For the Foxboro 870ITPH
Page 55
Appendix B. Configuration Table This appendix contains information that will help you configure your transmitter. The information is presented in the form of a table containing each prompt/parameter, its limits, and its factory default value. It also provides space for you to record your specific configuration and notations. Identifies the structure diagram that contains a specific prompt/parameter.
Limits of each parameter.
Prompts to parameters in the order in which they are displayed when you step through the menu structure.
Structure Diagram 1
Prompt/Parameter
Parameter Limits
PASSCODE
0001 AND 9999
CONFIG DISPLAY DISPLAY PRIMARY PRIMARY SENSOR SENSOR PRIMARY RESOLUTN RESOLUTN PRIMARY SCALE SCALE
------pH, ISE, ORP ---
0.1, 0.01 pH --9.999, 99.99, 999.9, 9999 ppm DISPLAY SECOND - - SECOND TEMP, ABSOLUTE, mA CONFIG OUTPUT --MODE ANALOG, DIGITAL OUTPUT ANALOG - - ANALOG ABSOLUTE, MEASURE, TEMP OUTPUT RANGE ---
Space for you to record your specific configuration. Standard factory configuration as shipped from Foxboro.
Factory Configuration 0800
Additional information and space for your notations
User Configuration
Remarks and Notes If PASSCODE MAJOR ≠ 0000 pH If pH 0.01 pH If ISE
TEMP
ANALOG If ANALOG MEASURE
49
Page 56
MI 611-211 – October 2005
Structure Diagram 1
2
50
Prompt/Parameter MIN MAX OUTPUT FAIL SIG FAIL SIG DC mA
Appendix B. Configuration Table
Parameter Limits
Factory Configuration
User Configuration
Remarks and Notes
Depends on Analog 0 pH Output 14 pH --OFF, ON, PULSE OFF 3.8 and 20.5 mA 3.8
CONFIG ELECTROD - - GLASS, ELECTROD ANTIMONY, OTHER POSITIVE, NEGATIVE CONFIG TEMPCOMP TEMPCOMP STANDARD, AMMONIA, CUSTOM (CUSTOM) --REF TEMP -99.9 and 999.9 NUM PTS 02 and 21 TEMPx -99.9 and 999.9 VALx -2.00 and +14.00 pH TEMPCOMP STANDARD, CUSTOM (CUSTOM) --REF TEMP -99.9 and 999.9 NUM PTS 02 and 21 TEMPx -99.9 and 999.9 VALx 0 AND 2000 ppm
GLASS
If FAIL SIG = ON If pH If ISE
STANDARD
If pH
0.0 02 0.0 0.00
° C or ° F ° C or ° F
If ISE
° C or ° F ° C or ° F
Page 57
Appendix B. Configuration Table
Structure Diagram 3
Prompt/Parameter CONFIG TEMP TEMP UNITS UNITS TEMP MODE MODE DEFAULT MANUAL TEMP TYPE TYPE RTD 2 WIRE 3 WIRE CONFIG DAMPING DAMPING CONFIG CHEMCOMP CHEMCOMP
4
MI 611-211 – October 2005
Parameter Limits
Factory Configuration
----CELSIUS, CELSIUS FAHREN --AUTO, MANUAL AUTO -99.9 and 999.9 -99.9 and 999.9 - -RTD, BALCO 3K RTD 2 WIRE, 3 WIRE 3 WIRE 100 OHM, 1000 OHM 100 OHM, 1000 OHM 1000 OHM --1, 5, 10, 20, 40, or 5 200 ---
STANDARD, CUSTOM (CUSTOM) --NUM PTS 02 and 21 ABSx 0 and 9999 ppm NEWx 0 and 9999 ppm CONFIG INSTRMNT - - INSTRMNT mA CAL - - mA CAL 4 mA 4 mA mA CAL 20 mA 20 mA
User Configuration
Remarks and Notes
If AUTO If MANUAL
If RTD If 2 WIRE If 3 WIRE
SECond If ISE
STANDARD
02
51
Page 58
MI 611-211 – October 2005
Structure Diagram 4
52
Appendix B. Configuration Table
Parameter Limits
Prompt/Parameter INSTRMNT STAB IND STAB IND MEAS MEAS ON TIME TIME ON VAR VAR STAB IND TEMP TEMP ON TIME TIME ON VAR VAR CONFIG TIMEOUT TIMEOUT
Factory Configuration
User Configuration
Remarks and Notes
----ON, OFF
ON
5 - 60
5
1 and 9 --ON, OFF
9
5 - 60
5
1 and 9 --0 and 999
9
If ON Increments of 5 s
ON
600
Increments of 5 s
Seconds
Page 59
Appendix B. Configuration Table
Structure Diagram 5
6
Prompt/Parameter CONFIG BUFFERS BUFFERS
MI 611-211 – October 2005
Parameter Limits
--AMERICAN, NIST, EUROPEAN, SPECIAL SPECIAL BUFFER 1 - - NUM PTS 02 AND 11 TEMPx -99.9 and 999.9 VALx -2.00 and 14.00 pH SPECIAL BUFFER 2 - - NUM PTS 02 AND 11 TEMPx -99.9 and 999.9 VALx -2.00 and 14.00 pH SPECIAL BUFFER 3 - - NUM PTS 02 AND 11 TEMPx -99.9 and 999.9 VALx -2.00 and 14.00 pH CONFIG PASSCODE - - PASSCODE MINOR - - MINOR 0 AND 9999 PASSCODE MAJOR - - MAJOR 0 AND 9999 CONFIG STRINGS --STRINGS TAG NUM - - TAG NUM STRINGS TAG --NAME TAG NAME STRINGS LOCATE - - LOCATE STRINGS DEV --NAME DEV NAME CONFIG AC PWR --60, 50 Hz AC PWR CONFIG DIAGS ---
Factory Configuration
User Configuration
Remarks and Notes If pH
AMERICAN
If SPECIAL 02 0.00 0.00
° C or ° F
02 0.00 0.00
° C or ° F
02 0.00 0.00
° C or ° F
0800 0800
60
53
Page 60
MI 611-211 – October 2005
Structure Diagram 6
Prompt/Parameter ALLDIAGS GLASS LIMIT CUTOFF T LEAKAGE LOWSLOPE LIMIT PREAMP ATCSHORT ATC OPEN COAT LIMIT AGING 4 - 20 RNG COMP RNG MEAS RNG
54
Appendix B. Configuration Table
Parameter Limits
Factory Configuration
CLEAR, ENABLE, DISABLE DISABLE CLEAR, ENABLE, DISABLE, LIMIT, CUTOFF T 0 - 1100 kOHM 0.0 and 100.0 CLEAR, ENABLE, DISABLE CLEAR, ENABLE, DISABLE, LIMIT 0.0 and 100.0 80.0 CLEAR, ENABLE, DISABLE CLEAR, ENABLE, DISABLE CLEAR, ENABLE, DISABLE CLEAR, ENABLE, DISABLE, LIMIT 0 - 100 kOHM 80 kOHM CLEAR, ENABLE, DISABLE CLEAR, ENABLE, DISABLE CLEAR, ENABLE, DISABLE CLEAR, ENABLE, DISABLE
User Configuration
Remarks and Notes
Increments of 100 50.0
Increments of 10
Page 61
Index A Ambient Temperature 4 Automatic Calibration 31 C Changing Data 13 Configuration Table 49 Controls and Indicators 11 E Error Messages 12 H Hold 29 I Identification 6 Installation 7 L Load Limitations 4 M Maintenance 35 Manual Calibration 31 Measure 14 Mounting 7 O Operation 11 P Panel Mounting 7 Passcode Access 13 Pipe Mounting 8 Power Supply Requirements 4
55
Page 62
MI 611-211 – October 2005
Q Quick Start
Index
1
R Reference 3 S Sensors 4 Status 16 Structure Diagrams 37 Surface Mounting 7 U Unpacking 7 W Wiring 8
ISSUE DATES JUN 1995 OCT 1996 JUN 1999 JUL 2004 OCT 2005 Vertical lines to right of text or illustrations indicate areas changed at last issue date. 33 Commercial Street Foxboro, MA 02035-2099 United States of America http://www.foxboro.com Inside U.S.: 1-866-746-6477 Outside U.S.: 1-508-549-2424 or contact your local Foxboro Representative. Facsimile: (508) 549-4492
Invensys, Foxboro, and I/A Series are trademarks of Invensys plc, its subsidiaries, and affiliates. All other brand names may be trademarks of their respective owners.
Copyright 1995-2005 Invensys Systems, Inc. All rights reserved MB 123