Page 31
Section 2-2
Expansion Units Input Specifications Number of inputs
Data inputs: 8; control inputs: 4
Input voltage
12 to 24 VDC +10%
ON current
3 to 15 mA
ON voltage
8.8 V max.
OFF current
0.1 mA max.
Input OFF voltage
4.5 V min.
ON delay
RESET input: 10 ms max.; other inputs: 0.5 ms max.
OFF delay
RESET input: 15 ms max.; other inputs: 0.7 ms max.
DI 0 to 7
Parallel data inputs
RESET
Resets the F300
STEP
Runs one measurement
DSA
Requests data output with handshaking
Internal circuit
COM IN
Input terminals
Output Specifications Number of outputs
Data outputs: 8 AND/OR outputs: 2
Output voltage
12 to 24 VDC +10%
Load current
100 mA max.
ON residual voltage
2 V max.
OFF leak current
0.1 mA max.
DO 0 to 7
Parallel data outputs
BUSY
Disables inputs during measurement
GATE
For data output timing
AND
AND output of evaluations on all windows
OR
OR output of evaluations on all windows
Internal circuit
Output terminals
COM OUT
26
Page 32
Section 2-2
Expansion Units Terminal Allocations Input indicators (orange) Output indicators (orange)
Pin no.
Pins 2 4
Pins 1 3
. . .
. . .
26
25
Name
Pin no.
Name
1
COM IN (for input)
14
GATE
2
STEP
15
BUSY
3
RESET
16
DO 0
4
DSA
17
DO 1
5
DI 0
18
DO 2
6
DI 1
19
DO 3
7
DI 2
20
DO 4
8
DI 3
21
DO 5
9
DI 4
22
DO 6
10
DI 5
23
DO 7
11
DI 6
24
AND
12
DI 7
25
COM OUT (for output)
13
NC
26
OR
Data input
Data output
27
Page 33
Section 2-2
Expansion Units Dimensions and Weight
The following diagram provides the external dimensions and weight of the F300-D Terminal Block Unit. Unless otherwise noted, values are in millimeters (mm). 14
34.6
133.5
250
Base Unit
Front cover
Weight: approx. 400 g
2-2-4 F300-DC Parallel I/O Unit The F300-DC Parallel I/O Unit is connected to a programmable controller or other device to input measurement commands and output measurement results. The following illustration shows the essential components of the F300-DC Parallel I/O Unit.
Components Input indicators (orange) Operation indicators
Output indicators (orange)
I/O connectors
Specifications General Specifications
28
Ambient operating temperature
0%C to 50%C
Operating humidity
35% to 85% RH (no condensation)
Operating environment
No corrosive gases
Storage temperature
–25%C to 65%C
Enclosure rating
IP20 (in-panel)
Page 34
Section 2-2
Expansion Units Input Specifications Number of inputs
Data inputs: 8
Input voltage
12 to 24 VDC +10%
ON current
3 to 15 mA
ON voltage
8.8 V max.
OFF current
0.1 mA max.
OFF voltage
4.5 V min
ON delay
RESET input: 10 ms max.; other inputs: 0.5 ms max.
OFF delay
RESET input: 15 ms max.; other inputs: 0.7 ms max.
Internal circuit COM IN
Input terminals
Output Specifications Number of outputs
Data outputs: 32
Output voltage
12 to 24 VDC +10%
Load current
100 mA max.
ON residual voltage
2 V max.
OFF leak current
0.1 mA max.
Internal circuit
Output terminals COM OUT
29
Page 35
Section 2-2
Expansion Units Terminal Allocation No.
Signal name
No.
Signal name
1
COM OUT (for output)
26
COM OUT (for output)
2
DO 31
27
DO 30
3
DO 29
28
DO 28
4
DO 27
29
DO 26
5
DO 25
30
DO 24
6
DO 23
31
DO 22
7
DO 21
32
DO 20
8
DO 19
33
DO 18
9
DO 17
34
DO 16
10
DO 15
35
DO 14
11
DO 13
36
DO 12
12
DO 11
37
DO 10
13
DO 09
38
DO 08
14
DO 07
39
DO 06
15
DO 05
40
DO 04
16
DO 03
41
DO 02
17
DO 01
42
DO 00
18
BUSY
43
GATE
19
NC
44
DI 7
20
DI 6
45
DI 5
21
DI 4
46
DI 3
22
DI 2
47
DI 1
23
DI 0
48
DSA
24
RESET
49
STEP
25
COM IN (for input)
50
COM IN (for input)
Data outputs
Data inputs
50
25
26
1
Inputs
30
DI 0 to 7
Parallel data inputs
RESET
Resets the F300
STEP
Runs one measurement
DSA
Requests data output with handshaking
Data outputs
Data inputs
Page 36
Section 2-2
Expansion Units Outputs DO 0 to 31
Parallel data outputs
BUSY
Disables inputs during measurement
GATE
For data output timing
Dimensions and Weight The following diagram provides the external dimensions and weight of the F300-DC Parallel I/O Unit. Unless otherwise noted, values are in millimeters (mm). 69.5
133.5
14
250
Front cover
Base Unit
Weight: approx. 800 g
2-2-5 F300-E RS-232C I/F Unit The RS-232C exchanges data with a personal computer or other device via the RS-232C interface. Menu operation and data settings are possible directly from the computer. The following illustration shows the essential components of the F300-E RS-232C I/F Unit.
Components Operation Indicators Channel 0 (orange) Channel 1 (orange)
RS-232C Connectors Channel 0
Channel 1
31
Page 37
Section 2-2
Expansion Units Specifications General Specifications
Communications Specifications
Ambient operating temperature
0%C to 50%C
Operating humidity
35% to 85% RH (no condensation)
Operating environment
No corrosive gases
Storage temperature
–25%C to 65%C
Enclosure rating
IP20 (in-panel)
Number of channels
2 channels
Electrical ratings
Conform to EIA RS-232C
Synchronization
Non-synchronous
Transmission speed
19,200 bps max.
Connector Pin Allocation
Pin no.
Connecting a Computer with a RS-232C I/F Unit
25
13
14
1
Signal name
Name
1
FG (GND)
Frame ground
2
SD (TXD)
Send data
3
RD (RXD)
Receive data
4
RS (RTS)
Request to send
5
CS (CTS)
Clear to send
6
DR (DSR)
Data set ready
7
SG (GND)
Signal ground
20
ER (DTR)
Data terminal ready
Connect the F300-VR as shown below. SD (TXD)
2
2
SD (TXD)
RD (RXD)
3
3
RD (RXD)
RS (RTS)
4
4
RS (RTS)
CS (CTS)
5
5
CS (CTS)
DR (DSR)
6
6
DR (DSR)
SG (GND)
7
7
SG (GND)
CD (DCD)
8
8
CD (DCD)
ER (DTR)
20
20
ER (DTR)
Note The F300-VR uses a 25-pin connector. Be sure to check the pin arrangement of the computer before wiring.
32
Page 38
Section 2-2
Expansion Units Dimensions and Weight
The following diagram provides the external dimensions and weight of the F300-E RS-232C I/F Unit. Unless otherwise noted, values are in millimeters (mm). 34.6 133.5
14
250
Front cover
Base Unit
Weight: approx. 400 g
2-2-6 F300-G Dummy Unit The F300-G Dummy Unit is inserted into empty slots to improve external appearance, protect connectors, and maintain the strength of the Rack. The Dummy Unit plays no role in measurements and consumes no power.
Specifications General Specifications Ambient operating temperature
0%C to 50%C
Operating humidity
35% to 85% RH (no condensation)
Operating environment
No corrosive gases
Storage temperature
–25%C to 65%C
Enclosure rating
IP20 (in-panel)
33
Page 39
Section 2-3
F300 Camera Lens Dimensions and Weight
The following diagram provides the external dimensions and weight of the F300-G Dummy Unit. Unless otherwise noted, values are in millimeters (mm). 34.6 133.5
14
250
Front cover
Base Unit
Weight: approx. 200 g
2-3
F300 Camera Lenses
2-3-1 Types of Lens The camera used with the F300 System can use a 12.5 to 75 mm CCTV lens or a 2x to 6x magnification lens. The selection of the correct lens between wide-angle and macro from this range allows optimum camera field-of-view and mounting distance. The camera requires a C-mount lens. C-mount lenses other than those listed below can also be used. When using Cmount lenses other than those recommended, determine the field-of-view and mounting distance by trial and error. CCTV Lens Model
Focal length (mm)
3Z4S-LE B1214D-2
12.5
3Z4S-LE C1614A
F number (F) F 1.4
Max. external diameter
Overall length
42 mm dia.
50 mm
16
30 mm dia.
33 mm
3Z4S-LE B2514D
25
30 mm dia.
37.3 mm
3Z4S-LE B5014A
50
48 mm dia.
48 mm
3Z4S-LE B7514C
75
62 mm dia.
79 mm
Magnifying Lens Model
Magnification
F number (F)
End diameter
3Z4S-LE ELE-008-2.5
2.5
F12
3Z4S-LE ELE-008-4
4
F17
67.0 mm
3Z4S-LE ELE-008-6
6
F24
91.5 mm
3Z4S-LE ELE-009-2
2
F12
3Z4S-LE ELE-009-4
4
F20
103.3 mm
3Z4S-LE ELE-009-6
6
F27
139.3 mm
34
14 mm dia.
Overall length
22 mm dia.
48.6 mm
67.4 mm
Page 40
Section 2-3
F300 Camera Lens
2-3-2 CCTV Lens, Field of Vision and Mounting Distance In this table, A = camera distance, WD = working distance, and t = extension tube thickness. The figures are approximate, when the 2/3-inch CCD dimension I = 6.546 mm. (Unit: mm) Field of vision 3Z4S-LE C815B
CCTV lens
3Z4S-LE B1214D-2
3Z4S-LE B2514D
3Z4S-LE B5014A
3Z4S-LE B7514C
100
150
200
300
500
A
5
7
10
15
20
30
50
70
153.2
218.2
283.3
413.7
674.4
WD
112.6
177.8
243.0
373.5
634.3 t
---
---
---
---
---
A
320.1
416.6
609.9
996.5
WD
269.5
366.2
559.6
946.4 t
---
---
---
---
A
101.0
152.8
302.2
415.9
606.1
796.5
1,177.6
1,940.1
WD
52.8
110.0
262.6
377.0
567.7
758.4
1,139.8
1,902.4 t
10
5
---
---
---
---
---
---
A
183.3
179.9
188.7
216.0
248.7
319.5
467.7
1,226.2
1,607.1
2,369.3
3,894.6
WD
69.9
85.2
108.0
146.2
184.3
260.6
413.2
1,176.0
1,557.4
2,320.2
3,845.9 t
65
45
30
20
15
10
5
---
---
---
---
A
217.7
258.6
307.7
414.1
636.7
863.0
1,204.6
1,775.8
2,347.8
3,492.6
5,783.2
WD
89.6
146.9
204.2
318.7
547.9
777.0
1,120.7
1,693.5
2,266.3
3,411.9
5,703.2 t
45
30
20
15
5
5
---
---
---
---
---
Flange surface (front of camera) 17.526 mm Camera mounting distance A (mm) Working distance, WD (mm) Camera body
CCTV lens f Field of vision L (mm)
Image sensor
H1 H2 a mm
Max. external radius
1”-32UN 30 dia. -2A
6.546 mm b mm
Total length
The above figures are approximate standards. Allow for some fine adjustment at the time of installation.