For the Wedgewood Model 702

Page 8

Model 702 pH Monitor Operation and Installation Manual

Rev 2 July 13, 2005

4.0 pH Sensor and Electrode Installation The sensor should be located in the piping system so that the pH probe is at least 15 degrees above a horizontal plane. The pH probe contains liquid solutions that must remain in contact with their respective junctions hence the reason for placement of the sensor so that the pH probe is above the horizontal plane.

Vertical Pipe Mounting 15° pH Probe

Horizontal Pipe Mounting 15° pH Probe

Fig. 6. Minimum Angle of pH Electrode Installation on Typical Piping Systems Since the pH electrode must remain wetted at all times, the sensor should be positioned so that the flowcell will be filled with a liquid during standby or that the pH electrode is easily removed for storage.

M4x10MM SKT HD AND WASHER (4 REQ,D)

PROBE HOUSING

TYPICAL 12 X 120 MM COMBINATION pH PROBE

Fig. 7. pH Probe Installation and Assembly The pH electrode is not shipped mounted, but protected in the original Manufacturer's packaging. The electrode is immersed in a buffer solution of 4.0 pH. Whenever the pH electrode is not in use, storage in a 4.0 pH buffer solution is advised. The installation of the pH electrode and its hardware are done insitu. Before attempting installation please refer figure 7. Carefully clean all parts prior to assembly. To install the probe, use the following procedure. 1. Slip an 'O' Ring over the electrode up to the back of the threaded cable connector. 2. Carefully insert the electrode into the housing.

Wedgewood Analytical Inc.

Page 6

Page 9

Model 702 pH Monitor Operation and Installation Manual

Rev 2 July 13, 2005

3. Hand tighten the threaded connector end of the probe into the probe housing, securing the electrode in place. 4. Slip the second 'O' Ring over the electrode into the seal seat at the free end of the probe housing. Position the 'O' ring in the seat. 5. Insert the electrode and the housing in the pH probe adapter and tighten clockwise until the housing seats. Note: All seals are designed for controlled compression on the 'O' rings. Therefore only hand tightening is needed.

5.0 Model 702 pH Monitor Operation and Calibration 5.1 Model 702 Front Panel Control Functions The front panel controls and their location are shown in Figure 8.

5

6

7

8

1

10 pH Input

Model 702 pH Monitor 40 50 60 30 70 20 10 0

.1

90

.4

100

11

0

.1

.2 60

.3 oC W edgewood Technology, Inc

1

59 58

80 57

61 mV/pH

.2 .3 .4 .5

.5 pH

20

4 3 2 Fig. 8. Model 702 Front Panel Controls and Location 1. The BNC input connector for the pH probe. 2. The 'oC' control manually adjusts for temperature compensation to cancel the Nerstian effects on the pH signal. Set the operating temperature on this dial. 3. The 'mV/pH' adjusts the slope response of the pH probe. Adjustment should be made with the probe immersed in either a 4.00 pH or a 10.0 pH solution. 4. The 'Delta pH' control compensates for the probe response at 7.00 pH when the 'oC' (Temperature control is at 25oC and the mV/pH control is at 59.2 mV/pH. Compensation configured for Manual) 5. Connect Term. 1 and 2 for Automatic Temperature compensation.

Wedgewood Analytical Inc.

Page 7

Page 10

Model 702 pH Monitor Operation and Installation Manual

Rev 2 July 13, 2005

6. Connect Term. 3 and 4 for Manual Temperature compensation. 7. Connect Term. 5 and 6 for 0-14 pH Full Scale. 8. Connect Term. 7 and 8 for 0-10 pH Full Scale. 9. Term. 5, 6, 7, and 8 open for 2-12 pH Full Scale. 5.2 pH Calibration and Standard Solutions Before attempting to calibrate the pH sensor, using standard buffer solutions, clean the flowcell with distilled water and cap one end. To calibrate the sensor use the following procedure. 5.3 pH Sensor Calibration Note: Each buffer solution is specified from the manufacturer with a temperature correction pH value. It is important to use this corrected value when calibrating the instrument. Standard values are based on a temperature of 25° C. 1. Rinse the sensor and flowcell with distilled water several times. 2. Rinse the sensor with a buffer solution of 7.00pH to remove any remaining distilled water. Discard the buffer used to rinse the sensor. Do this several times, then fill the sensor with this buffer. Adjust the 'Delta pH' (Standard Adjust) control until the voltage/current analog outputs read as in Table 1. Note: At temperatures other than 25° C, a correction is required for the buffer solutions (manufacturer supplied). See output calculation for voltage/current values. pH Standard 4.00pH 7.00pH 10.00pH

0-10pH Range VDC mA 0.800 10.40 1.400 15.20 2.000 20.00

2-12pH Range VDC mA 0.400 7.20 1.000 12.00 1.600 16.80

0-14pH Range VDC mA 0.571 8.57 1.000 12.00 1.429 15.43

Table 1. Recorder Voltage and Current Output Vs. Selected Full Scale and pH Standards 1. Pour out the buffer solution and rinse the sensor with a pH 4.00 buffer solution. Do this several times, then fill the sensor with this buffer and adjust the mV/pH control until the voltage/current analog outputs read as in Table 1. 2. Repeat steps 2 and 3 until no further adjustments are necessary. 3. When the above steps are completed, rinse the sensor with a buffer solution of 10.00 pH. Do this several times and then fill the sensor with this solution. If the reading deviates from that in Table 1 significantly, pH electrode cleaning or replacement is indicated. 4. The response time of the measurements should be less than 30 seconds. If during the above procedure, the response of the electrode is longer, pH electrode cleaning or replacement is indicated. 5. This completes the calibration of the Model 702 pH Monitor with the pH probe. The voltage/current analog output values may be calculated as follows:

Wedgewood Analytical Inc.

Page 8

Page 11

Model 702 pH Monitor Operation and Installation Manual

Solution Value ( pH(Temp Corrected)

- Min Scale

(Full Scale - Min Scale)

Rev 2 July 13, 2005

) X 2 = VDC Output

(VDC Output X 8) + 4 = mA Output Example: 7.00pH Solution, Range 0-10pH ( 7.00 - 0 )

x 2 = 1.400 VDC

( 10 - 0 ) ( 1.400 x 8 ) + 4 = 15.20mA

6.0 Model 702 Transmitter and Sensor Maintenance 6.1 Model 702 pH Monitor Maintenance The Model 702's electronics have been designed and manufactured for a long and trouble free life. Failure of the system in most cases is due to electrical connection failure. If, however, it is determined that one of the printed circuit boards has failed, return the defective unit to Wedgewood Analytical for repair and re-calibration. 6.1.1 Removing Module Case The rail mount terminals and top lid are fastened to the enclosure body with four (4) screws and side friction latches. To remove the enclosure lid, remove the 4 screws and then with a blade screw-driver, depress the latch and lifting the lid until the latch is disengaged. Repeat for the opposite side. When the lid is free, carefully remove the enclosure box.

Typical DIN 35 mm Rail Mount Enclosure

Top Mounting Screws (4 Pl.)

Module Latch Access Hole (each side)

Fig. 9. Accessing Rail Mount Enclosure Interior PCB 6.1.2 Fuse Replacement

Wedgewood Analytical Inc.

Page 9

Page 12

Model 702 pH Monitor Operation and Installation Manual

Rev 2 July 13, 2005

The instrument's fuses are located on the main PCB. Fuse failure is normally caused by improper voltage selection and/or faulty wiring. Replace the fuse with the type and rating installed. If after fuse replacement and if the installation is correct, the fuse fails again, this is an indication that an internal fault has occurred. Return the instrument to Wedgewood for repair

6.1.3 Changing Input Power For AC powered instruments, refer to the Terminal Identification label on the enclosure side panel for wiring of 115Vac and 230Vac. DC powered units will operate with a supply Voltage of between 20 and 28Vdc. 6.1.4 Analog Output Interfacing. A 0-2Vdc output is available and independent of the 4 - 20mA analog output. This allows both current and voltage input devices to be used simultaneously. The voltage output is referenced to circuit common. The current limit of the 0-2Vdc recorder output is 5mA. Therefore the output should be used with high impedance measuring instruments (i.e. 10,000 ohm/volt minimum).

Schurter Microfuse, MSF-250, Fast Acting Part No. MSF-034-6002, 80mA, 250v 20

19

18

JP101

17

16

15

1 4

13

12

Fuses F101

11

Edge Card Connector

VIN C119

F102

Q101 C101 C118

6 4

4-20 mA ZERO ADJ

3 1

U106

4-20 mA Recorder Zero Adj. (4.00 mA)

R106

+10VR R105

C117

VOUT

SIG_OUT

T101

C107

SIG_IN

C106

C103

C102

AC

Main PCB Power Power to and Signal From pH PCB

C116 C109 JP103

C110

GND C112

C108 L101

AC C104

D101

C105

U103

L102

C114 GNDI

C113 C111 U104

C115

Fig. 10. Power Supply and Recorder Output PCB and Fuse Location

6.2 Sensor Maintenance

Wedgewood Analytical Inc.

Page 10

Page 13

Model 702 pH Monitor Operation and Installation Manual

Rev 2 July 13, 2005

Under normal operating conditions, the Inline Sensors require no routine maintenance. What maintenance is required is usually restricted to 'housekeeping'. Inspect from time to time for signs of leakage, cable fraying, or any abnormal conditions. If sensor cleaning and/or maintenance is required, remove the sensor from service and perform the needed maintenance.

7.0 Replacement Parts Description pH Probes and Cables Standard ISI pH probe, Quick Disconnect Standard ISI Cable, 10 feet long Standard ISI Cable, 25 feet long Ingold DPAS pH Probe, 12 mm Diameter by 120 mm Long Ingold Cable, 1 meter, BNC Connector Ingold Cable, 2 meter, BNC Connector Ingold Cable, 3 meter, BNC Connector Ingold Cable, 5 meter, BNC Connector Ingold Cable ,10 meter, BNC Connector Ingold Cable ,20 meter, BNC Connector

1600-1200-00 1957-0130-00 1957-0140-00 1600-1002-00 A050-0600-01 A050-0600-02 A050-0600-03 A050-0600-05 A050-0600-10 A050-0600-20

Flowcell Seal Kits Silicon Rubber Seal Kit Viton Seal Kit Kalrez Seal Kit EPDM Seal Kit

A000-0605-00 A000-0605-01 A000-0605-04 A000-0605-05

Fuses Fuse, AC Version, 80mA - 250V Fuse, DC Version, 160mA - 250V

1678-0018-80 1678-0019-00

Wedgewood Analytical Inc.

Wedgewood Part Number

Page 11

Page 14

Model 702 pH Monitor Operation and Installation Manual

Rev 2 July 13, 2005

WARRANTY Wedgewood Analytical Inc. warrants its products to be free from defects in workmanship and materials. Wedgewood's liability is limited to replacing the instrument or any part thereof that is returned by the original purchaser, transportation paid, to the factory within one (1) year after the date of shipment, provided that a defect existed under proper and normal use. Wedgewood Analytical Inc. shall not be liable for consequential or incidental damages.