870ITPH Intelligent Electrochemical Transmitter for pH, ORP, and Ion Selective Electrode Measurements: 3. Operation, pages 33 to 40
Pages 33 to 40 of 62, from 870ITPH Intelligent Electrochemical Transmitter for pH, ORP, and Ion Selective Electrode Measurements
The original document is the PDF (605 KB). This is its text as extracted, unedited; tables and drawings read best in the PDF, and each page heading opens it at that page.
For the Foxboro 870ITPH
Page 33
3. Operation
MI 611-211 – October 2005 following individually: GLASS, LEAKAGE, LOWSLOPE, PREAMP, ATCSHORT, ATC OPEN, COAT, AGING, 4-20 RNG, COMP RNG, and MEAS RNG. Each of these is discussed below. A typical approach is to initially select ENABLE ALLDIAGS and to DISABLE select messages later when you find out that you do not require them. Further, when you have messages on your display, you may CLEAR all of the messages or CLEAR specific messages. This feature then disables the reporting of those messages for an hour to allow time for corrective maintenance. NOTE
Any sensor diagnostic that has been set to ENABLE causes the output to go to the fail-safe DC mA value when the diagnostic condition exists. Therefore, if sensor diagnostics are enabled, it is important to set the fail-safe DC mA value to an acceptable safe output level (see “Config Output” on page 18). If you do not want the output to respond to some or all of the sensor diagnostics, then you must set the selected diagnostics to DISABLE. GLASS
This message reports a problem in the resistance of a glass electrode in the probe. Resistance of a glass electrode is checked versus solution ground. An error message appears if the resistance is less than user set limit. Besides clearing, enabling, and disabling the message, you can LIMIT the message to resistances below a set value. The LIMIT value may be set between 0 and 1100 kΩ in increments of 100 kΩ. The default value is 500 kΩ. You may also specify a CUTOFF Temperature above which this feature is disabled. CUTOFF T can be set between 0 and 100 ( ° C or ° F). The default value is 50. The resistance of a glass bulb decreases (roughly is halved) for every 10°C increase in temperature. NOTE
Disable this message for ORP and ISE measurement. LEAKAGE
This message reports a problem of liquid leakage into the sensor. The diagnostics checks the resistance between solution ground and the RTD. An error message appears if the resistance is anything but infinite. The message may be enabled, disabled, or cleared. LOWSLOPE
This message reports a problem usually associated with an aging glass electrode. An error message appears after a 2-point calibration in buffer solutions when the Nernst slope of the sensor becomes less than the user set limit. The limit may be set between 0 and 100 percent. The default value is 80. The message may be enabled, disabled, or cleared. NOTE
Disable this message for ORP and ISE measurement. PREAMP
This message reports a failed preamplifier. The diagnostics checks the preamplifier by monitoring the voltage coming out of it. An error message appears when the voltage is greater than 2 volts. 27
Page 34
MI 611-211 – October 2005
3. Operation
The message may be enabled, disabled, or cleared. The following Foxboro sensors contain a preamplifier: 871A-4, 871PH-3, 871PH-4, PH10-.P and ORP10-.P. ATCSHORT, ATC OPEN
This message reports a problem if the resistance of the temperature compensator is greater or less than the expected resistance of the device configured. The message for each may be enabled, disabled, or cleared. COAT
This message reports an increase in reference junction resistance. This can result from a fouled reference junction or if the electrode is not immersed in the process solution. The diagnostic checks the resistance against solution ground. An error message appears if the resistance is greater than the user set limit. The limit may be set between 0 and 100 kΩ in increments of 10 kΩ. The default value is 80. The message may be enabled, disabled, or cleared. NOTE
1. During normal operation, this diagnostic may indicate that the pipe or tank has gone dry and the sensor is no longer in solution. 2. During sensor maintenance such as cleaning or calibration, this diagnostic may appear during the times the sensor is out of solution. When the sensor is again immersed, the diagnostic should turn itself off in a few minutes. If the diagnostic remains on, then the sensor should be treated as having a coating. AGING
This message reports an aging glass electrode. An error message appears after a 2-point calibration in pH buffer solutions when the Nernst slope of the sensor has decreased sequentially five times. The message may be enabled, disabled, or cleared. 4-20 RNG
This message reports that the measurement that is tied to the analog output is outside the range that is configured for the application. The message may be enabled, disabled, or cleared. COMP RNG
This message reports that the measured temperature or the absolute measurement is outside the temperature or chemical compensation curve that is configured for the application. The message may be enabled, disabled, or cleared. MEAS RNG
This message reports that the measurement cannot be displayed accurately in the precision/scale that is configured for the application. The message may be enabled, disabled, or cleared. Table 11 shows what diagnostics are available with various Foxboro sensors if enabled and configured with reasonable limits.
28
Page 35
3. Operation
MI 611-211 – October 2005
Table 11. Sensor Diagnostics Diagnostic(a) Sensor
Glass
Leakage
Lowslope
Preamp
Coat
Aging
ATC
PH10 ORP10 871PH-1 871PH-2 871PH-3 871PH-4 871PH-5 871PH-6 871A-1 871A-2 871A-3 871A-4 EP460-.N EP460-.P EP462 EP466
X X
X X
X X X
X (b) X (b)
X X
X X X
X X X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X X X
X X X
X X
X X
X X X X X X X X
X X X X X X X X
X X X X X X X X
X
X X X X X
(a)X = applicable; blank = not applicable (b)For PH10-.P or ORP10-.P.
Hold (Refer to “Structure Diagram 7” on page 44.) The Hold mode enables you to hold the output at a determined value, thus allowing you to configure or calibrate the transmitter or perform sensor maintenance without disturbing the control loop. If you do not put the transmitter in Hold mode before configuring or calibrating, the output will go to 4 mA if you have configured the output as ANALOG or it will be held at the present value if you have configured the output as DIGITAL. While you are in the Hold mode, the bar graph will flash on and off. To enter the Hold mode, press the HOLD key. You will be prompted to enter the Minor Passcode. If you enter the incorrect passcode, the transmitter will revert to the Measurement (normal) mode. If you enter the correct passcode, the display will read HOLD OFF or HOLD ON. If it reads HOLD OFF, use the Down arrow key to change it to HOLD ON. Then specify whether you want to hold the output at the PRESENT value or at another value that you will MANUALly set. If you selected MANUAL and your transmitter was specified with analog communications (870ITPH-A), your display will read MANUAL ANALOG. Specify the value between 4.00 and 20.00 mA at which you want to hold the output while in the Hold mode. The default value is 4.00.
29
Page 36
MI 611-211 – October 2005
3. Operation
If you selected MANUAL and your transmitter was specified with digital communications (870ITPH-F), and you had previously selected ANALOG under CONFIG OUTPUT, your display will read MANUAL ANALOG. 1. Specify the value between 4.00 and 20.00 mA at which you want to hold the output while in the Hold mode. The default value is 4.00. 2. Next, go to MANUAL DIGITAL and specify the MEASUREment units, TEMPerature, and ABSOLUTE value at which you want to hold the output while in the Hold mode. a. The TEMPerature can be specified between -99.9 and +999.9 ( ° C or ° F). The default value is 0.0. b. The ABSOLUTE value can be specified between -2000.0 and +2000.0 mV. The default value is 0.0. If you selected MANUAL and your transmitter was specified with digital communications (870ITPH-F), and you had previously selected DIGITAL under CONFIG OUTPUT, your display will prompt you to specify the MEASUREment units, TEMPerature, and ABSOLUTE value at which you want to hold the output while in the Hold mode. 1. The TEMPerature can be specified between -99.9 and +999.9 ( ° C or ° F). The default value is 0.0. 2. The ABSOLUTE value can be specified between -2000.0 and +2000.0 mV. The default value is 0.0.
Calibration (Refer to “Structure Diagram 9” on page 46 and “Structure Diagram 10” on page 47.) Calibration routines in your transmitter were designed for simplicity and the elimination of user errors. The procedures are self-prompting. Two routines are available if you are measuring pH. A manual calibration enables you to enter or edit the value(s) of your solution. A single point or a 2-point manual calibration may be performed. However, a single point calibration may only be used if you have previously done a 2-point calibration with your sensor. An automatic calibration provides a buffer-recognition mechanism which locks in on the buffer value most closely representing the combination of millivolts and temperature values being reported from the sensor. Only manual calibration is available if you are measuring ORP or ISE. Only single point calibration is available if you are measuring ORP. Note that damping is not performed during either manual or automatic calibration. NOTE
Before doing any calibration, check that the temperature on the secondary display in the Measure mode agrees with the actual temperature of the solution. If there is a difference, the actual temperature should be entered under CAL TEMP in the Calibration mode before beginning your calibration. The temperature may be between -99.9 and + 999.9 ( ° C or ° F). While you are in the Calibration mode, two segments of the bar graph will appear under the legend CAL at the top of your front panel. These segments will flash if you are also in a Hold mode. To enter the Calibration mode, press the CAL key. You will be prompted to enter the Minor passcode. If you enter the incorrect passcode, the transmitter will revert to the 30
Page 37
3. Operation
MI 611-211 – October 2005
Measurement (normal) mode. If you enter the correct passcode, the display will read CAL MANUAL. If you need to adjust the temperature per the note above, go to CAL TEMP in the structure and enter the correct value. Then go back to either CAL MANUAL or to CAL AUTO. If you are measuring ORP, instead of a MANUAL SINGLEPT calibration, you can use a MANUAL ABSOLUTE calibration. This is an internal calibration that rids the transmitter of nonapplicable calibration data. To do this, go from MANUAL ABSOLUTE to PERFORM ABSOLUTE and press ENTER. If your diagnostics were configured to enable a COAT message, this message appears if the sensor is out of the solution for a sufficient period of time. This is a normal occurrence. The message disappears when you immerse your sensor. NOTE
Accurate calibration requires that the sensor and the standard solution must be stable at the same temperature. A less than optimum calibration results if either the sensor or the solution is changing temperature during the calibration procedure. For example, when a sensor is removed from a hot process, the sensor takes several minutes to cool to ambient temperature. Calibration should not be performed until the sensor has reached a stable temperature. You can monitor the temperature of the sensor in Status mode.
Manual Calibration To perform a manual calibration (grab calibration): 1. Specify SINGLEPT or TWO PT calibration. SINGLEPT must be used for ORP. 2. Immerse your sensor in Solution 1 and press ENTER. If stability parameters were configured ON, the display READING XXXXXXXX will flash until stability is achieved and then alternate with the message ENTER TO EDIT. Wait for this message and then press ENTER again. If stability parameters were configured OFF, watch for the measurement to stabilize before proceeding to press ENTER. 3. If the reading is not correct, edit it to the correct value under VALUE 1. Press ENTER. Wait for a display of CAL COMPLETE (for SINGLEPT) or IMMERSE IN SOL 2 (for TWO PT). If calibration was not successfully completed, the display will read CAL ERROR and return you to MANUAL SINGLEPT or MANUAL TWO PT. 4. If you are doing a 2-point calibration, place the sensor in the second solution and repeat the procedure. Upon completion a display of CAL COMPLETE will appear. 5. Press MEASURE to return to Measurement mode. NOTE
A manual calibration of the transmitter without its sensor can also be done with theoretical millivolt inputs. See “Electronic Calibration” below.
Automatic Calibration In Automatic Calibration, the pH value is first measured with the last calibration constants entered to decide which buffer is in use. The algorithm checks each buffer starting with Buffer 1 31
Page 38
MI 611-211 – October 2005
3. Operation and selects the first one for which this pH is within 0.5 pH of the average pH for the buffer. If this pH is not within 0.5 pH unit of any buffer table, a CAL ERROR message is displayed and no buffer is selected. To perform an automatic calibration: 1. Press ENTER at the prompt CAL AUTO. 2. At the prompt IMMERSE IN BUF 1, immerse your sensor in the first buffer solution and press ENTER after you have done so. 3. If stability parameters were configured ON, the display READING XX.XX pH will flash until stability is achieved and then alternate with the message ENTER TO CONTINUE. Wait for this message and then press ENTER again. If stability parameters were configured OFF, watch for the measurement to stabilize before proceeding to press ENTER. The instrument will then find the nearest buffer value and display that value on the display under the heading BUF 1 IS. Press ENTER if acceptable. 4. Wait for the first calibration to be completed. At the prompt IMMERSE IN BUF 2, immerse your sensor in the second solution and repeat the procedure just completed for BUF 1. Completion will be indicated with the display CAL COMPLETE.
Documentation of Calibration Procedure Next, optionally follow the prompts to enter the calibration date, and name of the calibrator. The date should be in the form mm/dd/yy. The name of the calibrator can contain up to six characters. NOTE
Invensys Foxboro recommends keeping a record of your calibrations. Recording the mV values and slope can help you establish maintenance and replacement information on your sensors. See Table 12 to calculate the expected mV value generated by a sensor in contact with a buffer of known pH value. Slope and mV values together provide useful data for troubleshooting the pH sensor.
Electronic Calibration This calibration procedure can be used to verify the calibration of the transmitter. It is not required at startup or for routine sensor/transmitter loop calibration. This procedure is recommended for troubleshooting a measurement loop to isolate a problem to the transmitter or eliminate the transmitter as a source of a problem. A millivolt source is used to supply theoretical millivolt inputs in place of the sensor. 1. Disconnect all sensor leads from your transmitter. 2. Connect a millivolt source between sensor input terminals as follows: positive to terminal 3 and negative to terminal 4. Jumper terminals 4 and 5. 3. In Configuration mode, go to CONFIG TEMP and set the temperature in MANUAL to 25 ° C or 77 ° F. 4. Go to Calibration mode. In CAL MANUAL, specify SINGLEPT or TWO PT calibration. SINGLEPT must be used for ORP. 32
Page 39
3. Operation
MI 611-211 – October 2005
5. Adjust the millivolt supply to your first value as determined by the formula in Table 12 and press ENTER. Wait for the display READING XXXXXXXX to alternate with the message ENTER TO EDIT and then press ENTER again. Table 12. Millivolt Supply Formulas Measurement Mode mV Power Supply Formula
Glass pH
(pH-7)x (-59.16)
Antimony pH
(pH-1)x (-55)
ORP ISE
(mV) -59 x log (expected Fin ppm) + 118 mV
Examples pH0: (0-7) x (-59.16) = +414.1 mV pH14: (14-7) x (-59.16) = -414.1 mV pH0: (0-1) x (-55.00) = +55.0 mV pH7: (7-1) x (-55.00) = -330.0 mV Input absolute millivolt Fluoride (F-) Charge is negative and monovalent (n=-1) Typical mV values: 1 ppm: (-59)log(1) +118 = 118 mV 2 ppm: (-59)log(2) +118 = 100 mV 10 ppm: (-59)log(10) = 118 = 59 mV 100 ppm: (-59)log(100) +118 = 0 mV
6. If the reading is not correct, enter the correct value under VALUE 1. Press ENTER. Wait for a display of CAL COMPLETE (for SINGLEPT) or IMMERSE IN SOL 2 (for TWO PT). If calibration was not successfully completed, The display will read CAL ERROR and return you to MANUAL SINGLEPT or MANUAL TWO PT. 7. If you are doing a 2-point calibration, adjust the millivolt supply to the second value and repeat the procedure. 8. Go to Measure mode. Adjust the millivolt source to a known value between the calibration points and verify the measurement as displayed on the transmitter per Table 12. 9. Go to Configuration mode to reset the temperature to AUTO or desired MANUAL setting as necessary. 10. Lastly, reconnect your sensor wires to the transmitter.
33
Page 40
MI 611-211 – October 2005
34
3. Operation