Spec sheet · Signet
Signet 8550 Spec Sheet
Spec sheet4 pagesPDF 420 KB
All 4 pages are below, each with its text.
Covers · 1 model
The models it covers
From the sheet · read page by page
Key facts from this sheet
Each line names its page in the PDF.
This is a four-page data sheet for the Signet 8550 ProcessPro flow transmitter. It describes the features, specifications, dimensions and mounting options, and lists the order numbers of the dual-channel field and panel mount packages and their accessories.
- Device type
- Converts the signal from frequency and digital (S3L) flow sensors into a 4 to 20 mA signal p. 1
- Features
- 2 or 4 wire power; dual input and output; two optional relays; permanent and resettable totalizers; output simulation p. 1
- Order numbers
- 3-8550-3 field mount package; 3-8550-3P panel mount package; both two inputs, two 4 to 20 mA outputs, two open collectors p. 4
- Power
- 12 to 24 VDC ±10%, regulated p. 2
- Current output
- 4 to 20 mA, isolated, passive, fully adjustable and reversible p. 2
- Display
- Alphanumeric 2 x 16 LCD p. 2
- Keypad
- Sealed 4-key silicon rubber p. 2
- Enclosure
- PBT resin; NEMA 4X/IP65 front face only on panel mount, field mount 100% NEMA 4X/IP65 p. 2
- Operating temperature
- -10 °C to 70 °C p. 2
- Shipping weight
- 0.325 kg (0.7 lb) p. 2
- Integral mount kits
- 3-8051 NPT, Valox; 3-8051-1 NPT, PP; 3-8051-2 NPT, PVDF p. 4
- Mounting accessories
- 3-8050 universal mounting kit; 3-0000.596 heavy duty wall mount bracket and 3-5000.598 surface mount bracket (panel mount only) p. 4
- Connectors, filter
- 3-9000.392 kit for rear cover (3 connectors); 3-9000.392-1 NPT; 3-9000.392-2 PG 13.5; 3-8050.396 RC filter kit for relay use p. 4
- Integral mount sensors
- With the 3-8051-X kit: Signet integral mount sensors 3-8510-XX and 3-8512-XX p. 3
What this sheet walks you through
- Choosing panel, field (pipe, tank, wall) or integral mounting p. 3
- Dimensions of the panel and field versions p. 3
- Sizing the panel cutout p. 3
- Ordering the transmitter packages, mounting kits and connector kits p. 4
- Typical applications such as flow control, filtration regeneration and leak detection p. 1
Good to know
- On the panel mount version only the front face is NEMA 4X/IP65. p. 2
- The splashproof rear cover for the panel mount version is ordered separately. p. 3
- The sensors and fittings in the system overview are sold separately. p. 3
Questions this sheet answers
What does the Signet 8550 flow transmitter do?
It converts the signal from frequency and digital (S3L) flow sensors into a 4 to 20 mA signal for long distance transmission. p. 1
What is the difference between the 3-8550-3 and 3-8550-3P?
The 3-8550-3 is the field mount package and the 3-8550-3P the panel mount package; both have two inputs and two 4 to 20 mA outputs. p. 4
What size panel cutout does the 8550 need?
92 mm X 92 mm (3.62 in X 3.62 in.). p. 3
Which kit mounts the 8550 on a pipe, tank or wall?
The field mount version uses the 3-8050 Universal Mount Kit. p. 3
The file
Size and text
Every page · 4
Read it here
Each page as it is printed, and its words beside it, so a phone can search and copy them.
Page1of 4
Open in the PDF
The text of page 1, from the PDF itself
Signet 8550 Flow Transmitters Check out the 9900 Transmitter for your single channel needs
Features Member of the ProcessPro Family of Instruments ®
- 2 or 4 wire power
- Available with dual input & output
- 4 to 20 mA scaleable outputs
- Permanent & resettable totalizers
- Relay options available
- NEMA 4X enclosure with self-healing window
Panel Mount
Pipe, Wall, Tank and Integral Mount
- Output simulation for complete system testing
MEASURING EQUIPMENT 77CJ
Signet 8550 Flow Transmitters are advanced instruments that convert the signal from frequency and digital (S3L) flow sensors into a 4 to 20 mA signal for long distance transmission. Configuration flexibility is maximized with dual input/output, two optional relays for process control, two packaging options for integral/ pipe mount or panel installation, and scalability for virtually any flow range or engineering unit. State-ofthe-art electronic design ensures long-term reliability, signal stability, and simple user setup and operation.
Applications
- Flow Control and Monitoring
- Filtration or Softener Regeneration
- Effluent Totalization
- Pump Protection
- Feed Pump Pulsing
- Ratio Control
- Water Distribution
- Leak Detection
Page2of 4
Open in the PDF
The text of page 2, from the PDF itself
Specifications
General Compatibility
Signet Flow Sensors with frequency outputs
Accuracy Display
± 0.5% of reading Alphanumeric 2 x 16 LCD
Update Rate Contrast
1 second User selectable, 5 levels
Materials EnclosurePBT resinKeypad
Sealed 4-key silicon rubber
Panel and Case Gasket
Neoprene
Window Electrical Power
Polyurethane coated polycarbonate
Sensor Input Range Sensor Power
Current Output Max. Loop Impedance
Update Rate Accuracy Relay Output Mechanical SPDT contacts Maximum Voltage Rating Hysteresis Maximum 400 pulses/min Open-Collector Output
12 to 24 VDC ±10%, regulated 100 mA max. 0.5 to 1500 Hz 2-wire: 5 VDC ± 1% @ 1.5 mA 3 or 4 wire: 5 VDC ± 1% @ 20 mA Optically isolated from current loop short circuit protected 4 to 20 mA, isolated, passive, fully adjustable and reversible 50 Ω max. @ 12 V 325 Ω max. @ 18 V 600 Ω max. @ 24 V 100 ms ±0.03 mA High, Low, Pulse, Off 30 VDC @ 5 A , 250 VAC @ 5 A resistive load User selectable
High, Low, Pulse, Off Optically isolated, 50 mA max. sink, 30 VDC max. pull-up voltage. User selectable for exiting alarm condition
Hysteresis Maximum 400 pulses/min. Environmental Operating Temperature
-10 °C to 70 °C
Storage Temperature Relative Humidity Enclosure
-15 °C to 80 °C 5 °F to 176 °F 0 to 95%, non-condensing NEMA 4X/IP65 (front face only on panel mount); field mount is 100% NEMA 4X/IP65
14 °F to 158 °F
Shipping Weight 0.325 kg
0.7 lb
Standards and Approvals CE, FCC, UL, CUL RoHS compliant, China RoHS Manufactured under ISO 9001 for Quality and ISO 14001 for Environmental Management and OHSAS 18001 for Occupational Health and Safety
www.gfsignet.com
Page3of 4
Open in the PDF
The text of page 3, from the PDF itself
Dimensions 3-8550-XP 96 mm (3.8 in.)
Field version with universal mount
96 mm (3.8 in.)
82mm (3.23 in.)
Optional Splashproof Rear Cover
41mm (1.6 in.)
92 mm (3.6 in.)
64mm (2.5 in.) 106mm (4.2 in.)
56 mm (2.2 in.) 97mm (3.8 in.)
Panel Mount Signet 8550 Flow Transmitter (Includes mounting bracket and panel gasket)
System Overview
42mm (1.7 in.)
Field Mount - Pipe, Tank, Wall Signet 8550 Flow Transmitter with 3-8050 Universal Mount Kit
Integral Mount Signet 8550 Flow Transmitter with 3-8051-X Integral Mount Kit
+ Signet Sensor 515 525 2100 2507 2537 2540 2552
2000 2536 2551
Signet Sensor 515 2507 2540 525 2536 2537 2000 2551 2552 2100
+ Signet Integral Mount Sensor 3-8510-XX 3-8512-XX
Signet Fittings
All sold separately
Ordering Notes 1) Use the field mount version to directly mount the instrument to the Model 515 or 2536 integral mount sensor. See sensor data sheet for more information. 2) Field mount and sensor can be ordered in a package. See Integral Mount for more information.
3) Panel cutout should be 92 mm X 92 mm (3.62 in X 3.62 in.). 4) An optional splashproof rear cover for the panel mount version can be ordered separately if needed.
Please refer to Wiring, Installation, and Accessories sections for more information. www.gfsignet.com
Page4of 4
Open in the PDF
The text of page 4, from the PDF itself
Ordering Information Mfr. Part No.
CodeInputOutputPowerTwo
Two 4 to 20 mA outputs and two open collectors for Hi, Lo, Pulse or Frequency
2 or 4 wire
Two
Two 4 to 20 mA outputs and two open collectors for Hi, Lo, Pulse or Frequency
2 or 4 wire
ProcessPro Flow Transmitter Field mount package 3-8550-3
159 000 051
Panel mount package 3-8550-3P
159 000 052
Accessories and Replacement Parts Mfr. Part No.
Code
Description
Mounting Accessories 3-8050
159 000 184Universal mounting kit3-8051159 000 187
Flow Sensor integral mount kit, NPT, Valox
3-8051-1
159 001 755
Flow Sensor integral mount kit, NPT, PP
3-8051-2
159 001 756
Flow Sensor integral mount kit, NPT, PVDF
3-0000.596
159 000 641
Heavy duty wall mount bracket (panel mount only)
3-5000.598
198 840 225
Surface mount bracket (panel mount only)
3-8050.395 3-9900.396
159 000 186 159 001 701
Splashproof rear cover (panel mount only) Angle adjustment adapter kit
Liquid Tight Connectors and Other 3-9000.392
159 000 368
Liquid tight connector kit for rear cover (includes 3 connectors)
3-9000.392-1
159 000 839
Liquid tight connector kit, NPT (1 connector)
3-9000.392-2
159 000 841
Liquid tight connector kit, PG 13.5 (1 connector)
3-8050.396
159 000 617
RC filter kit (for relay use)
3-8550.099 Rev F (4/13) © Georg Fischer Signet LLC 3401 Aero Jet Avenue, El Monte, CA 91731-2882 U.S.A. • Tel. (626) 571-2770 • Fax (626) 573-2057 • www.gfsignet.com • e-mail: signet.ps@georgfischer.com Specifications subject to change without notice. All rights reserved. All corporate names and trademarks stated herein are the property of their respective companies.