Page 41

4. Maintenance ! WARNING This product contains components that have critical safety characteristics. Do not substitute components. Replace components only with identical factory supplied components. Component substitution may impair the electrical safety of this equipment and its suitability for use in hazardous locations. The maintenance of the 870IT Transmitter is limited to replacement of the keypad, the core printed wiring assembly (PWA), and the sensor PWA. See PL 611-156 for Style A or PL 611-153 for Style B transmitter part numbers. Attempts to repair printed wiring assemblies could result in damage and voiding of the warranty. The recommended repair procedure is replacement of the PWA or return of the PWA to the factory for repair.

35

Page 42

MI 611-211 – October 2005

36

4. Maintenance

Page 43

Appendix A. Structure Diagrams Top Level Structure Diagram

CONFIG DISPLAY CONFIG OUTPUT

CONFIG ELECTROD CONFIG TEMPCOMP CONFIG TEMP

CONFIG DAMPING CONFIG CHEMCOMP CONFIG INSTRMNT

CONFIG TIMEOUT CONFIG BUFFERS CONFIG PASSCODE CONFIG STRINGS CONFIG AC PWR CONFIG DIAGS

37

Page 44

MI 611-211 – October 2005

Appendix A. Structure Diagrams

Structure Diagram 1 CONFIGURATION If PASSCODE MAJOR ≠ 0000 PASSCODE

CONFIG DISPLAY

DISPLAY PRIMARY

PRIMARY SENSOR

SENSOR pH ISE ORP

If pH PRIMARY RESOLUTN

RESOLUTN 0.1 pH 0.01 pH

If ISE PRIMARY SCALE

SCALE 9.999 ppm 99.99 ppm 999.9 ppm 9999. ppm

DISPLAY SECOND

SECOND TEMP ABSOLUTE mA

CONFIG OUTPUT

MODE

OUTPUT ANALOG

DIGITAL ANALOG

ANALOG ABSOLUTE MEASURE TEMP

OUTPUT RANGE

MIN

OUTPUT FAIL SIG

FAIL SIG OFF ON PULSE

CONFIG ELECTROD (see next page)

38

MAX

DC mA

Page 45

Appendix A. Structure Diagrams

MI 611-211 – October 2005

Structure Diagram 2 CONFIGURATION (cont.) CONFIG OUTPUT (see previous page)

CONFIG ELECTROD

ELECTROD

If pH GLASS ANTIMONY OTHER

ISO POT

If ISE POSITIVE NEGATIVE

CONFIG TEMPCOMP

TEMPCOMP

If pH STANDARD AMMONIA CUSTOM

REF TEMP

NUM PTS

TEMPx

VALx

TEMPx

VALx

If ISE STANDARD CUSTOM

REF TEMP

NUM PTS

TEMPx

VALx

TEMPx

VALx

CONFIG TEMP (see next page)

39

Page 46

MI 611-211 – October 2005

Appendix A. Structure Diagrams

Structure Diagram 3 CONFIGURATION (cont.) CONFIG TEMPCOMP (see previous page)

CONFIG TEMP

TEMP UNITS

UNITS CELSIUS FAHREN

TEMP MODE

MODE AUTO

MANUAL

TEMP TYPE

DEFAULT

MANUAL

TYPE RTD BALCO

RTD 2 WIRE

2 WIRE 100 OHM 1000 OHM

3 WIRE

3 WIRE 100 OHM 1000 OHM

CONFIG DAMPING

DAMPING 1 SEC 5 SEC 10 SEC 20 SEC 40 SEC 120 SEC

If ISE CONFIG CHEMCOMP

CHEMCOMP STANDARD CUSTOM

CONFIG INSTRMNT (see next page)

40

NUM PTS

ABSx

NEWx

ABSx+1

NEWx+1

Page 47

Appendix A. Structure Diagrams

MI 611-211 – October 2005

Structure Diagram 4 CONFIGURATION (cont.) CONFIG CHEMCOMP (see previous page)

If ANALOG CONFIG INSTRMNT

INSTRMNT mA CAL

INSTRMNT STAB IND mA CAL 4 mA

4 mA mA CAL 20 mA

20 mA

STAB IND MEAS

MEAS OFF ON

STAB IND TEMP

TIME

ON VAR

VAR

ON TIME

TIME

ON VAR

VAR

TEMP OFF ON

CONFIG TIMEOUT

ON TIME

TIMEOUT

CONFIG BUFFERS (see next page)

41

Page 48

MI 611-211 – October 2005

Appendix A. Structure Diagrams

Structure Diagram 5 CONFIGURATION (cont.) CONFIG TIMEOUT (see previous page)

If pH CONFIG BUFFERS

BUFFERS AMERICAN NIST EUROPEAN SPECIAL

SPECIAL BUFFER 1

SPECIAL BUFFER 2

SPECIAL BUFFER 3

CONFIG PASSCODE

CONFIG STRINGS

CONFIG ac PWR (see next page)

42

PASSCODE MINOR

MINOR

PASSCODE MAJOR

MAJOR

STRINGS TAG NUM

TAG NUM

STRINGS TAG NAME

TAG NAME

STRINGS LOCATE

LOCATE

STRINGS DEV NAME

DEV NAME

NUM PTS

NUM PTS

NUM PTS

TEMPx

VALx

TEMPx+1

VALx+1

TEMPx

VALx

TEMPx+1

VALx+1

TEMPx

VALx

TEMPx+1

VALx+1

Only on 870ITPH-F Transmitters

Page 49

Appendix A. Structure Diagrams

MI 611-211 – October 2005

Structure Diagram 6 CONFIGURATION (cont.) CONFIG STRINGS (see previous page)

CONFIG AC PWR

AC PWR 60 Hz 50 Hz

CONFIG DIAGS

DIAGS ALLDIAGS

ALLDIAGS CLEAR ENABLE DISABLE

DIAGS GLASS

GLASS CLEAR ENABLE DISABLE LIMIT CUTOFF T

DIAGS LEAKAGE

LIMIT

CUTOFF T

LEAKAGE CLEAR ENABLE DISABLE

DIAGS LOWSLOPE

LOWSLOPE CLEAR ENABLE DISABLE LIMIT

DIAGS PREAMP

LIMIT

PREAMP

DIAGS AGING

CLEAR ENABLE DISABLE DIAGS ATCSHORT

CLEAR ENABLE DISABLE

ATCSHORT

DIAGS 4 - 20 RNG

CLEAR ENABLE DISABLE DIAGS ATC OPEN

ATC OPEN

DIAGS COMP RNG

Continued

COMP RNG CLEAR ENABLE DISABLE

COAT CLEAR ENABLE DISABLE LIMIT

4 - 20 RNG CLEAR ENABLE DISABLE

CLEAR ENABLE DISABLE DIAGS COAT

AGING

DIAGS MEAS RNG

LIMIT

MEAS RNG CLEAR ENABLE DISABLE

43

Page 50

MI 611-211 – October 2005

Appendix A. Structure Diagrams

Structure Diagram 7 HOLD If PASSCODE MINOR ≠ 0000 PASSCODE

HOLD OFF

HOLD ON

ON PRESENT MANUAL

If ANALOG MODE MANUAL ANALOG

ANALOG

MANUAL DIGITAL

DIGITAL MEASURE

MEASURE

DIGITAL TEMP

TEMP

DIGITAL ABSOLUTE

ABSOLUTE

If DIGITAL MODE

44

MANUAL MEASURE

MEASURE

MANUAL TEMP

TEMP

MANUAL ABSOLUTE

ABSOLUTE

Page 51

Appendix A. Structure Diagrams

MI 611-211 – October 2005

Structure Diagram 8 STATUS TEMP

SLOPE ABSOLUTE mA

GLASSRES

ASYMM

REF RES

ATC RES

CAL NAME

LAST CAL

TAG NAME* LOCATION*

SONUMBER MS CODE

S/W VER

DEV NAME*

DEV TYPE*

TAG NUM

* Appears only on devices coded 870ITPH-F

45

Page 52

MI 611-211 – October 2005

Appendix A. Structure Diagrams

Structure Diagram 9 CALIBRATION If PASSCODE MINOR ≠ 0000 PASSCODE

If pH CAL MANUAL

MANUAL SINGLEPT

MANUAL TWO PT

READING XXXXXXX

CAL ERROR

If error during calibration

IMMERSE IN SOL 1

READING XXXXXXX

CAL ERROR

IMMERSE IN SOL 2 CAL ERROR CAL AUTO

IMMERSE IN BUF 1

ENTER TO EDIT

IMMERSE IN SOL

READING xx.xx pH

VALUE 1

CAL COMPLETE ENTER TO EDIT

VALUE 1

If error during calibration

READING XXXXXXX

ENTER TO EDIT

VALUE 2

If error during calibration CAL COMPLETE ENTER TO CONTINUE

BUF 1 IS xx.xx pH

BUF NOT FOUND CAL ERROR IMMERSE IN BUF 2

If error during calibration

READING xx.xx pH

ENTER TO CONTINUE

BUF NOT FOUND CAL ERROR CAL TEMP

TEMP

CAL DATE

DATE

CAL PERSON

PERSON

(If ISE)

46

If error during calibration

BUF 2 IS xx.xx pH

CAL COMPLETE

Page 53

Appendix A. Structure Diagrams

MI 611-211 – October 2005

Structure Diagram 10 CALIBRATION (cont.) If ISE CAL MANUAL

MANUAL SINGLEPT

MANUAL TWO PT

READING XXXXXXX

CAL ERROR

If error during calibration

IMMERSE IN SOL 1

READING XXXXXXX

CAL ERROR

IMMERSE IN SOL 2

CAL TEMP

TEMP

CAL DATE

DATE

CAL PERSON

PERSON

ENTER TO EDIT

IMMERSE IN SOL

VALUE 1

CAL COMPLETE ENTER TO EDIT

VALUE 1

If error during calibration

READING XXXXXXX

ENTER TO EDIT

CAL ERROR

If error during calibration

IMMERSE IN SOL

READING XXXXXXX

CAL ERROR

If error during calibration

VALUE 2

CAL COMPLETE

If ORP CAL MANUAL

MANUAL SINGLEPT

MANUAL ABSOLUTE

PERFORM ABSOLUTE

CAL TEMP

TEMP

CAL DATE

DATE

CAL PERSON

PERSON

ENTER TO EDIT

VALUE 1

CAL COMPLETE

CAL COMPLETE

47

Page 54

MI 611-211 – October 2005

Appendix A. Structure Diagrams

Structure Diagram 11 DIAGNOSTICS If no error DIAGS OK

If transmitter error ERROR XMTR

If probe error(s) ERROR PROBE

PROBE ER GLASS

GLASS REPLACE

PROBE ER LOWSLOPE

LOWSLOPE REPLACE

PROBE ER COAT

COAT CLEAN

PROBE ER LEAKAGE

LEAKAGE REPLACE

PROBE ER PREAMP

PREAMP REPLACE CONFIG 4 - 20 RNG

4 - 20 mA OVER RNG

PROBE ER AGING 4 - 20 mA UNDR RNG PROBE ER ATCSHORT

ATCSHORT REPLACE

PROBE ER ATC OPEN

ATC COAT REPLACE

CONFIG CAL REQ

CAL REQUIRED

CONFIG MEAS RNG

MEASURE OVER RNG

MEASURE UNDR RNG

If configuration error ERROR CONFIG

CONFIG COMP RNG

COMPTABL OVER RNG

COMPTABL UNDR RNG

48