The text of a manual · part 7 of 8
870ITPH Intelligent Electrochemical Transmitter for pH, ORP, and Ion Selective Electrode Measurements, pages 41 to 54
Pages 41 to 54 of 62: 4. Maintenance; Appendix A. Structure Diagrams
The original document is the PDF (605 KB). This is its text as extracted, unedited; tables and drawings read best in the PDF, and each page heading opens it at that page.
For the Foxboro 870ITPH
Page 41
4. Maintenance ! WARNING This product contains components that have critical safety characteristics. Do not substitute components. Replace components only with identical factory supplied components. Component substitution may impair the electrical safety of this equipment and its suitability for use in hazardous locations. The maintenance of the 870IT Transmitter is limited to replacement of the keypad, the core printed wiring assembly (PWA), and the sensor PWA. See PL 611-156 for Style A or PL 611-153 for Style B transmitter part numbers. Attempts to repair printed wiring assemblies could result in damage and voiding of the warranty. The recommended repair procedure is replacement of the PWA or return of the PWA to the factory for repair.
35
Page 42
MI 611-211 – October 2005
36
4. Maintenance
Page 43
Appendix A. Structure Diagrams Top Level Structure Diagram
CONFIG DISPLAY CONFIG OUTPUT
CONFIG ELECTROD CONFIG TEMPCOMP CONFIG TEMP
CONFIG DAMPING CONFIG CHEMCOMP CONFIG INSTRMNT
CONFIG TIMEOUT CONFIG BUFFERS CONFIG PASSCODE CONFIG STRINGS CONFIG AC PWR CONFIG DIAGS
37
Page 44
MI 611-211 – October 2005
Appendix A. Structure Diagrams
Structure Diagram 1 CONFIGURATION If PASSCODE MAJOR ≠ 0000 PASSCODE
CONFIG DISPLAY
DISPLAY PRIMARY
PRIMARY SENSOR
SENSOR pH ISE ORP
If pH PRIMARY RESOLUTN
RESOLUTN 0.1 pH 0.01 pH
If ISE PRIMARY SCALE
SCALE 9.999 ppm 99.99 ppm 999.9 ppm 9999. ppm
DISPLAY SECOND
SECOND TEMP ABSOLUTE mA
CONFIG OUTPUT
MODE
OUTPUT ANALOG
DIGITAL ANALOG
ANALOG ABSOLUTE MEASURE TEMP
OUTPUT RANGE
MIN
OUTPUT FAIL SIG
FAIL SIG OFF ON PULSE
CONFIG ELECTROD (see next page)
38
MAX
DC mA
Page 45
Appendix A. Structure Diagrams
MI 611-211 – October 2005
Structure Diagram 2 CONFIGURATION (cont.) CONFIG OUTPUT (see previous page)
CONFIG ELECTROD
ELECTROD
If pH GLASS ANTIMONY OTHER
ISO POT
If ISE POSITIVE NEGATIVE
CONFIG TEMPCOMP
TEMPCOMP
If pH STANDARD AMMONIA CUSTOM
REF TEMP
NUM PTS
TEMPx
VALx
TEMPx
VALx
If ISE STANDARD CUSTOM
REF TEMP
NUM PTS
TEMPx
VALx
TEMPx
VALx
CONFIG TEMP (see next page)
39
Page 46
MI 611-211 – October 2005
Appendix A. Structure Diagrams
Structure Diagram 3 CONFIGURATION (cont.) CONFIG TEMPCOMP (see previous page)
CONFIG TEMP
TEMP UNITS
UNITS CELSIUS FAHREN
TEMP MODE
MODE AUTO
MANUAL
TEMP TYPE
DEFAULT
MANUAL
TYPE RTD BALCO
RTD 2 WIRE
2 WIRE 100 OHM 1000 OHM
3 WIRE
3 WIRE 100 OHM 1000 OHM
CONFIG DAMPING
DAMPING 1 SEC 5 SEC 10 SEC 20 SEC 40 SEC 120 SEC
If ISE CONFIG CHEMCOMP
CHEMCOMP STANDARD CUSTOM
CONFIG INSTRMNT (see next page)
40
NUM PTS
ABSx
NEWx
ABSx+1
NEWx+1
Page 47
Appendix A. Structure Diagrams
MI 611-211 – October 2005
Structure Diagram 4 CONFIGURATION (cont.) CONFIG CHEMCOMP (see previous page)
If ANALOG CONFIG INSTRMNT
INSTRMNT mA CAL
INSTRMNT STAB IND mA CAL 4 mA
4 mA mA CAL 20 mA
20 mA
STAB IND MEAS
MEAS OFF ON
STAB IND TEMP
TIME
ON VAR
VAR
ON TIME
TIME
ON VAR
VAR
TEMP OFF ON
CONFIG TIMEOUT
ON TIME
TIMEOUT
CONFIG BUFFERS (see next page)
41
Page 48
MI 611-211 – October 2005
Appendix A. Structure Diagrams
Structure Diagram 5 CONFIGURATION (cont.) CONFIG TIMEOUT (see previous page)
If pH CONFIG BUFFERS
BUFFERS AMERICAN NIST EUROPEAN SPECIAL
SPECIAL BUFFER 1
SPECIAL BUFFER 2
SPECIAL BUFFER 3
CONFIG PASSCODE
CONFIG STRINGS
CONFIG ac PWR (see next page)
42
PASSCODE MINOR
MINOR
PASSCODE MAJOR
MAJOR
STRINGS TAG NUM
TAG NUM
STRINGS TAG NAME
TAG NAME
STRINGS LOCATE
LOCATE
STRINGS DEV NAME
DEV NAME
NUM PTS
NUM PTS
NUM PTS
TEMPx
VALx
TEMPx+1
VALx+1
TEMPx
VALx
TEMPx+1
VALx+1
TEMPx
VALx
TEMPx+1
VALx+1
Only on 870ITPH-F Transmitters
Page 49
Appendix A. Structure Diagrams
MI 611-211 – October 2005
Structure Diagram 6 CONFIGURATION (cont.) CONFIG STRINGS (see previous page)
CONFIG AC PWR
AC PWR 60 Hz 50 Hz
CONFIG DIAGS
DIAGS ALLDIAGS
ALLDIAGS CLEAR ENABLE DISABLE
DIAGS GLASS
GLASS CLEAR ENABLE DISABLE LIMIT CUTOFF T
DIAGS LEAKAGE
LIMIT
CUTOFF T
LEAKAGE CLEAR ENABLE DISABLE
DIAGS LOWSLOPE
LOWSLOPE CLEAR ENABLE DISABLE LIMIT
DIAGS PREAMP
LIMIT
PREAMP
DIAGS AGING
CLEAR ENABLE DISABLE DIAGS ATCSHORT
CLEAR ENABLE DISABLE
ATCSHORT
DIAGS 4 - 20 RNG
CLEAR ENABLE DISABLE DIAGS ATC OPEN
ATC OPEN
DIAGS COMP RNG
Continued
COMP RNG CLEAR ENABLE DISABLE
COAT CLEAR ENABLE DISABLE LIMIT
4 - 20 RNG CLEAR ENABLE DISABLE
CLEAR ENABLE DISABLE DIAGS COAT
AGING
DIAGS MEAS RNG
LIMIT
MEAS RNG CLEAR ENABLE DISABLE
43
Page 50
MI 611-211 – October 2005
Appendix A. Structure Diagrams
Structure Diagram 7 HOLD If PASSCODE MINOR ≠ 0000 PASSCODE
HOLD OFF
HOLD ON
ON PRESENT MANUAL
If ANALOG MODE MANUAL ANALOG
ANALOG
MANUAL DIGITAL
DIGITAL MEASURE
MEASURE
DIGITAL TEMP
TEMP
DIGITAL ABSOLUTE
ABSOLUTE
If DIGITAL MODE
44
MANUAL MEASURE
MEASURE
MANUAL TEMP
TEMP
MANUAL ABSOLUTE
ABSOLUTE
Page 51
Appendix A. Structure Diagrams
MI 611-211 – October 2005
Structure Diagram 8 STATUS TEMP
SLOPE ABSOLUTE mA
GLASSRES
ASYMM
REF RES
ATC RES
CAL NAME
LAST CAL
TAG NAME* LOCATION*
SONUMBER MS CODE
S/W VER
DEV NAME*
DEV TYPE*
TAG NUM
* Appears only on devices coded 870ITPH-F
45
Page 52
MI 611-211 – October 2005
Appendix A. Structure Diagrams
Structure Diagram 9 CALIBRATION If PASSCODE MINOR ≠ 0000 PASSCODE
If pH CAL MANUAL
MANUAL SINGLEPT
MANUAL TWO PT
READING XXXXXXX
CAL ERROR
If error during calibration
IMMERSE IN SOL 1
READING XXXXXXX
CAL ERROR
IMMERSE IN SOL 2 CAL ERROR CAL AUTO
IMMERSE IN BUF 1
ENTER TO EDIT
IMMERSE IN SOL
READING xx.xx pH
VALUE 1
CAL COMPLETE ENTER TO EDIT
VALUE 1
If error during calibration
READING XXXXXXX
ENTER TO EDIT
VALUE 2
If error during calibration CAL COMPLETE ENTER TO CONTINUE
BUF 1 IS xx.xx pH
BUF NOT FOUND CAL ERROR IMMERSE IN BUF 2
If error during calibration
READING xx.xx pH
ENTER TO CONTINUE
BUF NOT FOUND CAL ERROR CAL TEMP
TEMP
CAL DATE
DATE
CAL PERSON
PERSON
(If ISE)
46
If error during calibration
BUF 2 IS xx.xx pH
CAL COMPLETE
Page 53
Appendix A. Structure Diagrams
MI 611-211 – October 2005
Structure Diagram 10 CALIBRATION (cont.) If ISE CAL MANUAL
MANUAL SINGLEPT
MANUAL TWO PT
READING XXXXXXX
CAL ERROR
If error during calibration
IMMERSE IN SOL 1
READING XXXXXXX
CAL ERROR
IMMERSE IN SOL 2
CAL TEMP
TEMP
CAL DATE
DATE
CAL PERSON
PERSON
ENTER TO EDIT
IMMERSE IN SOL
VALUE 1
CAL COMPLETE ENTER TO EDIT
VALUE 1
If error during calibration
READING XXXXXXX
ENTER TO EDIT
CAL ERROR
If error during calibration
IMMERSE IN SOL
READING XXXXXXX
CAL ERROR
If error during calibration
VALUE 2
CAL COMPLETE
If ORP CAL MANUAL
MANUAL SINGLEPT
MANUAL ABSOLUTE
PERFORM ABSOLUTE
CAL TEMP
TEMP
CAL DATE
DATE
CAL PERSON
PERSON
ENTER TO EDIT
VALUE 1
CAL COMPLETE
CAL COMPLETE
47
Page 54
MI 611-211 – October 2005
Appendix A. Structure Diagrams
Structure Diagram 11 DIAGNOSTICS If no error DIAGS OK
If transmitter error ERROR XMTR
If probe error(s) ERROR PROBE
PROBE ER GLASS
GLASS REPLACE
PROBE ER LOWSLOPE
LOWSLOPE REPLACE
PROBE ER COAT
COAT CLEAN
PROBE ER LEAKAGE
LEAKAGE REPLACE
PROBE ER PREAMP
PREAMP REPLACE CONFIG 4 - 20 RNG
4 - 20 mA OVER RNG
PROBE ER AGING 4 - 20 mA UNDR RNG PROBE ER ATCSHORT
ATCSHORT REPLACE
PROBE ER ATC OPEN
ATC COAT REPLACE
CONFIG CAL REQ
CAL REQUIRED
CONFIG MEAS RNG
MEASURE OVER RNG
MEASURE UNDR RNG
If configuration error ERROR CONFIG
CONFIG COMP RNG
COMPTABL OVER RNG
COMPTABL UNDR RNG
48