Page 1
RAM 4021-DPX
Operation Manual
Gi GfG Instrumentation
Worldwide Manufacturer of Gas Detection Solutions
Page 2
Page 3
ABL 4021-DPX /RAM 4021-DPX TABLE OF CONTENTS DESCRIPTION ...........................................................................................1 General 1 SET UP PROCEDURE.................................................................................2 Set-up modes 2 OPERATION..............................................................................................4 CALIBRATION ...........................................................................................5 Calibration carbon monoxide sensor with co test gas 5 Dew point sensor calibration 6 “ZERO/CAL” adjustment for carbon monoxide 6 MAINTENANCE (qualified technicians only) ........................................7 Dew point sensor replacement 8 Carbon monoxide sensor replacement 8 TROUBLESHOOTING SECTION ..................................................................9 Zero fail during CO calibration 9 Calibration gas fail during co calibration 9 ACCESSORIES AND FIELD REPLACEMENT PARTS ..................................10 ACCESSORIES 10 REPLACEMENT PARTS 10 EQUIPMENT TECHNICAL DATA ..............................................................11 4021-DPX DIAGRAM ..............................................................................12 WARRANTY............................................................................................13 WIRING DIAGRAM ..................................................................................14
Page 4
ABL-4021-DPX FOR YOUR SAFETY Like any piece of complex equipment, the GfG Instrumentation ABL-4021 series will do the job it was designed to do, only if it is used and serviced in accordance with the manufacturer’s instructions. All individuals who have or will have the responsibility of servicing the equipment must carefully read this manual. The warranties made by GfG Instrumentation with regards to this instrument are voided, if the product is not used and serviced in accordance with the instructions in this manual. Please protect yourself and others who depend on this instrument by following these instructions. The above does not alter statements regarding GfG Instrumentation’s warranties and conditions of sale and delivery. DESCRIPTION General This unit is designed to provide continuous monitoring of the carbon monoxide and dew point in the levels of breathing air. The instrument’s electronics are enclosed in a NEMA-4X polycarbonate case. The case is corrosion resistant, positively pressurized by the compressor supply line, and sealed except for a bleed hole to exhaust the compressor’s air. The unit operates on 110 V AC power. It comes equipped with a case mounted horn that may be disconnected if not required.
1
Page 5
Calibration adjustment controls are not necessary since they are automatically performed by the microprocessor. The carbon monoxide chemical cell has a life expectancy of two to three years with a 30-day calibration check recommended. The dew point sensor has an expected life of two years and requires no calibration due to its auto calibration feature. SET UP PROCEDURE By briefly depressing the top button (see Figure 1) it is possible to cycle through the set up parameters. Each time the top button is depressed the display screen will identify the selected mode and will indicate the current set up. The lower button will change the parameter. Set-up modes The following outlines the set up modes and options available. In each mode it is possible to change the set up using the lower button. DPalm xx.x Dew point high relay setting (dryer failure alarm) DLalm xx.x Dew point low relay setting (dryer control) COalm xx
The alarm point in parts per million is displayed
CO XX
The part per million of carbon monoxide to be used for calibration is displayed (i.e. CO - 20).
2
Page 6
The carbon monoxide test gas concentration may be set from 5 to 100 PPM and is factory set to 20 PPM. It is important to use the same test gas as the “CAL” setting, otherwise the calibration will fail because of the error protection feature. NOTE: This set-up must be identical to the PPM concentration of the carbon monoxide calibration gas (5 50 PPM CO). CO rly
The carbon monoxide alarm relay (remote) can be made operational (ON) or disabled (OFF) in this mode. Setting the relay ON or OFF permits any auxiliary device connected to them to be shut off if not needed. During calibration it will be automatically shut off.
DP rly
Setting the relay ON or OFF permits any auxiliary device connected to them to be shut off if not needed. Factory set to 10 degrees F. (Dryer failure relay)
DL rly
Used for controlling a dryer to cycle on/off at the set point determined-factory set to 4 degrees F.
Test Mode
Activates alarms and solid squares on LED readout when lower button is pushed. The alarm horn and light relays may be tested by pressing and holding the bottom button while in alarm test Set-Up mode. When testing an auxiliary horn, be sure the relay is turned ON. 3
Page 7
OPERATION Plug the unit into 110 V AC and the display will show introductory messages and a warm-up countdown. If the unit does not power up, check the electrical connections and try replugging the unit. If the start up does not occur, call the factory. After the warm-up countdown, the instrument will display CO readings. The air regulator may be adjusted at any time to set the flow level from 0.5 to 0.9 CFH. If the low flow indication shows on the display, increase the flow to 1 CFH; and then drop the flow to operational range of .5 to .8 CFH. CAUTION: If the unit is reading a carbon monoxide gas level, do not make any adjustment for a few hours until the unit has settled in. If calibration is attempted within the first fifteen minutes, a TOO SOON message will occur on the display. If the carbon monoxide gas readings remain high or below zero (-0), recalibration may be needed. We also recommend checking the compressor’s air intake, which may be the cause for high readings. Outside air intake is recommended - but can easily be contaminated by furnace exhaust, building vents, parking lots, etc. Intakes within the building are to be avoided because they often contain low PPM level carbon monoxide.
4
Page 8
CALIBRATION Calibration carbon monoxide sensor with co test gas (recommended every 30 days) Although the unit was calibrated at the factory, it may require carbon monoxide recalibration due to handling. The only way to assure the sensor is operating properly is to place gas on it. To calibrate the unit with carbon monoxide test gas, shut off the air from the compressor supply line with the regulator, and after a few moments, a “LOW FLOW” message will appear. Assemble the calibration kit and connect the tank of test gas to the cal port connector on the instrument. Open the gas valve (use only the fixed flow regulator provided with the calibration kit). The display will read “CAL GAS DETECTED” then “WAIT” and a 60 second count down will begin. If the compressor air supply line was not shut off, a message will appear “SUPPLY OFF”. If such a message occurs, shut off the supply air; and then begin the cal process again to activate the cal port switch. After 60 seconds a gas reading will appear. With 20 PPM test gas applied the gas reading should be 20. The number will move around as it’s self adjusting, then will automatically set. After the unit auto calibrates, a message will appear “CAL PASSED”, then “REMOVE CAL GAS”, indicating that the CO gain adjustment has been set for 20 PPM. This prompts the calibrator to remove test gas. Then the message “LOW FLOW” appears, prompting the calibrator to turn the supply on at the regulator. 5
Page 9
If an incorrect gas concentration is used or the sensor and/or instrument is not properly functioning, the auto calibration process will not finish. If the instrument fails to complete the auto calibration process within 5-6 minutes, remove the calibration gas, and the message will appear “CAL FAILED”. This affords improper calibration protection and an effort should be made to understand why it did not calibrate (see troubleshooting section for assistance). Dew point sensor calibration The dew point sensor does not require external calibration, and will self calibrate periodically, or after a severe humidity change. During the dew point sensor auto-calibration, the dew point reading will remain constant on the display until auto calibration is complete. The auto calibration process will take 2-3 minutes. “ZERO/CAL” adjustment for carbon monoxide The carbon monoxide read-out (PPM) will be “0” in the absence of carbon monoxide. This “0” can be calibrated by flowing clean air over the sensor and activating the zero/cal set switch. The “ZERO/CAL” set switch is activated through a small hole in the face of the unit using a wire activator (bent paper clip). Two methods are available to check or set the carbon monoxide ZERO/CAL. With normal flow through the unit from a clean air supply, depress the ZERO/CAL switch and the message “ZEROING” will appear. Hold closed until the “ZEROING” message disappears, immediately release the switch, and the instrument will self zero. 6
Page 10
If the supply air is clean the display will show a 0 reading for CO which indicates that a zero setting has been accepted and is now in use. If the supply line has more than a trace of carbon monoxide the following message will appear “ZERO CO”, alternating with a flashing CO alarm light. The instrument is informing the user that it will not zero calibrate because of bad Zero air, and it will use its previously zeroed cal setting. CAUTION: Be sure that the cylinder in use is “zero gas” impurity free air (standard air) when zero initializing. A second method for zeroing is to place zero test gas (Impurity Free Air) in the cal port in the same manner described previously with calibration gas, and the unit will initiate its calibration gas routine. However, the unit expects that 20 PPM CO is being applied in the cal port unless the zero set switch is pushed. Check to see that the message now says “ZEROING” instead of “WAIT” with the 60 second countdown. At any time during the countdown the zero may be pushed to program the unit for zero gas. If the set switch is not pushed, the instrument will fail to calibrate. This is the error protection. After the zero button is depressed, a “ZEROING” message appears. Then release the zero set switch, and the instrument will self zero. MAINTENANCE (qualified technicians only) NOTE: Except for the sensors, all internal parts are to be serviced only by the factory or its authorized agents.
7