISE or ORP was selected as PRIMARY DISPLAY. This section of the structure asks you to specify the buffers to be used if you are measuring pH and will use the Automatic Calibration feature described on page 31. In Automatic Calibration, the pH value calculated with values from the last calibration is used to decide which buffer is in use. The algorithm checks each buffer starting with Buffer 1 and selects the first one for which this pH is within 0.5 pH of the average pH for the buffer. You may select between AMERICAN, NIST, EUROPEAN, and SPECIAL buffers as standard buffer table choices. If you select SPECIAL, you must specify the number of points (NUM PTS) between 02 and 11 for Buffer 1, 2, and 3. The default value is 02. Each point specifies a value at a certain temperature. The temperatures (TEMPx) can be specified from -99.9 to +999.9 ( ° C or ° F) and
22
Page 29
3. Operation
MI 611-211 – October 2005 have a default of 0.0. The values (VALx) can be specified from -2.00 to +16.00 pH and have a default of 0.00. If the average pH values of two special buffers are less than 1 pH unit apart, using this procedure might result in the wrong buffer being selected. In this case, the buffer selection is based on the midpoint between the two buffers. It is not necessary to configure all three special buffers. If only two are used, they should be Buffers 1 and 2. If one is used, it should be Buffer 1. Enter Special buffer tables in the increasing order of the average pH for the buffer. The average pH is estimated as (first entry + last entry) / 2. Tables of buffer values for nominal 4, 7, and 10 pH for American, NIST, and European buffers are listed below: Table 7. American Buffer Tables Temperature °C
4.00 pH
7.00 pH
10.01 pH pH Value pH Value pH Value
0 5 10 15 20 25 30 35 40 45 50
4.00 4.00 4.00 4.00 4.00 4.00 4.01 4.02 4.03 4.04 4.06
7.12 7.09 7.06 7.04 7.02 7.00 6.99 6.99 6.98 6.98 6.97
10.32 10.25 10.18 10.12 10.06 10.01 9.97 9.93 9.89 9.86 9.83
NOTE
Values in Table 7 are based on common (nominal value) technical buffer systems. pH values are at 25°C. Chemical formulas are as follows: 4.00 pH: 0.05 molar Potassium Hydrogen Phthalate 7.00 pH: 0.041 molar Disodium Hydrogen Phosphate; 0.026 molar Potassium Dihydrogen Phosphate 10.01 pH: 0.025 molar Sodium Carbonate; 0.025 molar Sodium Bicarbonate
Table 8. NIST Buffer Tables Temperature °C
4.01 pH
6.86 pH
9.18 pH pH Value pH Value pH Value
0 5 10
4.00 4.00 4.00
6.98 6.95 6.92
9.46 9.40 9.33 23
Page 30
MI 611-211 – October 2005
3. Operation
Table 8. NIST Buffer Tables (Continued) Temperature °C
4.01 pH
6.86 pH
9.18 pH pH Value pH Value pH Value
15 20 25 30 35 40 45 50
4.00 4.00 4.01 4.02 4.03 4.04 4.05 4.06
6.90 6.88 6.86 6.85 6.84 6.84 6.83 6.83
9.28 9.23 9.18 9.14 9.10 9.07 9.04 9.01
NOTE
Values in Table 8 are based on NIST Primary Buffer Systems. pH values are at 25°C. Chemical formulas are as follows: 4.01 pH: 0.05 molar Potassium Hydrogen Phthalate 6.86 pH: 0.025 molar Potassium Dihydrogen Phosphate; 0.025 molar Disodium Hydrogen Phosphate 9.18 pH: 0.01 molar Sodium Tetraborate Decahydrate (Borax)
24
Page 31
3. Operation
MI 611-211 – October 2005
Table 9. European Buffer Tables Temperature °C 0 5 10 15 20 25 30 35 40 45 50 55 60 70 80 90 95
4.61 pH
7.00 pH
9.21 pH pH Value pH Value pH Value
7.12 7.09 7.06 7.04 7.02 7.00 6.99 6.98 6.97 6.97 6.97 6.97 6.98 7.00 7.04 7.09 7.12
9.52 9.45 9.38 9.32 9.26 9.21 9.16 9.11 9.07 9.03 8.99 8.96 8.93 8.88 8.83 8.79 8.77
4.64 4.62 4.61 4.61 4.61 4.62 4.63 4.64 4.66 4.67 4.69 4.71
NOTE
Values in Table 9 are based on commercially available technical buffers used in some parts of Europe. pH values are at 25°C. Chemical formulas are as follows: 4.61 pH: 0.1 molar Sodium Acetate; 0.1 molar Acetic Acid 7.00 pH: 0.026 molar Potassium Dihydrogen Phosphate; 0.041 molar Disodium Hydrogen Phosphate 9.21 pH: 0.05 molar Sodium Tetraborate Decahydrate (Borax)
Config Passcode (Refer to “Structure Diagram 5” on page 42.) NOTE
This parameter is not displayed if the transmitter is locked (i.e., the correct passcode was not entered for Configuration). This section of the structure asks you to specify the Minor and Major Passcode. The Minor passcode limits access to routine maintenance functions including CALIBRATION and output HOLD. The Major passcode prohibits unauthorized personnel from changing the configuration parameters. Each is a 4-character string between 0000 and 9999. They may or may not be the
25
Page 32
MI 611-211 – October 2005
3. Operation same. The default passcodes are both 0800. To disable the passcode feature configure a passcode as 0000.
Config Strings (Refer to “Structure Diagram 5” on page 42.) This section of the structure asks you to specify the designation of the Tag Number (TAG NUM), Tag Name (TAG NAME), Location (LOCATE), and Device Name (DEV NAME). Each can be made up of characters from Table 10 and can be of up to the following number of characters in length: TAG NUM TAG NAME LOCATE DEV NAME
12 14 14 6
Only eight characters will be shown on your display at a time but the display will scroll as you enter additional characters. NOTE
Only TAG NUM appears if your transmitter was specified with analog output only (870ITPH-A).
Table 10. List of Characters Characters 0 through 9 . (period) –(dash) (blank) A through Z _ (underscore)
Config AC Power (Refer to “Structure Diagram 6” on page 43.) This section of the structure asks you to specify, under AC PWR, the frequency that is used in your location. This information is used to eliminate interference in the DIAGnostic mode. If you specify 50 Hz, the transmitter performs its diagnostic tests using 60 Hz. If you specify 60 Hz, the diagnostics uses 50 Hz.
Config Diags (Refer to “Structure Diagram 6” on page 43.) This section of the structure allows configuring error messages that can appear on your display. You can elect to ENABLE all messages, DISABLE all messages or enable or disable each of the
26