For the Wedgewood Model 702

Page 1

Model 702 pH Monitor Installation and Operation Manual

Wedgewood Analytical Inc. 4123 E. La Palma Avenue, Suite 200 Anaheim CA 92807-1813 USA Phone: 714.577.5600 Toll Free: 800.835.5474 Fax: 714.577.5688 www.WedgewoodAnalytical.com

Page 2

Model 702 pH Monitor Operation and Installation Manual

Rev 2 July 13, 2005

Table of Contents 1.0 GENERAL INFORMATION MODEL 702 PH MONITOR ..................................................................... 1 2.0 SPECIFICATIONS ................................................................................................................................ 2 2.1 MODEL 702 PH MONITOR ......................................................................................................................... 2 Fig. 1. Model 702 Mounting Dimensions ................................................................................................. 2 2.2 PH PROBE AND INLINE SANITARY SENSOR................................................................................................. 3 Fig. 2. Typical B6xx Inline pH Sensor...................................................................................................... 3 3.0 INSTALLATION .................................................................................................................................... 3 3.1 TRANSMITTER INSTALLATION .................................................................................................................... 3 Fig. 3. Model 702 (AC Input Version) Wiring Diagram ............................................................................ 4 Fig. 4. Model 702 (DC Input Version) Wiring Diagram ............................................................................ 5 Fig. 5. Terminal Preparation and Connector Assembly........................................................................... 5 4.0 PH SENSOR AND ELECTRODE INSTALLATION.............................................................................. 6 Fig. 6. Minimum Angle of pH Electrode Installation on Typical Piping Systems ..................................... 6 Fig. 7. pH Probe Installation and Assembly............................................................................................. 6 5.0 MODEL 702 PH MONITOR OPERATION AND CALIBRATION ......................................................... 7 5.1 MODEL 702 FRONT PANEL CONTROL FUNCTIONS ...................................................................................... 7 Fig. 8. Model 702 Front Panel Controls and Location ............................................................................. 7 5.2 PH CALIBRATION AND STANDARD SOLUTIONS ............................................................................................ 8 5.3 PH SENSOR CALIBRATION ........................................................................................................................ 8 Table 1. Recorder Voltage and Current Output Vs. Selected Full Scale and pH Standards ................. 8 6.0 MODEL 702 TRANSMITTER AND SENSOR MAINTENANCE........................................................... 9 6.1 MODEL 702 PH MONITOR MAINTENANCE .................................................................................................. 9 6.1.1 Removing Module Case ................................................................................................................. 9 Fig. 9. Accessing Rail Mount Enclosure Interior PCB ............................................................................. 9 6.1.2 Fuse Replacement.......................................................................................................................... 9 6.1.3 Changing Input Power .................................................................................................................. 10 6.1.4 Analog Output Interfacing. ............................................................................................................ 10 Fig. 10. Power Supply and Recorder Output PCB and Fuse Location ................................................. 10 6.2 SENSOR MAINTENANCE .......................................................................................................................... 10 7.0 REPLACEMENT PARTS .................................................................................................................... 11

Wedgewood Analytical Inc.

Page i

Page 3

Model 702 pH Monitor Operation and Installation Manual

Rev 2 July 13, 2005

1.0 General Information Model 702 pH Monitor The Model 702 pH Monitor together with a Model B605 Inline sensor (available in 1/2" to 2" line sizes) is specifically designed for sanitary applications. The sensor is CIP and SIP sterilizable. The materials of construction, 316 Stainless Steel, Kynartm and application specified gasket materials are ideally suited for biopharmaceutical processes. The pH electrode supplied, as standard with the Model 702 pH Monitor, is manufactured by Innovative Sensors, Inc. This electrode is specifically designed for biotech applications The standard pH probe supplied is a 12 mm diameter probe with a 120 mm insertion length. Probes by other manufacturers can be used if they are mechanically compatible. The Model 702 is typically used for inline, sanitary pH readings of bulk or transfer bioprocess fluids when pH control is critical. Chromatography, fermentation and bulk storage of biological solutions are examples of operations where maintaining pH is most important. The Model 702 ensures the viability of biological species and products during these operations or any bioprocessing step. The easy electrical interface allows overall process schemes and more sophisticated control strategies to be designed and used. The Model 702 pH Monitor has a user selectable full range of 0 -10 pH, 2 -12 pH or 0 -14 pH. The monitor's voltage and current outputs track the selected pH range. The resolution is +/- 0.01 pH with an accuracy of 0.5% FS at 25oC. Temperature compensation is user selectable to be automatically or manually adjustable from 0 to 100oC. When used in conjunction with the Model 722 combined Conductivity and Temperature Monitor, the signal for the pH temperature compensation is provided by the Model 722 temperature output. The Model 702 is housed in a DIN 35 mm Rail mount enclosure.

Wedgewood Analytical Inc.

Page 1

Page 4

Model 702 pH Monitor Operation and Installation Manual

Rev 2 July 13, 2005

2.0 Specifications 2.1 Model 702 pH Monitor 0-14 +/-0.01pH +/- 0.5% FS at 25oC Manual or Automatic (user selectable), 0-100oC >1015 ohms 0-10, 2-12 or 0-14 pH Full Scale (user selectable) Voltage: 0-2 vdc, 10,000 ohms load Current: 4-20 mA, 600 ohm max. 115/230 VAC +/- 15%, 50-60 Hz., 5.0 watts Optional 20-28VDC 3W 119 mm(4.69”) high, 75 mm(2.91”) deep by 100 mm(3.94”) wide rail mount enclosure constructed of ABS and glass filled polycarbonate Module less than one kilogram pH Range: Resolution: Accuracy: Temp. Compensation: pH Input Impedance: Output Range: Analog Outputs: Power: Dimensions: Weight:

3.94" 100.0 mm

1

10 pH Input

Model 702 pH Monitor 40 30

2.91" 74.0 mm

50

20

60 70 80

10

90 0

59

.1 60

.1 .2

.3

.3 .4

.4 57

100 oC Wedgewood Technology, Inc

61 mV/pH

11

4.69" 119.0 mm

0

.2 58

.5

.5 pH

20

Model 702 1. 2. See 3. Ta ble 4. Below 5. For Terminal 6. Function 7. Selection 8. 9. 10.

Termina l W iring 11. T emp Sig + 12. T emp Sig 13. Rec 0-2 vdc 14. Com 15. Rec 4-20 mA 16. Com L1 115 vac 17. 18. 230 vac 19. 115 vac L2 20.

Jumper 1 - 2 for A uto Temp Comp Jumper 3 - 4 for Man Temp Comp Jumper 5 - 6 for 0-14 pH Full Scale Jumper 7 - 8 for 0-10 pH Full Scale Ter 5,6,7 and 8 Open for 0-12 pH Wedgew ood Technology, Inc I N C O R P O R A T E D

Fig. 1. Model 702 Mounting Dimensions

Wedgewood Analytical Inc.

Page 2

Page 5

Model 702 pH Monitor Operation and Installation Manual

Rev 2 July 13, 2005

2.2 pH Probe and Inline Sanitary Sensor 1/2" to 2" Tri-Clover Sanitary Flange (up to 4” available) 316L Stainless Steel, other materials available Viton, EDPM, Silicon, EPR 0 to 130oC 12 mm Diameter., 120 mm Long, Steam Sterilizable Combination, Gel Filled

Line Size: Flowcell Material: Gaskets: Temperature Range: pH Probe: Type:

3.25"

1/2" THRU 2" TRI-CLOVER SANITARY INLINE SENSORS (Up to 4" available) PROBE HOUSING

OPTIONAL CONDUCTIVITY OR TEMPERATURE SENSOR TYPICAL 12 X 120 MM COMBINATION pH PROBE

Fig. 2. Typical B6xx Inline pH Sensor

3.0 Installation 3.1 Transmitter Installation Inspect the modules, sensors and cable for any signs of shipping damage. Report any visual damage or discrepancies to Wedgewood Analytical and the Shipper. The pH combination electrode is in the manufacturer's original packaging. It is shipped in a 4.0 pH buffer solution and should not be removed from this storage until the electrode is placed in operation. The Model 702 pH Monitor is a DIN 35MM Rail mount enclosure which can be mounted or installed on a vertical or horizontal surface. Refer to figure 1 for mounting dimensions. Mount or install the transmitter on a surface that is not subject to excessive vibration or shock. Allow enough clearance for easy cable entry.

3.2 Electrical Connection All electrical and sensor input connections to the Model 722 are located on the front of the module. A Terminal Wiring Label can be found on the side of the module. Refer to this label and the wiring diagram (figure 3 for AC input model) (figure 4 for DC input model) for electrical installation. When routing cables, we recommend always separating the pH sensor cable from input power wiring. All pH sensor cables supplied are terminated at both the sensor and transmitter ends.

Wedgewood Analytical Inc.

Page 3

Page 6

Model 702 pH Monitor Operation and Installation Manual

Rev 2 July 13, 2005

Connect input power according to type available. Units are configured at the factory as AC or DC versions. Check configuration before connecting. Select 0 -14pH, 2 - 12pH or 0 - 10pH Full Scale and Automatic or Manual temperature compensation by installing a hard wire jumper or dry contact closure across the appropriate terminals. When connecting to analog outputs, we recommend the use of shielded twisted pair cable. Terminate cables ends as shown in figure 5 and connect as in figure 3 / figure 4. S1 Closed, S2 Open, Auto Temp Compensation S1 Open, S2 Closed, Manual Temp Compensation S3 Closed, S4 Open, 0 - 14 pH Full Scale S3 Open, S4 Open, 2 - 12 pH Full Scale S3 Open, S4 Closed, 0 - 10 pH Full Scale S1 S2 S3 S4

1

10

Model 702 pH Monitor pH Input

40 50 60 30 70 80 20

59

.1

60

90

10 0

100 oC Wedgewood Technology, Inc

0

.1

.2

58

61

57

.2

.3

.3

.4

.4 .5

.5 pH mV/pH

11

20

+ VDC Temp Input Temp Input Com.

230 vac Jumper

0-2 vdc pH Rec Output L1

L2 pH Com 4-20 mA pH Output pH Com

115 vac

Jumpers L1

L2

Fig. 3. Model 702 (AC Input Version) Wiring Diagram On models equipped with the 24 VDC option, the input connections may be “daisy chained” to other 700 series equipment. Both pairs of terminals, 17-18, 19-20 are connected together to allow in-out power connections. Refer to figure 4.

Wedgewood Analytical Inc.

Page 4

Page 7

Model 702 pH Monitor Operation and Installation Manual

Rev 2 July 13, 2005

S1 Closed, S2 Open, Auto Temp Compensation S1 Open, S2 Closed, Manual Temp Compensation S3 Closed, S4 Open, 0 - 14 pH Full Scale S3 Open, S4 Open, 2 - 12 pH Full Scale S3 Open, S4 Closed, 0 - 10 pH Full Scale S1

Model 702 Terminal Wiring 1. 11. Temp Sig + 2. 12. Temp Sig See 3. 13. Rec 0-2 vdc Table 4. 14. Com Below 5. For Terminal 15. Rec 4-20 mA 6. 16. Com Function 7. 17. +24VDC Selection DC 8. 18. DC RTN INPUT +24VDC 9. 19. POWER 10. 20. DC RTN Jumper 1 - 2 for Auto Temp Comp Jumper 3 - 4 for Man Temp Comp Jumper 5 - 6 for 0-14 pH Full Scale Jumper 7 - 8 for 0-10 pH Full Scale Ter 5,6,7 and 8 Open for 2-12 pH

S2

S3

S4

1

10

Model 702 pH Monitor pH Input

40 50 60 30 70 80 20

59

.1

60

0

.2 .3 .4

.4

57

100 oC Wedgewood Technology, Inc

.1

.3

90

10

0

.2

58

.5

61

.5 pH mV/pH

11

20

Wedgewood Technology INCORPORATED S h ie ld e d T w is t e d P a ir s 11

12

13

14

15

16

17

18

19

20 D C R T N (O u t )

Temp In +

Analog Input

Analog Outputs

{ { {

+ 2 4 V D C (O u t )

Com

0-2VDC

D C R T N ( In )

Com

+ 2 4 V D C ( In )

4-20mA Com

Fig. 4. Model 702 (DC Input Version) Wiring Diagram The preparation of the wiring before connecting to the module terminals is shown in Figure 5. It is recommended that all wires be terminated with wire pins. The wire and the pins should be soldered after crimping to insure a good electrical connection.

Wire pin, crimp and solder Term. Number ## Insert prepared wire and tighten screw terminal

Fig. 5. Terminal Preparation and Connector Assembly

Wedgewood Analytical Inc.

Page 5