Page 1 of RAM 4021-DPX Operation Manual

Manual · GfG Instrumentation

RAM 4021-DPX Operation Manual

Manual20 pagesPDF 291 KB

Operation manual for the GfG Instrumentation RAM 4021-DPX monitor for carbon monoxide and dew point in compressed breathing air.

Download the PDF291 KB

Filed with · 1 model

Also filed with these models on the old site

Summary

What is in it

Operation manual for the GfG Instrumentation ABL/RAM 4021-DPX, a monitor that continuously checks carbon monoxide and dew point in breathing air from a compressor. It describes the unit, the set up modes for alarm points and relays, operation and air flow, calibrating the CO sensor with test gas and zeroing it, dew point sensor auto calibration, replacing the sensors, troubleshooting failed calibrations, accessories and replacement parts, technical data, a diagram, warranty and wiring diagram.

Its parts

  1. Description
  2. Set Up Procedure
  3. Operation
  4. Calibration
  5. Maintenance (qualified technicians only)
  6. Troubleshooting Section
  7. Accessories and Field Replacement Parts
  8. Equipment Technical Data
  9. 4021-DPX Diagram
  10. Warranty
  11. Wiring Diagram

From the manual · read page by page

Key facts from this manual

Some pages are scans or cannot be read; this says only what the readable text supports. Each line names its page in the PDF.

This is the GfG Instrumentation operation manual for the RAM 4021-DPX (also printed as ABL 4021-DPX), a breathing air monitor for carbon monoxide and dew point. The site files it under the 4021, but it covers the 4021-DPX version. It explains setup, start-up, calibration, sensor replacement and troubleshooting, and lists accessories and replacement parts.

What it is
Continuous monitor of carbon monoxide and dew point in breathing air p. 4
Power
110 V AC p. 4
Enclosure
NEMA-4X polycarbonate case, pressurized by the compressor supply line p. 4
Relays
CO alarm relay; dew point high relay (dryer failure alarm); dew point low relay (dryer control) p. 5, 6
CO range
0 to 100 ppm p. 14
Dimensions
7.3 x 5.9 x 4 inches (L x W x H) (185 x 150 x 101 mm) p. 14
Air inlet
½” feed, 80 to 120 psig p. 14
Sample flow rate
0.5 cubic feet of air per hour (scfh) p. 14
CO sensor
5503-020 chemical cell p. 13
Dew point sensor
1702-DPX; expected life two years; calibrates itself p. 5, 13
Calibration kit
7750-001, with calibration connector and 20 ppm test gas 34 L p. 13
Test gases
7802-001 20 ppm carbon monoxide 34 L; 7802-006 impurity free air 34 L (zero gas) p. 13
Other parts
2605-002 high pressure regulator (5,000 PSI); 7750-004 calibration regulator; 1301-002 remote horn, 110 V AC p. 13
CO test gas setting
Factory set to 20 PPM; must match the test gas used p. 6

What this manual walks you through

  • Setting alarm points and relays with the two front buttons p. 5, 6
  • Testing the alarms and relays p. 6
  • Starting up and setting the air flow p. 7
  • Calibrating the CO sensor with test gas p. 8, 9
  • Zeroing the CO sensor p. 9, 10
  • How the dew point sensor calibrates itself p. 9
  • Replacing the dew point sensor p. 11
  • Replacing the CO sensor p. 11
  • Troubleshooting a failed zero or gas calibration p. 12

Good to know

  • Apart from the sensors, internal parts are to be serviced only by the factory or its authorized agents. p. 10
  • Calibrating within the first fifteen minutes after start-up gives a TOO SOON message. p. 7
  • A new CO sensor must settle for at least an hour and be zeroed before a gas calibration. p. 11
  • The test gas must match the CAL setting, or the calibration fails. p. 6
  • Compressor air intakes inside the building should be avoided because they often carry low levels of carbon monoxide. p. 7

Questions this manual answers

How often should the RAM 4021-DPX be calibrated?

A CO calibration with test gas is recommended every 30 days; the dew point sensor calibrates itself. p. 8

What does TOO SOON mean on the 4021-DPX display?

Calibration was tried within the first fifteen minutes after start-up. p. 7

What is the replacement CO sensor part number for the 4021-DPX?

5503-020, the chemical cell sensor. p. 13

What does CAL FAILED mean on the 4021-DPX?

The auto calibration did not finish within 5-6 minutes, for example because of a wrong gas concentration or a faulty sensor or instrument. p. 9

What power does the 4021-DPX need?

110 V AC. p. 4

The whole text · 2 parts

Read the manual here as text

Its text as extracted from the PDF, unedited, in parts; each page opens the PDF at that page.

  1. Pages 1 to 101-10
  2. Pages 11 to 2011-20

The file

Size and text

PublisherGfG Instrumentation
Pages20
PDF291 KB
TextText from the PDF itself; pages 1, 15, 17 and 20 are scans, read by OCR, which may hold reading errors.