The text of a manual · part 4 of 8
DARTScanner Process Data Collector Operating Instructions: Installation, pages 25 to 33
Pages 25 to 33 of 52, from DARTScanner Process Data Collector Operating Instructions
The original document is the PDF (5.7 MB). This is its text as extracted, unedited; tables and drawings read best in the PDF, and each page heading opens it at that page.
Filed in the catalog under the RJG DS-3000
Page 25
RJG, Inc.
High Gain (500) Indirect Sensor Scaling Notes The DARTScannerTM will provide either a 0 to 20mV signal or a 0 to 10V signal relating to 0 to 1/2 of the full scale rating of the transducer. If the pressure on the transducer goes above half scale, the output will stay at 10 Volts (20mV). Refer to Figure 6 for a graphical representation. Example: A DARTScannerTM utilizing an indirect sensor with a full scale rating of 2,000 lb. sitting under a 1/4" ejector pin will provide a voltage of 5V or 10mV when the indirect sensor’s cavity is pressurized to 10,186psi. Pressure = (Signal Voltage/Maximum Signal Voltage) / (1/2 of Full scale Pressure/Ejector Pin Area) Pressure = 10,186psi = (5V/10V)*(1000lb./0.04909in²) (Conditioned High-Level Analog Signal Voltage) Pressure = 10,186psi = (10mV/20mV)*(1000lb./0.04909in²) (Conditioned Low-Level Analog Signal Voltage) Standard Mid Gain (250) Direct Sensor Scaling Notes The DARTScannerTM will provide either a 0 to 20mV signal or a 0 to 10V signal relating to 0 to the full scale rating of the transducer. If the pressure on the transducer goes above half scale, the output will follow it all the way to the full scale of the transducer. Refer to Figure 7 for a graphical representation. Example: a DARTScannerTM utilizing a direct sensor with a full scale rating of 20,000psi will provide a Voltage of 2.5V (Conditioned High-Level Analog) or 5mV (Conditioned Low-Level Analog) when the direct sensor’s cavity is pressurized to 5,000psi. Pressure = (Signal Voltage/Maximum Signal Voltage)*(Full scale Pressure) Pressure = 5,000psi = (2.5V/10V)*20,000psi (Conditioned High-Level Analog Signal Voltage) Pressure = 5,000psi = (5mV/20mV)*20,000psi (Conditioned Low-Level Analog Signal Voltage)
Standard Mid Gain (250) Indirect Sensor Scaling Notes The full scale entered for an indirect sensor will be its full scale force rating. The DARTScannerTM will provide either a 0 to 20mV signal or a 0 to 10V signal relating to 0 to 1/2 of the full scale rating of the transducer. Example: A DARTScannerTM utilizing an indirect sensor with a full scale rating of 2,000 lb. sitting under a 1/4" ejector pin will provide a voltage of 2.5V or 5mV when the indirect sensor’s cavity is pressurized to 10,186psi.
DARTScanner™ Rev 9.10
~28~
Page 26
RJG, Inc.
Pressure = (Signal Voltage/Maximum Signal Voltage) / (Full Scale Force/Ejector Pin Area) Pressure = 10,186psi = (2.5V/10V)*(2,000lb./0.04909in²) (Conditioned High-Level Analog Signal Voltage) Pressure = 10,186psi = (5mV/20mV)*(2,000lb./0.04909in²) (Conditioned Low-Level Analog Signal Voltage) Important Installation Notes Refer to Table 6 for a Machine Control Connector pin and pigtail cable guide. If only one control channel is being wired use those control signals labeled MPT1 (pins G, H, A, B, J and M). If the machine is to be wired for more than two channels of control, additional Machine Control Connectors can be added. Please refer to Tables 9 and 10 for important installation ratings. M i nimum control voltage output re s is t a n c e Conditio ned high- level a nalog output
1k Ohms
C o nd itioned low- level a nalog output
100k Ohms
Table 9: Minimum control voltage output resistance NOTE: The DARTScannerTM supplies its own excitation to the transducer. Do not wire the machine controller’s transducer excitation voltage to the
Control relay contact ratings Ite m
R ating
M a x imum switching power
50 watts
M a ximum switching voltage
500 VDC or Peak AC
Maximum switching and carry currents
1 A (limite d b y c ur r e nt limit fuse)
Maximum co n t r o l c o n t a c t s w i t c h i n g time after seeing set-point
13mS
DARTScannerTM. Table 10: Control relay contact ratings NOTE: The control relay is not horsepower rated and should not be used to control the machine solenoids directly.
~29~
DARTScanner™ Rev 9.10
Page 27
RJG, Inc.
4.4 Transducer Connector Interface Installation The DARTScannerTM is configured with multiple channels of signal conditioning for the following types of signals: •
• • • •
Strain Gage Type Hydraulic Transducers Strain Gage Type Mold Pressure Transducers Potentiometric Type Displacement Transducers LVDT Mold Deflection Transducers “J” or “K” Type Thermocouples (Extra signal conditioning may be required)
4.4.1 Transducer Inputs The bank of five Bendix connectors is used to input high or low level analog signals to the DARTScannerTM. These 6 pin female connectors are labeled HYD, MP1, MP2, MP3 and MP4, respectively. Typically, low level strain gage based signals from mold pressure sensors or hydraulic pressure sensors are input at these connectors. If properly configured, high level voltage signals may be inputted to any or all channels by switching the channel gain on the powered down DARTScannerTM, setting up the channel in DARTVisionTM and re-calibrating the channel. See Section 4.4.3 for more information on the gain switches. Figure 8: DARTScannerTM transducer input connectors
Pin
S ignal
A
EXC +
F ro m DARTS c a n n e r ™ to t r a n s d u c e r
B
EXC -
F ro m DARTS c a n n e r ™ to t r a n s d u c e r
C
S IG+
From transducer to D A R TS c a n n e r ™
D
S IG-
From transducer to D A R TS c a n n e r ™
E
Signal Flow
A B C
F ro m DARTS c a n n e r ™ SHIELD to t r a n s d u c e r
F
E D
N/A
Table 11: Transducer input connector pin-out guide DARTScanner™ Rev 9.10
F
~30~
Figure 9: Pin-out guide
Page 28
RJG, Inc.
NOTE: No metal tools should be used when working inside this box. Always power down this unit before opening the front panel door. Contact RJG, Inc. for technical assistance before attempting to reconfigure any channels in the field (231-947-3111).
4.4.2 Initial Factory Settings Unless otherwise specified, the DARTScannerTM is configured with the following standard channel assignments and associated gain settings: Cha nne l
As s ignme nt
Ga i n s e t ting
1
Hydraulic pressure
250
2
C avity pressure
250
3
C avity pressure
250
4
C avity pressure
250
5
C avity pressure
250
Table 12: Channel assignments & gain settings A gain of 250 works for most hydraulic and mold pressure applications. However, in low-pressure applications, some loss of resolution may be experienced. If transducer pressures are running under half scale for the entire cycle you may switch to a gain of 500 which will provide more resolution. In this case you must be careful of going over half-scale. An example of going over 50% of the full-scale force rating is: Using a Model T-405, which is a 500 lb. rated mold pressure transducer, under a 1/4" ejector pin. If there is force on the ejector pin equal to 490 lbs. (10,000 psi), which exceeds ½ of the rated transducer force, the gain should be set back to 250 for this channel on the DARTScannerTM box.
NOTE: Contact a Customer Support representative at RJG, Inc. if you are unsure as to what gain setting is required for your mold pressure application (231-947-3111).
~31~
DARTScanner™ Rev 9.10
Page 29
RJG, Inc.
4.4.3 Gain Set Switches Access to the Gain Switches is found inside the DARTScannerTM enclosure on the circuit board, which is mounted to the front panel door. Each channel on the DARTScannerTM can be configured with a gain of 1,250 or 500 depending on the specific application. CABLE BUNDLE (TRANSDUCER INPUT)
TB7
TB5
Channel 5
TB4
Channel 4 TB3 HINGE
Channel 3
500 250
1
TB2
Channel 2
GAIN SETTINGS
TB1
Channel 1 J6
TB7
Figure 10: DARTScanner™ Inside Front Panel Refer to Table 13 for the corresponding input signal level designations for the respective gain settings. Ty pical tra ns duc e r ty pe
Gain s witch pos i tion
Input range*
Position, temp e ra t ure or deflection sensors
Low gain (1)
0 to 5V, 0 to 10V
Hydraulic p ressure transducers
Mid gain (250)
- 2 0 mV to +20mV
Mold pressure transducers
Mid gain (250)
- 2 0 mV to +20mV
*Maximum DC common mode rejectio n to ±10V Table13: Gain Switch Positions, Transducer Input Ranges DARTScanner™ Rev 9.10
~32~
Page 30
RJG, Inc.
NOTE: The standard Gain switch position for a DARTScannerTM channel with an MPT input is Mid Gain (250). If you switch to High Gain (500) you must be watchful of your sensor going over half scale. If your pressure climbs above the transducer’s half scale, the measurement will “peg out” or plateau at the half scale level. In the software this will manifest itself in a curve that seems to have its top cut off. If this happens, changing that channel’s Gain switch to Mid Gain (250) and re-calibrating that channel will give you the full scale range of the transducer. If the problem still exists, then you may have to reevaluate your pin size or you may need an RJG mold pressure transducer with a higher rated load.
4.4.4 Interfacing With An STX-25V Stroke/Velocity Transducer The following describes how to interface an STX-25V Stroke/Velocity Transducer to the DARTScannerTM: 1. Power down the DARTScannerTM. 2. Change the gain settings for the stroke and velocity channels to 1. 3. Power up the DARTScannerTM. 4. In DARTVisionTM select the STX-25V and STX-VEL measurements for transducer scaling. 5. Recalibrate the channels.
4.4.5 Interfacing With A Mold Deflection Transducer The following describes how to interface a T-250 Mold Deflection Transducer to the DARTScannerTM unit. Properly interfacing the transducer and DARTScannerTM requires the selected channel to be rewired using a pair of internally mounted signal selection jumpers. Failure to do so will result in incorrect or non operation of the mold deflection transducer. If you are at all hesitant about which jumpers to move or are reluctant to make the required changes to your DARTScannerTM unit, please call RJG’s Customer Service for assistance (231-947-3111). 1. Turn the power off to the DARTScannerTM. 2. Attach a T-520MD cable to the T-250 transducer. 3. Attach the other end of the cable to the channel you have designated for Mold Deflection. The channel’s Bendix connector is located on the top panel of the DARTScannerTM.
~33~
DARTScanner™ Rev 9.10
Page 31
RJG, Inc.
4. Open the Rear Access door of the DARTScannerTM unit to change the position of the following jumpers (See Figure 11). Move only those channel jumpers which you have selected to interface with the mold pressure transducer. (i.e. If you will be using Channel 3 for mold deflection, move only those jumpers for Channel 3.)
TB3
Channel 3 C
1
3
2
+
--
Positions 1-2 & 1-2 3-2 & 3-2
1
R
Settings Standard Mold Deflection
Channel 4
Channel 5
+
R
C
C
--
+
R
C
--
+
TB5
+
TB4
Channel 3
TB3
Channel 2
TB2
TB1
Channel 1
2
3
R
C
--
+
R
C
-R
--
Jumper Positions
TB7 TB7
TM
DARTScanner Front Panel D5 Board Layout
Figure 11: Board with jumpers
Channel
+EXC selection jumper
-EXC/Ground selection jumper
1
JP2
JP1
2
JP4
JP3
3
JP6
JP5
4
JP8
JP7
5
JP10
JP9
Table 14: Mold deflection jumper assignments Move your selected channel jumpers to the position A-B, respectively. Be sure to include the mold deflection transducer in your DARTVisionTM software setup.
DARTScanner™ Rev 9.10
~34~
Page 32
RJG, Inc.
4.5 Computer and Network Interface Installation A DARTScannerTM can communicate with a computer either as a stand-alone unit via the RS-232 connector or on a network with other DARTScannersTM through the DARTNET (RS-485) connector.
4.5.1 RS-232 (Stand-alone Computer Interface) A computer can be made to communicate with one stand alone DARTScannerTM by using the connector and the proper cable (see Figure 12). No converters are needed but remember that using the RS-232 connection limits the number of DARTScannersTM connected to the computer to one.
TM
DARTScanner
RS-232 to Computer's COM Port DB-9 Male to DB-9 Female Cable
Figure 12: Interfacing a DARTScanner™ to a stand-alone computer using the RS-232 Computer Hook-up on the DARTScanner’s side panel A standard male DB-9 to a female DB-9 serial cable provides the connection between a DARTScannerTM and a standard RS-232 serial port. Refer to Table 15 for a pin-out guide. TM
Pin
Signal (RS-232)
2
Receive Data
3
Transmit Data
7
Request to Send
5
Ground - Shield
Table 15: RS-232 pin-outs
~35~
DARTScanner™ Rev 9.10
Page 33
RJG, Inc.
4.5.2 DARTNET Connector (RS-485 Computer Interface) A computer can be made to communicate with multiple DARTScannersTM by using the DARTNET connector and the proper cables (see Figure 12). An RS-485 network utilizes a two wire communications “Bus” with terminating resistors on each end. The DARTScannersTM are attached to the bus at points called “Nodes” that are attached to the bus using “Stubs”. These stubs are best kept as short as possible. In order to keep the stubs electrically short, the bus is run to the DARTScannerTM and returned on another pair. Figure 13 details this setup using a SP-321M splitter and related cabling.
TO COM PORT ON COMPUTER
"BUS" SP-321
SP-321 TP-321
T-320
CN-485
"STUB"
"STUB" TM
TM
DARTScanner
DARTScanner TPJ-361
TPJ-361
"NODE"
"NODE"
Figure 13: Interfacing a Multi-Machine DARTScanner Network to a computer using Network Cabling and the RS-485 DARTNET Connector on the DARTScanner TM
TM
Pin
Signal (RS-485)
Direction
1
System ground
--
4
Transmit/Receive+
BIDIR
5
Transmit/Receive-
BIDIR
Table 16: DARTNET pin-outs Important Note: Either the RS-232 or the RS-485 DARTNET connector can be used at one time. They cannot be used together.
DARTScanner™ Rev 9.10
~36~