Model 702 pH Monitor Installation and Operation Manual
Manual14 pagesPDF 398 KB
Installation and operation manual for the Wedgewood Analytical Model 702 pH Monitor with B6xx inline sanitary pH sensor.
Covers · 1 model
The models it covers
Summary
What is in it
Wedgewood Analytical manual for the Model 702 pH Monitor, a DIN rail mount monitor used with the B605 inline sanitary pH sensor for bioprocess fluids. It gives specifications for the monitor and sensor, transmitter mounting and AC or DC wiring, pH probe installation, front panel controls, range and temperature compensation selection, calibration with standard buffer solutions, maintenance (case removal, fuses, input power, analog outputs) and replacement parts.
Its parts
- General Information Model 702 pH Monitor
- Specifications
- Installation
- pH Sensor and Electrode Installation
- Model 702 pH Monitor Operation and Calibration
- Model 702 Transmitter and Sensor Maintenance
- Replacement Parts
From the manual · read page by page
Key facts from this manual
Each line names its page in the PDF.
The installation and operation manual (Rev 2, July 13, 2005) for the Wedgewood Model 702 pH Monitor, a DIN rail pH monitor used with inline sanitary pH sensors such as the Model B605. It covers specifications, mounting, wiring of the AC and DC versions, probe installation, front panel calibration with buffer solutions, maintenance, fuse replacement and replacement parts.
- Type
- pH monitor for inline sanitary pH readings, used with a Model B605 inline sensor p. 3
- pH ranges
- User selectable 0-10, 2-12 or 0-14 pH full scale p. 3
- Resolution and accuracy
- +/- 0.01 pH resolution; 0.5% FS accuracy at 25°C p. 3
- Temperature compensation
- Automatic or manual, 0 to 100°C; the Model 722 temperature output can supply the signal p. 3
- Analog outputs
- 0-2 vdc (10,000 ohms load) and 4-20 mA (600 ohm max) p. 4
- Power
- 115/230 VAC +/- 15%, 50-60 Hz, 5.0 watts; optional 20-28VDC, 3W p. 4
- Enclosure
- DIN 35 mm rail mount, ABS and glass filled polycarbonate; under one kilogram p. 3, 4
- Standard pH probe
- Innovative Sensors combination probe, 12 mm diameter, 120 mm insertion length, gel filled p. 3, 5
- Inline sensor
- 1/2" to 2" Tri-Clover sanitary flange (up to 4" available), 316L stainless steel p. 5
- Selection terminals
- Terminals 1-2: automatic temperature compensation; 3-4: manual; 5-6: 0-14 pH; 7-8: 0-10 pH p. 9, 10
- Output terminals
- 11-12 temperature signal; 13 0-2 vdc recorder; 14 common; 15 4-20 mA recorder; 16 common p. 4
- Fuses
- AC version 80mA 250V (Schurter MSF-250 fast acting, MSF-034-6002); DC version 160mA 250V p. 12, 13
- DC power chaining
- DC units can be daisy chained to other 700 series equipment through terminals 17-18 and 19-20 p. 6
What this manual walks you through
- Mounting the monitor on a DIN rail p. 5
- Wiring the AC and DC input versions p. 6, 7
- Preparing wires with crimped and soldered pins p. 7
- Positioning the sensor in the piping p. 8
- Installing the pH probe in its housing p. 8, 9
- Using the front panel controls p. 9
- Calibrating with 7.00, 4.00 and 10.00 pH buffers p. 10
- Calculating the expected voltage and current outputs p. 11
- Removing the module case p. 11
- Replacing the fuses p. 12
Good to know
- Store the pH electrode in 4.0 pH buffer solution whenever it is not in use. p. 8
- The pH electrode must stay wet at all times. p. 8
- Units are factory configured as AC or DC versions; check before connecting power. p. 6
- A response slower than 30 seconds during calibration means the electrode needs cleaning or replacement. p. 10
- The 0-2Vdc recorder output is limited to 5mA, so use it with high impedance instruments. p. 12
Questions this manual answers
How do I calibrate a Wedgewood Model 702 pH monitor?
Fill the rinsed sensor with 7.00 pH buffer and set the Delta pH control, then 4.00 pH buffer and set the mV/pH control, repeat, and check with 10.00 pH buffer. p. 10
What fuse does the Model 702 use?
The AC version uses an 80mA 250V fuse and the DC version a 160mA 250V fuse. p. 13
How do I set the pH range on the Model 702?
Fit a jumper or dry contact across terminals 5 and 6 for 0-14 pH, or 7 and 8 for 0-10 pH. p. 10
What should the Model 702 outputs read in 7.00 pH buffer on the 0-10 pH range?
1.400 VDC and 15.20 mA. p. 11
How should the Model 702 pH electrode be stored?
In a 4.0 pH buffer solution whenever it is not in use; it must stay wet. p. 8