Page 42

RJG, Inc.

6.3 Appendix C: Using Accessories 6.3.1 Sensor Simulator use instructions Introduction: The Sensor Simulator is a transducer simulator capable of checking the operations of DARTScanner™ and DARTPAK™ data collecting devices and accompanying cables. This test device utilizes a 350Ω star bridge built with precision resistors. The Simulator provides the following outputs and corresponding load (or psi) simulation for various RJG sensor models: Strain-Gage Switch Positions (mVolts) Output*

T-405/T412 T-425

T-406/T-413 T-426

T-410 T-414

T-3000U

LOW: 4mV 20% FS

100 lb.

400 lb.

25 lb.

600psi

MID: 8mV 40% FS

200 lb.

800 lb.

50 lb.

1200psi

HIGH: 16mV 80% FS

400 lb.

1600 lb

100 lb.

2400psi

Table 19: Strain-gage positions High-Level Switch Positions (Volts) Output*

STX-25V (Stroke)

STX-25V (Velocity)

LOW 1V

5" (20% FS)

1.6 IPS (20% FS)

MID 2V

10" (40% FS)

3.2 IPS (40% FS)

HIGH 4V

20" (80% FS)

5.4 IPS (80% FS)

Table 20: High-level switch positions Checking RJG Cables and Data Collecting Devices: 1. Unplug the sensor and cable from the DARTScanner™ or DARTPAK™ that you want to check. (If you are using older RJG equipment, call RJG Technical Support for assistance.) 2. Plug the Sensor Simulator into the sensor input channel on the DARTScanner™ or DARTPAK™ device with the provided 1-foot cable. 3. Set the Sensor Simulator output to 0V. For complete accuracy, let the Sensor Simulator warm-up for about 5 minutes. 4. Zero and calibrate the DARTScanner™ or DARTPAK™ per the unit’s instructions. 5. Set the Sensor Simulator output to LOW for either a high level or strain gage Input channel. 6. If the DARTScanner™ or DARTPAK™ display reads the correct value, the data acquisition device is working correctly (reference the tables above for the percentage value). The correct value should be within ± 1% of full scale. For example: for a full scale of 40,744, your expected 20% value range would be 8,067psi to 8,230psi. 7. If the DARTScanner™ or DARTPAK™ is working correctly, then test the suspect cable by replacing the 1-foot Sensor Simulator cable with the cable to test. 8. Repeat steps 3-5.

DARTScanner™ Rev 9.10

~46~

Page 43

RJG, Inc.

1. If the display reads the correct value, then the cable is working correctly also. 2. Hook the cable back up to the transducer. NOTE: After completing Steps 1-8 and if the problem still exists, call Tech Support @ 1-231-947-3111 and reference document Tech Note #105.doc. If you have a Sensor Tester, use it to verify proper sensor operation and reference document Tech Note #112.doc.

6.3.2 Transducer Cable Tester instructions Introduction The RJG, Inc. Transducer Cable Tester is a portable device capable of testing all pins of a standard RJG T-5XX transducer cable (Pin F not applicable). It can also test those cables used by other manufacturers that utilize a Bendix connector shell size of 10 and a male pin configuration of 6. Testing Cables Follow these steps to test a transducer cable: 1. Plug the ends of the cable to be tested into the female cable connectors on the Transducer Cable Tester. 2. Press the PIN A/+EXC button. If the Pin A LED does not light then that wire of the cable is open. If the Pin A LED is lit along with one or more other LEDs, there is a short between these pins. 3. Repeat step 2 for all Pins (Pin F is not supported on most cables). 4. If any test fails, have the cable repaired immediately to prevent possible damage to equipment. Dimens i o n s Environmental Power Requirements C urrent Drain Connectors Buttons Continuity Lights

5 . 8 "x3 . 4 "x1 . 5 " (1 4 7 mm x 8 6 mm x 38mm) N EMA 1 3 Vo l t s p r o v i d e d b y 2 A A A b a t t e r i e s 2 5 mA Bendix bayonet style female She ll size 10 P in configuration 6 Momentary membrane switches Green LEDs

Table 21: Specifications

Figure 23: Transducer cable tester ~47~

DARTScanner™ Rev 9.10

Page 44

RJG, Inc.

6.4 Appendix D: Troubleshooting Guide Table 22: Troubleshooting Guide

1

2

Proble m

P o s s i b l e C a us e (s )

Solution

N o dis pla y whe n t h e u n i t po we r s u p

This could be due to the power wires not being c o n n e c t e d p r o p e r l y.

D o u b l e c h e c k i n s t a l la t i o n .

N o p o w e r d ue t o a b l o w n fus e i n s i d e o f t h e p o w e r s up p ly.

Re p l a c e t h e b l o w n f u s e .

Is the machine T-INT14-D ma c hine inte r f a c e c a b l e c o n n e c t e d t o t he D A R TS c a n n e r ™ unit?

C o nnect the cable

A b a d c o m munication (C O M M ) c a b l e .

Re p l a c e t he c a b l e .

I s t h e C O M M c a b le c o n n e c t e d b e t w e e n t he D A R TS c a n n e r ™ a n d t h e serial port?

C o nne c t t he c a b le .

I s t h e C O M M p o r t t ha t is b e ing us e d s e le c t e d in the D A R TV i s i o n ™ s o f t w a r e ?

Select the correct COMM port.

Do the D A R T V i s i o n ™ software settings correspond to the D A R TS c a n n e r ™ s e t t i n g s ? i.e. Correct DART addresses.

C o rrect the settings. (See the DARTV i s i o n ™ s o ft w a r e m a n u a l f o r m o r e in f o r m a t i o n . )

N o c o m munic a t i o n b e t we e n the c o m p u t e r and the DARTScanner™ unit

More than one machine in C he c k t he a d d r e s s e s o f a l l the DART V i s i o n ™ s o ftware the machines in the ma y b e t r y i n g t o t a l k t o t h e D A R TV i s i o n ™ so f t w a r e . s a me D A R TS c a n n e r ™ M o r e t ha n o ne D A R TS c a n n e r ™ ma y b e r e s p o nd i n g t o t h e s a m e address.

DARTScanner™ Rev 9.10

~48~

C he c k t he a d d r e s s e s o f a l l t he D A R TS c a n n e r s ™ o n t he ne t w o r k .

Page 45

RJG, Inc.

3

4

5

Proble m

Pos s ible Ca u s e ( s )

Solution

Trigge r LED Indicators are not ope rating correctly

Are the triggers installed/wired to the machine correctly?

C h e c k t he installation.

Is the machine T-INT14-D machine interface cable connected to the DARTScanner ™ unit?

Check the cable connection.

Are the machine interface cable and the mating harness cable connected together correctly?

C h e c k t he connection.

N o data being displayed on Are the transducer cables the compute r screen connected to the correct corresponding channel inputs?

D isplay won't zero or calibrate

Check the transducer's output. See the Mold Pressure Transducer Manual for more information on RJG' s mo ld pressure transducers.

Is the trigger 1 (INJ FWD) LED coming on?

Refer to Problem 3 .

Possibly a d e fective cable.

A repla c e ment cable is needed.

Transducer has a large amount Remove the pre-load from the o f p r e - lo a d . transducer.

Wrong gain switch settings.

Check the gain switch settings.

Possibly a defective transducer.

Check the transducer.

~49~

DARTScanner™ Rev 9.10

Page 46

RJG, Inc.

6

7

Problem

Possible Cause(s)

Solution

Process alarm LEDs are not lighting up

The alarms may not be "ON" in the DARTVision™ software.

Check the alarm settings in the DARTVision™ software.

Are the alarm values in the DARTVision™ software correct for the corresponding channel assignments?

Make corredtions to the alarm values.

The standard gain switch position for a DARTScanner™ channel with MPT input is Mid Gain (250). However, if you are using High Gain (500) and your presssure climbs above the transducer's half scale, the measurement will "peg out" or plateau at the half scale level. In the software this will manifest itself in a curve that seems to have its top cut off.

change that channel's gain switch vack to the Standard Mid Gain (250) and recalibrate that channel. This will give you the full-scale range of the transducer.

When using a DARTScanner™ to monitor hydraulic or mold pressure, the curve displayed in the DARTVision™ software seems to be pressure limited, or "chopped off" at the half acale value of the transducer.

DARTScanner™ Rev 9.10

~50~

If the problem still exists, then you may have to re-evaluate your pin size or you may need an RJG mold pressure transducer with a higher rated load.