The text of a manual · part 5 of 9

Thermocouple/Millivolt Input Module (Cat. No. 1771-IXE Series D) User Manual, pages 41 to 50

Pages 41 to 50 of 85

The original document is the PDF (708 KB). This is its text as extracted, unedited; tables and drawings read best in the PDF, and each page heading opens it at that page.

For the Allen-Bradley 1771-IXE

Page 41

Chapter

6

Chapter Objective Tools and Equipment Tool or Equipment

In this chapter we tell you how to calibrate your module.

In order to calibrate your input module you will need the following tools and equipment: Description

Model/Type

+ ,(%.-$('

(+ '-

'-

- !(+ ,,(+,

- (+ -

Calibrating Your Input Module

Available from:

0

The thermocouple/millivolt input module is shipped already calibrated. If it becomes necessary to recalibrate the module, you must calibrate the module in an I/O chassis. The module must communicate with the processor and industrial terminal. Before calibrating the module, you must enter ladder logic into the processor memory, so that you can initiate BTWs to the module, and the processor can read inputs from the module. Calibration can be accomplished using either of two methods:

• auto–calibration • manual calibration

About Auto-Calibration

Auto–calibration calibrates the input by generating offset and gain correction values and storing them in EEPROM. These values are read out of EEPROM and placed in RAM memory at initialization of the module. The auto–calibration routine operates as follows:

• Whenever a block transfer write (BTW) of length 28 is performed to the module (any time after the module has been powered up), it interrogates word 28 for a request for auto–calibration. • The request can be for the following: offset calibration, gain calibration, save operation (save to EEPROM). When using auto–calibration, write transfer calibration words 20 through 27 must contain zeroes.

- -1 1

Page 42

6–2

Module Calibration

Performing Auto-Calibration

Calibration of the module consists of applying 0.000mV across each input channel for offset calibration, and +100.000mV across each input channel for gain correction.

Offset Calibration Normally all inputs are calibrated together. To calibrate the offset of an input, proceed as follows: 1. Apply power to the module. 2. Connect shorting links, or apply 0.000mV across each input channel on the 1771–WI field wiring arm as shown in Figure 6.1. Figure 6.1 Shorting Inputs for Offset Calibration

* 1(-,

1 %-/ $ "' "' ,,$*

1

- ,-1 20$

$ "' (,.21 -/ ..*3 "/-00 $ "' (,.21 "'

*$ *$ *$ *$ *$ *$ *$ *$

20$# *$ *$ *$ *$ *$ *$ *$ *$

-2,"1(-,

*

..*3

/+

3. After the connections stabilize, request the offset calibration by setting bit 00 in block transfer write word 28 and sending a block transfer write (BTW) to the module. Refer to Table 6.A. When the BTW is sent, all channels are calibrated to 0.000mV.

1(-, 4 3

Page 43

Module Calibration

6–3

Table 6.A Write Block Transfer Word 28

Word Bit

17 16 15 14 13 12 11 10 07 06 05 04 03

-

02

01

00

-

-#!,! -(

NOTE: Normally, all channels are calibrated simultaneously (bits 10–17 of word 28 are octal 0). To disable calibration on any channel, set the corresponding bit 10 through 17 of word 28. 4. Queue block transfer reads (BTRs) to monitor for offset calibration complete and any channels which may have not calibrated successfully. Refer to Table 6.B. Table 6.B Read Block Transfer Word 13

Word Bit

17 16 15 14 13 12 11 10

07

06

05 04 03

02

01

00

-

-(- .,!

-(

5. Proceed to Gain Calibration below.

Gain Calibration Calibrating gain requires that you apply +100.000mV across each input channel. Normally all inputs are calibrated together. To calibrate the gain of an input, proceed as follows: 1. Apply +100.000mV across each input channel as shown in Figure 6.2.

1 1

Page 44

6–4

Module Calibration

Figure 6.2 Applying 100.00mV for Gain Calibration

-

1

'(, -+!

-')-, -')-, -')-, -')-, -')-, -')-, -')-, -')-,

-+! -')-, -')-, -')-, -')-, -')-, -')-, -')-, -')-,

+

*&

))%.

2. After the connections stabilize, request the gain calibration by setting bit 01 in BTW word 28 and sending a block transfer write (BTW) to the module. Refer to Table 6.A. When the BTW is sent, all channels are calibrated to +100.00mV. NOTE: Normally, all channels are calibrated simultaneously (bits 10–17 of word 28 are octal 0). To disable calibration on any channel, set the corresponding bit 10 through 17 of word 28. 3. Queue BTRs to monitor for gain calibration complete and any channels which may not have calibrated successfully.

Save Calibration Values If any ”uncalibrated channel” bits (bits 10–17 of word 13) are set, a save cannot occur. Auto–calibration should be performed again, starting with offset calibration. If the module has a faulty channel, the remaining functioning channels can be calibrated by inhibiting calibration on the faulty channel. - /

Page 45

Module Calibration

6–5

The module can be run with the new calibration values, but will lose them on power down. To save these values, proceed as follows: 1. Request a ”save to EEPROM” by setting bit 02 in BTW word 28 and sending the BTW to the module. Refer to Table 6.A. 2. Queue BTRs to monitor for ”save complete”, ”EEPROM fault” and ”calibration fault.” An EEPROM fault indicates a nonoperative EEPROM; a calibration fault indicates at least one channel was not properly offset or gain calibrated and a save did not occur.

Performing Manual Calibration

You calibrate each channel by applying a precision voltage to the input terminals, comparing correct with actual results, and entering correction into the corresponding calibration word for that channel. The correction takes affect after it is transferred to the module by the corresponding BTW instruction in your ladder diagram program. Always start with offset adjustment followed by gain adjustment. Before calibrating the module, you must enter ladder logic into processor memory, so that you can initiate write block transfers to the module, and the processor can read inputs from the module. Write transfers will contain calibration values in words 20 through 27 for the channel you are calibrating. Use a precision voltage source, such as Data Precision 8200 or equivalent, for your calibration input voltage.

Setting Channel Offset Calibration 1. Select the millivolt range and binary data format. (BCD could be used, but it cannot display values larger than 100mV required during gain calibration.) 2. Apply 0.000 millivolts to the channel input as shown in Figure 6.3.

-

Page 46

6–6

Module Calibration

Figure 6.4 Shorting Inputs for Offset Calibration

* 1(-,

1 %-/ $ "' "' ,,$*

1

- ,-1 20$

$ "' (,.21 -/ ..*3 "/-00 $ "' (,.21 "'

-2,"1(-, *$ *$ *$ *$ *$ *$ *$ *$

20$# *$ *$ *$ *$ *$ *$ *$ *$

*

..*3

/+

3. Observe the input value read by the processor (word 4 of the BTR file for channel 1). It should be 0000. 4. Multiply the difference between your observed value and 0.000 by 3.0933. Determine the magnitude and sign of the required correction. You can adjust the correction up to +127 binary counts (+410.56µV). A negative correction means that the reading was too high and you want to subtract a corrective amount from that reading. A positive correction means that the reading was too low and you want to add a corrective amount to that reading. 5. Enter the magnitude and sign of the correction in binary code into the upper (offset correction) byte of the calibration word for that channel. (BTW file, word 20, bits 17–10 for channel 1.)

1(-, 4 3

Page 47

Module Calibration

6–7

For example, if the observed value was 17, enter –53 [(0 – 17) x 3.0933 = –53] in signed magnitude binary into the upper byte of the calibration word for that channel. Enter 10110101 in bits 17–10 of word 20. The lower byte will remain zero at this time. 6. Repeat steps 3 through 5 for each of the remaining input channels. 7. Initiate a write block transfer to send the corrections to the module. The input value read by the processor should now be 0000 for all channels.

Setting Channel Gain Calibration 1. Now set the precision voltage source for +100.000 millivolts. Allow sufficient time (at least 10 seconds) for the input filter and voltage source to settle. Figure 6.5 Applying 100.000mV for Gain Calibration

-

1

'(, -+!

-')-, -')-, -')-, -')-, -')-, -')-, -')-, -')-,

-+! -')-, -')-, -')-, -')-, -')-, -')-, -')-, -')-,

+

*&

))%.

- /

-

Page 48

6–8

Module Calibration

2. Record the input value read by the processor in the BTR file (word 4 for channel 1). Determine the percentage difference from 10000 and the sign of the correction. You can adjust the correction up to +0.19379%. A negative correction means that the reading was too high and you want to subtract a corrective amount from that reading. A positive correction means that the reading was too low and you want to add a corrective amount to that reading. If programming in BCD, the upper limit for the display is A000. If the overrange bit is set, turn back the voltage reference until the overrange turns off. Use the difference for the calculation. For example, if the observed value was 10014, then 10000–10014 = –14, and –14 divided by 10000 = –0.14%. 3. Using the following table, select gain correction values that most nearly add up to the percentage that you determined in step 1. Select a value only once. Bit

Value

-

-

-

Enter the bit code representing the sum of the corrections into the lower byte (gain correction) of the calibration word for that channel. For example, to attain the value of 0.140%, you would add: Percentage

Bit Number

-

-

Enter 11011100 in the lower byte of the calibration word for that channel. This entry would set bits 07 (sign) and 06, 04, 03 and 02 which is –0.1403807, very close to the required –0.14. Remember to keep the upper byte the same as it was from step 5.

Page 49

Module Calibration

6–9

4. Repeat the above steps 2 and 3 for channels 2 through 8. 5. Initiate a write block transfer to send the corrections to the module. The input value read by the processor should now be 10000 (A000 for BCD) for all channels. 6. If the correction changes the result in the wrong direction, change the sign and reenter it. Important:

Chapter Summary

If the % correction required is larger than +0.19379, check your reference voltage. If the reference voltage is correct, perform auto–calibration.

In this chapter, you learned how to calibrate your input module.

-

Page 50 read from a scan by OCR; it may hold reading errors

6-10 Module Calibration

Publication 1771-6.5.130 - May 1999