Page 71
6.4
Input #1 Set Up Wiring
Thermocouple inputs using an ice bath
Refer to Figure 6-2 and wire the controller according to the procedure given in Table 6-3. Table 6-3
Set Up Wiring Procedure for Thermocouple Inputs Using An Ice Bath
Step
Action
1
Connect the copper leads to the calibrator.
2
Connect a length of thermocouple extension wire to the end of each copper lead and insert the junction points into the ice bath.
3
Connect the thermocouple extension wires to the terminals for Input #1. See Figure 6-2.
4
Connect a cold junction resistor to terminals 25 and 27 for Input #1. See Figure 6-2.
Figure 6-2
Wiring Connections for Thermocouple Inputs Using an Ice Bath
+ –
10
1
L1
11
2
L2/N
12
3
22
13
4
23
14
5
24
15
6
25
16
7
26
17
8
27
9
T/C Extension Wire C/J Resistor
+ –
Millivolt Source
Copper Leads Ice Bath 24205
Continued on next page
1/01
UDC3300 Limit Controller Product Manual
61
Page 72
6.4
Input #1 Set Up Wiring, Continued
Thermocouple inputs using a precision resistor
Refer to Figure 6-3 and wire the controller according to the procedure given in Table 6-4. Table 6-4
Set Up Wiring Procedure for Thermocouple Inputs Using a Precision Resistor
Step
Action
1
Connect the copper leads to the calibrator.
2
Disconnect the cold junction resistor.
3
Install a 500-ohm precision resistor across terminal 25 (R) and terminal 27 (–). See Figure 6-3.
4
Subtract the millivolt value for 77°F (25°C) from the zero and span value for your range (see Table 6-1 for zero and span values) and use the adjusted value when calibrating.
Figure 6-3
Wiring Connections for Thermocouple Inputs Using a Precision Resistor
R
+ –
10
1
L1
11
2
L2/N
12
3
22
13
4
23
14
5
24
15
6
25
16
7
26
17
8
27
Copper Leads
9 Precision 500 Ohm Resistor
– + 1
1
Millivolt Source
Caution: The accuracy of the controller is directly affected by the accuracy of this resistor. At a minimum, use a 0.1%, 25ppm resistor. 22628
Continued on next page
62
UDC3300 Limit Controller Product Manual
1/01
Page 73
6.4 RTD inputs
Input #1 Set Up Wiring, Continued Use the copper leads and connect the calibrator to the rear terminals of Input #1. See Figure 6-4. Figure 6-4
Wiring Connections for RTD (Resistance Thermometer Device)
R
+ –
10
1
L1
11
2
L2/N
12
3
22
13
4
23
14
5
24
15
6
25
16
7
26
17
8
27
9
Decade Resistance Box Copper Leads equal length
22629
Continued on next page
1/01
UDC3300 Limit Controller Product Manual
63
Page 74
6.4
Input #1 Set Up Wiring, Continued
Radiamatic, millivolts, or volts (except 0 to 10 volts) inputs
Use the copper leads and connect the calibrator to the rear terminals of Input #1. See Figure 6-5. Figure 6-5
Wiring Connections for Radiamatic, Millivolts, or Volts (except 0 to 10 Volts)
+ –
10
1
L1
11
2
L2/N
12
3
22
13
4
23
14
5
24
15
6
25
16
7
26
17
8
27
9
– Millivolt/Volt + Source Copper Leads
22630
Continued on next page
64
UDC3300 Limit Controller Product Manual
1/01
Page 75
6.4
Input #1 Set Up Wiring, Continued
0 to 10 volt inputs
Use the copper leads and connect the calibrator to the rear terminals of Input #1. See Figure 6-6. Figure 6-6
Wiring Connections for 0 to 10 Volt Inputs
100K
10
1
L1
11
2
L2/N
12
3
22
13
4
23
14
5
24
15
6
25
16
7
26
17
8
100K
27
9
– Copper Leads
+
0 to 10 Volt Source
22631
Continued on next page
1/01
UDC3300 Limit Controller Product Manual
65
Page 76
6.4
Input #1 Set Up Wiring, Continued
4 to 20 mA inputs
Use the copper leads and connect the calibrator to the rear terminals of Input #1. See Figure 6-7. Figure 6-7
Wiring Connections for 4 to 20 mA Inputs
250Ω resistor assy
Copper Leads
66
+ –
10
1
L1
11
2
L2/N
12
3
22
13
4
23
14
5
24
15
6
25
16
7
26
17
8
27
9
– 4 to 20 mA + Source
UDC3300 Limit Controller Product Manual
22632
1/01
Page 77
6.5
Input #1 Calibration Procedure
Introduction
Apply power and allow the controller to warm up for 15 minutes before you calibrate. Read “Set Up Wiring” before beginning the procedure. Make sure you have “LOCKOUT” set to “NONE.” See Section 3 – Configuration. CAUTION For linear inputs, avoid step changes in inputs. Vary smoothly from initial value to final 100% value. Procedure
Table 6-5 Step 1
The Calibration procedure for Input #1 is listed in Table 6-5. Input #1 Calibration Procedure Description
Press
Enter Calibration Mode
Action Upper Display
SET UP until you see
CALIB Lower Display INPUT 1
You will see:
FUNCTION
Upper Display DISABL Lower Display CAL IN1
The calibration sequence is enabled and you will see: Upper Display BEGIN Lower Display CAL IN1
At the completion of the sequence, the selection automatically reverts to disable. 2
Calibrate 0%
You will see:
FUNCTION
Upper Display APPLY Lower Display IN1 ZERO
Adjust your calibration device to an output signal equal to the 0% range value for your particular input sensor. See Table 6-1 for Voltage or Resistance equivalents. Wait 15 seconds, then go to the next step.
Continued on next page
1/01
UDC3300 Limit Controller Product Manual
67
Page 78
6.5
Input #1 Calibration Procedure, Continued
Procedure, continued
Table 6-5 Step 3
Input #1 Calibration Procedure, Continued Description
Press
Calibrate 100%
Action You will see:
FUNCTION
Upper Display APPLY Lower Display IN1 SPAN
Adjust your calibration device to an output signal equal to the 100% range value for your particular input sensor. See Table 6-1 for Voltage or Resistance equivalents. Wait 15 seconds, and If… you are calibrating a Thermocouple input you are calibrating other than a Thermocouple input
4
Check the Cold Junction Temperature
FUNCTION
Go to step 5
The calculations for zero and span are now stored and you will see: Upper Display
ATTENTION The accuracy of the controller is directly affected by the accuracy of this value. Change this value only if the zero and span calibration procedures did not bring the controller within the specified accuracy requirements.
Then… Go to step 4
Lower Display
The cold junction temperature at the rear terminals
C-J TEMP
The value in the upper display is in the tenths of a degree. It is the current reading of the temperature as measured at the thermocouple terminals and recognized by the controller. You can change this value, if it is in error, using the s or t key.
ATTENTION When calibrating T/C inputs using a precision resistor, calibrate the cold junction as 77°F(25°C). 5
Exit the Calibration Mode
FUNCTION
68
The controller will store the calibration constants and exit calibration mode.
UDC3300 Limit Controller Product Manual
1/01
Page 79
Section 7 – Troubleshooting / Service 7.1
Overview
Introduction
Instrument performance can be adversely affected by installation and application problems as well as hardware problems. We recommend that you investigate the problems in the following order: • Installation related problems • Application related problems • Hardware and software related problems and use the information presented in this section to solve them. If a replacement of any part is required, follow the procedures listed under “Replacement Procedures”.
!
WARNING - SHOCK HAZARD INPUT CALIBRATION MAY REQUIRE ACCESS TO HAZARDOUS LIVE CIRCUITS, AND SHOULD ONLY BE PERFORMED BY QUALIFIED SERVICE PERSONNEL. MORE THAN ONE SWITCH MAY BE REQUIRED TO DEENERGIZE UNIT BEFORE CALIBRATION.
What’s in this section?
The following topics are covered in this section. Topic
See Page
7.1
Overview
69
7.2
Troubleshooting Aids
71
7.3
Power-up Tests
73
7.4
Status Tests
74
7.5
Background Tests
76
7.6
Controller Failure Symptoms
78
7.7
Troubleshooting Procedures
79
7.8
Parts Replacement Procedures
83
7.9
Maintenance (Cleaning)
92
Continued on next page
1/01
UDC3300 Limit Controller Product Manual
69
Page 80
7.1
Overview, Continued
Installation related problems
Read the Installation section in this manual to make sure the UDC3300 has been properly installed. The installation section provides information on protection against electrical noise, connecting external equipment to the controller, and shielding and routing external wiring. ATTENTION
System noise induced into the controller will result in diagnostic error messages recurring. If the diagnostic error messages can be cleared, it indicates a “soft” failure and is probably noise related. If system noise is suspected, completely isolate the controller from all field wiring. Use calibration sources to simulate PV and check all controller functions; i.e. Inputs, Alarms, etc. Application related problems
Review the application of the controller; then, if necessary, direct your questions to the local sales office.
Hardware and software related problems
Use the troubleshooting error message prompts and controller failure symptoms to identify typical failures which may occur in the controller. Follow the troubleshooting procedures to correct them.
70
UDC3300 Limit Controller Product Manual
1/01