Page 11

Dew point sensor replacement The dew point sensor is shipped to you pre-calibrated, no user calibration is required. The dew point sensor will auto calibrate periodically during use. To replace, disconnect the power to the instrument, unplug and remove the sensor from the dew point sensor chamber, replace sensor and reassemble. Carbon monoxide sensor replacement When CO test gas fails to show a gas response during calibration, or the instrument will not complete calibration, a new sensor is required. Most CO sensors will last from one and one-half years to three years. To replace the sensor, disconnect the power to the unit; remove the four corner screws, and the electronics front cover. Next remove the three screws that hold the grey sensor block and unplug the CO sensor from its socket. Replace with a new sensor after being sure that the shorting wire is removed from the new sensor. Reassemble the unit and reconnect to power. Let the new sensor settle in for at least an hour prior to calibrating. A NEW CO SENSOR MUST BE ZEROED When calibrating a new CO sensor, perform a zero operation prior to a gas calibration procedure. Caution needs to be taken that the air is free of carbon monoxide. If the supply line is not CO free, then obtain a tank of Impurity Free Air test gas from GfG (stock number 7802-006). After a successful zero operation is completed, supply 20 PPM cal gas (or another value selected in the Set Up mode) and perform a gas calibration on the instrument (see calibration section). 8

Page 12

TROUBLESHOOTING SECTION The CO zero cal will not set if the there is CO gas present, or if the sensor or electronics are bad. Also, the instrument will not calibrate if the incorrect concentration of CO gas is used, or the sensor is bad or beyond its useable life. With 20 PPM CO cal gas the instrument will read close to 20 during calibration. If the computer cannot auto adjust the readout to 0 (zero) or 20 (cal), the zero or calibration will fail. Zero fail during CO calibration If a failure occurs during CO zero cal, the zero air should be questioned for CO content. Another zero operation with known good zero air should be performed. If the instrument still fails to zero, the unit is malfunctioning, call the factory for further instructions. Calibration gas fail during co calibration With 20 CO PPM gas on the unit the reading should reach 20. If the reading does not reach 20, use the appropriate following procedure. 1.

If the gas reading does not show any increase when the gas is applied, probably the sensor is expired or the test gas is zero concentration. Check hose connections to assure gas is flowing into the sensor chamber.

2.

If the reading is close to 20, a zero calibration procedure with known zero gas will probably correct the problem. If the unit was zeroed with more than a couple PPM of CO, the cal will fail when 20 PPM test gas is applied. 9

Page 13

ACCESSORIES AND FIELD REPLACEMENT PARTS ACCESSORIES 7750-001

Calibration kit (includes calibration connector and 20 ppm test gas 34 L)

2605-002

High pressure regulator (5,000 PSI)

1301-002

Remote horn – 110 V AC

REPLACEMENT PARTS 5503-020

Sensor – chemical cell

1702-DPX

Dew point sensor

7750-004

Calibration regulator

7802-001

Test gas - 20 ppm carbon monoxide 34 L

7802-006

Test gas - impurity free Air 34 L (zero gas)

10

Page 14

EQUIPMENT TECHNICAL DATA Gases detected Sensor Meter scale Response Accuracy Expected sensor life CO sensor warranty Operating temperature range (sample air) Factory set adjustable alarm points Standard inlet pressure Sample flow rate Relays 1 – CO alarm 2 – dew point Power source Dimensions

Carbon monoxide Electro chemical cell 0 to 100 ppm CO 90% max. in 20 sec +/- 1ppm 3 years

Dew point Solid state Thin film type -80 qF to +40 qF N/A +/- 3 qF 2 years

1 year

+32 qF to 105 qF

10ppm CO

(H) +10 qF(L) -4 qF

½” feed - 80 to 120 psig 0.5 cubic feet of air per hour (scfh) 250 V AC/30 V DC @7A 110 V AC @ 1A 7.3 x 5.9 x 4 inches (L x W x H) (185 x 150 x 101 mm)

11

Page 15 read from a scan by OCR; it may hold reading errors

FIGURE 11 4021-DPX DIAGRAM setup up BPRS:

buttons

1/8" NPT INPUT FOR 80-120 PSI SAMPLE ~

Page 16

WARRANTY GFG INSTRUMENTATION, INC., warrants each new electrical product manufactured by it to be free from defective material and workmanship for the purpose intended, for a period of one (1) year from the date of sale to the original purchaser, and agrees to remedy any such defect or to furnish a new part (at the company’s option) in exchange for any part of any product of its manufacture which under normal use, and service disclosed such defect; provided the product is delivered by the purchaser to GFG’S factory, intact, for our examination, with all transportation prepaid to our factory, provided that such examination discloses, in our judgment, that it is thus defective. This warranty does not extend to any products that have been subjected to misuse, neglect, accident, unauthorized modifications, or to use in violation of instructions furnished by us, nor does it extend to products that have been repaired or altered outside of our factory. This warranty is in lieu of all warranties express or implied and no representative or person is authorized to assume for us any other liability in connection with the sale of our products. All implied warranties are limited to the duration of this written warranty. In no event is GfG INSTRUMENTATION, INC., liable for special, incidental, or consequential damages arising from any breach of warranty of product.

13

Page 17 read from a scan by OCR; it may hold reading errors

WIRING DIAGRAM

EXTERNAL POWER FOR AUXILIARY ALARM AUXIALARY ALARM Cloococodod oco|O = '++O0S002E00E : Org22502 202256 5O9sac dt asaas uzl7qSz a3aa5 sili sto ofaoa ft S MO0j76 560000 ofa a 5 a a= Z fy [>, =x* TT = RED [> RECORDER = GUTPUT |_¢) NIC JUMPER = =: RED a oo a oa e 820928 TRANSFORMER =e ras

Page 18

This page is a scan with no text; see it in the PDF.

Page 19

This page is a scan with no text; see it in the PDF.

Page 20 read from a scan by OCR; it may hold reading errors